The G7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode is a high pressure liquid chromatography system intended for IN VITRO DIAGNOSTIC USE ONLY. Glycosylated hemoglobin measurements obtained by this device are used in the management and treatment of diabetes.
Device Story
Automated HPLC system for whole blood analysis; separates hemoglobin components (A1a, A1b, F, L-A1c, SA1c, A0) using cation exchange column and gradient elution buffers. Input: 3 uL whole blood sample. Processing: internal microprocessors control sampling, pumping, degas, and detection; software compares retention times to known windows to identify fractions and variants (P0X or H-VX). Output: printed report and digital data (sample ID, retention times, SA1c %, total A1 %, chromatogram). Used in clinical laboratory settings by trained personnel. Results assist clinicians in diabetes management and treatment monitoring.
Clinical Evidence
Bench testing only. Comparative analysis of 230 patient samples between G7 and A1c 2.2 Plus yielded a slope of 1.001, intercept of -0.12, and correlation coefficient of 0.9987 across a range of 3.4% - 15.9% HbA1c. Precision studies showed intra-run CVs ≤0.9% and inter-run CVs ≤1.6%. Total precision was 1.12% at 5.79% HbA1c and 0.71% at 10.90% HbA1c. Recovery studies demonstrated 100% to 102% recovery.
Technological Characteristics
HPLC system using non-porous cation exchange column. Detection via visible wavelength detector (415 nm sample, 500 nm reference). Automated sample loading (90/290 capacity) for primary tubes or cups. Microprocessor-controlled with thermal printer and RS-232C bi-directional connectivity. Software-driven elution gradient control. Dimensions/form factor not specified. Sterilization not applicable (IVD).
Indications for Use
Indicated for the quantitative measurement of HbA1c and Total HbA in whole blood samples to assist in the management and treatment of diabetes.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
Predicate Devices
A1c 2.2 Plus Automated Glycohemoglobin Analyzer (K972265)
Submission Summary (Full Text)
{0}------------------------------------------------
KD11434
SEP 1 8 2001
### 510(k) Summary
### Tosoh G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode
Submitter:
Tosoh Medics, Inc. 347 Oyster Point Blvd., Suite 201 South San Francisco, CA 94080 Phone: (800) 248-6764 Fax:
Contact Person:
Lois Nakayama Manager, Quality Assurance
Date of Summary Preparation:
Device Name:
Classification Name:
Predicate Device:
August 23, 2001
G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode
Class II, LCP 21 CFR 864.7470 Glycosylated Hemoglobin Assay
A1c 2.2 Plus Automated Glycohemoglobin Analyzer Tosoh Medics, Inc. South San Francisco, CA K972265
#### Device Description:
The G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports stable A1c (SA1c) percentage in whole blood. The operational portion of the HPLC G7 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, floppy disk drive unit, operation panel and a printer.
The G7 uses a cation exchange column and separates the usual hemoglobin components in the blood into six fractions, A1a, A1b, F, L-A1c, SA1c and A0. The separation is done by eluting the hemoglobins from the column with a gradient of three elution buffers containing different salt concentrations. The result report is printed out from the on-board printer and can be stored on a floppy disk from the on-board floppy disk drive. The data can be transmitted to a host computer. The result report includes a sample ID, date, percentage and retention time of each fraction, SA1c percentage and total A1 percentage (A1a + A1b + SA1c) along with a chromatogram of the elution
{1}------------------------------------------------
pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the material depending upon its charge. The software compares the retention times to "known windows" and designates the material as P0X or H-VX if it does not match a defined window.
All automated processes in the G7 are controlled by internal microprocessors, using software downloaded via the on-board floppy disk drive.
#### Statement of Intended Use:
The G7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode is a high pressure liquid chromatography system intended for IN VITRO DIAGNOSTIC USE ONLY. Glycosylated hemoglobin measurements obtained by this device are used in the management and treatment of diabetes.
#### Substantial Equivalence:
#### Comparison Data:
The G7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode is substantially equivalent in intended use and technological features to instrument systems in commercial distribution, that are used to measure the glycohemoglobin level in the blood by HPLC methodology.
Specifically, the G7 Automated HPLC Analyzer – HbA1c Variant Analysis Mode is substantially equivalent to the Tosoh A1c 2.2 Plus Automated Glycohemoglobin Analyzer (see K972265) which has been reviewed and cleared to market by FDA. A comparison of the capabilities and specifications of these two systems is provided in Table 1. Pertinent similarities and differences between the analyzers are presented in detail below.
