K031819 · Msq (M2) , Ltd. · GEX · Oct 30, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031819
Device Name
DIO-DENT 10 DENTAL DIODE LASER
Applicant
Msq (M2) , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 30, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dio-Dent 10 Dental Diode Laser System is intended for incision, excision, ablation, vaporization and/or coagulation (hemostasis) of oral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva). In addition, the system is intended for teeth whitening.
Device Story
GaAlAs solid-state diode laser; delivers up to 10W energy via fiber optic handpiece. Visible aiming beam targets treatment area. Operator adjusts optical power and pulse settings. Used by dental professionals for oral soft tissue surgery (incision, excision, ablation, coagulation) and aesthetic teeth whitening. Provides precise tissue interaction; facilitates hemostasis; enables minimally invasive surgical access and decontamination. Benefits include reduced bleeding, improved healing, and efficient soft tissue management.
Clinical Evidence
No clinical data provided. Substantial equivalence based on bench testing, performance standards compliance (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 601-2-22), and comparison to predicate devices.
Technological Characteristics
Gallium Aluminum Arsenide (GaAlAs) solid-state diode laser; 10W maximum output; fiber optic delivery system; visible aiming beam. Complies with 21 CFR 1040.10/1040.11 (Class IV laser) and IEC standards. Adjustable power and pulse settings.
Indications for Use
Indicated for oral soft tissue procedures including biopsy, gingivectomy, frenectomy, lesion removal, and sulcular debridement; endodontic procedures including pulpotomy; and cosmetic dentistry including light-activated teeth whitening. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
OCT 3 0 2003
Page 1 of 3
## 510(K) SUMMARY
# DIO-DENT 10 DENTAL DIODE LASER SYSTEM
510(k) Number K 031819
Applicant's Name: MSq(M2) Ltd. 7 Haeshel St. P.O.B 3021 Caesarea Industrial Park 38900 Israel Tel: 04-6275357 Fax: 04-6275368 e-mail: zivk@MSg.co.il
Arava Hacohen Contact Person: Push-med Ltd. 117 Ahuzah St. Ra'anana 43373, Israel Tel: 972-9-7718130 Fax: 972-9-7718131 e-mail: arava(@push-med.com
Date Prepared: May 2003
Dio-Dent 10 Dental Diode Laser System Trade Name:
Classification Name: Laser Instrument. Surgical, Powered
FDA has classified laser device as a class II device (product code Classification: GEX) and it is reviewed by the General & Plastic Surgery Panel.
Predicate Devices: The Dio-Dent 10 Dental Diode Laser System substantially equivalent to the Ceralase D 980nm Diode Laser System, Model D15 (Ceramoptec, Inc.) cleared under K983058, K991891, K993002, K993911 and K022351 in terms of technological characteristics, performance, intended use, indications for use and user interface.
> In addition, the Dio-Dent 10 is substantially equivalent to the Opus 10 Dental Diode Laser System (OpusDent Ltd.) cleared under K000990 and K011769 in terms of intended use. indication for use. technological characteristics. performance and user interface.
{1}------------------------------------------------
Performance Standards: The Dio-Dent 10 Diode Laser complies with:
U.S. Federal Performance Standards 21 CFR 1040.10 and 21 CFR 1040.11 for class IV Laser Products.
In addition, the device complies with the European Medical Directive 93/42/EEC concerning medical devices (Annex II) and with the voluntary standards IEC 60601-1, IEC 60601-1-2, IEC-60825-1 and IEC 601-2-22.
- Intended Use / Indication for Use: The Dio-Dent 10 Dental Diode Laser System is intended for incision, excision, ablation, vaporization and/or coagulation (hemostasis) of oral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva). In addition, the system is intended for teeth whitening.
