CT Injector: Administration of contrast and flushing media in conjunction with computed tomography (CT) scanning of the body. EDA: The Extravasation Detection Accessory is indicated for the detection of extravasations of contrast media during CT using a power injector.
Device Story
EmpowerCTA Injector System is a motor-driven, software-controlled, venous-side, low-pressure injector. It uses one or two consumable syringes to displace contrast and flushing media into a peripheral vein. A linear actuator controls syringe piston displacement. The system includes an Extravasation Detection Accessory (EDA) using bio-impedance sensing via an electrode patch to detect contrast media leakage. Operated by clinicians in a CT suite, the device features a remote control with a touchscreen and an injector head with membrane keys. The system provides automated flow rate, volume, and pressure control. Output is displayed on the remote console, allowing the clinician to monitor injection progress and extravasation status. The device benefits patients by automating contrast delivery and providing real-time extravasation detection, potentially reducing complications associated with contrast administration.
Clinical Evidence
Bench testing only. Performance testing verified flow rate accuracy (±5%), delivery volume accuracy (±2%), and pressure limiting capabilities. Bio-impedance sensing accuracy was validated against calibrated endpoints. Electromagnetic compatibility and electrical safety were confirmed per IEC/EN 60601-1 and 60601-1-2 standards.
Technological Characteristics
Syringe-based injector; motor-driven linear actuator; bio-impedance sensing for extravasation detection. Connectivity includes external CT trigger port. Power: 100/240 VAC auto-seeking. Dimensions/form factor: Floor or overhead mounted injector head with remote console. Software-controlled. Safety standards: IEC/EN 60601-1, IEC/EN 60601-1-2, UL 2601-1/CSA C22.2 No. 601.1.
Indications for Use
Indicated for administration of ionic/non-ionic contrast and flushing media during CT scanning of the body. EDA indicated for detection of contrast media extravasation during powered CT injections.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for the Department of Health and Human Services. The logo features a stylized eagle head with three lines representing its feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular fashion around the eagle head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 9 2003
Mr. Steven Hartman Senior Project Engineer, CT Products E-Z-EM, Inc. 717 Main Street WESTBURY NY 11590
Re: K031571
Trade/Device Name: E-Z-EM EmpowerCTA Injector System with Optional EDA Regulation Number: 21 CFR 870.1650 Regulation Name: Automatic contrast medium injector Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 74 IZQ and 90 JAK Dated: May 19, 2003 Received: May 22, 2003
Dear Mr. Hartman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### هاء سه سه سه مسه
# INDICATIONS FOR USE
| 510(k) Application: | Special 510(k) Device Modification |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | E-Z-EM EmpowerCTA Injector System with optional EDA |
| Indications for Use: | |
| | CT Injector: Administration of contrast and flushing media in conjunction with computed tomography (CT) scanning of the body. |
| | EDA: The Extravasation Detection Accessory is indicated for the detection of extravasations of contrast media during CT using a power injector. |
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | KC31571 |
| Prescription Use (Per 21 CFR 801.109) | OR | Over-the-Counter Use |
|---------------------------------------|----|----------------------|
|---------------------------------------|----|----------------------|
.
{4}------------------------------------------------
JUN 1 9 2003
K031571
ECEMI
Visualize a healthier world
## EmpowerCTA Injector System 510(k) Summary
#### Submitter's Information:
Telephone Number: Fax Number: Contact: Date Prepared:
Trade Name: Common Name: Classification Name:
Predicate Device:
E-Z-EM, Inc. 717 Main Street Westbury, NY 11590 (516) 333-8230, ext. 2345 (516) 333-8278 Steven Hartman, Senior Project Engineer, CT Injectors May 19, 2003
EmpowerCTA Injector System (Trademark Pending) CT Iniector Injector, Syringe, Extra-Luminal Plethysmograph, Impedance
EmpowerCT Injector System, K011160 (Internal Project Name, and Name Used in the 510(k) Application: E-Z-EM PercuPump 2001 CT Injector)
### Indications for Use:
CT Injector: Administration of nonionic and ionic compounds and flushing media in conjunction with computed tomography (CT) scanning of the body.
The EDA is intended to detect extravasations of ionic and non-ionic contrast media during powered CT contrast injections.
#### Device Description:
The EmpowerCTA Injector System is an injection system that uses one or two consumable syringe(s) to displace contrast media and flushing media to the patient. A motor driven linear actuator controls displacement of the syringe piston. This method of contrast injection is consistent with predicate devices from E-Z-EM for over 12 years. This new version of the EmpowerCT Injector family addresses the addition of a second syringe for the device for purposes of displacing flushing media.
