GEMINI 16

K032036 · Philips Medical Systems (Cleveland), Inc. · KPS · Jul 11, 2003 · Radiology

Device Facts

Record IDK032036
Device NameGEMINI 16
ApplicantPhilips Medical Systems (Cleveland), Inc.
Product CodeKPS · Radiology
Decision DateJul 11, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1200
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The device is an imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT). The device produces attenuation and nonattenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as x-ray transmission images of these areas. The PET and CT images are registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined PET and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. The device can provide CT data suitable for use in attenuation correction.

Device Story

Gemini 16 is a combined PET/CT imaging system; integrates PET and multislice CT gantries within a common housing; utilizes single workstation for operator control. Inputs: PET radiopharmaceutical distribution data and X-ray transmission data. Operation: PET and CT images registered and fused for spatial alignment; CT data used for PET attenuation correction. Used in clinical oncology settings; operated by trained health professionals. Output: Fused PET/CT images for anatomical and functional interpretation. Benefits: Combined anatomical/functional data; reduced scan time; improved attenuation correction speed compared to predicate. System supports integrated or stand-alone PET/CT operation.

Clinical Evidence

Bench testing only. Performance measured in accordance with NEMA NU-2 standards. Compliance with 21 CFR 1020.30 and 1020.33 for radiation safety and IEC 60601-1 series for electrical/mechanical safety.

Technological Characteristics

Combined PET/CT scanner; GSO-based PET detectors; multislice CT capability. Integrated operator console. Compliance: NEMA NU-2, 21 CFR 1020.30/33, IEC 60601-1. Fixed or mobile installation capability.

Indications for Use

Indicated for clinical whole body oncology imaging in patients requiring PET and/or CT diagnostic scans. Used for head, body, and total body imaging by trained health professionals.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS General Information | A. | Submitter/ Contact Person:<br>Philips Medical Systems (Cleveland), Inc.<br>595 Miner Rd.<br>Cleveland, OH 44143 | Christof Littwitz<br>Tel: (440) 483-3585<br>Fax: (732) 352-6897 | |----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | B. | Device Trade Name: Gemini 16<br>Common Name: Positron Emission Tomography<br>Computed Tomography X-Ray<br>Classification Name: System, Emission Computed Tomography, (892.1200)<br>System, Computed Tomography X-Ray, (892.1750)<br>Device Class: 21CFR 892.1200, Class II<br>21 CFR 892.1750, Class II<br>Product Code: 90 KPS and 90 JAK | | | C. | Date prepared: June 20, 2003 | | | D. | Predicate Device: Gemini (K013521)<br>Mx8000 IDT (K012009) | | ## Performance Standards E. - 21 CFR 1020.30 1020.33 Performance Standards for Ionizing Radiation . Emitting Products, Computed Tomography Equipment (Applicable Sections) NEMA NU-2 . - F. Intended Use: The device is an imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT). The device produces attenuation and nonattenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as x-ray transmission images of these areas. The PET and CT images are registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined PET and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. The device can provide CT data suitable for use in attenuation correction. Device Description/ Comparison with Predicate Device: G. The device is a combination Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) scanner that can be utilized in both conventional, fixed installations or mobile environments. The system integrates the two system operator consoles into a single workstation to allow straightforward planning and system operation. The individual PET and the CT gantries remain intact as major subsystem located within a common integrated housing. It can be used in clinical protocols and procedures, which have been clinically conducted in a separate CT system and/or a PET system. {1}------------------------------------------------ The system is intended for use primarily as a clinical whole body oncology scanner with high-end multislice CT capability and high-end GSO based PET performance. It mechanically separates to allow for the greater flexibility for above three operation modes. System Performance Test/ Summary of Studies: H. To minimize electrical, mechanical and radiation hazards, Philips Medical System adheres to recognized and established industry practice. Radiation safety is assured by compliance and certification to the performance standards for ionizing radiation emitting product, 21CFR 1020.30 and 21CFR 1020.33. The radiation safety product report will be filed in accordance with 21CFR 1002.10 with the Center for Device and Radiological Health. Electrical and mechanical safety is assured by adherence and certification to the applicable standards in the IEC 60601-1 series. The device performance was measured in accordance with NEMA-NU2 standard. I. Comparison to Predicate Device The device is a modification of the currently marketed Gemini PET/CT System (K013521) with design modifications resulting in decreasing patient scan time and improvement in attenuation correction speed. The similarities and differences between the Gemini 16 and the predicate devices are described in detail in Section VI of the premarket notification. In conclusion, the device is substantially equivalent to the predicate device Gemini (K013521) and the Mx8000IDT (K012009) systems based upon similar intended use, technological comparison, and system performance. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 1 2003 Re: K032036 Trade/Device Name: Gemini 16 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 KPS and JAK Dated: June 27, 2003 Received: July 1, 2003 Dear Mr. Steneberg: Philips Medical System, Inc. % Mr. Heinz Joerg Steneberg Primary Third Party Reviewer 12 Commerce Road NEWTOWN CT 06470 TÜV Rheinland of North America We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure . # {4}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510 (k) NUMBER (IF KNOWN): K032036 DEVICE NAME: GEMINI 16 ## INDICATIONS FOR USE: The device is an imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT). The device produces attenuation and nonattenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as x-ray transmission images of these areas. The PET and CT images are registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined PET and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. The device can provide CT data suitable for use in attenuation correction. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109 OR Over-The-Counter-Use (Optional Format 1-2-96) . ** David R. Seymann
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...