HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact lenses are indicated for the correction of visual acuity in persons with not-aphakic, non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters or less that does not interfere with visual acuity. The HYDROGENIC® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are indicated for daily wear. The eye care practitioner may prescribe the contact lenses for either single use disposable wear or for frequent replacement wear. When prescribed for frequent replacement the lenses may be disinfected using a chemical or hydrogen peroxide disinfecting systems. The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Device Story
Hydrophilic soft contact lens; ocufilcon F material; 60% water content. Features aspheric anterior surface to control longitudinal spherical aberration; spherical posterior surface; lenticular edge design. Incorporates Vat Blue 6 tint and benzophenone-based UV-absorbing additive. Daily wear modality; prescribed by eye care practitioners. Corrects refractive errors; provides UV protection. Manufactured via cast molding. No clinical data required; equivalence established via comparison of physical, chemical, and optical parameters to predicate devices.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and comparison of physical/optical parameters to predicate devices.
Technological Characteristics
Material: ocufilcon F (FDA Group IV, 60% water). Additives: Vat Blue 6 dye, benzophenone-based UV absorber. Design: Aspheric anterior, spherical posterior, lenticular edge. Dk: 25.3 x 10^-11 (cm²/sec)(ml O₂/ml x mm Hg). Manufacturing: Cast molded. Power range: +10.00 D to -10.00 D.
Indications for Use
Indicated for correction of visual acuity in non-aphakic, non-diseased eyes with myopia or hyperopia and refractive astigmatism ≤ 2.00 D. For daily wear; prescribed for single use or frequent replacement.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
# JUL 1 8 2003
510(K) PREMARKET NOTIFICATION
SUMMARY
| Reference | OSLA60 |
|-----------|--------|
| Section | 3 |
| Version | 1 |
| Page | 1 / 5 |
(ocufilcon F) soft (hydrophilic) contact lens
SECTION 3:
## 510(K) SUMMARY
## APPLICANT'S NAME AND ADDRESS
Ocular Sciences Inc. 1855 Gateway Blvd. Suite 700 Concord, CA 94520 USA
Richard Lippman, OD FAAO Vice President Ophthalmic Regulatory Medical Products R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Road Suite 30 Rockville, Maryland 20850 Telephone: (240) 683-3738 (240)683-9236 Fax:
## 1. IDENTIFICATION OF DEVICE
Common Name: Trade Name:
Contact Person
Classification: Device classification: Hydrophilic Soft Contact Lens HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lens Daily Wear Soft (hydrophilic) Contact Lens Class II (21 CFR 886.5925)
#### 2. DESCRIPTION OF DEVICE
HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are available with in monomer tint (Vat Blue 6) and with ultraviolet absorbing additive (benzophenone based):
- +10.00 D to -10.00 D Asphere
- with center thickness from 0.025mm to 0.40mm
- . with base curves of 8.00mm to 9.20mm
- . with diameter of 12.00mm to 18.00mm
The lens material for the HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) is the same as the Biomedics® 60 UV (ocufileon F) lens in spherical design as described in submission K992264 cleared on November 24, 1999.
The aspheric lens design for the HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) is the same design construct as the BIOMEDICS® UV ASPHERE (ocufilcon D) as cleared in K020193 on February 28, 2002.
The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lens has a spherical posterior surface. The anterior (convex) surface is constructed in lenticular form to provide optimum edge thickness and contour. This front optical surface allows for correction of visual acuity in non-aphakic persons with non-diseased eyes and has been aspherized to control longitudinal spherical aberration of the lens across the power range.
