K033137 · Wada, Inc. · JPA · Dec 8, 2003 · Hematology
Device Facts
Record ID
K033137
Device Name
POTENS +
Applicant
Wada, Inc.
Product Code
JPA · Hematology
Decision Date
Dec 8, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
The POTENS+ is a multipurpose system for use in performing clot based, photo-optical in-vitro coagulation studies including Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
Device Story
POTENS+ is a semi-automated, photo-optical system for in vitro coagulation studies. It accepts citrated human plasma samples; utilizes a potentiophotometer, reagent dispenser, pipettor, and heating block to perform clot-based assays. The device detects changes in light transmittance at 660nm via an LED and silicon photovoltaic cell during clot formation. It processes signals using linear digital electronics to calculate PT, FBG, and INR. Operated by medical professionals in laboratory settings, the system provides quantitative results to clinicians. These outputs assist in monitoring anticoagulation therapy, assessing clotting pathway status, and evaluating cardiovascular risk. The device benefits patients by providing rapid, accurate coagulation profiles for clinical decision-making.
Clinical Evidence
Clinical comparison study using patient specimens across normal, diagnostic, and therapeutic ranges. Correlation coefficients (r-values): Fibrinogen 0.973-0.980; PT 0.95-0.974; INR 0.919-0.963. Precision studies followed NCCLS EP5A; total CVs <6% for fibrinogen, <5% for PT, and <8% for INR using human/bovine control plasmas. Results demonstrate performance comparable to predicate devices.
Technological Characteristics
Semi-automated potentiophotometer; optical detection via 660nm LED and silicon photovoltaic cell. Operates at 37 ± 0.3°C. Components include reagent dispenser, pipettor, heating block, computer, and printer. Performs clot-based assays using linear digital signal processing. Requires citrated plasma; no sample dilution needed. Complies with NCCLS standards for precision and interference testing.
Indications for Use
Indicated for medical professionals in laboratory settings to determine the coagulation profile of citrated human plasma specimens, including Prothrombin Time (PT), derived Fibrinogen (FBG), and International Normalization Ratio (INR).
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE AND INSTRUMENT TEMPLATE
A. 510(k) Number:
K033137
B. Analyte:
Prothrombin Time, Derived Fibrinogen, INR
C. Type of Test:
Semi-quantitative, Fibrin clot based and photo-optical
D. Applicant:
WADA, Inc.
E. Proprietary and Established Names:
POTENS+
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5425
Multipurpose System for In Vitro Coagulation Studies
2. Classification:
Class II
3. Product Code:
JPA
4. Panel:
Hematology (81)
G. Intended Use:
1. Intended use:
The POTENS+ is a multipurpose system for use in performing clot based, photo-optical in-vitro coagulation studies including Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
2. Indication for use:
The POTENS+ an in vitro medical device to determine the coagulation profile of citrated human plasma specimens to be used by a medical professional in a laboratory setting. This profile includes Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
3. Special condition for use statement(s):
Not applicable
4. Special instrument Requirements:
Not applicable
{2}
Page 2 of 8
# H. Device Description
POTENS+ is a semi-automated system for *in vitro* coagulation studies, utilizing a clot based, photo-optical method. It is an arithmetically linear, digital coagulation device that determines the Prothrombin Time (PT) and calculates the International Normalization Ratio (INR) of a patient’s plasma sample. It determines the derived Fibrinogen (FBG) in which the Optical Density (O.D.) change of the optically clear FBG in plasma converts to turbid fibrin clot to determine the FBG content of the specimen. Also, included in this system are two Fibrinogen standards and three Fibrinogen controls.
