K031852 · Micro Therapeutics, Inc. · HCG · Aug 20, 2003 · Neurology
Device Facts
Record ID
K031852
Device Name
SAPPHIRE DETACHABLE FIBER COILS
Applicant
Micro Therapeutics, Inc.
Product Code
HCG · Neurology
Decision Date
Aug 20, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sapphire Detachable Coils are intended for the endovascular embolization of intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) be inoperable. The intended uses of the Sapphire fiber coils are narrowed relative to the predicate device GDC, but still fall within the 510(k) cleared intended use. Sapphire Detachable Fibered Coil - The Sapphire fibered coil is intended for embolization of neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Fibered Sapphire is also intended for arterial and venous embolizations in the peripheral vasculature.
Device Story
Platinum alloy wire wound into primary coil and secondary helical shape; attached to stainless steel positioning wire with Teflon jacket. Detachment achieved via battery-operated Sapphire Detachment System (SDS) which dissolves detachment element. Used by neurosurgeons in clinical/OR settings for endovascular embolization. Device provides mechanical occlusion of aneurysms/vascular abnormalities; nylon fibers on coil promote embolization. Physician monitors placement via fluoroscopy; detachment confirmed by SDS. Benefits include minimally invasive treatment of high-risk/inoperable intracranial aneurysms and vascular malformations.
Clinical Evidence
Bench testing (coil strength, fatigue, tensile strength, radiopacity, shelf life, fiber endurance) met acceptance criteria. Animal testing compared to predicate showed comparable coil-per-aneurysm, detachment time, and angiographic assessment at 1-3 months; Sapphire demonstrated higher detachment reliability. Physician evaluation confirmed comparable ease of use, trackability, and fluoroscopic visibility.
Technological Characteristics
Platinum alloy wire coil; stainless steel positioning wire with Teflon jacket; nylon fibers. Detachment via electrolytic dissolution (SDS). Available in Helix and Cyclone shapes. MRI compatible (1.5 Tesla).
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms in patients at high risk for traditional surgery or inoperable; also indicated for embolization of neurovascular abnormalities (AVMs, arteriovenous fistulae) and arterial/venous embolizations in peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
AUG 2 0 2003
K03 i852/P.13
Premarket Notification (510(k) for Sapphire Detachable Fibered C
| 510(k) Summary | |
|-----------------|---------------------------------|
| Trade Name: | Sapphire Detachable Fiber Coils |
| Generic Name: | Artificial Embolization Coil |
| Classification: | Class III, 21 CFR 882.5950 |
| Submitted By: | Micro Therapeutics, Inc. |
| | 2 Goodyear |
| | Irvine, California 92618 |
| Contact: | Florin Truuvert |
#### Predicate Device:
Micro Therapeutics, Inc.
| Number | Description | Predicate For | Clearance Date |
|---------|------------------------------------------------------|---------------------------------------|----------------|
| K993418 | Fibered GDC VortX Shape<br>Guglielmi Detachable Coil | Sapphire Fibered Helix<br>and Cyclone | 1/21/2000 |
### Device Description
The Sapphire Detachable Coil (SDC) is manufactured from a platinum alloy wire which is first wound into primary coil and then formed into a secondary helical shape. The coil is welded to a positioning wire, which consist of ground stainless steel core wire with a stainless steel coil laser welded at the distal end and a Teflon outer jacket. The coil is detached by the battery operated power supply (Sapphire Detachment System, SDS), which dissolves a small detachment element between the emboliozation coil and the positioning wire. The fibered coil is manufactured with nylon fibers secured into the primary coil. The fibered coils are available in two shapes (Helix and Cyclone) and different sizes.
#### Indication For Use
The Sapphire Detachable Coils are intended for the endovascular embolization of intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) be inoperable.
The intended uses of the Sapphire fiber coils are narrowed relative to the predicate device GDC, but still fall within the 510(k) cleared intended use.
Sapphire Detachable Fibered Coil - The Sapphire fibered coil is intended for embolization of neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Fibered Sapphire is also intended for arterial and venous embolizations in the peripheral vasculature.
