VITREA2, VERSION 3.4 MEDICAL IMAGE PROCESSING SOFTWARE
K032748 · Vital Images, Inc. · LLZ · Oct 29, 2003 · Radiology
Device Facts
Record ID
K032748
Device Name
VITREA2, VERSION 3.4 MEDICAL IMAGE PROCESSING SOFTWARE
Applicant
Vital Images, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 29, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.2050
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Coronary Artery Calcification Scoring
—
—
—
—
—
—
—
Vascular Stenosis Analysis
—
—
—
—
—
—
—
Tumor Volume Measurement
—
—
—
—
—
—
—
Cerebral Perfusion Mapping
—
—
—
—
—
—
—
Chest CT Computer Assisted Detection
—
—
—
—
—
—
—
Indications for Use
Vitrea™2 is a medical diagnostic workstation that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. In addition, Vitrea™2 has the following specific indications. VScore™ (K990442) is an option within the Vitrea™2 application and is intended for cardiac scoring from whole body CT derived measurements, including non-invasive detection and quantification of atherosclerotic plaque. Two image processing options, EKG Gate (K001682) and Auto Gate (K003230), allow the operator to select images with reduced motion artifacts when processing data for Coronary Artery Calcification Scoring. Automated Vascular Measurement (K002519) is an option within the Vitrea™2 application and is intended for study/analysis of selected vessels for stenosis analysis, pre/post stent planning and directional vessel tortuosity evaluation. Tumor Volume Measurement (K002519) is an option within the Vitrea™2 application and is intended for the analysis/quantification of tumor volumes obtained from MR brain series scans. CT Brain Perfusion (K003639) is an option within the Vitrea™2 application and is intended for post processing based on dynamic CT images continuously acquired during the injection of contrast for the visualization of apparent blood flow in brain tissue and pictorial illustration of perfusion related parameters to aid in the type and extent of cerebral perfusion disturbances. ImageCheckerCT is an option within the Vitrea™2 application and is intended for the display of a composite view of 2D crosssections, and 3D volumes of Chest CT images, including findings of regions of interest ("ROI") identified by the radiologist, or Computer Assisted Detection ("CAD") findings.
Device Story
Vitrea™2 is a medical diagnostic workstation for processing, reviewing, and analyzing multi-dimensional digital images from various imaging modalities. Input data is retrieved via DICOM network protocols. The device provides dedicated protocols for exam-specific image adaptation, gallery-based selection of 2D/3D views, real-time interactive manipulation, annotation, and measurement. It supports specialized clinical modules: VScore (cardiac/plaque scoring), EKG/Auto Gate (motion artifact reduction), Automated Vascular Measurement (stenosis/stent planning), Tumor Volume Measurement (MR brain), CT Brain Perfusion (blood flow visualization), and ImageCheckerCT (CAD findings). Operated by clinicians in clinical settings, the workstation outputs processed views to printers, web servers, or other DICOM devices. Output aids clinicians in visualizing anatomy and pathology, facilitating diagnostic decision-making and treatment planning.
Clinical Evidence
Bench testing only. Software was designed, developed, and validated according to written Design Control procedures. Integration testing, verification, and Beta validation were performed. Potential hazards were addressed via a Risk Management Plan.
Technological Characteristics
Medical diagnostic workstation software. Supports DICOM network connectivity. Features include multi-dimensional visualization, automated exam-specific protocols, and specialized modules for cardiac, vascular, tumor, and perfusion analysis. Operates on standard workstation hardware.
Indications for Use
Indicated for clinicians requiring processing, review, and analysis of multi-dimensional digital images (CT, MR) for cardiac scoring, atherosclerotic plaque detection, vascular stenosis/tortuosity analysis, tumor volume quantification, cerebral perfusion assessment, and chest CT CAD findings.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Vital Images, Inc., Vitrea2, Version 2.1 (K002519)
{0}------------------------------------------------
## Attachment 1 - 510(k) Summary
## Submitter's Name / Contact Person
Stephen S. Andersen Sr. Director of Quality and Customer Satisfaction Vital Images, Inc. 3300 Fernbrook Lane N, Suite 200 Plymouth, MN 55447
#### General Information
| Trade Name | Vitrea2, Version 3.4 Medical Image Processing Software | |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Common / Usual Name | System, Image Processing, Radiological | |
| Classification Name | LLZ, Class II, CFR 21 892.2050 | |
| Predicate Devices | Vital Images, Inc., Vitrea2, Version 2.1 (K002519)<br>GE, Advantage Workstation 4.1 (K020483)<br>R2 Technologies, Inc., ImageChecker-CT Workstation (K023003)<br>GE, Advanced Lung Analysis (K013381) | |
## Device Description
Vitrea™2 is a medical diagnostic workstation that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea™2, version 3.4 is an upgrade to Vitrea™2, version 2.1 initially released for commercial distribution by FDA on K002519.
Vitrea 2 provides multi-dimensional visualization of digital images to aid clinicians in their analysis of anatomy and pathology. The Vitrea2 user interface follows typical clinical workflow patterns to process, review, and analyze digital images, including:
- Retrieve image data over the network via DICOM .