# I. Substantial Equivalence Information:
1. Predicate device name(s):
Organon Tecknika MDA-180 for PT and INR
Organon Tecknika Coag-A-Mate (CAM) MTX for Fibrinogen
2. Predicate K number(s):
K962857, K924453
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | POTENS+ | MDA-180 | CAM -MTX |
| Intended Use | Determine coagulation profile of clinical specimens | Determine coagulation profile of clinical specimens | Determine coagulation profile of clinical specimens |
| Analytes | PT, FBG, INR | PT, FBG, INR | FBG |
| Methodology | Fibrin Clot based assay; optical measurement of direct light transmission during clot formation. | Clauss method: Fibrin Clot based assay; optical measurement of relative light transmission to the time of clot formation | Clauss method: Fibrin Clot based assay; optical measurement of relative light transmission to the time of clot formation |
| Principles of operation | Detects change in light transmittance at 660nm | Detects change in light transmittance in the spectral range 405 to 710nm. FBG @425 & PT @ 580nm | Detects change in light transmittance at 405nm |
| Method of Detection | Optical (LED with silicon photovoltaic cell | Optical (tungsten lamp) using photodiode detectors | Optical using a photometer and detectors |
| Target User | Clinical Laboratory | Clinical Laboratory | Clinical Laboratory |
| Operating Temp | 37±0.3°C | 37±1°C | 37 |
| Test Sample | Citrate plasma | Citrate plasma | Citrate plasma |
| Expression of Results | Fibrinogen: mg/dl
PT: seconds
INR: INR units | Fibrinogen: mg/dl
PT: seconds
INR: INR units | Fibrinogen: mg/dl |
| Differences | | | |
| Item | POTENS+ | MDA-180 | CAM -MTX |
{3}
Page 3 of 8
| Instrument System | Potentiophotometer, Reagent Dispenser, Pipettor, Heating block, computer and printer | Multi-Channel Discrete Analyzer | Analyzer (measuring rotor and photometer), computer and printer |
| --- | --- | --- | --- |
| Signal Processing | Linear handling of light transmission using digital electronics | Logarithmic handling of light transmission using analog electronics | Logarithmic handling of light transmission using analog electronics |
| Reagents/ Accessories | Calibration standards (2)
Controls (3)
Reagent Water
Test Cuvetts
Cuvette rack
Normal Saline
Thromboplastin (Dade® Thromboplastin C Plus) | VeriCal Calibrator Set
Verify Controls
Reagent Water
Buffer
Test Cuvetts
Clauss Reagents
Thrombin, Imidiazole buffer) | Reference Plasma (for calibration)
Verify Controls: Veronal Buffer
Sample cups
Cuvette Rotor
Fibriquik (Thrombin) |
| Test Sample Dilution | None | Required | Required |
| Reportable Range | FBC: linear range of FBG
This is from high to low standards (HS & LS),
beyond this to be established by user (99-851 mg/dl obtained in comparative study)
PT: 7.1 to 90 seconds
INR: ≥0.5 with limited clinical utility >4.5. Upper limit dictated by ISI of thromboplastin used. | FBG II: 35-700 mg/dl depending on dilution set up
PT: 8-150
INR 0.8-8.0 with limited clinical utility >4.5 | FBG: 40-650 mg/dl depending on dilution set up |
| Quality Control | Use control plasmas to monitor performance; run once every 55 samples and every 8 hours, whichever occurs first as specified in operators manual. | Use control plasmas and other parameters at frequency and limits specified in operators manual | Use control plasmas and other parameters at frequency and limits specified in operators manual |
J. Standard/Guidance Document Referenced (if applicable):
WHO Expert Committee on Biological Standardization 33rd Report; WHO
Technical Report Series 687, Geneva, 1983.
EP5-A “Evaluation of Precision Performance of Clinical Chemistry Devices,” NCCLS
EP6-P “Evaluation of the Linearity of Quantitative Analytical Methods,” NCCLS
EP7-P “Interference Testing in Clinical Chemistry,” NCCLS
{4}
Page 4 of 8
K. Test Principle:
POTENS+ is an arithmetically linear, digital coagulation instrument that simultaneously determines the Prothrombin time (PT), the International Normalization Ratio (INR) and Fibrinogen (FBG) level in plasma samples.