{1}------------------------------------------------
K031852/P2/3
## Micro Therapeutics, Inc. Premarket Notification (510(k) for Sapphire Detachable Fibered Coil System
## Verification and Test Summary Table
| Bench Testing | Sapphire Fibered Coils |
|--------------------------------------------------------|----------------------------------------------|
| Coil Strength (Coil Deformation and Softness Testing) | Meet established acceptance criteria |
| Ease of Delivery | Meet established acceptance criteria |
| Reliability After Fatigue & Premature Detachment | Meet established acceptance criteria |
| Tensile Strength of Coil & Detachment Zone | Meet established acceptance criteria |
| Tensile Strength of Delivery Wire | Meet established acceptance criteria |
| Delivery Wire Flexibility | Meet established acceptance criteria |
| Detachment Time | Meet established acceptance criteria |
| Radiopacity | Meet established acceptance criteria |
| Particulate Generation of Detachment Zone | Meet established acceptance criteria |
| Aging and Shelf Life (3 year) | Meet established acceptance criteria |
| Fiber Pull Out | Meet established acceptance criteria |
| Fiber Endurance | Meet established acceptance criteria |
| Packaging validation after 3-year accelerated aging | Meet established acceptance criteria |
| Animal Testing | Comparison to the Predicate Device |
| Coil Per Aneurysm (Avg./S.D.) | Comparable |
| Detachment Time (seconds) | Comparable |
| Coil Detachment Reliability (detachment attempts/coil) | The Sapphire demonstrated higher reliability |
| Post Embolization Angiographic Assessment | Comparable |
| Angiographic Assessment One-Month Follow-Up | Comparable |
| Angiographic Assessment Three-Month Follow-Up | Comparable |
| Physician Device Evaluation | Acceptance Criteria |
| Ease of Coil Preparation | Comparable |
| Ability to move coil within Catheter | Comparable |
| Access | Acceptance Criteria |
| Ability to Place catheter tip in desired location | Comparable |
| Trackability, Friction of coil through catheter | Comparable |
| Coil Visualization | Acceptance Criteria |
| Fluoroscopic visibility of Coil | Comparable |
| Fluoroscopic visibility of Detachment Markers | Comparable |
| Coil Delivery Procedure | Acceptance Criteria |
| Ability to Position Coil in Sac | Comparable |
| Detachment Time | Comparable |
| Ability to Pack Aneurysm Sac | Comparable |
| Ability to Reposition | Comparable |
| Ease of use of the SDS Detachment System | Comparable |
| Coil Positional Stability/Aneurysm Occlusion | Comparable |
| Overall | Acceptance Criteria |
| Ease of Use | Comparable |
| Overall performance | Comparable |
| MRI Compatibility | Compatible with 1.5 Tesla |
. •
{2}------------------------------------------------
K031852/P3/3
#### Micro Therapeutics, Inc. Premarket Notification (510(k) for Sapphire Detachable Fibered Coil System
#### Summary of Substantial Equivalence
The above comparison table demonstrates the technological similarity and equivalency of the Sapphire fibered coils compared with the predicate device, Target GDC fibered coils. The two devices have the same intended use,
- Use the same operating principle,
- Incorporate the same basic design, .
- Have the same Intended Use, ●
- Are packaged and sterilized using similar materials and processes.
In summary, the Sapphire fibered coils described in this submission are, in our opinion, substantially equivalent to the predicate device.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 0 2003
Ms. Florin Truuvert Manager, Regulatory Affairs Micro Therapeutics, Inc. 2 Goodyear Irvine, California 92618
Re: K031852
Trade/Device Name: Sapphire Detachable Fibered Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Artificial embolization device Regulatory Class: III Product Code: HCG Dated: June 13, 2003 Received: June 24, 2003
Dear Ms. Truuvert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Florin Truuvert
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark N. Melson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K031852
0
# Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
Sapphire Detachable Fibered Coil System
Indications for Use:
#### Sapphire Detachable Coils (Non-Fiber)
The Sapphire Detachable Coils are intended for the endovascular embolization of intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) be inoperable.
The intended uses of the Sapphire fiber coils are narrowed relative to the predicate device GDC, but still fall within the 510(k) cleared intended use.
Sapphire Detachable Fibered Coil - The Sapphire fibered coil is intended for embolization of neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Fibered Sapphire is also intended for arterial and venous embolizations in the peripheral vasculature.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div>OR</div> <div>(Per 21 CFR 801.109)</div> | Over the Counter Use |
|------------------|-----------------------------------------------|----------------------|
|------------------|-----------------------------------------------|----------------------|
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
| 510(k) Number | K031852 |
|---------------|---------|
|---------------|---------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.