SURGASSIST CIRCULAR STAPLER DIGITAL LOADING UNITS (CS21, CS25, CS29, CS33)
K032701 · Power Medical Interventions, Inc. · GDW · Sep 30, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032701
Device Name
SURGASSIST CIRCULAR STAPLER DIGITAL LOADING UNITS (CS21, CS25, CS29, CS33)
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Sep 30, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SurgASSIST™ Circular Stapler Digital Loading Units® have applications throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.
Device Story
SurgASSIST™ Circular Stapler Digital Loading Units (DLUs) are single-patient-use surgical instruments for alimentary tract anastomoses. DLUs contain titanium staples and integral circular cutting blades. Device operates via remote electrically powered console; provides computer-mediated steering and stapling. Input: surgeon control via console. Transformation: device applies double staggered circular row of staples to secure tissue layers; simultaneously cuts excess tissue at controlled diameter. Output: stapled, anastomosed tissue. Safety feature: prevents deployment unless within appropriate range for desired closed staple height. Modification (K032701) addresses latching mechanism failure by modifying spline tube and trocar to improve latch finger engagement. Benefits: provides controlled staple closure and tissue cutting in a steerable, flexible format.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: ASTM F-67 unalloyed titanium staples, stainless steel knife, polymeric components. Energy: Electrically powered via remote console. Form factor: Flexible, steerable circular stapler (21, 25, 29, 33 mm). Connectivity: Memory module for digital data/identification. Sterilization: Ethylene Oxide (ETO). Software: Embedded, includes safety interlocks for deployment range.
Indications for Use
Indicated for patients requiring end-to-end, end-to-side, or side-to-side anastomoses throughout the alimentary tract.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST™ System with Circular Stapler Digital Loading Units® (K003277)
Endopath ILS Endoscopic Circular Staplers (K920752)
Submission Summary (Full Text)
{0}------------------------------------------------
Power Medical Interventions, Inc SurgASSIST™ Circular Stapler DLUs Special 510(k) Corrective Action Being Effected - August 29, 2003
ψ32701 (P.1.AS)
### Special 510(k) CORRECTIVE ACTION BEING EFFECTED SAFETY AND EFFECTIVENESS SUMMARY
## SurgASSIST Circular Stapler Digital Loading Units®
In Accordance with 21 CFR section 807.92 Power Medical Interventions, Inc., is submitting the following safety and effectiveness summary.
- 1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 267-775-8151 Ph 267-775-8123 Fax
Barbara J. Whitman Applicant:
August 29, 2003 Date of Notification:
- 2) Name of Device:
| Trade Name: | SurgASSIST™ |
|-------------|----------------------------|
| | Circular Stapler DLUs |
| | 21 mm, 25 mm, 29 mm, 33 mm |
Circular Stapler with Implantable Staples Common Name:
Classification Name: Staple, Implantable, GDW
- 3) Predicate Devices:
- A. SurqASSIST™ System with Circular Stapler Digital Loading Units®, (21 mm, 25 mm, 29 mm, 33 mm), with Titanium Implantable Staple, Power Medical Interventions, Inc., New Hope, PA. REF CS21, CS25, CS29, CS33 (K003277).
- B. Endopath ILS Endoscopic Circular Staplers, 21 mm, 25 mm, 29 mm, 33 mm, Ethicon Endo-Surgery, Cincinnati, OH. (K920752).
{1}------------------------------------------------
## 4) Device Description:
The SurqASSIST™ System with Circular Stapler Digital Loading Units® (DLUs) offers computer mediated steering and stapling. The DLUs contain implantable, titanium staples and integral cutting blades. The DLUs are used to anastomose tubular structures by applying a double staggered circular row of staples through the tissue. The staples form to controlled closed conditions to secure the lavers of tissue together. The DLUs also cut away the excess tissue at a controlled diameter inside the ring of staples. The Circular Stapler DLUs are available in the following four sizes: 21 mm, 25 mm, 29 mm, and 33 mm. The DLU is supplied sterilized and ready for use upon removal from its packaging.
- 5) Device Modification
Modifications were made to the predicate SurgASSIST™ Circular Stapler Digital Loading Units® (originally cleared under K003277) to address the root cause of the voluntary recall, which was a potential for latching mechanism failure, which could result in staple line failure and/or anvil jam. In order to further minimize the already low risk of latching mechanism failure, the spline tube and trocar were modified to improve engagement of the latch fingers. Further details of this modification can be found in Section H of this submission, under the "Device Modification" heading.
- 6) Indications For Use
The SurqASSIST™ Circular Stapler Digital Loading Units® have applications throughout the alimentary tract for end-to-end, end-toside, and side-to-side anastomoses.
- 7) Comparison to Predicate Devices
The following table compares the subject Circular Stapler to the previously cleared predicates, Circular Stapler Cutter Digital Loading Units® (K003277) and Endopath ILS Endoscopic Circular Staplers (K920752).