POTENS+ measures PT and Fibrinogen following the principles of PT-derived methods. When Thromboplastin-Calcium reagent is added to a citrated plasma sample, enzymatic reactions occur leading to the conversion of Prothrombin to Thrombin. Thrombin then converts Fibrinogen to a fibrin clot. POTENS+ detects the beginning of the fibrin clot formation that indicates the PT of the sample. The change in optical density (OD) during fibrin clot formation is detected and compared to the results from similarly treated Fibrinogen standards. POTENS+ detects these OD changes using patented potentiophotometric technology.
POTENS+ directly displays the concentration of the substance being measured because the output of the POTENS+ is inherently linear, in the true digital sense and does not require a logarithmic-linear converting device, circuit, or computers which are required
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-run precision studies were performed according to NCCLS Guideline EP5-A using three levels of control plasma (Control 1, Control 2, and Control 3) and two instruments.
| Analyte | Sample | Inst.# 1/2 | N | Mean | CV |
| --- | --- | --- | --- | --- | --- |
| Fibrinogen | CTL 1 | combined | 40 | 322.2 | 2.8 % |
| | CTL 2 | combined | 40 | 189.7 | 3.2 % |
| | CTL 3 | combined | 40 | 101.4 | 5.7 % |
| PT | CTL 1 | combined | 40 | 20.4 | 10.7 % |
| | CTL 2 | combined | 40 | 11.5 | 2.8 % |
| | CTL 3 | combined | 40 | 16.7 | 6.2 % |
| INR | CTL 1 | combined | 40 | 3.1 | 18.4 % |
| | CTL 2 | combined | 40 | 1.2 | 6.6 % |
| | CTL 3 | combined | 40 | 2.2 | 12.3 % |
{5}
Page 5 of 8
Day to day precision studies were performed according to NCCLS Guideline EP5-A using three levels of control plasma (Control 1, Control 2, and Control 3) and two instruments on at least 20 days.
| Analyte | Sample | Inst.# 1/2 | Number of days | Number of results | Mean | CV |
| --- | --- | --- | --- | --- | --- | --- |
| Fibrinogen | CTL 1 | combined | 41 | 82 | 320.9 | 4.5% |
| | CTL 2 | combined | 41 | 82 | 191.5 | 4.4% |
| | CTL 3 | combined | 41 | 82 | 106.9 | 5.6% |
| PT | CTL 1 | combined | 41 | 82 | 20.5 | 6.4% |
| | CTL 2 | combined | 41 | 82 | 11.63 | 2.3% |
| | CTL 3 | combined | 41 | 82 | 17.2 | 4.6% |
| INR | CTL 1 | combined | 41 | 82 | 3.07 | 11.0% |
| | CTL 2 | combined | 41 | 82 | 1.18 | 5.5% |
| | CTL 3 | combined | 41 | 82 | 2.28 | 7.6% |
b. Linearity/assay reportable range:
The linearity was established according to NCCLS Guideline EP6-P. POTENS+ is designed to be linear from the Low Standard through the High Standard. A plasma pool was spiked with Fibrinogen cryoprecipitate and a portion diluted 1:9 with normal saline.
Mixtures of these two solutions were made to obtain a total of five levels. Four replicates of the five levels were evaluated. The results were compared against the theoretical results. These results indicate that the POTENS+ was linear from 99 to 851 mg/dl. Each laboratory should verify the linearity of the instrument below and above the Standard for each Lot of Standards and Thromboplastin used. Only patient results in the linear range should be reported.
Reportable range:
Method comparison studies were used to verify that the instrument reports results within the range of:
Fibrinogen 136-851 mg/dl
PT 8.8-41.2 seconds
INR 0.7-9.9
However, each laboratory should verify the reportable range of each analyte.
c. Traceability (controls, calibrators, or method):
Indirect traceability was demonstrated by including plasma fibrinogen (98/614) standard in the Ware-Ratnoff method currently used to determine the Fibrinogen concentration of the POTEN+ Standards. Communication from the NIBSC indicated that the 1st International Standard (IS) (98/614) could be used in the Ratnoff method.
