Q.STEPS BIOMETER G BLOOD GLUCOSE MONITORING SYSTEM

K033627 · Biomedix, Inc. · CGA · May 21, 2004 · Clinical Chemistry

Device Facts

Record IDK033627
Device NameQ.STEPS BIOMETER G BLOOD GLUCOSE MONITORING SYSTEM
ApplicantBiomedix, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateMay 21, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The Q.STEPS™ Biometer G Blood Glucose Monitoring System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertip for all ages (excluding neonates.) It is intended for use outside the body (for in vitro diagnostic use) by health care professionals in settings such as clinical laboratories and physician offices laboratories (POLs) as an aid to monitor the effectiveness of diabetes control.

Device Story

System consists of meter, glucose test strips, control solution, and lancet device. Operates as electrochemical biosensor; uses glucose oxidase enzyme to catalyze oxidation of D-glucose in capillary blood samples. Electron transfer at electrode surfaces generates current proportional to glucose concentration. Used by healthcare professionals in clinical settings to monitor diabetes control. Provides quantitative glucose results to assist clinical decision-making regarding patient glycemic management.

Clinical Evidence

Clinical and internal studies compared Q.STEPS system results against YSI 2300 STST reference analyzer. Accuracy: >98% of clinical data fell into zones A & B of Clarke Error Grid; bias within 20mg/dL for glucose <100mg/dL. Precision: CV <8% across 20-day testing. Linearity: 50-400 mg/dL. Hematocrit interference: acceptable within 30-60% range. Interference testing: no significant interference observed for 23 common substances at physiological/therapeutic levels, except Uric Acid and high-dose compounds.

Technological Characteristics

Enzymatic electrochemical biosensor; glucose oxidase catalyst. Amperometric sensing principle. Measures current proportional to glucose concentration. Sample volume 5 µL; test time 15 seconds. Measurement range 50–400 mg/dL. Hematocrit range 30–60%.