- Display images that are automatically adapted to exam type via dedicated protocols .
- Select images for closer examination from a gallery of up to six 2D or 3D views .
- . Interactively manipulate an image in real-time to visualize anatomy and pathology
- Annotate, tag, measure, and record selected views
- Output selected views to standard film or paper printers, or post a report to an Intranet Web server or . export views to another DICOM device
- Retrieve reports that are archived on a Web server .
## Intended Use
Vitrea™2, Version 3.4 is a medical diagnostic workstation that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. In addition, VitreaTM2 has the following specific indications.
{1}------------------------------------------------
VScore™ (K990442) is an option within the Vitrea™2 application and is intended for cardiac scoring from whole body CT derived measurements, including non-invasive detection of atherosclerotic plaque. Two image processing options. EKG Gate (K001682) and Auto Gate (K003230), allow the operator to select images with reduced motion artifacts when processing data for Coronary Artery Calcification Scoring.
Automated Vascular Measurement (K002519) is an option within the Vitrea™2 application and is intended for study/analysis of selected vessels for stenosis analysis, pre/post stent planning and directional vessel tortuosity evaluation.
Tumor Volume Measurement (K002519) is an option within the Vitrea™2 application and is intended for the analysis/quantification of tumor volumes obtained from MR brain series scans.
CT Brain Perfusion (K003639) is an option within the Vitrea™2 application and is intended for post processing based on dynamic CT images continuously acquired during the injection of contrast for the visualization of apparent blood flow in brain tissue and pictorial illustration related parameters to aid in the assessment of the type and extent of cerebral perfusion disturbances.
ImageCheckerCT is an option within the Vitea™2 application and is intended for the display of a composite view of 2D cross-sections, and 3D volumes of Chest CT images, including findings of regions of interest ("ROI") identified by the radiologist, or Computer Assisted Detection ("CAD") findings.
### Predicate Device Comparison
The Vitrea2, Version 3.4 Workstation and the predicate devices allow for the analysis, communication and media interchange of digital images acquired from a variety of acquisition devices support the DICOM protocol for communication of images with other medical imaging devices.
#### Summary of Studies
The software utilized was designed, developed, tested, and validated according to written Design Control procedures. These procedures specify individuals within the organization responsible for developing and approving product specifications, coding, testing, validating and maintenance.
The Vitrea2, Version 3.4 Workstation will successfully complete Integration testing/verification prior to Beta validation. The software Beta testing validating the workstation will be successfully completed prior to release. In addition, potential hazards have been studied and controlled by a Risk Management Plan.
## Conclusion
The Vitrea2, Version 3.4 Workstation has the same intended use as the predicate devices and very similar indications and technological characteristics. Any technological differences do not raise any new questions regarding safety or effectiveness. Thus, the Vitrea2, Version 3.4 Workstation is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular logo with an abstract depiction of an eagle in the center. The eagle is composed of three curved lines that suggest its head, body, and wings. Encircling the eagle is text that reads 'DEPARTMENT OF HEALTH & HUMAN SVC'. The text is arranged along the circumference of the circle, with 'DEPARTMENT' at the bottom left and 'SVC' at the top right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 9 2003
Mr. Stephen S. Anderson Sr. Director, Quality and Customer Satisfaction Vital Images, Inc. 3300 Fernbrook Lane North, Suite 200 PLYMOUTH MN 55447-5341
#### Re: K032748
Trade/Device Name: Vitrea2, Version 3.4 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II Product Code: 90 LLZ Dated: September 4, 2003
Received: September 5, 2003
Dear Mr. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
# 510(k) Number (if known): _ 11 6 3 27 48
#### Device Name: Vitrea™2, Version 3.4 Medical Image Processing Software
#### Indications for Use:
Vitrea™2 is a medical diagnostic workstation that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. In addition, Vitrea™2 has the following specific indications.
VScore™ (K990442) is an option within the Vitrea™2 application and is intended for cardiac scoring from whole body CT derived measurements, including non-invasive detection and quantification of atherosclerotic plaque. Two image processing options, EKG Gate (K001682) and Auto Gate (K003230), allow the operator to select images with reduced motion artifacts when processing data for Coronary Artery Calcification Scoring.
Automated Vascular Measurement (K002519) is an option within the Vitrea™2 application and is intended for study/analysis of selected vessels for stenosis analysis, pre/post stent planning and directional vessel tortuosity evaluation.
Tumor Volume Measurement (K002519) is an option within the Vitrea™2 application and is intended for the analysis/quantification of tumor volumes obtained from MR brain series scans.
CT Brain Perfusion (K003639) is an option within the Vitrea™2 application and is intended for post processing based on dynamic CT images continuously acquired during the injection of contrast for the visualization of apparent blood flow in brain tissue and pictorial illustration of perfusion related parameters to aid in the type and extent of cerebral perfusion disturbances.
ImageCheckerCT is an option within the Vitrea™2 application and is intended for the display of a composite view of 2D crosssections, and 3D volumes of Chest CT images, including findings of regions of interest ("ROI") identified by the radiologist, or Computer Assisted Detection ("CAD") findings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Prescription Use* ✓
David A. Symm
Division of Reproductive, Abdom and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.