STERICHEK BICARB PH REAGENT STRIPS
Device Facts
| Record ID | K031267 |
|---|---|
| Device Name | STERICHEK BICARB PH REAGENT STRIPS |
| Applicant | Environmental Test Systems, Inc. |
| Product Code | MNV · Gastroenterology, Urology |
| Decision Date | Dec 23, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
Indications for Use
SteriChek® Bicarb pH Reagent Strips provide a quick convenient means of checking the pH of the bicarbonate solution of used to prepare dialysate or concentrated bicarbonate solution used in the preparation of dialysate. The color that develops in the pad after exposure to the sample according to the directions is compared to a color chart to determine the pH of the sample.
Device Story
Device consists of 0.20-inch square reagent pad affixed to 3.25-inch white opaque polystyrene strip. Reagent pad contains Cresol Red and Phenol Red indicators. User exposes pad to bicarbonate solution sample; color change occurs based on pH/hydrogen ion concentration. User visually compares pad color to calibrated color chart on label to determine pH. Used in clinical settings for monitoring dialysate preparation. Provides rapid, convenient pH verification to ensure proper solution concentration.
Clinical Evidence
Bench testing only. Performance compared against Serim™ Bicarb pH II Test Strips using acid/bicarbonate and concentrated bicarbonate solutions. Reference pH values determined via standard method calibrated with NIST traceable reference material. Performance found to be equivalent.
Technological Characteristics
Reagent pad on polystyrene strip. Indicators: Cresol Red and Phenol Red. Sensing principle: colorimetric pH-dependent indicator reaction. Form factor: 3.25" x 0.20" strip. Manual visual readout via color chart comparison.
Indications for Use
Indicated for checking the pH of bicarbonate solutions and concentrated bicarbonate solutions used in the preparation of dialysate.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
- Serim™ Bicarb pH II Test Strips