MODIFICATION TO BIO-LOGIC EVOKED POTENTIAL FOR ABAER/ NAVIGATOR PRO.
K994149 · Bio-Logic Systems Corp. · GWJ · Dec 27, 1999 · Neurology
Device Facts
Record ID
K994149
Device Name
MODIFICATION TO BIO-LOGIC EVOKED POTENTIAL FOR ABAER/ NAVIGATOR PRO.
Applicant
Bio-Logic Systems Corp.
Product Code
GWJ · Neurology
Decision Date
Dec 27, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. An auditory stimulus (click, tone, etc.) is presented to the patient's ear through an earphone or headphones, and the Brainstem Auditory Evoked Response from the patient is recorded using EEG electrodes placed on the scalp. Although this Brainstem Response is very low in amplitude (with respect to surrounding EEG "noise"), the stimulus-response cycle is repeated many times and the resulting responses are averaged from the time of the stimuli. The random noise averages to zero, but if the Brainstem Response signal will be easily determined in the averaged signal. The Bio-logic EP System can be used for patients of all ages, from children infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
Device Story
Device records/analyzes human physiological data for neurological diagnosis of sensory disorders; specifically Auditory Evoked Potentials (AEP). Input: auditory stimuli (clicks/tones) via headphones; EEG scalp electrode signals. Operation: stimulus-response cycles repeated thousands of times; EEG responses amplified, digitized, and averaged to isolate signal from noise. Artifacts (high-voltage noise) automatically discarded. Output: averaged AEP signal for clinician review. Used in clinical settings by physicians or trained healthcare professionals. Navigator Pro hardware (external module) replaces internal ISA boards; Windows-based software replaces DOS-based software. POVR algorithm monitors signal-to-noise ratio to optimize test speed/quality. Benefits: improved ease-of-use, portability, and flexibility for hearing screening in infants and difficult-to-test populations.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, design control procedures (ISO-9001, FDA QSR), and risk assessment (EN-1441/Fault Tree Analysis) comparing the modified hardware/software to the predicate.
Technological Characteristics
Standalone hardware module with medical-grade power supply; connects via serial port. Uses EEG scalp electrodes for signal acquisition. Software is Windows-based with DSP-based hardware control. Implements POVR algorithm for signal-to-noise monitoring. Patient isolation circuits identical to predicate. Designed to meet ISO-9001 and FDA QSR standards.
Indications for Use
Indicated for patients of all ages (infants, children, adults, geriatric) requiring auditory and hearing-related disorder diagnosis. Specifically for individuals where behavioral audiometric results are unreliable, including infants, young children, and cognitively impaired or uncooperative adults.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Navigator and Traveler Evoked Potential Product (K930328)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 2 7 1999
K994149
One Bio-logic Plaza Mundelein, Illinois 60060-3700 1-800-323-8326 Fax: 847-949-8615
# SECTION 2: SUMMARY AND CERTIFICATION
### 510(K) SUMMARY
Safety and effectiveness information concerning the Bio-logic Evoked Potential product and this device modification is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
| PREPARED BY: | Bio-logic Systems Corp |
|--------------|------------------------|
| | One Bio-logic Plaza |
| | Mundelein, IL 60060 |
(847)-949-5200 TELEPHONE:
Norman E. Brunner CONTACT PERSON:
DATE ON WHICH THE SUMMARY WAS PREPARED: December 3, 1999
NAME OF DEVICE: Bio-logic ABAER / Navigator Pro.
COMMON NAME: Evoked Response System.
CLASSIFICATION NAME: Evoked Response Auditory Stimulator (per CFR 882-1900).
PREDICATE DEVICE: Bio-logic Evoked Potential for ABAER I, reference 510(k) #K992807.
## DESCRIPTION OF THE DEVICE:
The Bio-logic Evoked Potential family of products is intended to be used for the recording and analysis of human physiological data for the purpose of neurological diagnosis and treatment of sensory disorders. The predicate device referenced above is the latest in a series of systems of this type marketed by Bio-logic. Other related devices comprising the Evoked Potential family include:
- 510(k) #K803226 Bio-logic Evoked Response Stimulators. 1.