{6}
Page 6 of 8
d. Detection limit:
The lower detection limit was measured in comparative studies:
Lowest Level measured:
Fibrinogen 79 mg/dl
PT 8.8 seconds (software limit 7.1)
INR 0.7 (INR calculations are limited by the lower PT limit of 7.1 and by the ISI of the thromboplastin used)
e. Analytical specificity:
Interference studies were performed by spiking human plasma or plasma pools with bilirubin, hemoglobin, heparin triglycerides, and fibrinogen degradation products. Although bilirubin, hemoglobin and triglyceride showed no affect above the normal range typically found in plasma, visibly icteric, hemolyzed or lipemic specimens should not be used. Heparin at the upper therapeutic range (0.5 Units/ml) affected PT and INR but did not affect Fibrinogen result at the highest level tested (1.7 Units/ml). Fibrinogen Degradation Product digest did not affect the PT or INR results, however, precautions should be taken when analyzing specimens with this elevation until additional studies have been completed determined the level of Fibrinogen Degradation Products that does not cause interference.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Comparison studies were performed at four sites to demonstrate that POTENS+ is equivalent to the MTX for determining Fibrinogen concentration and to the MDA-180 for determining PT and INR results using a minimum of 200 specimens. Equivalence was demonstrated using commercially available methods/product along with patient specimens covering the normal, therapeutic and diagnostic ranges. Correlation coefficients (r-value) ranged from 0.973 to 0.980 for Fibrinogen, 0.95 to 0.974 for PT and 0.919 to 0.963 for INR results.
b. Matrix comparison:
Not reported
3. Clinical studies:
a. Clinical sensitivity:
Not provided.
b. Clinical specificity:
Not provided.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
Not applicable
{7}
Page 7 of 8
5. Expected values/Reference range:
From the literature, the generally accepted fibrinogen reference interval is 200mg/dl to 400mg/dl. This is only a guide to the user. Each laboratory should establish specific fibrinogen reference intervals.
## M. Instrument Name: POTENS+
## N. System Descriptions:
1. Modes of Operation:
Semi-automatic
2. Software:
The POTENS+ software is PC based with the primary responsibility of providing an interface to the user, performing the testing, and handling the results data.
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Sample Identification:
Sample identification is by manual entry. The software accepts input from both a scanner and a keyboard.
4. Specimen Sampling and Handling:
The POTENS+ uses citrate plasma samples and the software controls interface with the pipettor so that the user can transfer the sample to a cuvette.
5. Assay Types:
Clot based assays
6. Reaction Types:
Photo-optical
7. Calibration:
The system includes two Fibrinogen standards. A four standards test is required (2 “high” and 2 “low”). The software shall recognize via the Specimen ID that the standard is being run. If it is the first standard, POTENS+ shall reset all parameters to indicate that a new batch is being started. When the instrument has successfully completed the fourth standard (second low standard) test it will determine the mean values of the PT and the Fibrinogen (FBG) calibration curve. All future FBG calculations for the batch uses this equation.
8. Quality Control:
This system includes three Fibrinogen controls. Whenever the POTENS+ encounters a Specimen ID of CTL1 (Control 1), CTL2 (Control 2) OR CTL3 (Control 2), it regards the test as a control test. The software will use the appropriate maximum/minimum control values from the parameter table to assure that the instrument is properly calibrated. If the checks fail, the software will report the value as “OUT of control.” Otherwise, the software will report the value as “In control.”
{8}
Page 8 of 8
POTENS+ Control 1 is not suitable for use as a control for Prothrombin Time (PT) or INR assays
O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "L. Performance Characteristics" Section Of the SE Determination Decision Summary.
P. Conclusion:
Based upon review of the information and labeling provided, this device is found to be substantially equivalent (SE) to 21 CFR 864.5425 JPA, System, Multipurpose for In Vitro Coagulation Studies, Class II.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.