A summary of the device as compared to the predicate device is as follows:
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 1 of 9
{5}------------------------------------------------
### Comparative Summary Tables
| | Proposed Device | Predicate Device |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| | EmpowerCTA Injector | EmpowerCT Injector<br>(PercuPump 2001) |
| Indication for Use | Administration of nonionic<br>and ionic compounds and<br>flushing media in conjunction<br>with computed tomography<br>(CT) scanning of the body | Administration of<br>nonionic and ionic<br>compounds in<br>conjunction with<br>computed tomography<br>(CT) scanning of the<br>body |
| Design | Syringe type injector,<br>software controlled, venous<br>side, low pressure injector. | Same |
| Anatomical Sites | Inject contrast and flushing<br>media into a peripheral vein | Inject contrast media into<br>a peripheral vein |
| Radiation | No ionizing radiation emitted | Same |
| Thermal | No thermal energy introduced<br>into patient | Same |
.
:
:
:
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 2 of 9
{6}------------------------------------------------
| | Proposed Device<br>EmpowerCTA Injector | Predicate Device<br>EmpowerCT Injector<br>(Percupump 2001) |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Injector Head | Located Adjacent to CT Scanner<br>Gantry via Floor or Overhead<br>Mounting | Same |
| Power Supply | Switching Power Supply in<br>Dedicated Enclosure | Switching Power Supply<br>in Dedicated Enclosure |
| Remote Control | Located Adjacent to CT Scanner<br>Console via Table or Wall<br>Mounting | Same |
| EDA (Extravasation<br>Detection Accessory) | Located in Dedicated Enclosure<br>mounted to same Floor or<br>Overhead Mounting as Injector<br>Head | Same |
| Pendant | Dedicated Single Button,<br>Start/Pause connected to Injector<br>Head and/or Remote | Same |
| | Proposed Device<br>EmpowerCTA Injector | Predicate Device<br>EmpowerCT Injector<br>(Percupump 2001) |
|----------------------------------------|---------------------------------------------------------------------|------------------------------------------------------------|
| Syringe | Existing 200 ml, Proprietary E-Z-EM<br>Interface (Catalog No. 6720) | Same |
| EDA Electrode Patch | Existing Proprietary E-Z-EM<br>Electrode Patch (Catalog No. 7881) | Same |
| EDA Patient Cable<br>(Semi-Disposable) | Existing Patient Cable (Catalog No.<br>9871) | Same |
E-Z-EM EmpowerCTA 510(k) Submission Summary
Page 3 of 9
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{7}------------------------------------------------
| | Proposed Device | Predicate Device |
|---------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Performance<br>(Injector) | EmpowerCTA Injector | EmpowerCT Injector<br>(Percupump 2001) |
| Flow Rate | 0.1 to 10 ml/sec in user specified<br>increments of 0.1 ml/sec<br>Accuracy: ±5% of programmed<br>rate +0.1 ml/sec | 0.1 to 10 ml/sec in user<br>specified increments of<br>0.1 ml/sec<br>Accuracy: ±5% of<br>programmed rate +0.1<br>ml/sec) |
| Delivery Volume | 1 to 200 ml in user specified<br>increments of 1 ml<br>Accuracy: ±2% of programmed<br>volume +1ml) | 1 to 200 ml in user<br>specified increments of 1<br>ml<br>Accuracy: ±2% of<br>programmed volume<br>+1ml |
| Maximum Pressure | 50 to 300 psi in user specified<br>increments of 1 psi<br>Accuracy: ±10% of programmed<br>pressure limit + 10 psi) | 20 to 300 psi in user<br>specified increments of 1<br>psi<br>Accuracy: ±10% of<br>programmed pressure<br>limit + 10 psi) |
| Pressure Limiting | Yes | Yes |
100 - 100 -
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 4 of 9 .
