The MULTIGENT™ Hb A1c assay is used in clinical laboratories for the quantitative in vitro measurement of percent Hb A1c (hemoglobin fraction) in human whole blood on the AEROSET® System and ARCHITECT® c8000™ System. The Hb A1c assay is intended to aid in the monitoring of long-term blood glucose control and compliance in individuals with diabetes mellitus. The MULTIGENT™ Hb A1c assay is not intended for use in diagnosing diabetes mellitus. The MULTIGENT™ Hb A1c Calibrators are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for the calibration of the assays in the MULTIGENTTM Hb A1c Reagent Kit. The MULTIGENT™ Hb Alc Controls are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for quality control of the assays in the MULTIGENTTM Hb A1c Reagent Kit.
Device Story
Device comprises reagents, calibrators, and controls for use on Abbott AEROSET and ARCHITECT c8000 systems. Input: human whole blood specimen. Process: erythrocytes lysed; hemoglobin degraded by pepsin to form hemolysate. Total hemoglobin (THb) measured colorimetrically at 604 nm. Stable HbA1c measured via immunoturbidimetric microparticle agglutination inhibition at 700 nm; HbA1c in sample competes with hapten-bound polymer for antibody-coated microparticles, slowing agglutination rate. Output: % HbA1c calculated as HbA1c/THb ratio. Used in clinical laboratories by trained personnel. Results aid clinicians in monitoring long-term glycemic control and patient compliance in diabetes management.
Clinical Evidence
Bench testing only. Precision studies (n=80) per NCCLS EP5-A showed total precision of 1.46% CV at 5.0% HbA1c and 1.31% CV at 10.5% HbA1c. Method comparison (n=117) against predicate yielded correlation coefficients of 0.993 (Aeroset) and 0.994 (c8000). Linearity established from 2% to 20% HbA1c.
Technological Characteristics
Immunoturbidimetric microparticle agglutination inhibition assay for HbA1c; colorimetric assay for total hemoglobin. Reagents include mouse monoclonal anti-HbA1c antibodies on microparticles and hapten-bound polymer agglutinator. System: Abbott AEROSET/ARCHITECT c8000. Wavelengths: 604 nm (THb), 700 nm (HbA1c).
Indications for Use
Indicated for quantitative in vitro measurement of percent HbA1c in human whole blood to monitor long-term blood glucose control and compliance in individuals with diabetes mellitus. Not for use in diagnosing diabetes.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Seradyn. The logo consists of the word "Seradyn" in a bold, sans-serif font, with the "S" slightly larger than the other letters. Above the word is a stylized globe with curved lines wrapping around it. The logo is black and white.
Suite 1000 Indianapolis, IN 46268
### 510K SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K0336
### 1. COMPANY/CONTACT PERSON:
Seradyn, Inc 7998 Georgetown Road, Suite 1000 Indianapolis, IN 46268
Establishment registration No: 1836010
Les Padilla Technical Product Manager Telephone: (317) 610-3823 Fax: (317) 610-0018 e-mail: lpadilla@seradyn.com
#### 2. DATE PREPARED:
November 18, 2003
#### 3. DEVICE NAME:
| a. Proprietary Name: | MULTIGENTTM Hemoglobin A1c Reagents<br>MULTIGENTTM Hb A1c Calibrators<br>MULTIGENTTM Hb A1c Controls |
|-------------------------|------------------------------------------------------------------------------------------------------|
| b. Common Name: | Hemoglobin A1c (Hb A1c); Glycated / Glycosylated<br>Hemoglobin |
| c. Classification Name: | Class II, LCP, 21 CFR 864.7470 Glycosylated<br>Hemoglobin assay |
#### 4. LEGALLY MARKETED DEVICES TO WHICH EQUIVALENCY IS CLAIMED:
Tosoh Medics Inc., G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode, cleared under K011434.
{1}------------------------------------------------
#### 5. DESCRIPTION OF DEVICE:
The Device consists of the MULTIGENT™ Hemoglobin A 1 c Reagents, MULTIGENT™ Hemoglobin Alc Calibrators , and MULTIGENT™ Hemoglobin A 1 c Controls, intended for use on AEROSET® System and ARCHITECT® c8000™ System for determination of stable % HbAlc.
The assay consists of two separate concentration measurements, the stable form of glycated hemoglobin (Hb A 1 c) and the total hemoglobin (THb), which are used only to determine the percent Hb Alc. and must not be used individually for diagnostic purposes.
The whole blood specimen is pre-treated to lyse the erythrocytes. The hemoglobin is degraded by the proteolytic enzyme, pepsin, to form a hemolysate. Both the THb and the Hb A1c concentrations are determined from the same hemolysate.
The concentration of total hemoglobin is determined colorimetrically using a wavelength of 604 mm. The sample's measured absorbance is compared to a two-point calibration curve for total hemoglobin.