Indications for Use

Indicated for quantitative measurement of glucose in fresh capillary whole blood from the fingertip for all ages, excluding neonates. Intended for use by healthcare professionals in clinical laboratories and physician office laboratories to monitor diabetes control.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a symbol that resembles three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 2 5 2005 Paul Shieh, Ph.D. President Biomedix, Inc. 40471 Encyclopedia Circle Fremont, CA 94538 Re: k033627 Trade/Device Name: Q.STEPS (TM) Biometer G Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, JJX Dated: February 24, 2004 Received: February 25, 2004 Dear Dr. Shieh: This letter corrects our substantially equivalent letter of May 21, 2004 regarding the incorrect product code of CBA verses the correct product code of CGA - Glucose Oxidase, Glucose. We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Jean M. Cooper MS DUM Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # MAY 2 1 2004 Ko336.27 # 510 K SUMMARY | Introduction: | According to the requirements of 21 CFR 807.92, the following information provides<br>sufficient detail to understand the basis for a determination of substantial<br>equivalence. | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter name,<br>address: | Biomedix, Inc., U.S.A.<br>40471 Encyclopedia Circle<br>Fremont, CA 94538 | | Contact Person: | Judy Shieh Chen, Ph.D.<br>Phone:(510) 438-9500, ext.414<br>Fax: (510) 438-9141<br>E-mail: jchen@biomedixusa.com | | Date Prepared: | March 29th 2004 | | Device Name: | Proprietary name:<br>Q.STEPS™ Biometer G Blood Glucose Monitoring System | | | Common Name:<br>Whole Blood Glucose Test System | | | Classification<br>1. Q.STEPS™ Biometer G and Q.STEPS™ Glucose Test Strip - Class II devices (21<br>C.F.R. §862.1345, Glucose Test System)<br>2. Q.STEPS™ Glucose Control Solution - Class I device (21 C.F.R. §862.1660,<br>Quality Control Material)<br>3. Sterile lancet, Lancing Device and Accessories - Class I (exempt) devices (21<br>C.F.R. §878.4800, Lancet, Blood). | | | Classification Name:<br>Glucose Oxidase, Glucose | | Predicate Device: | We claim substantial equivalent to the Life Scan OneTouch Basic Blood Glucose<br>Monitoring System (Test Strips K#031472) | | Intended Use: | The Q.STEPS™ Biometer G Blood Glucose Monitoring System is intended to be<br>used for quantitative measurement of glucose in fresh capillary whole blood from the<br>fingertip for all ages (excluding neonates.) It is intended for use outside the body (for<br>in vitro diagnostic use) by health care professionals in settings such as clinical<br>laboratories and physician offices laboratories (POLs) as an aid to monitor the<br>effectiveness of diabetes control. | | Test Principle: | The Q.STEPS™ Biometer G Blood Glucose Monitoring System is an in vitro device<br>designed for measuring the concentration of glucose in capillary whole blood from<br>the fingertip. The system uses electrochemical methodology. The system quantifies<br>glucose amperometrically by measuring the current that is produced when glucose<br>oxidase catalyzes the oxidation of glucose to gluconic acid. The electrons generated<br>during this reaction are transferred from the blood to the electrodes. The magnitude<br>of the resultant current is proportional to the concentration of glucose in the<br>specimen and is converted to a readout displayed on the monitor. | {3}------------------------------------------------ #### Summary of Performance Studies #### Safety No adverse health effects were observed in the pre-clinical & clinical study. No postclinical trial reports of any health effect were seen. #### Linearity The linearity of the Q.STEPS™ Biometer G System is good through a wide range of The lineanty of the or s . 50 to 400 mg/dl. This is confirmed with internal laboratory and external clinical studies. #### Accuracy internal studies and clinical trial method comparison studies showed that the whole micrhal stadios and others from fingertips determined by Q.STEPS™ System correlated very well with the reference laboratory method, i.e., Yellow Spring Instrument (YSI) 2300 STST Glucose/Lactate Analyzer. All data collected fall well mistiument (101) 2000 oriteria of a bias within 20mg/dl (Blood Glucose <100mg/dl) or whillin the assopeanse on the reference method. Furthermore, over 98% of all clinical data fall into acceptable zones A & B in the Clarke Error Grid analysis. #### Precision Precision evaluation performed in the internal laboratory and clinical environment shows that the Q.STEPS™ System can determine very precise glucose concentrations. For commercial controls, readings were taken twice in the morning (Run1) and twice in the evening (Run2). Each reading obtained from each run was taken from two different Q.STEPS™ Biometers. This was performed every day for 20 consecutive days. For venous whole blood specimens spiked with five different 20 concentrations of glucose, 20 readings were taken from each concentration on the same day. Overall, the standard deviation values are very low and CV values are well within the acceptance criteria of less than 8%. #### Stability In both room temperature and accelerated stability studies, the Q.STEPS™ Test Strips provide a constant glucose reading over time. From the accelerated stability evaluation, the Q.STEPS™ Test Strips have been determined to have an approximate shelf life of 2 years. ### Hematocrit Interference The degree of the hematocrit effect depends on the glucose concentration. The Q.STEPS™ System was tested within the hematocrit range from 30% to 60%. With an acceptance criteria of ±20% of the glucose concentration determined with a set of hematocrit standard (45+3%), the results of this study show that hematocrit levels of 30% to 60% is acceptable for use with the Q.STEPS™ Biometer G System. #### Interference Substances - Twenty-three commonly tested interference substances were examined in . this study. No interference was observed in the Bilirubin, Cholesterol, Creatinine and Citrate at physiological levels except for Uric Acid. - No interference was observed in the Ibuprofen, Aspirin and Tetracycline. . - No interference was observed at therapeutical ranges of Acetaminophen, . Ascorbic Acid (Vit.C.), L-Dopa, Dopamine, Methyldopa and Tolazamide. Some interference was observed at higher dosages of these compounds. - No interference was observed in EDTA (in blood drawing tubes), D-. Galactose, K3Fe (CN)g, D-Mannose, D-Xylose and Maltose. {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K033627 Device Name: Q.STEPS Indications For Use: The Q.STEPS™ Biometer G Blood Glucose Monitoring System is intended to be used for quantitative s and the security of a free been ill muchale hood from the fingerin for all a The Q.STEPS™ Biometer G Blood Glocker in the finger in for all ages (excluding neonates.) It is measurement of glucose in tresh opliary who included in an artisesionals in sgrings such as intended for use outside the body (for in vitro diagnostic use professionals in s intended for use outside the body (for in vitro diagnosic use) by health on a subcontrol. Ruth A. Clarke (for Carol Benson) I)ivision Sign-C office of In Vitro Diagnostic Devi Svaluation and Safety \$\sin(k) \frac{k_{0}33627}{/}\$ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...