{1}------------------------------------------------
| <b>Ocular Sciences</b> | <b>510(K) PREMARKET NOTIFICATION</b> | Reference : OSLA60 | | | | | | |
|-----------------------------------------------|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------|-----|---------|-----|------|---------|
| (ocufilcon F) soft (hydrophilic) contact lens | SUMMARY | <table><tr><td>Section</td><td>: 3</td></tr><tr><td>Version</td><td>: 1</td></tr><tr><td>Page</td><td>: 2 / 5</td></tr></table> | Section | : 3 | Version | : 1 | Page | : 2 / 5 |
| Section | : 3 | | | | | | | |
| Version | : 1 | | | | | | | |
| Page | : 2 / 5 | | | | | | | |
#### 3. INTENDED USE
HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact lenses are indicated for the correction of visual acuity in persons with not-aphakic, non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
The HYDROGENIC® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are indicated for daily wear. The eye care practitioner may prescribe the contact lenses for either single use disposable wear or for frequent replacement wear. When prescribed for frequent replacement the lenses may be disinfected using a chemical or hydrogen peroxide disinfecting systems.
The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
#### 4. PREDICATE DEVICES
The following predicate lenses were selected to address material (FDA Group IV: high water, ionic polymer), intended use (daily wear) and lens designs (Asphere).
#### Lens material, spherical lens design and intended use:
BIOMEDICS® UV 60 (ocufilcon F) Hydrophilic Contact Lenses for Daily Wear, FDA Group IV, high water content, ionic soft contact lenses are marketed nationally by OCULAR SCIENCES Inc. under PMA K992264 cleared for marketing on November 24, 1999.
#### Asphere lens design:
BIOMEDICS® UV ASPHERE (ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear, FDA Group IV, high water content, ionic soft contact lenses are marketed nationally and internationally by OCULAR SCIENCES Inc. under K020193 cleared for marketing on February 28, 2002.
{2}------------------------------------------------
| Ocular Sciences | 510(K) PREMARKET NOTIFICATION | Reference : OSLA60 |
|-----------------------------------------------|-------------------------------|--------------------|
| | SUMMARY | Section : 3 |
| (ocufilcon F) soft (hydrophilic) contact lens | | Version : 1 |
| | | Page : 3 / 5 |
## 5. CHARACTERISTICS
The characteristics of the HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are compared to the characteristics of the predicate device HYDROGENICS 60 UV Sphere in the following table.
#### TABLE I
| Material Comparison | | |
|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------|
| | Predicate Device<br>HYDROGENICS® 60<br>UV Sphere and Toric<br>(K992264) | Subject device<br>HYDROGENICS® 2.0 60<br>UV Asphere |
| PRODUCTION METHOD | Cast molded process | Cast molded process |
| INTENDED USE | Daily wear<br>Correction of ametropia | Daily wear<br>Correction of ametropia |
| MATERIAL | ocufilcon F<br>60% | ocufilcon F<br>60% |
| Type | Group IV | Group IV |
| Color additive | Vat Blue 6 Dye<br>CFR 130-20-1 | Vat Blue 6 Dye<br>CFR 130-20-1 |
| UV additive | Yes | Yes |
| Dk permeability, ISO 9913-1 Polarimetric method<br>with edge correction @ 35°C<br>x $10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg) | $25.3 x 10^{-11}$ | $25.3 x 10^{-11}$ |
| Light transmittance | Tv, % @ 20°C<br>Illuminant A<br>with a 2°<br>observer<br>(between 380<br>and 780 nm)<br>97.7% | (between 400 and<br>800 nm)<br>>95% |
| | Tv, % @ 20°C<br>Illuminant A<br>with a 2°<br>observer<br>(between 380<br>and 780 nm)<br>97.7% | (between 400 and<br>800 nm)<br>>95% |
{3}------------------------------------------------
## 510(K) PREMARKET NOTIFICATION
## SUMMARY
(ocufilcon F) soft (hydrophilic) contact lens
## TABLE 2
| | Lenses design comparison | | | |
|----------------------------------------|-------------------------------------------------------------------|---------|--------------------------------------------------------------------------|---------|
| | Subject Device<br>HYDROGENICS® 2.0 60 UV ASPHERE<br>(ocufilcon F) | | Predicate Device<br>BIOMEDICS® 55 UV Asphere<br>(ocufilcon D)<br>K020193 | |
| Characteristics comparison,<br>-2.00 D | Measured | Labeled | Measured | Labeled |
| Base Curve, mm | 8.5 | 8.5 | 8.4 | 8.6 |
| Diameter, mm | 14.08 | 14.1 | 14.1 | 14.2 |
| Power, D | -1.97 | -2.00 | 2.13 | -2.00 |
## 6. NON CLINICAL STUDIES
In support to this application we are providing
- Additional Manufacturing information .