{2}------------------------------------------------
### Table 1
# G7 Automated HPLC vs. A1c 2.2 Plus Comparison
| | G7 - Hb A1c Variant Analysis Mode | A1c 2.2 Plus |
|-------------------------|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Intended Use | Quantitative measurement of<br>Hb A1c and Total Hb A | Quantitative measurement of<br>Hb A1c and Total Hb A |
| Methodology | HPLC | HPLC |
| Column | Non-porous cation exchange | Non-porous cation exchange |
| Reported result | Peak resolution:<br>A1a, A1b, F, LA1c, SA1c, A0<br>and variant (as P0X or H-VX))<br>Percentage of SA1c and Total A1 | Peak resolution:<br>A1a, A1b, F, LA1c, SA1c. A0<br>and variant (as HbS+)<br>Percentage of SA1c and Total A1 |
| Detection method | Visible wavelength detector<br>415 nm (sample)<br>500 nm (reference) | Visible wavelength detector<br>415 nm (sample)<br>500 nm (reference) |
| Microprocessor | Yes | Yes |
| User Input | Pressure sensitive LCD | Pressure sensitive LCD |
| Visual display | LCD display | LCD display |
| Throughput | 2.2 minutes per sample | 3.0 minutes per sample |
| Automated | Yes | Yes |
| Calibration | 2 point | 2 point |
| Sample type | Whole blood<br>(undiluted or diluted) | Whole blood<br>(undiluted or diluted) |
| Sample volume | 3 uL | 5 uL |
| Auto Dilution | Yes | Yes |
| Sample loading capacity | 90 or 290 (automatic) | 50 or 90 or 290 (automatic) |
| Sample holder | Primary tube or Sample cup | Primary tube or Sample cup |
| Printer | Thermal | Thermal |
| Bar-Code Capability | Yes | Yes |
| RS-232C | Yes, Bi-directional | Yes, Uni-directional |
{3}------------------------------------------------
#### Comparative Analysis
Comparative analysis studies were performed with patient samples obtained from 230 patients by assaying the samples with the G7 Automated HPLC Analyzer in the HbA1c Variant Mode and the A1c 2.2 Plus Automated Glycohemoglobin Analyzer. Least squares linear regression analysis yields the following statistics:
Comparative Analysis: G7 Automated HPLC vs. A1c 2.2 Plus
| Slope: | 1.001 |
|--------------------------|------------|
| Intercept: | -0.12 |
| Correlation Coefficient: | 0.9987 |
| Range of Samples (%): | 3.4 - 15.9 |
| Number of Samples: | 230 |
Precision studies demonstrated intra-run precision %CVs of 0.9% or less and inter-run precision %CVs of 1.6% or less. Total precision at a HbA1c mean of 5.79% was 1.12% while at a mean value of HbA1c of 10.90%, the total precision was 0.71%. Recovery studies performed with samples mixed in specified ratios generated recoveries of 100% to 102%.
#### Conclusion:
Considering the excellent correlation between the Tosoh G7 Automated HPLC - HbA1c Variant Analysis Mode and the A1c 2.2 Plus, it can be concluded that the G7 Automated HPLC Analyzer -HbA1c Variant Analysis Mode is substantially equivalent to the A1c 2.2 Plus Automated Glycohemoglobin Analyzer, which has been 510(k) cleared. Based on the establishment of substantial equivalence, the safety and effectiveness of the Tosoh G7 HPLC is confirmed.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird in flight, composed of three curved lines that suggest feathers or wings. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the circumference of the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## SEP 1 8 2001
Ms. Lois Nakayama TOSOH Medics, Inc. 347 Oyster Point Boulevard Suite 201 South San Francisco, California 94080
Re: K011434
Trade Name: G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode Regulation Number: 21 CFR § 864.7470 Regulatory Class: II Product Code: LCP Dated: July 13, 2001 Received: July 19, 2001
Dear Ms. Nakayama:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrlv/dsma/dsmamain.html".
Sincerely yours.
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Tosoh Medics, Inc. The logo consists of a stylized four-petal flower shape with a square in the center, followed by the text "TOSOH MEDICS, INC." to the right. Below the flower shape, the text "TOSOH" is printed in a smaller font size.
### PREMARKET NOTIFICATION
### INDICATION FOR USE STATEMENT
# G7 Automated HPLC Analyzer: HbA1c Variant Analysis Mode
The G7 Automated HPLC Analyzer is a high pressure liquid chromatography system intended for IN VITRO DIAGNOSTIC USE ONLY. Glycosylated system intended for wents obtained by this device are used in the management and treatment of diabetes.
Josephine Bautista
(Division Sign-Off) (Division of Clinical Laboratory Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.