K040064 · Newmedical Technology, Inc. · KTT · Mar 23, 2004 · Orthopedic
Device Facts
Record ID
K040064
Device Name
NEWMEDICAL MICRO BONE DISTRACTOR/FIXATOR
Applicant
Newmedical Technology, Inc.
Product Code
KTT · Orthopedic
Decision Date
Mar 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
indicated for use for pediatric as well adult Metacarpal-Phalangeal fractures. For external fixation of hand fractures or for distraction of osteotomies for the purpose of distraction osteogenesis as fractures.
Device Story
Micro distractor/external fixator for stabilization/lengthening of bone and bone fragments; used for metacarpal-phalangeal fractures and distraction osteogenesis. Device attaches to bone via pins/threaded wires held by clamps; surgeon manually adjusts recessed distraction screw to control compression/distraction length. Tamper-resistant design requires custom actuator provided in kit; beneficial for pediatric patients. Features lightweight construction; alloy metals provide stability. Device functions as both dynamic and static fixator without part adjustments. Used by surgeons in clinical settings to stabilize fractures or lengthen bone; output is mechanical force applied to bone segments to facilitate healing or osteogenesis.
Clinical Evidence
No clinical data. Substantial equivalence based on design, material, and operational principle comparisons to legally marketed predicate devices.
Technological Characteristics
Materials: Stainless steel (ASTM F 138-97), gold plating, electro-polishing. Sensing/Actuation: Manual screw-driven mechanical distraction/compression. Form factor: Micro-profile external fixator with guide rods and clamps. Connectivity: None (mechanical). Sterilization: Not specified. Software: None.
Indications for Use
Indicated for pediatric and adult patients requiring external fixation of metacarpal-phalangeal fractures or distraction of osteotomies for distraction osteogenesis.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
NewMedical Micro Bone Distractor/Fixator (K030485)
Reference Devices
Hoffman Mini Fixator
Synthes Mini-Lengthener
Hand Bone Lengthener
Related Devices
K030485 — NEWMEDICAL MICRO BONE DISTRACTOR/FIXATOR · Newmedical Technology, Inc. · Dec 3, 2003
{0}------------------------------------------------
MAR 2 3 2004
K040064
page 1 of 2
# 510(k) Summary
Image /page/0/Picture/3 description: The image contains the words "New medical Technology". The word "New" is in a larger, bold font. The word "medical" is in a smaller, regular font. The word "Technology" is in a very small, italicized font and is located below the word "medical".
#### NewMedical Technology, Inc
N.14 W.23833 Stone Ridge Dr. Suite G100 Waukesha, WI 53188 USA 262- 523 0521 Main 262- 513 0799 Fax 262-391 7899 Mobile
Date of Submission: December 10 , 2003
Submitted by:
Haitham Matloub, Senior Engineer, Research and Development
The micro distractor/external fixator introduced by NewMedical Technology, Inc., is a device which is substantially equivalent to other marketed external fixators /distractors. Its indications for use are substantially the same, that is, external fracture fixation and/or distraction. The NewMedical Micro Distractor/External Fixator described here is completely identical to the NewMedical Micro Bone Distractor/Fixator that has already been approved by the (FDA/Dental Devices Committee) for use in external fixation of mandibular fractures or for mandibular distraction, referring to NewMedical Technology, Inc. product (K030485).
The main advantages over other marketed devices from other vendors include its' smaller profile and lightweight construction. The recessed distraction actuating screw is tamper resistant and hence better for use in pediatric patients. The device can only be actuated with the use of a custom actuator which is provided in the kit with the device. Hence, this tamper resistant design is useful when the device is used as an external bone fixation device for the treatment of fractures. Similarly, the conversion to a static from a dynamic fixator does not reguire any adjustment or addition of parts. The alloy metals used provide for excellent stability.
In summary, the features incorporated into the NewMedical Micro Distractor/External Fixator have specific advantages over other substantially equivalent products to enhance ease of placement along with the care and comfort of the patient requiring external fixation/distraction of the mandible or small bones. Neither one of the above features will raise any safety concern because they don't change any of the basic design concepts compared to other similar approved devices.
{1}------------------------------------------------
04004
page 2 of 2
### Equivalence To Marketed Devices:
Equivalency of the NewMedical Micro Distractor/Fixator Device, is based on similarities in intend use, materials, design and operational principles to the predicate external distractor ffixation devices.
The design features of the NewMedical Micro Distractor/Fixator Device, and the named predicate devices are substantially equivalent as these devices attach to bone and move the bone outward when mechanical distraction forces are applied. In addition to moving bones, the NewMedical Micro Distractor/ Fixator Device, is substantially equivalent to the named derives in stabilizing fractured bone structures.
#### 1-Intended Use:
All of the named devices, and the NewMedical Micro Distractor/Fixator Device, are intended to be used as external fixation/ distraction devices for the stabilization/lengthening of bone and bone fragments.
#### 2-Materials:
The materials used in the manufacturing of the NewMedical Micro Distractor/Fixator Device. are two grades of Stainless steel. The material conforms to ASTM F 138-97. Gold plating and Electro polishing are applied to all device components. Stainless steel and gold materials have long history of successful clinical usage. Stainless steel is used in the manufacture of the Hoffman Mini Fixator, Synthes Mini-Lengthener and the Hand Bone Lengthener.
Lengthening screw Stainless Steel 316L/gold plating Locking set screws: SS 18-8/gold plating Guide rods ( rails) SS 316L/Gold plating Body SS 316 L/Gold plating
#### 3- Design:
The basic design features of the NewMedical Micro Distractor/Fixator Device are substantially equivalent to the predicate devices. The subject device and the named derives accept various diameter pins. The subject device and the named devices operate using the same principle by utilizing screw mechanism that the surgeon adjusts to control the distraction/compression length. The subject device and the predicate devices use clamps attached to the distraction screw for holding the pins or threaded wires. Distraction/compression occurs by turning the screw shaft clockwise or counter clockwise causing attached clamps to slide along the axial axis of the screw shaft.
Lengthening screw Stainless Steel 316L,
Locking set screws: SS 18-8/gold Guide rods ( rails) SS 316L Body SS 316 L
#### 4- Operational Principles:
The basic operational principles for the named device and the subject device are similar - mainly in site preparation, pin placement and device operation. The bone distraction/compression is similar also.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 3 2004
Mr. Haitham Matloub Senior Engineer, Research and Development NewMedical Technology Inc. N.14 W.23833 Stone Ridge Drive Suite G100 (Mail Box # 1) Waukesha, Wisconsin 53188
Re: K040064
Trade/Device Name: NewMedical Micro Bone Distractor/Fixator Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: KTT Dated: December 23, 2003 Received: February 18, 2004
Dear Mr. Matloub:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Haitham Matloub
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mule N Mellekson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K040064
Device Name: NewMedica! Micro Bone Distractor/Fixator
Indications For Use: indicated for use for pediatric as well adult Metacarpal-Phalangeal fractures. For external fixation of hand fractures or for distraction of osteotomies for the purpose of distraction osteogenesis as fractures.
Mark N Milkeran
Division Sign Off
Division of General, Restorative, and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C.)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.