K030429 · Advanced Sterilization Products · MLR · Aug 18, 2003 · General Hospital
Device Facts
Record ID
K030429
Device Name
STERRAD 200 STERILIZER
Applicant
Advanced Sterilization Products
Product Code
MLR · General Hospital
Decision Date
Aug 18, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Indications for Use
The STERRAD® 200 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cvcle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. The STERRAD® 200 Sterilize can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with only a single stainless steel lumen with the following dimensions can be processed in the STERRAD® 200 Sterilizer. - An inside diameter of 1 mm or larger and a length of 125 mm or shorter* ● - An inside diameter of 2 mm or larger and a length of 250 mm or shorter* ● - An inside diameter of 3 mm or larger and a length of 400 mm or shorter . Teflon® and polyethylene lumen instruments with inside diameters of 6 mm or larger and lengths of 310 mm or shorter can be processed in the STERRAD® 200 Sterilizer. *The validation testing for this lumen size was conducted using a maximum of 12 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. Note: The validation studies for this sterilizer were performed using the STERRAD cycle consisting of a 6-minute 45-second injection phase followed by a 2-minute diffusion phase and a 2-minute plasma phase. The validation studies were performed using a validation load consisting of four instrument trays each weighing 9.12 lbs. (the total weight of the load was 36.48 lbs).
Device Story
STERRAD 200 Sterilizer is a stand-alone system for low-temperature sterilization of heat/moisture-sensitive medical instruments. Process: aqueous hydrogen peroxide injected into vaporizer; transformed into vapor; introduced into process chamber under negative pressure; converted to gas plasma via electrical energy. System includes sterilization chamber, carriage, disposable sterilant cassettes, and reusable trays. Operated in clinical settings to ensure instrument sterility. Hardware is identical to predicate STERRAD 100S. Benefits include effective sterilization of complex, diffusion-restricted devices without heat damage.
Clinical Evidence
Bench testing only. Validation studies utilized the "overkill" approach with B. stearothermophilus spores. Testing confirmed efficacy for various lumen sizes (1x125mm, 2x250mm, 3x400mm, 6x310mm), surface sterilization, mated surfaces, and pouched configurations. Additional testing included bacteriostasis/fungistasis, simulated use, toxicity, and chemical indicator functionality. All tests passed.
Technological Characteristics
Hydrogen peroxide gas plasma sterilization system. Operates under negative pressure with electrical energy for plasma generation. Hardware includes sterilization chamber, carriage, and disposable cassettes. Compatible with stainless steel, Teflon, and polyethylene materials. Stand-alone system.
Indications for Use
Indicated for sterilization of heat- and moisture-sensitive metal and nonmetal medical devices, including those with diffusion-restricted spaces (e.g., hinged instruments) and specific stainless steel, Teflon, or polyethylene lumens.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
Predicate Devices
STERRAD® 100S Sterilization System
Submission Summary (Full Text)
{0}------------------------------------------------
# AUG 1 8 2003 510(k) Summary
## K030429
## Applicant's Name, Address, Telephone, FAX, Contact Person
Advanced Sterilization Products Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618
## Contact Person
Kevin Corrigan Director of Regulatory Affairs Tel: (949) 453-6410 Fax: (949) 789-3900
August 12, 2003
#### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME 1.0
Classification Name: Sterilizer, Class II Common/Usual Name: Hydrogen Peroxide Gas Plasma Sterilization System Product Classification: Sterilizer, Class II Proprietary Name: STERRAD® 200 Sterilizer
#### 2.0 PREDICATE DEVICES
STERRAD® 100S Sterilization System
#### INDICATIONS FOR USE 3.0
The STERRAD® 200 Sterilizer is designed for sterilization of both metal and nonnetal medical devices at low temperatures. Because the cvcle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.
The STERRAD® 200 Sterilize can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with only a single stainless steel lumen with the following dimensions can be processed in the STERRAD® 200 Sterilizer.
- An inside diameter of 1 mm or larger and a length of 125 mm or shorter* ●
- An inside diameter of 2 mm or larger and a length of 250 mm or shorter* ●
- An inside diameter of 3 mm or larger and a length of 400 mm or shorter .
{1}------------------------------------------------
Teflon® and polyethylene lumen instruments with inside diameters of 6 mm or larger and lengths of 310 mm or shorter can be processed in the STERRAD® 200 Sterilizer.
*The validation testing for this lumen size was conducted using a maximum of 12 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.
Note: The validation studies for this sterilizer were performed using the STERRAD cycle consisting of a 6-minute 45-second injection phase followed by a 2-minute diffusion phase and a 2-minute plasma phase. The validation studies were performed using a validation load consisting of four instrument trays each weighing 9.12 lbs. (the total weight of the load was 36.48 lbs).