- 510(k) #K842543 Bio-logic Evoked Potential System. 2.
- 510(k) #K844992 Bio-logic Portable Evoked Response System. 3.
- 510(k) #K862690 Bio-logic Traveler LT System. 4.
- 5. 510(k) #K930328 - Navigator and Traveler Evoked Potential Product.
{1}------------------------------------------------
Section 2
K994149
The Predicate Device performs Evoked Potential recording and analysis functions, provides two channels of simultaneous data recording, and includes the option of applying the Point Optimized Variance Ratio (POVR) algorithm for optimizing signal quality and speed of test completion. This new device has both hardware and software modifications and improvements over the Predicate and related devices. Related Device #1 above is the first Evoked Potential device marketed by Bio-logic. It provides for up to 4 channels of data recording. Related Device #2 above is a hardware/software modification to the first device. Related Device #3 above is similar to #2, but utilizes a "portable" computer for ease of use and transportability. Device #4 utilizes hardware and software variations over devices #1, #2 and #3, primarily to enhance size-reduction and portability. It provides for a maximum of 2 channels of data recording. Trade names of "Traveler", "Express" and "LT" are associated with these transportable devices. Device #5 incorporates the "Explorer" hardware with corresponding hardware control and analysis features in the software. Data recording hardware is available in three variations: the "E" Series, for up to 2 channels of data recording; the "SE" Series, for up to 4 channels of data recording; and the "Explorer" Series, for up to 8 channels of data recording. The Evoked Potential software has stayed essentially the same for all of these products, with variations in models to accommodate differences in the hardware.
Evoked Potential systems can be used for three different kinds of tests: Auditory Evoked Potentials (AEP), Visual Evoked Potentials (VEP), and Somatosensory Evoked Potentials (SEP). These variations are called "modalities", and are offered as options in all three models of hardware marketed by Bio-logic. Each modality has its own unique hardware requirements. The changes described in this 510(k) are to incorporate a repackaged variation of the previouslydescribed "E" Series hardware (2-channel), along with a new Windows-based software program for the control of this new hardware. This hardware performs the AEP test only, so, correspondingly, the software contains only AEP control and analysis features.
The AEP test works on the basis of repeating a stimulus-response cycle. An auditory stimulation (click, tone, etc.) is presented to the patient through the use of an earphone or headphones. The EEG response from the brain is read through the use of one or more scalp electrodes placed on the patient. The response time of interest is approximately from 1 – 20 milliseconds following the stimulus. The response voltage readings for this time period are amplified, digitized and stored in the AEP system computer's memory. The stimulation is then repeated, the EEG response is read again, and this cycle is repeated many times . Each time the response is read, it is averaged together with all previous responses. The final data record is the result of averaging several thousand (usually 2000-3000) responses. This averaging process is necessary because the EEG signal is very small, much lower in voltage than the surrounding EEG "noise" present in the recording. The noise is averaged out over the many readings, because the noise will have an average net value of zero. The result from the averaging process will be the signal.
Some of the EEG responses may have large amounts of noise or other artifact caused by random events such as patient movement or externally-generated electrical noise. These artifacts are usually characterized by very high amplitude voltages (relative to normal EEG levels). The EP program automatically monitors the response for abnormally-high voltage levels. When a response contains such artifacts, the response is discarded, and not counted in the cycle count.
{2}------------------------------------------------
*Section 2*
K994149
The Predicate Device describes the use of the POVR algorithm to monitor the signal-to-noise ratio of the averaged response signal every 256 stimulation-response cycles. This modified device, ABAER / Navigator Pro, incorporates the same POVR algorithm for the same purpose, to yield high-quality test results in the shortest possible time. In addition, the feature modifications for this new device include the use of a redesigned hardware package (called the Navigator Pro) and a Windows-based software program to control the hardware. The functions of the hardware are substantially the same as those of the Predicate Device hardware. The software for control of this device is a simplified subset of the DOS-based Predicate Device software. Together, they implement the same functionality and perform the same intended use as the Predicate Device, but with improved ease-of-use.
