FRESENIUS NATURALYTE GRANUFLO DRY ACID CONCENTRATE 10XX SERIES, 24XX SERIES, 30XX SERIES

K030497 · Fresenius Medical Care · KPO · May 20, 2003 · Gastroenterology, Urology

Device Facts

Record IDK030497
Device NameFRESENIUS NATURALYTE GRANUFLO DRY ACID CONCENTRATE 10XX SERIES, 24XX SERIES, 30XX SERIES
ApplicantFresenius Medical Care
Product CodeKPO · Gastroenterology, Urology
Decision DateMay 20, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is indicated in the treatment of acute and chronic renal failure during the hemodialysis procedure. This concentrate is formulated to be used with a three-stream hemodialysis machine which is calibrated for acid and bicarbonate concentrates.

Device Story

Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is a non-granulated powder dialysate concentrate for hemodialysis; used as a direct replacement for existing Granuflo® Series 1000, 2400, and 3000 concentrates. The device is mixed with water and bicarbonate concentrate by a three-stream hemodialysis machine to create the final dialysate solution. It is intended for single use in clinical settings for patients with acute or chronic renal failure. The concentrate maintains the same chemical composition as predicate devices, ensuring compatibility with existing hemodialysis equipment. It is supplied non-sterile and non-pyrogenic.

Clinical Evidence

Bench testing only.

Technological Characteristics

Non-granulated powder dialysate concentrate; chemical composition identical to predicate devices; supplied non-sterile and non-pyrogenic; intended for use with three-stream hemodialysis machines.

Indications for Use

Indicated for patients with acute and chronic renal failure requiring hemodialysis treatment. Used with three-stream hemodialysis machines calibrated for acid and bicarbonate concentrates.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K030497 Page 1 of 1 #### Fresenius Naturalyte® Granuflo® Dry Acid Concentrate 510(k) Premarket Notification MAY 2 0 2003 # 510K Summary In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the Fresenius Naturalyte® Granuflo® Dry Acid Concentrate. Company: Art Eilinsfeld, Director of Regulatory. Affairs Fresenius Medical Care North America 95 Hayden Ave. Lexington, MA 02420 1-800-662-1237 January 14, 2003 Date: Trade Name: Fresenius Naturalyte® Granuflo® Dry Acid Concentrate Common Name: Dialysate concentrate for hemodialysis (liquid or powder) Classification Name and Reference: 21 CFR \$876.5820 Dialysate concentrate for hemodialysis (liquid or powder) - Class II Device Product Code and Panel Code: KPO, 78 Predicate Device(s): Granulyte Powder Dialysate Concentrates and Mixer; (K911459, SE 07/17/1991); Granulyte Dialysate Concentrate; (K922005, SE 03/30/94) #### Description: The Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is designed to be used as direct product replacement for the current Granuflo@ Concentrate (Series 1000, 2400, and 3000). The new product will be available in a non-granulated formula. It is used only during hemodialysis. It is manufactured using the same raw materials. The new Fresenius Naturalyte® Granuflo® Dry Acid Concentrate has the same chemical composition as the predicate devices. It is for single use only. It is supplied non-sterile and is non-pyrogenic. #### Intended Use: The Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is indicated in the treatment of acute and chronic renal failure during the hemodialysis procedure. This concentrate is formulated to be used with a three-stream hemodialysis machine which is calibrated for acid and bicarbonate concentrates. #### Safety and Performance: The intended use, technological characteristics, design features, and materials are substantially equivalent to the predicate devices previously cleared for market. The safety and effectiveness of the Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is supported by the substantial equivalence information, materials data, device description, and performance testing. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written along the border of the circle. MAY 2 0 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Arthur Eilinsfeld Director of Regulatory Affairs Fresenius Medical Care North America 95 Hayden Avenue LEXINGTON MA 02173 Re: K030497 Trade/Device Name: Fresenius Naturalyte® Granuflo® Dry Acid Concentrate Regulatory Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO Dated: February 14, 2003 Received: February 19, 2003 Dear Mr. Eilinsfeld: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy Crogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Fresenius Medical Care. The logo consists of a geometric shape resembling a funnel or inverted pyramid, positioned above the text "Fresenius Medical Care". The geometric shape is made up of several horizontal lines that decrease in length as they descend, culminating in a pointed apex. # Indications for Use Statement Device Name: Fresenius Naturalyte® Granuflo® Dry Acid Concentrate Indications for Use: The Fresenius Naturalyte® Granuflo® Dry Acid Concentrate is indicated in the treatment of acute and chronic renal failure during the hemodialysis procedure. This concentrate is formulated to be used with a three-stream hemodialysis machine which is calibrated for acid and bicarbonate concentrates. ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use (Per 21 CFR 801.109) | <div style="display:flex; align-items:center;">✓</div> | |---------------------------------------|--------------------------------------------------------| |---------------------------------------|--------------------------------------------------------| OR Over-The-Counter Use Nancy C Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K030497 | |---------------|---------| |---------------|---------| ### Fresenius Medical Care North America Curporate Headquarters: 95 Hayden Avenue Lexington, MA 02420 (781) 402-9800
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...