- Finished Lens Parameters comparison with the predicate device ●
- Labeling Information ●
By reference: K992264, K020193 and K984046 data can be found concerning:
- Predicate Device ●
- . Chemistry
- Manufacturing process .
- Toxicology: lens and packaging materials
- Residual (leachables) Monomer
- Shelf life data
- Microbiology .
- . Packaging
## 7. CLINICAL DATA
It was determined that Clinical Studies were not necessary to establish the safety and efficacy of the HYDROGENICS®2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses, This determination was based on the following:
- The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses were . demonstrated to be substantial equivalent to the predicate BIOMEDICS/HYDROGENICS® UV Sphere lenses (K99264 and K984046). All demonstrated substantial equivalence in parameters.
- . The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lens design was demonstrated to be equivalent to the predicate Asphere lenses: BIOMEDICS® UV Asphere (ocuffilcon D) approved under K020193.
{4}------------------------------------------------
| Ocular | Sciences |
|--------|----------|
|--------|----------|
## 510(K) PREMARKET NOTIFICATION
(ocufilcon F) soft (hydrophilic) contact lens
## 8. CONCLUSIONS DRAWN FROM STUDIES
#### Substantial Equivalence:
The information provided in this 510(k) establishes that the HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are equivalent in optical, chemical and physical properties of the predicate devices and do not raise any questions of safety and effectiveness. Therefore the device is substantially equivalent to the predicate device.
#### Risk and Benefits:
The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear base. The patient are the same as those for other soft (hydrophilic) contact lenses.
## 9. ROUTE CHOSEN IN THE FLOW CHART FOR 510 (K) DAILY WEAR CONTACT LENS
#### FIGURE 1
## HYDROGENICS® 2.0 60 UV ASPHERE
Image /page/4/Figure/13 description: This image is a flowchart that outlines a decision-making process. It starts with "510(k)" and branches out based on questions like "Clear?" and "FDA Approved Care Regimen?". The flowchart then leads to different categories such as "Same USAN with Same Suffix", "Same USAN with Different Suffix", and "New USAN", each requiring specific data like Physicochemical Properties, Toxicology, Microbiology, and Clinical Performance Data to determine if it is SE or NSE.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol resembling three abstract human figures or birds in flight, arranged in a layered, overlapping manner.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ocular Sciences, Inc.
c/o Richard E. Lippman, O.D., F.A.A.O. R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Road, Suite 30 Rockville, MD 20850
Re: K031477
Trade/Device Name: Hydrogenics® 2.0 60 UV Asphere (ocufilcon F) Soft (hydrophilic) Contact Lenses indicated for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL; MVN Dated: May 9, 2003 Received: May 9, 2003
JUL 1 8 2003
Dear Dr. Lippman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
## Page 2 - Richard E. Lippman, O.D., F.A.A.O.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Reipi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### INDICATION FOR USE STATEMENT
# SECTION 2: INDICATION FOR USE STATEMENT
KC31477 510(k) Number (if known)
HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses Device Name:
#### Indications for Use:
HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact lenses are indicated for the correction of visual acuity in persons with not-aphakic, non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism 2.00 diopters or less that does not interfere with visual acuity.
The HYDROGENIC® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses are indicated for daily wear. The eye care practitioner may prescribe the contact lenses for either single use disposable wear or for frequent replacement wear. When prescribed for frequent replacement the lenses may be disinfected using a chemical or hydrogen peroxide disinfecting systems.
The HYDROGENICS® 2.0 60 UV ASPHERE (ocufilcon F) Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR Over-the-counter-use
(Division Sign-off)
Division of Ophthalmic Devices
510(k) Number K031477
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.