The concentration of stable Hb Alc is measured immunoturbidimetrically using a microparticle agglutination inhibition method. The Hb A1c antibody reagent (R1) contains specific anti-Hb A 1c mouse monoclonal antibodies coupled to microparticles. The Hb A1c agglutinator reagent (R2) contains several copies of the immunoreactive portion of Hb A1c (hapten), covalently bound to a polymer.
In the absence of Hb Alc in the sample, the hapten in the R2 reagent binds with the antibodycoated microparticles in the R1 antibody reagent and results in an increase in the rate of agglutination and results in an increase in measured absorbance. In the presence of HD A Ic in the sample, the Hb Alc competes with the hapten in the R2 reagent for binding sites on the antibody-coated microparticles in the R1 antibody reagent and will slow the rate of agglutination as it competes with the Hb Alc agglutinator for antibody binding sites.
The increase in concentration of Hb Alc in the sample is inversely proportional to the rate of agglutination and the measured absorbance. The absorbance is measured using a wavelength of 700 nm. The measured absorbance of the sample is compared to the measured absorbance of known Hb Alc concentrations (g/dL) of a six-level calibration curve, and the concentration of the sample is interpolated. The percent Hb A1c is the Hb A1c /THb ratio, calculated automatically by the AEROSET® System and ARCHITECT® c8000™ System, using a conversion factor to correlate the result with an NGSP-certified method.
The calibrators are supplied in liquid form and are ready to use without pretreatment. The controls are supplied in lyophilized form and are to be reconstituted with the supplied reconstitution fluid.
{2}------------------------------------------------
#### 6. INTENDED USE:
The MULTIGENT™ Hb A1c assay is used in clinical laboratories for the quantitative in vitro measurement of percent Hb Alc (hemoglobin fraction) in human whole blood on the AEROSET® System and ARCHITECT® c8000™ System. The Hb A 1 c assay is intended to aid in the monitoring of long-term blood glucose control and compliance in individuals with diabetes mellitus. The MULTIGENT™ Hb A1c assay is not intended for use in diagnosing diabetes mellitus.
The MULTIGENT™ Hb A1c Calibrators are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for the calibration of the assays in the MULTIGENTTM Hb A1c Reagent Kit.
The MULTIGENT™ Hb Alc Controls are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for quality control of the assays in the MULTIGENTTM Hb A1c Reagent Kit.
#### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The MULTIGENT™ Hemoglobin A1c system for measuring % Hb A1c is based on an immunoturbidimetrically microparticle agglutination inhibition method that is specific for stable Hb Alc and is used in conjunction with a colorimetric method to determine the Total Hemoglobin.
The Predicate device, Tosoh Medics Inc., G7 Automated HPLC Analyzer: Hb A1c Vanant Analysis Mode, uses an automated High Performance Liquid Chromatography (HPLC) system, a cation exchange column and gradient elution buffers which separates the stable Hb A1c from other hemoglobin components
Although the MULTIGENT™ Hemoglobin A1c system has technological differences from the predicate device, the intended use are similar and the results from performance characteristics data from non-clinical studies supports a claim of substantial equivalence to the predicate device.
The performance data is summarized below.
#### 8. SUMMARY OF NON-CLINICAL TESTING (COMPARATIVE ANALYSIS TO PREDICATE DEVICE LABEL CLAIMS:
#### Linearity (% HbA1c) and Limit of Detection
Linearity of the MULTIGENT™ HbA1c assay for the reportable % Hb A1c is from 2% to 20%. The label claim for the predicate device is 4.2% to 20.8%
{3}------------------------------------------------
#### Specificity and Interfering Substances:
The MULTIGENT™ Hb A1c assay has no interference as defined by acceptance criteria of difference of ≤ 1% Hb Alc when compared to an untreated sample. Below is a summary of the concentrations tested with the MULTIGENT™ HbA1c assay and the predicate device. Tosoh acceptance criteria is ± 10% of the untreated sample.
| Interfering<br>Substance | Concentrations of Interferent<br>Tested by MULTIGENT™<br>Hb A1c Assay with | Concentrations of Interferent Tested<br>By TOSOH G7 Automated HPLC |
|--------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------|
| Bilirubin | 50 mg/dL | 20 mg/dL |
| Triglyceride | 1600 mg/dL | 2000 mg/dL |
| Rheumatoid Factor | 3100 U/mL | None reported |
| Acetyl Salicylate | 50.8 mg/dL | None reported |
| Sodium Cyanate | 50 mg/dL | 20 mg/dL |
| Ascorbic Acid | 50 mg/dL | None reported |
| Urea (Carbamyl GHb) | 667 mg/dL | None reported |
| Gamma Globulin | 5 g/dL | None reported |
| HAMA Type 1 | Plasma 100% replaced | None reported |
| HAMA Type 2 | Plasma 100% replaced | None reported |
| Labile Hb Alc | 14 mg/mL of glucose | Separates LA1c |
| Ala | 2.85 % | Chromatographically Separates out Ala |
| Alb | 1.25% | Chromatographically Separates out Alb |
#### Precision
Precision studies for the MULTIGENT™ Hemoglobin A1c assay was performed on both the AEROSET® System and ARCHITECT® c8000™ System using the NCCLS EP5-A protocol (n=80). The studies demonstrated that a whole blood sample with a mean of approximately 5.0% HbA1c, a within run precision of 1.17 % CV, a between run precision of 0.40 %CV, and a total precision of 1.46 %CV was achieved. For a whole blood sample with approximately 10.5% HbAlc, a within run precision of 1.07 % CV, a between run precision of 0.73 %CV, and a total precision of 1.31 %CV was achieved.