#### 4.0 DESCRIPTION OF DEVICE
The STERRAD® 200 Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices, using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer bowl where the solution is heated and transformed into a vapor, introducing the vapor into the process chamber under negative pressure and transforming the vapor into a gas plasma with electrical energy.
The equipment (hardware) for the STERRAD® 200 Sterilizer is the same as that of the predicate device, the STERRAD® 100S Sterilizer. (Note: The STERRAD® 200 Sterilizer is available in either a one-door or two-door model, depending on customer preference). The hardware consists of a sterilization chamber onto which is mounted a variety of instruments and components, housed in a covered frame. The system also uses accessories such as a movable carriage for transport of the cart, disposable sterilant cassettes, reusable instrument trays, printer paper and ink cartridges.
#### 5.0 SUMMARY OF NONCLINICAL TESTS
#### Validation Testing 5.1
Testing was performed using the "overkill" approach utilizing B. stearothermophilus spores. Table one on the following page identifies the validation studies performed and the results obtained.
{2}------------------------------------------------
| | | Table 1: Validation Studies | |
|--|--|-----------------------------|--|
|--|--|-----------------------------|--|
| Study | Results |
|----------------------------------------------------------|---------|
| Dose Response with 1 x 125mm Lumens | Passed |
| Dose Response with 2 x 250mm Lumens | Passed |
| Dose Response with 3 x 400mm Lumens | Passed |
| Surface Sterilization | Passed |
| Mated Surface Sterilization | Passed |
| 1 x 125mm Lumen Validation | Passed |
| 2 x 250mm Lumen Validation | Passed |
| 3 x 400mm Lumen Validation | Passed |
| 6 x 310mm Lumen Validation | Passed |
| Pouched 1 x 125mm Lumen Validation | Passed |
| Pouched 2 x 250mm Lumen Validation | Passed |
| Pouched 3 x 400mm Lumen Validation | Passed |
| Bacteriostasis Testing of Carrier Materials | Passed |
| Sporicidal Testing | Passed |
| In Use Testing | Passed |
| Bacteriostasis/Fungistasis Testing | Passed |
| Simulated Use Testing | Passed |
| Toxicity Testing of Materials | Passed |
| Chemical Indicator Functionality | Passed |
| Bacteriostasis Testing of CycleSure Biological Indicator | Passed |
#### 6.0 OVERALL PERFORMANCE CONCLUSIONS
The nonclinical studies demonstrate that the STERRAD® 200 Sterilizer is safe and effective for sterilization of medical devices within the indications for use for the sterilizer and establish equivalence of the STERRAD® 200 Sterilizer to the predicate device, the STERRAD® 100S Sterilizer.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a black and white seal. The seal is circular and contains text around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings and a wavy line representing its body. The text around the perimeter is not clearly legible.
Public Health Service
AUG 1 8 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Natalie Bennington Senior Regulatory Affairs Specialist Advanced Sterilization Products 33 Technology Drive Irvine, California 92618
Re: K030429
Trade/Device Name: STERRAD® 200 Sterilizer Regulation Number: 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: II Product Code: FLF, MRL Dated: May 21, 2003 Received: May 22, 2003
Dear Ms. Bennington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Bennington
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Patina Currente ffor
Susan Runner, DDS, Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
ADVANCED STERILIZATION PRODUCTS®
*Johnson-Johnson company*
REGULATORY AFFAIRS DEPARTMENT
## 510(k) Number (if known): K030429
STERRAD® 200 Sterilizer Device Name:
## Indications-For-Use:
Chio S. Lim
n of Anesthesiology, General Hospital. on Control Dental Davices
510(k) Number: K030429
The STERRAD® 200 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.
The STERRAD® 200 Sterilizer can sterilize instruments which have diffusion-restricted spaces. such as the hinged portion of forceps and scissors.
Medical devices with only a single stainless steel lumen with the following dimensions can be processed in the STERRAD® 200 Sterilizer.
- An inside diameter of 1 mm or larger and a length of 125 mm or shorter* .
- An inside diameter of 2 mm or larger and a length of 250 mm or shorter* .
- An inside diameter of 3 mm or larger and a length of 400 mm or shorter .
Teflon® and polyethylene luments with inside diameters of 6 mm or larger and lengths of 310 mm or shorter can be processed in the STERRAD® 200 Sterilizer.
*The validation testing for this lumen size was conducted using a maximum of 12 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.
Note: The validation studies for this sterilizer were performed using the STERRAD cycle consisting of a 6-minute 45-second injection phase followed by a 2-minute diffusion phase and a 2-minute plasma phase. The validation studies were performed using a validation load consisting of four instrument trays each weighing 9.12 lbs. (the total weight of the load was 36.48 lbs.).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (PER 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
DIVISION OF ETHICON, INC. . 33 TECHNOLOGY DRIVE . IRVINE, CA 92618 . (949) 581-5799 FAX (949) 789-3988
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.