K030086 · Biomet, Inc. · LOD · Dec 17, 2003 · Orthopedic
Device Facts
Record ID
K030086
Device Name
PALACOS R BONE CEMENT WITH GENTAMICIN
Applicant
Biomet, Inc.
Product Code
LOD · Orthopedic
Decision Date
Dec 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Palacos® G with gentamicin is indicated for use as bone cement in arthroplasty procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of revision of previous arthroplasty procedures due to joint infection. The cement is intended for use to affix a new prosthesis in the second phase of a two-stage revision after the initial infection has been cleared.
Device Story
Palacos® G is a two-component, fast-setting polymethylmethacrylate (PMMA) bone cement containing gentamicin sulfate. Components consist of a powder (PMMA copolymers, zirconium oxide, benzoyl peroxide, chlorophyll, gentamicin) and a liquid (methyl methacrylate monomer, N,N-dimethyl-p-toluidine, chlorophyll). Mixing initiates polymerization; accelerator (N,N-dimethyl-p-toluidine) speeds free radical generation; stabilizer prevents premature polymerization. Resulting paste anchors metal/plastic prostheses to bone, transferring mechanical stresses across the interface. Used in OR by surgeons during joint arthroplasty. Gentamicin provides broad-spectrum antibiotic properties to aid in revision settings. Hardens within 15 minutes. Benefits include stable prosthetic fixation and localized antibiotic delivery in revision arthroplasty.
Clinical Evidence
No clinical data. Substantial equivalence supported by comparative bench testing against Palacos® R and established clinical history of gentamicin-impregnated bone cement in Europe.
Indicated for patients undergoing hip, knee, or other joint arthroplasty requiring fixation of prosthetic components to living bone, specifically for two-stage revision procedures following joint infection.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{1}------------------------------------------------
K030086 page 17
# DEC 17 2003
## 510(k) Summary of Safety and Effectiveness Palacos® G Bone Cement with Gentamicin
Submitter's Name and Address: Biomet Inc., P.O. Box 587 S6 East Bell Drive Warsaw, IN 46581
Contact Person: Lonnie Witham Biomet Inc. P.O. Box 587 Warsaw, IN 46581
Name of the Device: Palacos® G Bone Cement (with Gentamicin)
Legally marketed device to which the submitter claims substantial equivalence: Legally marketod de needs to m. FDA PMA No. P810020 (1984) - Subsequently reclassified into class II.
### Description Of Palacos® G (with Gentamicin)
Description O1 1 anacos® O (With Gentamicin)
Palacos® G (with gentamicin) is a fast setting polymer (polymethylmethacy from don Fillacosw O (with gentamism) is a the two sterile components, consisting of a powder and a liquid, initially produces a paste, which is used to anchor the prosthesis or to fill an and a inquiry intrially produced bone cement allows stable fixation of the prosthesis and transfers stresses produced on movement to the bone via the large interface. Insoluble translers succioi in included in the cement powder as an x-ray contrast medium. The chlorophyll additive serves as optical marking of the bone cement at the site of the omorophyn Batalavamicin component is a broad-spectrum antibiotic.
#### Material used:
#### 40.8 grams powder
| Gentamicin sulfate (equivalent to 0.5 grams Gentamicin) | 0.835 grams |
|---------------------------------------------------------|-----------------------|
| Methyl acrylate-methyl methacrylate copolymer (6:94) | 27.77 grams |
| Methyl acrylate-methyl methacrylate copolymer (42:58) | 5.68 grams |
| Zirconium (IV) oxide (mono-clinic) | 6.13 grams |
| Benzoyl peroxide | 0.315 grams |
| Chlorophyll | 0.008 grams (200 ppm) |
Methylmethacrylate copolymer is the primary constituent of the powder component. Mediaci yiate coporymer is alle pacifier. Chlorophyll is added as a colorant to distinguish polymer from bone at the site of operation. Benzoylperoxide is a starter. All ulstinginsit poryiner from arment. Gentamicin® sulfate is an antibiotic that has been used in bone cement in Europe for approximately 20 years.
