MODIFICATION TO BIO-LOGIC EVOKED POTENTIAL (EP) SYSTEM
Applicant
Bio-Logic Systems Corp.
Product Code
GWJ · Neurology
Decision Date
Apr 28, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and display of human physiological data, for auditory screening purposes and to assist in determining possible auditory and hearing-related disorders. Auditory stimuli are presented to the patient's ear through an earphone or headphones, and the corresponding Auditory Brainstem Responses (ABR) from the patient are recorded using EEG electrodes placed on the scalp. Standard ABR testing is most used clinically for 2 reasons: (1) to predict behavioral audiometric thresholds, and (2) as an audiological testing tool to assist in the assessment of possible auditory nervous system abnormalities. The Bio-logic Navigator Pro EP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
Device Story
System records/displays human physiological data for auditory screening/assessment. Inputs: auditory stimuli (clicks/tones) via earphones; EEG signals via scalp electrodes. Operation: stimulus-response cycle repeated thousands of times; EEG responses (1-20ms post-stimulus) amplified, digitized, and averaged to extract ABR signal from noise. Hardware (Navigator Pro module) performs data acquisition via DSP; host PC runs Windows-based software for control, display, and user interface. Used in clinical settings by trained professionals. Output: averaged ABR waveforms for clinician review. Assists in predicting audiometric thresholds and identifying nervous system abnormalities. No automated diagnosis; clinician reviews raw data and makes final decisions.
Clinical Evidence
Bench testing only. No clinical data presented. Safety and effectiveness established via internal product development procedures (ISO-9001, ISO-13485, FDA QSR Design Control) and risk management (ISO-14971/Fault Tree Analysis).
Technological Characteristics
System consists of a patient connection module (Navigator Pro) and a host PC. Sensing: scalp EEG electrodes. Actuation: auditory transducers (earphones/headphones). Connectivity: serial port communication between PC and hardware module. Software: Windows-based application controlling DSP-based hardware. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for auditory screening and assessment of auditory/hearing-related disorders in patients of all ages (infants, children, adults, geriatric). Used to predict behavioral audiometric thresholds and assess auditory nervous system abnormalities. Requires prescription and supervision by physician or trained healthcare professional.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Bio-logic Systems Corp. The logo is in black and white. The word "Bio-logic" is in a bold, sans-serif font, with the word "SYSTEMS CORP" in a smaller, sans-serif font underneath. There is a registered trademark symbol to the right of the word "Bio-logic".
One Bio-logic Plaza Mundelein, IL 60060-3700 1-800-323-8326 Fax: 847-949-8615 www.bio-logic.com
# SECTION 2: SUMMARY AND CERTIFICATION 510(K) SUMMARY
Safety and effectiveness information concerning the Bio-logic Evoked Potential product and this device modification is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
| PREPARED BY: | Bio-logic Systems Corp |
|--------------|------------------------|
| | One Bio-logic Plaza |
| | Mundelein, IL 60060 |
(847)-949-5200 TELEPHONE:
Norman E. Brunner CONTACT PERSON:
## DATE ON WHICH THE SUMMARY WAS PREPARED: March 26, 2003
NAME OF DEVICE: Bio-logic Evoked Potential System.
COMMON NAME: Evoked Response System.
CLASSIFICATION NAME: Evoked Response Auditory Stimulator (per CFR 882-1900).
PREDICATE DEVICE: Bio-logic Navigator Pro, reference 510(k) #K994149.
## DESCRIPTION OF THE DEVICE:
The Bio-logic Auditory Evoked Potential (AEP) family of products is used for the recording, display, computation, and manipulation of human physiological data, for auditory screening purposes and to provide assistance in the neurological assessment and treatment of auditory sensory disorders. The predicate device referenced above is the latest in a series of complete systems of this type marketed by Bio-logic.
The Navigator Pro Evoked Potential Predicate Device performs Auditory Brainstem Response (ABR) recording functions with two channels of simultaneous data recording. The software for the Navigator Pro implements the standard ABR functions common to most similar systems on the market for many years.
{1}------------------------------------------------
Ko31009
The standard ABR test works on the basis of repeating a stimulus-response cycle. An auditory stimulation (click, tone, etc.) is presented to the patient through the use of an earphone or headphones. The EEG response from the brain is read through the use of one or more scalp electrodes placed on the patient. The response time of interest is approximately from 1 - 20 milliseconds following the stimulus. The response voltage readings for this time period are amplified, digitized and stored in the AEP system computer's memory. The stimulation is then repeated, the EEG response is read again, and this cycle is repeated many times . Each time the response is read, it is averaged together with all previous responses. The final data record is the result of averaging several thousand (usually 2000-3000) responses. This averaging process is necessary because the ABR signal is very small, much lower in voltage than the surrounding EEG "noise" present in the recording. The noise is averaged out over the many readings, because the noise will have a very low average net value. The result from the averaging process will be the signal.
