Lyphochek® Hemostasis Control is intended for use as a quality control plasma to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
Device Story
Lyphochek® Hemostasis Control is a lyophilized, human plasma-based quality control product used in clinical coagulation laboratories. It contains various biochemical analytes, including PT, APTT, Fibrinogen, Antithrombin III, Thrombin Time, and various clotting factors. The control is reconstituted by laboratory personnel and used to monitor the precision of laboratory testing procedures. By comparing the measured values of the control against expected ranges, clinicians and laboratory technicians verify the accuracy and reliability of their diagnostic testing equipment. This process ensures that patient test results are consistent and reliable, ultimately supporting accurate clinical decision-making for patients undergoing coagulation testing.
Clinical Evidence
No clinical data. Performance is supported by bench testing, specifically stability studies confirming reconstituted stability (8 hours at 2-25°C, except Protein S at 2-8°C) and shelf life (3 years at 2-8°C).
Technological Characteristics
Lyophilized human plasma-based control containing purified biochemicals and preservatives. Dimensions/form factor: Reconstituted liquid. Storage: 2-8°C. Analyte list includes PT, APTT, Fibrinogen, AT III, TT, Factors II, V, VII, VIII, IX, X, XI, XII, Protein S, Protein C, and Plasminogen.
Indications for Use
Indicated for use as a quality control plasma to monitor the precision of laboratory testing procedures for analytes listed in the package insert. No specific patient population, age, or gender is defined as the device is a laboratory control product.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{1}------------------------------------------------
K020878
Page 1 of 2
## Summary of Safety and Effectiveness Lyphochek® Hemostasis Control
APR 1 5 2002
#### 1.0 Submitter
Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1555 Fax:
## Contact Person
Maria Zeballos Regulatory Affairs Specialist Telephone: (949) 598-1367
### Date of Summary Preparation
March 15, 2002
#### 2.0 Device Identification
| Product Trade Name: | Lyphochek® Hemostasis Control |
|---------------------|--------------------------------------------------|
| Common Name: | Control Plasma Normal<br>Control Plasma Abnormal |
| Classifications: | Class II |
| Product Code: | GGC |
| Regulation Number: | 21 CFR 864.5425 |
#### 3.0 Device to Which Substantial Equivalence is Claimed
Coagulation S.A.C Assayed Control 1 Helena Laboratories Beaumont, Texas
Docket Number: K941737
Coaqulation S.A.C Assayed Control 2 Helena Laboratories Beaumont, Texas
Docket Number: K941872
#### 4.0 Description of Device
Lyphochek® Hemostasis Control is prepared from human plasma with added purified biochemicals and preservatives. The control is provided in lyophilized form for increased stability.
{2}------------------------------------------------
#### Statement of Intended Use 5.0
Lyphochek® Hemostasis Control is intended for use as a quality control plasma to Eyphooner - Fromootable aboratory analytes listed in the package insert.
#### Comparison of the new device with the Predicate Device 6.0
Lyphochek® Hemostasis Control claims substantial equivalence to the Coagulation S.A.C Assayed Control 1 and 2 currently in commercial distribution.