The predicate device precision labeling claims are: within run precision of 0.90 %CV, between run precision of 0.40%CV, and a total precision of 1.12 %CV for a whole blood sample with a mean of 5.8% HbA1c. A whole blood sample with a mean of 10.9% HbA1c had a within run precision of 0.53 %CV, between run precision of 0.46%CV, and a total precision of 0.71 %CV
The precision of the MULTIGENT " Hemoglobin A1c assay, performed on either the AEROSET® System and ARCHITECT® c8000™ System are within acceptable limits. None of the raw data values differed from a MIN/MAX of >1% Hb A1c.
{4}------------------------------------------------
#### Method Comparison
Correlation studies were performed by assaying whole blood patient samples on the Abbott AEROSET® System and the Abbott c8000 system using the MULTIGENT™ Hb A1c assay and the Tosoh G7 Automated HPLC - Hb A1c Variant Analysis Mode. The data was subjected to linear regression statistics (least squares method) and yielded the following results:
| Regression Statistics | | |
|--------------------------|---------|-------|
| Independent Variable | Tosoh | Tosoh |
| Dependent Variable | Aeroset | c8000 |
| Multiple R (Corr. Coef.) | 0.993 | 0.994 |
| Observations (n) | 117 | 117 |
| y-Intercept | 0.240 | 0.141 |
| Slope | 0.976 | 0.998 |
#### 9. CONCLUSIONS:
The results of non-clinical testing demonstrate that the performance and safety and effectiveness of the MULTIGENT™ Hb A1c assay on the Abbott AEROSET® System and the Abbott ARCHITECT® c8000 system are substantially equivalent to that of the Tosoh G7 Automated HPLC - Hb A1c Variant Analysis Mode.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the circle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB - 9 2004
Mr. Les Padilla Technical Product Manager Seradyn, Inc. 7998 Georgetown Road - Suite 1000 Indianapolis, IN 46268-5620
Re: k033674
> Trade/Device Name: Multigent™ Hemoglobin A1c on the Abbott Aeroset® System and the Abbott Architect® c8000TM System Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: Class II Product Code: LCP; GGM; KRZ Dated: November 19, 2003 Received: November 24, 2003
Dear Mr. Padilla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Jean M. Cooper, MS, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Seradyn. The logo consists of the word "Seradyn" in a stylized font, with a globe-like graphic above it. The globe is made up of lines and circles, and there are three curved lines that appear to be orbiting it.
7998 Georgetown Rd
Suite 1000
Indianapolis, IN 46268
## INDICATIONS FOR USE STATEMENT
# 510(k) Number (if known): K033 (g 74
#### MULTIGENT™ HEMOGLOBIN A1c ON THE ABBOTT AEROSET® Device Name: SYSTEM AND THE ABBOTT ARCHITECT® c8000™ SYSTEM
#### Indications For Use:
The MULTIGENT™ Ho Alc assay is used in clinical laboratories for the quantitative in vitro measurement of percent Hb A1c (hemoglobin fraction) in human whole blood on the AEROSET® System and ARCHITECT® c8000™ System. The Hb A1c assay is intended to aid in the monitoring of long-term blood glucose control and compliance in individuals with diabetes mellitus. The MULTIGENT™ Hb A1c assay is not intended for use in diagnosing diabetes mellitus.
The MULTIGENT™ Ho A1c Calibrators are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for the calibration of the assays in the MULTIGENT™ Hb Alc Reagent Kit.
The MULTIGENT™ Ho Alc Controls are intended for in vitro diagnostic use with the AEROSET® System and the ARCHITECT® c8000™ System for quality control of the assays in the MULTIGENTTM Hb A1c Reagent Kit.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <div> <span style="text-decoration: underline;"></span> </div> |
|---------------------------------------|----------------------------------------------------------------|
|---------------------------------------|----------------------------------------------------------------|
OR
| Over-The-Counter Use (Optional Format 1-2-96) | <div> <span style="text-decoration: underline;"></span> </div> |
|-----------------------------------------------|----------------------------------------------------------------|
|-----------------------------------------------|----------------------------------------------------------------|
Carol C Benson
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety510(k) K03 3674
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.