The Navigator Pro hardware is very similar in electronics design to the Navigator E hardware used in the Predicate Device, except that the electronics hardware has been re-packaged into a stand-alone box which derives its power from a separate medical-grade power supply and connects to the host computer through a serial port. In the Predicate Device, the Navigator E hardware consists of two ISA-bus computer boards mounted inside the host computer and connected to an external patient-connection module through a dedicated cable. The ISA computer bus is no longer fully supported by many computer makers, and many new computers do not have any ISA slots at all. Also, because computers are getting smaller and many (such as laptops) do not have any expandable internal bus at all, this new Navigator Pro hardware design offers more flexibility and options for use than does the present Navigator E hardware. Nearly all computers have at least one serial port, which is all that is necessary to connect to the Navigator Pro hardware.
The host software for the Predicate Device is DOS-based, whereas the host software for the ABAER - Navigator Pro (subject of this Special 510(k)) is WINDOWS-based. The WINDOWS program has much of the same functionality as that of the predicate device, but with improved user interfaces and overall ease-of-use. Because the primary intention of the ABAER program is one of simplified testing and screening, some of the more complex diagnostic functions of the Predicate Device were removed to make the device easier to use.
#### INTENDED USE:
The Bio-logic Evoked Potential (EP) product family is intended for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. An auditory stimulus (click, tone, etc.) is presented to the patient's ear through an earphone or headphones, and the Brainstem Auditory Evoked Response from the patient is recorded using EEG electrodes placed on the scalp. Although this Brainstem Response is very low in amplitude (with respect to surrounding EEG "noise"), the stimulus-response cycle is repeated many times and the resulting responses are averaged from the time of the stimuli. The random noise averages to zero, but if the Brainstem Response signal is present, it's signal will be easily determined in the averaged signal.
The Bio-logic AEP System can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
{3}------------------------------------------------
$$\forall \forall \forall \forall \neg$$
The ABAER with Navigator Pro device uses a redesigned hardware package and a Windowsbased software program to control the hardware. The functions of the hardware are substantially the same as those of the Predicate Device hardware. The software for control of this device is a simplified subset of the DOS-based Predicate Device software. Together, they implement the same functionality and perform the same intended use as the Predicate Device, but with improved ease-of-use. The POVR algorithm described in the Predicate Device is implemented in the new Windows software to generally assist in test data interpretation, and specifically assist in the assessment of signal-to-noise ratio and the quality of the Brainstem Auditory Evoked Response in infants. Based on this automatic assessment, the speed of testing may be reduced and/or the quality of the data recording may be improved, without compromising the quality of recorded data or limiting the control and flexibility of the health care professional administering the test.
## SAFETY AND EFFECTIVENESS SUMMARY
To establish the safety and effectiveness of this modification to the Bio-logic Evoked Potential software, the modification was designed and incorporated into the product in accordance with the Bio-logic internal Product Development procedures which are intended to meet ISO-9001 and FDA QSR Design Control specifications. A detailed Hazard/Risk analysis for the EP family of products was performed using the Fault Tree analysis (FTA) approach, and a detailed Risk Assessment was written in accordance with EN-1441, the International Standard for Hazard/Risk analysis. An addendum to this Hazard/Risk file was written based on a review of this new ABAER / Navigator Pro design.
The Navigator Pro patient-connection hardware utilizes many of the same design principles and circuit designs as are used in the Bio-logic Evoked Potential "E" series hardware. There are no newly-introduced hardware-related methods by which the patient can be harmed or injured through the use of this device. The same patient isolation methods are used in both products. The Navigator Pro utilizes a medical-grade power supply, whereas the "E" Series hardware used the computer's power. Direct hardware control of all Navigator Pro functions is provided from the Digital Signal Processor (DSP) and its program code located inside the Navigator Pro package, instead of directly from the host computer program. By distributing the hardwarespecific functions to the DSP, the Windows-based host computer program has fewer real-time demands, and performance and reliability are improved.
The ABAER / Navigator Pro software does not make any final decisions that result in any automatic forms of diagnosis or treatment. All program "recommendations" are subject to review by the EP Technologist or Physician, and may be modified, overridden or deleted as determined by a qualified user. The program provides additional functionality to allow the qualified user to review all raw data collected and perform other data analysis to suit his or her requirements.