The Indications for Use of the Dio-Dent 10 Dental Diode Laser System include:
Cosmetic Dentistry
- · Light activation for bleaching materials for teeth whitening
- · Laser-assisted bleaching/whitening of the teeth
## Endodontology
- · Pulpotomy
- · Pulpotomy as an adjunct to root canal therapy
# Periodontology
- · Sulcular debridement (removal of diseased or inflamed soft tissue in the periodental pocket)
# Oral Soft Tissue Surgery
- · Biopsy
- · Operculectomy
- · Gingivectomy
- · Gingivoplasty
- · Oral papillectomy
- Leukoplakia
- Treatment of aphthous ulcers
- · Fibroma removal
- · Frenectomy
- · Frenotomy
- · Tissue retraction for impressions
- · Incising and draining of abscesses
- · Draining fistulas
- · Exposure of unerupted partially erupted teeth
- · Lesion (tumor) removal
- · Implant recovery
{2}------------------------------------------------
Page 3 of 3
- · Implant uncovering
- Gingival troughing
- · Crown lengthening
- · Hemostasis of donor site
- · Coagulation and decontamination of extraction sites
- · Removal of granulation tissue
- · Degranulation of infrabony defects
- · Laser assisted flap surgery
- · Vestibuloplasty
- · Removal of hyperplastic tissues
- · Debridement of diseased epithelial lining
#### The Dio-Dent 10 Dental Diode Laser is designed to perform several Device Description: medical procedures in the oral soft tissue and to perform laser assisted aesthetic tooth whitening procedures.
A Gallium Aluminum Arsenide (GaAlAs) solid state laser diode provides optical energy to oral soft tissues.
A fiber optic held by a handpiece delivers up to 10Watts laser energy. A visible light emitted from the handpiece's distal end targets the area of treatment. The optical power output and pulse may be adjusted to specific use requirements.
Substantial Equivalence: There are no unique applications, material or specifications presented below. Evidence of equivalence has been demonstrated through:
- · The Dio-Dent 10 intended use and indications for use were previously cleared by FDA for the predicate devices.
- · The technical characteristics of the Dio-Dent 10 are similar to those of the cleared Ceralas™ and Opus 10TM.
- · Laser output values of the Dio-Dent 10 are well within previous cleared values of the predicate dental diode laser systems as described.
- · The predicate devices and other previous cleared lasers with similar energy output have a proven safety and effectiveness in the treatment of the claimed indications.
- · Safety and performance testing.
Therefore. the Dio-Dent 10 Dental Diode Laser System is substantially equivalent to its predicate devices cited above and raises no new safetv and/or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is simple and monochromatic.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 3 0 2003
MSq(M2) Ltd. c/o Mr. Arava Hacohen Push-med Ltd. 117 Ahuzah Street Ra'anana 43373. Israel
Re: K031819 Trade/Device Name: Dio-Dent 10 Dental Diode Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 1, 2003 Received: July 30, 2003
Dear Mr. Hacohen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Arava Hacohen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
## Device Name: Dio-Dent 10 Dental Diode Laser System
Indications for Use: The Dio-Dent 10 Dental Diode Laser System is intended for incision, excision, ablation, vaporization and/or coagulation (hemostasis) of oral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva). In addition, the system is intended for teeth whitening.
> The Indications for Use of the Dio-Dent 10 Dental Diode Laser System include:
## Cosmetic Dentistry
Light activation for bleaching materials for teeth whitening, Laser-assisted bleaching/whitening of the teeth
### Endodontology
Pulpotomy, Pulpotomy as an adjunct to root canal therapv
## Periodontology
Sulcular debridement (removal of diseased or inflamed soft tissue in the periodental pocket)
## Oral Soft Tissue Surgery
Biopsy, Operculectomy, Gingivectomy, Gingivoplasty, Oral papillectomy, Leukoplakia, Treatment of aphthous ulcers, Fibroma removal, Frenectomy, Frenotomy, Tissue retraction for impressions. Incising and draining of abscesses, Draining fistulas, Exposure of unerupted partially erupted teeth, Lesion (tumor) removal, Implant recovery, Implant uncovering, Gingival troughing, Crown lengthening, Hemostasis of donor site, Coagulation and decontamination of extraction sites, Removal of granulation tissue, Degranulation of infrabony defects, Laser assisted flap surgery, Vestibuloplasty, Removal of hyperplastic tissues, Debridement of diseased epithelial lining
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
| 510(k) Number | K031819 |
|------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR Over the Counter Use |
| | Miriam C. Provost<br>(Division Sign-Off)<br>Division of General, Restorative<br>and Neurological Devices |
| | 510(k) Number K031819 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.