The reason for this 510(k) is that the software for the host computer (PC) in the Evoked Potential system has been modified. The host computer software for the Predicate Device is DOS-based, whereas the host software for this modified Bio-logic AEP program is WINDOWS-based. The WINDOWS program has the same features and functionality as that of the predicate device, but with improved user interfaces and overall ease-of-use. The functions of the hardware are the same as those of the Predicate Device (Navigator Pro) hardware. The host computer software for control of the hardware has the same functionality as that of the DOS-based Predicate Device software. Direct hardware control of all Navigator Pro functions is provided from the Digital Signal Processor (DSP) and its program code located inside the Navigator Pro module, instead of directly from the host computer program. (There are no changes to the DSP program code in the Navigator Pro module as a result of this host computer software modification.) By distributing the hardware-specific functions to the DSP, the Windows-based host computer program has fewer real-time demands, and performance and reliability are improved. Together, the hardware and software implement the same functionality and perform the same intended use as the Predicate Device, but with improved ease-ofuse and flexibility for the user.
{2}------------------------------------------------
K031009
### INTENDED USE:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and displav of human physiological data, for auditory screening purposes and to assist in determining possible auditory and hearing-related disorders. Auditory stimuli are presented to the patient's ear through an earphone or headphones, and the corresponding Auditory Brainstem Responses (ABR) from the patient are recorded using EEG electrodes placed on the scalp. Standard ABR testing is most used clinically for 2 reasons: (1) to predict behavioral audiometric thresholds, and (2) as an audiological testing tool to assist in the assessment of possible auditory nervous system abnormalities.
The Bio-logic Navigator Pro EP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
# SAFETY AND EFFECTIVENESS SUMMARY
To establish the safety and effectiveness of this modification to the Bio-logic Evoked Potential software, the modification was designed and incorporated into the product in accordance with the Bio-logic internal Product Development procedures, which are intended to meet ISO-9001, ISO-13485 and FDA QSR Design Control specifications. A detailed Hazard/Risk analysis for the EP family of products was performed using the Fault Tree analysis (FTA) approach, and a detailed Risk Assessment for the modified Evoked Potential product was written in accordance with ISO-14971, the International Standard: Application of Risk Management to Medical Devices.
As with the Predicate Device, the modified software does not make any final decisions that result in any automatic forms of diagnosis or treatment. All program "recommendations" are subject to review by a qualified health care professional, and may be modified, overridden or deleted as determined by the qualified user. The program provides functions to allow the qualified user to review all raw data collected and perform other data analysis to suit his or her requirements.
The chart on the following page provides a summary comparison of the technological characteristics of the modified Evoked Potential device relative to the predicate Navigator Pro device. This is to demonstrate that this Modification to Bio-logic Evoked Potential for Navigator Pro has no significant differences which would adversely affect product safety and effectiveness.
{3}------------------------------------------------
| Parameter for comparison | Similarity or Difference |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | No differences. |
| Population | No differences. |
| Hardware Configuration | No differences. Both devices use the Navigator Pro patient<br>connection hardware module. |
| Computer Control Software | The modified Evoked Potential device has changes to the<br>AEP software program running in the PC (host computer).<br>These changes allow full use of the Windows operating<br>system functions and GUI features. The Predicate Device<br>uses the DOS operating system or runs in a "DOS Window"<br>with very little Windows compatibility. |
| Patient information and tracking. | No differences. |
| Patient connections (transducers and<br>electrodes) | No differences. |
| Host Computer to Patient Connection<br>Module Communications | No differences. The communication from the PC (host<br>computer) to the Navigator Pro patient connection module<br>is provided through a serial port cable for both devices. |
| Presentation of Data / User Interface | Changes have been made in the modified device software to<br>incorporate GUI features and functionality common to most<br>Windows programs on the market today. Computer system<br>users are very familiar with standard Windows features.<br>Therefore, these changes increase ease-of-use and provide<br>improved flexibility and efficiency for the user, with no<br>significant changes to the underlying AEP features and<br>functions. |
| Physical Characteristics | No differences. |
| Safety Characteristics | No differences. The patient connection and isolation<br>circuits are the same for both products. |
| Product Labeling | No differences. |
| Anatomical sites | No differences. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines that also suggest human profiles.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 8 2003
Mr. Norman E. Brunner Vice-President of Research and Development Bio-logic Systems Corporation One Bio-logic Plaza Mundelein, Illinois 60060-3700
Re: K031009
Trade/Device Name: Modification to Bio-logic Evoked Potential (EP) System Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked response auditory stimulator Regulatory Class: II Product Code: GWJ Dated: March 28, 2003 Received: March 31, 2003
Dear Mr. Brunner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Norman E. Brunner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
### 510(k) Number (if known):
#### K031009
### Device Name: Bio-logic Evoked Potential System.
#### Indications For Use:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and display of human physiological data, for auditory screening purposes and to assist in determining possible auditory and hearing-related disorders. Auditory stimuli are presented to the patient's ear through an earphone or headphones, and the corresponding Auditory Brainstem Responses (ABR) from the patient are recorded using EEG electrodes placed on the scalp. Standard ABR testing is most used clinically for 2 reasons: {{} to predict behavioral audiometric thresholds, and (2) as an audiological testing tool to assist in the assessment of possible auditory nervous system abnormalities.
The Bio-logic Navigator Pro EP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
-Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K631009
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.