The chart on the following page provides a summary comparison of the technological characteristics of the new modified device relative to the predicate Evoked Potential device. This is to demonstrate that this new ABAER / Navigator Pro device has no significant differences which would adversely affect product safety and effectiveness.
{4}------------------------------------------------
Section 2
K994/49
| Parameter for comparison | Similarity or Difference |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | No differences. |
| Population | No differences. |
| Hardware Configuration | Predicate Device uses Navigator E hardware, whereas<br>the new ABAER product uses Navigator Pro hardware.<br>The two hardware types are similar in functionality,<br>patient connection, and patient isolation circuits. |
| Computer Control Software | Predicate Device software is DOS-based, whereas the<br>new ABAER product uses a combination of Windows-<br>based host software and a DSP hardware control<br>program in the Navigator Pro module. |
| Patient information and tracking. | The new ABAER program incorporates an EXCEL-<br>compatible database for Patient and Test Information,<br>which represents an extensive improvement in<br>performance over the DOS filename-based method of<br>the Predicate Device. |
| Test Performance -- Time to Complete. | No significant differences. |
| Test Performance - Data Quality. | No significant differences. |
| Safety Characteristics | No differences. The basic patient connection and<br>isolation circuits are the same for both products. |
| Product Labeling | The labels used on the two products are different<br>primarily because of the different physical<br>characteristics of the two hardware types. Also, the<br>new ABAER / Navigator Pro label makes use of<br>standard international symbols for patient and device<br>connections. |
| Anatomical sites. | No differences. |
| Physical Characteristics | The new Navigator Pro hardware is physically different<br>from that of the Predicate Device. It is housed in a<br>separate module with its own separate medical-grade<br>power supply, and connected to the host computer via<br>serial port. Predicate Device Series "E" hardware has<br>boards inside the computer and a separate patient-<br>connection module. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an image of a stylized eagle with its wings spread. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 7 1999
Mr. Norman E. Brunner Vice President of Research and Development Bio-Logic Systems Corporation One Bio-Logic Plaza Mundelein, Illinois 60060
Re: K994149 Trade Name: ABAER/Navigator Pro Evoked Potential Regulatory Class: II Product Code: GWJ Dated: December 6, 1999 Received: December 8, 1999
Dear Mr. Brunner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
#### Page 2 – Mr. Norman E. Brunner
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Honeil/29
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K994149
510(k) Number (if known):
Not Assigned
#### Device Name: Modification to Bio-logic Evoked Potential for ABAER / Navigator Pro.
Indications For Use:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. An auditory stimulus (click, tone, etc.) is presented to the patient's ear through an earphone or headphones, and the Brainstem Auditory Evoked Response from the patient is recorded using EEG electrodes placed on the scalp. Although this Brainstem Response is very low in amplitude (with respect to surrounding EEG "noise"), the stimulus-response cycle is repeated many times and the resulting responses are averaged from the time of the stimuli. The random noise averages to zero, but if the Brainstem Response signal will be easily determined in the averaged signal.
The Bio-logic EP System can be used for patients of all ages, from children infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
The feature modifications represented in this Special 510(k) are for the use of a redesigned hardware package and a Windowsbased software program to control the hardware. The functions and electronic design of the hardware are substantially the same as those of the Predicate Device hardware. The software for control of this device is a simplified subset of the DOS-based Predicate Device software. Together, they implement the same infant hearing screening functions and perform the same intended use as the Predicate Device, but with improved ease-of-use. The POVR algorithm described in the Predicate Device is implemented in the new Windows software to generally assist in test data interpretation, and specifically assist in the assessment of signal-to-noise ratio and the quality of the Brainstem Auditory Evoked Response in infants. Based on this automatic assessment, the speed of testing may be reduced and/or the quality of the data recording may be improved, without compromising the quality of recorded data or limiting the control and flexibility of the health care professional administering the test.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kenne "J. Hay.. 4521
Division Sign-Off) Division of General Restorative 510(k) Number _
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.