{2}------------------------------------------------
K030086 page 72
20 ml liguid Methyl methacrylate N, N-Dimethyl-p-toluidine Chlorophyll
18.4 grams 0.38 grams 0.005 grams (200 ppm)
Methymethacrylate monomer is the primary constituent of the liquid component. In much Methyliedlaci yiate monomer 16 the primary over the stablidine, and the stabilizer, smaller qualifics are the accelerative in the stituents of PMMA bone cement.
## Scientific concepts, significant physical and performance characteristics:
Scientific concepts, significant puyscal and pixed, the accelerator speeds the generation When the powder and liquid components and many of the early free radicals,
of free radicals and the stabilizer in the liquid reacts with many of the early free radicals, of tree radicals and the Stabilized in the rigard reason formation of polymer chains.
but is soon consumed. Free radicals can then initiate formation of polymer chains.
Polymerization proceeds slowly over the first few minutes. Polymer chains at the surface Polymerizanon proceeds stowly over the fire rowly formed polymer chans, while
of the powder beads mingle with monomer and newly researce rises as set time of of the powder beads millgre with modiomer and now for rises as set time of the smaller beads may dissolve completely. The concern valiped and the bone cement hard within 15 minutes.
Palacos® G with gentamicin is made of the same materials as the approved bone cements Palacos® G with gemaintient is made of the same to be compatible and have a long
Palacos® R (P 810020). These materials have shown to be compatible and have a for Palacos® K (F 810020). These materials no retamicin has been used in Europe for approximately 20 years. Since Palacos® G (with gentamicin) is the same cement as approximately 20 years. Shice Falacose & (vines girstituents added nor removed),
Palacos® R approved in P810020 (1984) (no chemical vecco in Europe, no Palacos® R approved in P810020 (1964) (1964) (1984) (1984) 1984 119 11999) 1999 1999, 10 toxicological studies have been conducted.
### Statement of intended use of the device:
Palacos® G with gentamicin is indicated for use as bone cement in arthroplasty Palacosw of the hip, knee and other joints to fix plastic and metal prosthetic parts to procedures of the hip, Kice and outler joints to inx previous of previous arthroplasty living bone when reconstruction. The coment is intended for the new prositesis
procedures due to joint infection. The coment is intense has been cleared procedures and to John firection. The center is the initial infection has been cleared
## Summary of the technological characteristics of the new device in comparison to those of the predicate device:
those of the predicate device.
The components of Palacos® G (with gentamicin) are identical to the difference is the I he components of Falacos® O (with genearsed and sterilized. The only difference is the device Palacos® K and are identically proveder component. Palacos® R impregnated with gentamicin has a long clinical history of use in Europe.
{3}------------------------------------------------
K030086 page 373
The effectiveness and substantial equivalence of Palacos® G (with gentamicin) was The Creenveness and substantial equivparative testing to Palacos® R and by comparing the relevant data. The results showed that Palacos® G (with gentamicin) possesses mechanical and chemical characteristics to fulfill its intended use.
In summary, Palacos® G with gentamicin is safe and effective for use in the abovementioned indications. Palacos® G is substantially equivalent to Palacos® R for its primary intended use of fixation of prosthetic components.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 7 2003
Mr. Lonnie Witham Biomet, Inc. P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K030086
Trade/Device Name: Palacos® G Bone Cement with Gentamicin Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: LOD and MBB Dated: September 19, 2003 Received: September 22, 2003
Dear Mr. Witham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Mr. Lonnic Witham
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), plcase contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Mark A. Melkersen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### STATEMENT OF INDICATIONS FOR USE
510(k) Number
Device Name: Palacos® G Bone Cement with Gentamicin
#### Indications for Use:
Palacos® G with gentamicin is indicated for use as bone cement in arthroplasty procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of revision of previous arthroplasty procedures due to joint infection. The cement is intended for use to affix a new prosthesis in the second phase of a two-stage revision after the initial infection has been cleared.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use No
(Optional Format 1-2-96)
Mark A. Mellers
morel, Restorative
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.