MEDISISS REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS

K030598 · Medisiss · NLQ · Jan 23, 2004 · SU

Device Facts

Record IDK030598
Device NameMEDISISS REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS
ApplicantMedisiss
Product CodeNLQ · SU
Decision DateJan 23, 2004
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Pediatric

Intended Use

The SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments are intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel instruments. The instruments are used in general, pediatric, thoracic, gynecologic, urologic, and other open and endoscopic surgeries for the transection, dissection, and coagulation of tissue(s).

Device Story

Reprocessed ultrasonic surgical instruments; hand-manipulated devices with metal shaft and distal end-effector (scalpel configurations: flat, cutting, blunt). Operated via handpiece handles (suppress/release or button/switch activation); may include 360° rotation. Used in open or endoscopic surgery (via trocar/cannula) by surgeons. Device transforms electrical energy into ultrasonic mechanical energy at end-effector to transect and coagulate soft tissue. Reprocessing includes cleaning, sterilization, and 100% visual/mechanical testing to ensure functional equivalence to original single-use devices. Benefits include bleeding control and minimal thermal injury during tissue manipulation.

Clinical Evidence

Bench testing only. Evidence includes validation of cleaning, sterilization (ANSI/AAMI/ISO 11135-1994, ANSI/AAMI/ISO 10993-7:1995), and packaging. Mechanical testing performed to confirm functional performance post-reprocessing. No clinical data provided.

Technological Characteristics

Ultrasonic surgical instrument; metal shaft/barrel; hand-actuated end-effector. Energy source: ultrasonic (via handpiece). Connectivity: none. Sterilization: Ethylene Oxide (EO). Materials: identical to predicate. Testing: 100% visual and mechanical inspection.

Indications for Use

Indicated for soft tissue incision, transection, dissection, and coagulation in general, pediatric, thoracic, gynecologic, urologic, and other open and endoscopic surgeries where bleeding control and minimal thermal injury are required.

Regulatory Classification

Identification

Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code “LFL”). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird-like shape, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 1 2004 NOV Mike Sammon, Ph.D. Senior Director of Research and Development MediSISS. Inc. 2747 SW 6th Street Redmond, Oregon 97756 K030598 - Supplemental Validation Submission Re: Trade/Device Name: Ethicon, CS-14C, LCS-C5 (see enclosed list) Regulation Name: Ultrasonic Surgical Instruments Regulatory Class: Unclassified Product Code: NLQ Dated: December 30, 2003 Received: December 31, 2003 Dear Dr. Sammon: The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on January 23, 2004. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Mike Sammon, Ph.D. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Stupt Rhodes Colin M. Wittmer, M.D., Ph.D. fa . Witten, M.D., Ph.D. ia N Director Division of General, Restorative, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Reprocessed Ultrasonic Surgical Instruments found to be Substantially Equivalent: . . ::: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Ethicon, CS-14C Ethicon, LCS-C5 ( L. ా {3}------------------------------------------------ AND / OR Prescription Use Use (Part 21 CFR 801 Subpart D) Over-The-Counter (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED ) Concurrence of CDRH, Office of Device Evaluation (ODE) Stist ludu (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 1 of 1 510(k) Number K030598 {4}------------------------------------------------ ## Company 510(k) # | Ethicon Endo-Surgery (Ultracision) | K980099, K993054, and K010898 | |------------------------------------|-------------------------------| | U.S.Surgical | K971861 | | Olympus Optical | K972114 | They are also similar to the Reprocessed Ultrasonic Surgical Instruments reprocessed by SterilMed Corporation and legally marketed and distributed pursuant to Reprocessed Harmonic Scalpels 510(k) K012571 and likewise, Vanguard's 510(k) 022780 Vanguard Reprocessed Ultrasonic Scalpel. K030598 {5}------------------------------------------------ 510(k): SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments Page 41 of 42 # Device Description Reprocessed Ultrasonic Surgical Instruments consist of hand-manipulated devices, with or without rotation capability, with or without cutting ability. The handpiece handles are connected to the distal end-effector by a narrow-diameter metal barrel or shaft. The distal end of the device consists of a scalpel with a variety of end configurations including Flat, Cutting, and Blunt or a combination of the same. The devices are designed to be inserted through an appropriately sized trocar sleeve or cannula or used to be used in open surgery. The end-effectors are usually operated by the handpiece handles. The handles may be designed to be suppressed and released to activate the instruments end-effector. Similarly the handle may incorporate a button or switch to activate the end-effectors. The device's shaft may be designed to (depending on the device model and type) be rotated (up to 360°) either direction (by manipulating controls located on the handle.) ## Intended Use The SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments are intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel instruments. The instruments are used in general, pediatric, thoracic, gynecologic, urologic, and other open and endoscopic surgeries for the transection, dissection, and coagulation of tissue(s). # Summary of Technological Characteristics The intended use and technological features of the reprocessed devices do not differ from the legally marketed predicate device(s). Both the reprocessed devices(s) and the predicate device(s) have the same materials and product design. There are no changes to the claims, intended use, clinical applications, patient populations, performance specifications, or methods of operation. The technological characteristics of the Reprocessed UItrasonic Surgical Instruments are the same as those of the legally marketed predicate devices. In addition the SISS d.b.a. MediSISS™ manufacturing process includes 100 % visual and mechanical testing of all products prior to packaging, labeling, and sterilization. # Summary of Performance Data The SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments comply with the following standards, practices, and guidance's: ## Sterilization Validation and EO Residuals: - ANSI/AAMI/ISO 11135-1994, Medical Devices—Validation and Routine Control of � Ethylene Oxide Sterilization SISS d.b.a. MediSISS™ {6}------------------------------------------------ Page 42 of 42 510(k): SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments - ANSI/AAMI/ISO 10993-7:1995, Biological Evaluation of Medical Devices-Part 7: Ethylene oxide sterilization residual. # Cleaning Validation: . - AAMI RDS0TIR No. 12-1994. Designing, Testing, and Labeling Reusable Medical . Devices for Reprocessing in Health Care Facilities: A guide for Device Manufactures. Association for the Advancement of Medical Instrumentation, Arlington, VA.Food and Drug Administration. 1996. - Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA . Reviewer Guidance, Office of Device Evaluation. FDA, Washington, D.C. Cleaning, sterilization, packaging validations, and visual/mechanical testing demonstrate that the devices are equivalent and continue to be safe and effective for their intended use. SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments undergo mechanical testing to demonstrate that the parts do not change in function. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device's packaging. ## Conclusion Since the SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments meet the requirements of the stated standards and embody technological characteristics identical to the predicate device, we believe the device is safe and effective and performs as well as or better than the predicate device. The SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments will be reprocessed per specifications, good manufacturing practices, and QSR (Quality System Regulations) that ensure the device is safe and effective for its intended use. In Accordance with the Federal Food, Drug, and cosmetic Act, 21 CFR Part 807, and based on the 510(k) "Substantial Equivalence" Decision Making Process Chart and the information provided in this premarket notification, SISS d.b.a. MediSISS™ concludes that the device(s)(Reprocessed Ultrasonic Instruments) are safe, effective, and substantially equivalent to the predicate devices as described herein. This conclusion is based upon the SISS d.b.a. MediSISS™ Reprocessed Ultrasonic Surgical Instruments similarities in functional design, materials, indications for use, and methods of construction to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order