{2}------------------------------------------------
| Features & Description | SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm<br>REF: CS21, CS25, CS29, CS33 | Predicate<br>SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm | Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ILS Endoscopic Circular<br>Stapler<br>21 mm, 25 mm, 29 mm, 33 mm |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Name | Circular Stapler<br>Digital Loading Units®<br>21 mm, 25 mm, 29 mm, 33 mm | Circular Stapler<br>Digital Loading Units®<br>21 mm, 25 mm, 29 mm, 33 mm | Ethicon Endo-Surgery, Inc.<br>Endopath ILS Endoscopic<br>Circular Stapler<br>21 mm, 25 mm, 29 mm, 33 mm |
| Manufacturer of Record | Power Medical Interventions, Inc. | Power Medical Interventions, Inc. | Ethicon Endo-Surgery, Inc |
| Contract Manufacturer | Lacey Manufacturing<br>Bridgeport, CT | Lacey Manufacturing<br>Bridgeport, CT | Ethicon Endo-Surgery, Inc |
| 510(k) Clearance Numbers | Subject of this Notification | K003277 | K920752 |
| Intended use | Applications throughout the alimentary<br>tract for end-to-end, end-to-side and<br>side-to-side anastomoses. | Applications throughout the alimentary<br>tract for end-to-end, end-to-side and<br>side-to-side anastomoses. | Applications throughout the alimentary<br>tract for end-to-end, end-to-side and<br>side-to-side anastomoses. |
| Contraindications | Same, refer to labeling | Same, refer to labeling | Same, refer to labeling |
| FDA Class (System) | II | II | II |
| Sizes | 21 mm, 25 mm, 29 mm, 33 mm<br>Circular Staplers | 21 mm, 25 mm, 29 mm, 33 mm<br>Circular Staplers | 21 mm, 25 mm, 29 mm, 33 mm<br>Circular Staplers |
| Staple Shape | B-Shaped | B-Shaped | B-Shaped |
| Features & Description | SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm<br>REF: CS21, CS25, CS29, CS33 | Predicate<br>SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm | Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ILS Endoscopic Circular<br>Stapler<br>21 mm, 25 mm, 29 mm, 33 mm |
| Closed Staple Height | Approximately<br>1.5 mm - 2.3 mm | Approximately<br>1.5 mm - 2.3 mm | 1.0 mm - 2.5 mm |
| Staple Material | ASTM F-67 Unalloyed Titanium | ASTM F-67 Unalloyed Titanium | ASTM F-67 Unalloyed Titanium |
| Knife Material | Stainless steel | Stainless steel | Stainless steel |
| DLU Materials | Polymeric materials, surgical grade<br>stainless steels, adhesives, and<br>lubricants | Polymeric materials, surgical grade<br>stainless steels, adhesives, and<br>lubricants | Polymeric materials, surgical grade<br>stainless steels, adhesives, and<br>lubricants |
| Cutting Mechanism | Circular Knife | Circular Knife | Circular Knife |
| DLU Internal Power | None | None | None |
| Power | Electrically powered via a remote<br>Power Console | Electrically powered via a remote<br>Power Console | Manually powered |
| Software containing | Yes | Yes | No |
| Digital Information | Memory module containing digital data<br>for identification, etc. | Memory module containing digital data<br>for identification, etc. | None |
| How Supplied | Sterile - Single Patient Use | Sterile - Single Patient Use | Sterile - Single Patient Use |
| Features & Description | SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm<br>REF: CS21, CS25, CS29, CS33 | Predicate<br>SurgASSIST™<br>Circular Stapler<br>Digital Loading Units® (DLUs)<br>21 mm, 25 mm, 29 mm, 33 mm | Predicate<br>Ethicon Endo-Surgery, Inc.<br>Endopath ILS Endoscopic Circular<br>Stapler<br>21 mm, 25 mm, 29 mm, 33 mm |
| Safety Mechanism | Will not deploy until within an<br>appropriate range for desired closed<br>staple height | Will not deploy until within an<br>appropriate range for desired closed<br>staple height | Contains indicators for appropriate<br>range for desired closed staple height,<br>but can be deployed out of range |
| Insertion Mechanism | Flexible, steerable | Flexible, steerable | Rigid |
| Method of Sterilization | Ethylene Oxide Gas (ETO) | Ethylene Oxide Gas (ETO) | Irradiation |
| Packaging | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid |
# Circular Stapler DLU Product Features Comparison Chart
{3}------------------------------------------------
# Circular Stapler DLU Product Features Comparison Chart (continued from previous page)
{4}------------------------------------------------
## Circular Stapler DLU Product Features Comparison Chart (continued from previous page)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the caduceus symbol.
SEP 3 0 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Barbara J. Whitman Regulatory Affairs Manager Power Medical Interventions, Inc. 110 Union Square Drive New Hope, Pennsylvania 18938
Re: K032701
Trade/Device Name: SurgASSIST™ Circular Stapler Digital Loading Units® 21mm, 25mm, 29mm, 33mm Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: August 29, 2003 Received: September 2, 2003
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Ms. Barbara J. Whitman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
fed Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Power Medical Interventions, Inc. New Hope, PA 18938
510(k) No. K \$32741
Device Name:
SurgASSIST™ Circular Stapler Digital Loading Units® 21 mm, 25 mm, 29 mm, 33 mm
#### INDICATIONS FOR USE:
The SurgASSIST™ Circular Stapler Digital Loading Units® have applications throughout the alimentary tract for end-to-end, end-toside, and side-to-side anastomoses.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use
Per 21CFR §801.109
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Miriam C. Provost
OR
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K03270/
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.