| | Bio-Rad | Helena Laboratories | Helena Laboratories |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Lyphochek® Hemostasis Control<br>(New Device) | Coagulation S.A.C.-1 Assayed<br>Control 1<br>(Predicate Device - K941737) | Coagulation S.A.C.-2 Assayed<br>Control 2<br>(Predicate Device - K941872). |
| | Similarities | | |
| Intended Use | Lyphochek® Hemostasis Control<br>is intended for use as a quality<br>control plasma to monitor the<br>precision of laboratory testing<br>procedures for analytes listed in<br>the package insert. | S.A.C.-1 is assayed control<br>plasma, which may be used in<br>many facets of testing in the<br>coagulation laboratory. | S.A.C.-2 is assayed abnormal<br>control plasma, which may be<br>used in many facets of testing in<br>the coagulation laboratory. |
| Form | Lyophilized | Lyophilized | Lyophilized |
| Matrix | Human plasma based | Human plasma based | Human plasma based |
| | Differences | | |
| Storage | 2-8°C | 2-6°C | 2-6°C |
| (Unopened) | Until expiration date | Until expiration date | Until expiration date |
| Reconstituted<br>Vial Claim | 8 hours at 2°C to 25°C with the<br>following exception:<br>Protein S will be stable for 8 hours<br>at 2°C to 8°C. | 4 hours 2-6°C | 4 hours 2-6°C |
| Analytes | Contains: Prothrombin Time (PT),<br>Activated Partial Thromboplastin<br>Time (APTT), Fibrinogen,<br>Antithrombin III (AT III) Thrombin<br>Time (TT), Factor II, V, VII, VIII, IX,<br>X, XI, XII, Protein S (Functional)<br>and Protein C (Functional),<br>Plasminogen<br>Does not Contain: Alpha-<br>Antiplasmin, von Willebrand Factor<br>Antigen, Ristocetin cofactor | Contains: Prothrombin Time<br>(PT), Activated Partial<br>Thromboplastin Time (APTT),<br>Fibrinogen, Antithrombin III (AT<br>III), Factor II, V, VII, VIII, IX, X, XI,<br>XII, Total and Free Protein S,<br>Protein C, Plasminogen, von<br>Willebrand Factor Antigen, Alpha-<br>antiplasmin, Ristocetin cofactor<br>Does not Contain: Thrombin<br>Time (TT) | Contains: Prothrombin Time<br>(PT), Activated Partial<br>Thromboplastin Time (APTT),<br>Fibrinogen, Antithrombin III (AT<br>III), Factor II, V, VII, VIII, IX, X, XI,<br>XII, Total and Free Protein S,<br>Protein C, Plasminogen,<br>Ristocetin cofactor, Alpha-<br>antiplasmin, Monoclonal Free<br>Protein S, von Willebrand Factor<br>Antigen<br>Does not Contain: Thrombin<br>Time (TT) |
| | Table 1. Similarities and Differences between new and predicate device. | | | | | |
|--|-------------------------------------------------------------------------|--|--|--|--|--|
|--|-------------------------------------------------------------------------|--|--|--|--|--|
#### Summary of Performance Data 7.0
Stability studies have been performed to determine the reconstituted stability and shelf life for the Lyphochek® Hemostasis Control. Product claims are as follows:
{3}------------------------------------------------
- 7.1 Reconstituted Stability: Once the control material is reconstituted, all analytes will be stable for 8 hours at 2°C to 25°C with the exception of Protein S which will be stable for 8 hours at 2ºC to 8ºC.
- Shelf Life: Three years when stored at 2 to 8 °C. 7.2
Real time studies will be ongoing to support the shelf life of this product. All supporting data is retained on file at Bio-Rad Laboratories.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# APR 1 5 2002
Ms. Maria Zeballos Regulatory Affairs Specialist Bio-Rad Laboratories 9500 Jeronimo Road Irvine, California 92618-2017
Re: k020878
Trade/Device Name: Lyphochek® Hemostasis Control Regulation Number: 21 CFR § 864.5425 Regulation Name: Control Plasma Abnormal Regulatory Class: II Product Code: GGC Dated: March 15, 2002 Received: March 18, 2002
Dear Ms. Zeballos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it 11 your de roo to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510 (k) Number (if known):
Lyphochek® Hemostasis Control Device Name: Indications for Use:
Lyphochek® Hemostasis Control is intended for use as a quality control plasma to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription use | <span style="text-decoration: overline;">✓</span> | or | Over-the Counter use | <span style="text-decoration: overline;"></span> |
|------------------|---------------------------------------------------|----|----------------------|--------------------------------------------------|
|------------------|---------------------------------------------------|----|----------------------|--------------------------------------------------|

| (Division/Sign-Off) | |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K020878 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.