{8}------------------------------------------------
| | Proposed Device | Predicate Device |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Performance (EDA) | EmpowerCTA Injector | EmpowerCT Injector<br>(Percupump 2001) |
| Bio-Impedance Sensing | Range: 10 to 250<br>Resolution: 1/3 Ohm<br>Accuracy: +/-10%<br>Endpoints calibrated to 10 Ohm<br>+10% -0% and<br>250 Ohm -10% +0% | Same |
| Indicated Extravasation<br>Detection Threshold | 20 ml | Same |
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 5 of 9
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{9}------------------------------------------------
| | Proposed Device | Predicate Device |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Environmental<br>Requirements | EmpowerCTA Injector | EmpowerCT Injector<br>(Percupump 2001) |
| Operating Temperature<br>Operating Humidity,<br>Operating Altitude &<br>Storage Temperature | Meets Requirements set forth in<br>IEC/EN60601-1, Medical<br>Electrical Equipment Part 1:<br>General Requirements for Safety | Same |
| Electromagnetic<br>Compatibility (EMC) | Meets Requirements set forth in<br>IEC/EN 60601-1-2, Medical Electrical<br>Equipment Part 1: Collateral Standard,<br>Electromagnetic Compatibility<br>CISPR 11 (EN55011: 1998/1999)<br>IEC 60601-1-2: 1993/2001<br>EN61000-4-2: 1995<br>EN61000-4-3: 1996<br>EN61000-4-4: 1995<br>EN61000-4-5: 1995<br>EN61000-4-6: 1996<br>EN61000-4-8: 1993<br>EN61000-4-11: 1994 | Meets Requirements set forth in IEC/EN<br>60601-1-2, Medical Electrical Equipment<br>Part 1: Collateral Standard,<br>Electromagnetic Compatibility<br>CISPR 11<br>IEC 801-2<br>IEC 801-3<br>IEC 801-4<br>IEC 801-5 |
| UL/CSA | Medical Device Listing to UL<br>2601-1 / CSA C22.2 No. 601.1,<br>Electrical Class II, Type B Isolation<br>rating, Injector, Type CF, Isolation<br>rating, EDA | Same |
| Energy Requirements | 100/240 VAC 50/60 Hz<br>Auto Seeking | Same |
| Shock and Vibration | International Safe Transit Authority<br>(ISTA), Project 3C | Same |
:
60
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 6 of 9
・・・・・・・・・
{10}------------------------------------------------
| Injector Head<br>Displays and<br>Controls | Proposed Device<br>EmpowerCTA Injector | Predicate Device<br>EmpowerCT Injector<br>(PercuPump 2001) |
|-------------------------------------------------------|-----------------------------------------------|------------------------------------------------------------|
| Display | 240 x 180 Pixel<br>Electrolumenescent Display | Same |
| Syringe Advance<br>Slow | Membrane Key | Same |
| Syringe Advance<br>Fast | Membrane Key | Same |
| Syringe Retract<br>Slow | Membrane Key | Same |
| Syringe Retract Fast | Membrane Key | Same |
| Auto Initialize | Membrane Key | Same |
| Replace Syringe | Membrane Key | Same |
| Test Injection | Membrane Key | Same |
| Auto-Fill | Membrane Key | Same |
| Begin Fill | Membrane Key | Same |
| | Proposed Device | Predicate Device |
| Injector Head<br>Displays and Controls<br>"Continued" | EmpowerCTA Injector | EmpowerCT Injector<br>(PercuPump 2001) |
| Increase Flow Rate | Membrane Key | Membrane Key |
| Decrease Flow Rate | Membrane Key | Membrane Key |
| Hand Knob | Direct Connection to Drive<br>Mechanism | Same |
| Syringe Warmer | Snap On | Same |
| Illuminated<br>Stop/Arm/Run/Pause Status<br>Indicator | LED | LED |
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 7 of 9
1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
61
{11}------------------------------------------------
E-Z-EM EmpowerCTA 510(k) Submission Summary Page 8 of 9
-
:
62
:
:
{12}------------------------------------------------
| | Proposed Device | Predicate Device |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Remote Control<br>Display and<br>Controls | EmpowerCTA Injector | EmpowerCT Injector<br>(PercuPump 2001) |
| Display | 800 x 600 Color TFT LCD<br>with Touchscreen Overlay | 800 x 600 Color TFT<br>LCD<br>with Touchscreen<br>Overlay |
| Maximum Number of<br>Injection Phases per<br>Protocol | 8 (Contrast Only)<br>or<br>2 (1 Contrast followed by 1<br>Saline)<br>or<br>3 (2 Contrast followed by 1<br>Saline) | 8 |
| Maximum Number of<br>Stored Injection<br>Protocols | 50 | 50 |
| Programmed Pause | Yes | Yes |
| External CT Trigger<br>Port | Yes | Yes |
| Audio Speaker | Yes | Yes |
| EDA Display and<br>Controls | Proposed Device | Predicate Device |
|-----------------------------|----------------------------------------------------------------------------------|----------------------------------------|
| EDA Display and<br>Controls | EmpowerCTA Injector | EmpowerCT Injector<br>(PercuPump 2001) |
| Display | EDA Function Supported<br>within Remote Control Display<br>and Touchscreen Field | Same |
and the comments of the country of
E-Z-EM EmpowerCTA 510(k) Submission Summary
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.