The BIOMEDICS 55 (ocufilcon D) UV Blocking soft (hydrophilic) contact lens is indicated for daily wear in not-aphakic, normal eyes which manifest myopia, hyperopia, or astigmatism in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Spherical lenses may mask astigmatism in patients with up to 2.00 diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the eyecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.
Device Story
Soft, hydrophilic contact lens; hemispherical shell design; fits over corneal surface. Manufactured from polymerized HEMA and monomeric ingredients; includes proprietary UV absorber; crosslinked with EGDMA. Corrects refractive errors (myopia, hyperopia, astigmatism) by providing corrective power corresponding to eye's refractive needs. Features base curves, optical zone (>6mm), secondary/tertiary curves, and beveled edges for centration and comfort. Used by eyecare practitioners to prescribe for daily wear; patient performs daily disposal or cleaning/disinfection via chemical regimen. Provides UV protection; benefits patient by correcting vision and reducing ocular UV exposure.
Clinical Evidence
Bench testing only. Evidence consists of physical, optical, and chemical property comparisons and ultraviolet transmittance profiles (UVA/UVB) compared to human cornea and crystalline lens.
Technological Characteristics
Material: Ocufilcon D (HEMA, monomeric ingredients, UV absorber, EGDMA crosslinker). Form: Hemispherical soft contact lens. Dimensions: Base curves 6.5-10.8mm, diameter 12.5-18.0mm, center thickness 0.025-0.27mm. Power: -20.00 to +20.00 D. Tint: Blue visibility. Sterilization: Chemical disinfection regimen.
Indications for Use
Indicated for daily wear in non-aphakic, normal eyes with myopia, hyperopia, or astigmatism (-20.00 to +20.00 D; astigmatic correction -0.25 to -10.00 D). Spherical lenses may mask up to 2.00 D of astigmatism.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
BIOMEDICS 55 (ocufilcon D) In Monomer Tint (K972303)
{1}------------------------------------------------
JAN 2 1 1999
Image /page/0/Picture/1 description: The image shows a logo with the words "OCULAR SCIENCES" on the top line. Below that is a symbol that looks like a target, followed by the words "AMERICAN HYDRON" on the next two lines. The text and symbol are all in black and white.
K984046
# 510(k) Summary
SUBMITTER:
# Submitted on behalf of:
Company Name: Address: Phone: Fax:
Ocular Sciences, Inc. 475 Eccles Av. South San Francisco, CA 95014 (415) 583-1400 (415) 583-1108
CONTACT PERSON:
Richard E. Lippman, O.D., F.A.A.O. Official Representative and Correspondent C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Suite 625 Rockville, MD 20852
# DATE SUMMARY PREPARED: October 1, 1998
TRADE NAME:
BIOMEDICS® 55 (ocufilcon D) UV Blocking Daily Wear soft (Hydrophilic) Contact Lens
## COMMON NAME: contact lens
## SUBSTANTIALLY EQUIVALENT TO:
The BIOMEDICS 55 (ocufilcon D) UV Blocking Daily Wear Contact Lens with In-Monomer Tint is equivalent to the daily wear lens of the same material as cleared in K942214, P890023/S26, P890023/S4,and K972303 in-monomer visibility tint cast molded ocufilcon D for daily wear currently marketed by Ocular Sciences, Inc.
The BIOMEDICS 55 (ocufilcon D) UV Blocking Daily Wear contact lens is substantially equivalent to the indication for use of the BIOMEDICS 55 (ocufilcon D) In Monomer Tint ocufilcon D cast molded contact lens marketed for use in the U.S. by Ocular Sciences, Inc. cleared under K972303. Additionally, the subject contact lens is equivalent and has the same characteristics and properties as the clear BIOMEDICS® 55 (ocufilcon D) cast molded contact lens, cleared under K942214, and P890023/S26. The cast molding process was approved under P890023/S4. This lens is in the Lens Group IV high water ionic group as established by the FDA and located in the Guidance
# 000032
**Corporate Offices:**
475 Eccles Avenue
South San Francisco.
California 94080
800-628-5367
Ph: 650-583-1400
Canada:
110 Cochrane Drive
Markham, Ontario
L3R 9S1
Canada
Ph: 905-940-1998
England:
Reliant Close
Chandlers Ford
Eastleigh
Hampshire S053 4ND
England
Hungary:
Hársmajor u.8
1118 Budapest
Hungary
Ph: 36-1-319-6241
Fax: 36-1-319-3034
RICO:
P.O. Box 2801
Highway 538 km. 0.7
Santa Isabel Industrial Park
Santa Isabel,
Puerto Rico 00757
{2}------------------------------------------------
Document for Daily Wear Contact Lenses, Revised Edition May 1994. The physical, optical, and chemical properties of the BIOMEDICS® 55 (ocufilcon D) UV visibility tint cast molded Ocular Sciences, Inc.contact lens are substantially equivalent to both the clear and tinted versions of the BIOMEDICS 55 (ocufilcon D) ocufilcon D cast molded contact lenses.
#### DESCRIPTION of the DEVICE:
Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA and other monomeric ingredients including a proprietary ultraviolet absorber, crosslinked with EGDMA and other components which yield the appearance of lenses which are designed to fit over the corneal surface of the eye. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness), hyperopia (farsightedness) and astigmatism (multiple foci). Each lens provides corrective power which is to correspond to the refractive power of the eye to which it is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6mm. Secondary and tertiary curves as well as beveled edge configurations are built into the lens for the purpose of aiding in lens centration and comfort.
#### INDICATIONS FOR USE:
The BIOMEDICS® 55 (ocufilcon D) soft (hydrophilic) contact lens is indicated for daily wear in normal eyes whose eyes are either myopic, hyperopic, or astigmatic in powers from -20.00 to +20.00 Diopters and astigmatic corrections from -0.25 to -10.00 Diopters. Spherical lenses may mask astigmatism of up to 2.00 Diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the evecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.
{3}------------------------------------------------
# PARAMETERS AVAILABLE:
| Base Curves: | 6.50 mm to 10.8 mm |
|-------------------|----------------------------------------|
| Diameters: | 12.5 mm to 18.0 mm |
| Powers: | -20.00 to +20.00 Diopters sphere |
| Cylinder Power: | -00.25 to -10.00 Diopters |
| Center Thickness: | 0.025 mm to 0.27 mm depending on power |
| Tint: | Blue Visibility tint |
# ULTRAVIOLET TRANSMITTANCE CURVES
The following graph represents the transmittance of ultraviolet light in the UVA and UVB range as well as light transmittance through the visual spectrum, as compared to the human cornea and crystalline lens:
Image /page/2/Picture/4 description: The image shows a scattering of small, irregularly shaped black dots against a white background. The dots vary in size, with some appearing as tiny specks and others as slightly larger, more defined shapes. The distribution of the dots is uneven, with some areas of the image having a higher concentration of dots than others, creating a sense of randomness and dispersion.
Image /page/2/Picture/5 description: The image shows the number 000034 in a bold, sans-serif font. The numbers are evenly spaced and aligned horizontally. The image is simple and only contains the number.
{4}------------------------------------------------
#### Typical Transmittance Profile of Ocufilcon D Hydrophilic Contact Lens with UV versus a Human Cornea and a Human Lens
Image /page/3/Figure/1 description: The image is a graph comparing the percentage of light transmission through different types of lenses at different wavelengths. The x-axis represents the wavelength in nanometers, ranging from 200 to 750. The y-axis represents the percentage of transmission, ranging from 0 to 100. The graph shows the transmission of light through an Ocufilcon D contact lens with UV protection, a human cornea from a 24-year-old person, and a human crystalline lens from a 25-year-old person.
O
COOO(
{5}------------------------------------------------
# Typical Transmittance Profile of -2.50 D Ocufilcon D Hydrophilic Contact Lens with UVC versus a Human Cornea and a Human Lens
Image /page/4/Figure/1 description: The image is a graph that shows the percentage of transmission of light through different materials as a function of wavelength. The x-axis represents the wavelength of light in nanometers, ranging from 200 to 750 nm. The y-axis represents the percentage of transmission, ranging from 0 to 100%. There are three lines on the graph, representing the transmission of light through Ocufilcon D contact lens, the human cornea of a 24-year-old person, and the human crystalline lens of a 25-year-old person.
COOOO
{6}------------------------------------------------
# Typical Transmittance Profile of Ocufilcon D Hydrophilic Contact Lens with UVAM versus a Human Cornea and a Human Lens
Image /page/5/Figure/1 description: The image is a graph comparing the percentage of light transmission through different parts of the human eye and a contact lens. The x-axis represents the wavelength of light in nanometers, ranging from 200 to 750. The y-axis represents the percentage of light transmission, ranging from 0 to 100. The graph shows the transmission of light through the Ocufilcon D contact lens, the human cornea of a 24-year-old, and the human crystalline lens of a 25-year-old.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 1 1999
Ocular Sciences, American Hydron, Inc. c/o Richard E. Lippman, O.D., F.A.A.O. C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Rockville, Maryland 20852
Re: K984046 Trade Name: Biomedics ® 55 (ocufilcon D) UV Blocking Soft Contact Lens for Daily Wear Regulatory Class: II Product Code: 86 LPL Dated: November 6, 1998 Received: November 13, 1998
Dear Dr. Lippman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{8}------------------------------------------------
Page 2 - Richard E. Lippman, O.D., F.A.A.O.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of - Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### Indications Statement
510(k) Number (if known) K984046
Device Name; Biomedics 55 (ocufilcon D) UV Blocking Daily Wear Contact Lens
#### Indications for Use:
Additional Claims:
The BIOMEDICS 55 (ocufilcon D) UV Blocking soft (hydrophilic) contact lens is indicated for daily wear in not-aphakic, normal eyes which manifest myopia, hyperopia, or astigmatism in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Spherical lenses may mask astigmatism in patients with up to 2.00 diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the eyecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Presription Use | X |
|----------------------|---|
| Over-The-Counter Use | |
(Division Sign-Off)
Division of Ophthalmic Devices
(Oprtional Format 1-2-96)
510(k) Number_K984046
00001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.