The MIIG™ paste is intended to be injected into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremittes, spine, and pelvis) and to cure in-situ. These open bone voids may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The MIIGM paste provides a bone void filler that resorbs and is replaced with bone during the healing process. The MIIG™ paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g., K-Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. The MIIG™M II is provided sterile for single use only.
Device Story
MIIG II is a bone void filler consisting of surgical grade calcium sulfate hemihydrate powder and a sterile water mixing solution. Components are mixed into a paste and injected into non-structural osseous defects or gaps in the extremities, spine, or pelvis. The material cures in-situ to provide temporary intra-operative support for bone fragments, often augmenting provisional hardware like K-wires. The paste is resorbable and replaced by bone during healing. Intended for single-patient use in a surgical setting by a physician. It serves as a temporary support medium and does not provide permanent structural support.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by materials data and bench testing results.
Technological Characteristics
Material: Surgical grade calcium sulfate hemihydrate. Form: Powder and liquid mixing system. Principle: In-situ curing bone void filler. Sterilization: Provided sterile for single use.
Indications for Use
Indicated for patients with open bone voids or gaps in the skeletal system (extremities, spine, pelvis) not intrinsic to structural stability, including surgically created defects or traumatic osseous injuries. Not for structural support.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
K024336
Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" formed by three black triangles, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. A registered trademark symbol is located at the bottom left of the "W" symbol. A horizontal line is located underneath the text.
MAR 0 4 2003
A Wright Medical Group Company
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the MIIG™ II.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|----------------------------------------------------------------------------------------|
| Date: | February 22, 2001 |
| Contact Person: | Ehab M. Esmail<br>Manager, Regulatory Affairs<br>Phone: 901-867-4732 Fax: 901-867 4630 |
| Proprietary Name: | MIIG <sup>TM</sup> II |
| Common Name: | Bone Void Filler |
| Classification Name and Reference: | Unclassified |
| Device Product Code and Panel Code: | Orthopedics/87/MOV |
### DEVICE INFORMATION
#### INTENDED USES/ INDICATIONS A.
The MIIG™ paste is intended to be injected into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremittes, spine, and pelvis) and to cure in-situ. These open bone voids may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The MIIGM paste provides a bone void filler that resorbs and is replaced with bone during the healing process.
The MIIG™ paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g., K-Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.
The MIIG™M II is provided sterile for single use only.
Image /page/0/Picture/12 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the logo is the text "A Wright Medical Group Company".
Image /page/0/Picture/13 description: The image shows the text '510(K) SUMMARY.doc'. The text is in a serif font and is black. The image appears to be a document or file name.
Image /page/0/Picture/14 description: The image shows a black and white drawing of a globe. The globe is partially shaded, with the darker areas likely representing landmasses or geographical features. A curved line is visible above the globe, possibly indicating an atmospheric layer or a celestial arc. The bottom of the globe is cut off by the edge of the image.
{1}------------------------------------------------
#### B. DEVICE DESCRIPTION
The MIIGTM II paste consists of pre-measured surgical grade calcium sulfate, premeasured mixing solution, and the tools necessary to the components into a paste and inject the material into the defect site. These products are provided sterile for single patient use. When mixed and injected according to directions. MIIG™ II paste will harden in situ and provide temporary intra-operative support to a site prior to the fixation of final hardware. The MIIGTM II product has been developed to address the identical indications as the MIIGTM product.
#### C. MATERIALS
The materials used for the MIIG™ II are substantially equivalent to the previously submitted and cleared MIIGTM.
| | MIIGTM | MIIGTM II |
|--------|-------------------------------------------------------------------------|--------------------------------------------------------|
| POWDER | MIIGTM Powder | MIIGTM II Powder |
| | Surgical Grade Calcium Sulfate Hemihydrate Accelerator Setting Additive | Surgical Grade Calcium Sulfate Hemihydrate Accelerator |
| LIQUID | Diluent | Diluent |
| | Saline Irrigation | Sterile Water for Irrigation |
#### SUBSTANTIAL EQUIVALENCE INFORMATION D.
The intended use, material composition, and design features of the MIIGTM II are substantially equivalent to the intended use, material composition, and design features of the previously submitted and cleared MIIG.
The safety and effectiveness of the MIIGTM II is adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification.
Image /page/1/Picture/9 description: The image is a logo for Wright Medical Technology. The logo features a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below a horizontal line, the text "A Wright Medical Group Company" is present.
Image /page/1/Picture/10 description: The image shows a black and white photograph of what appears to be a cell or a planet. The object is circular with a dark upper right portion and a lighter, textured area in the lower left. There is a small, dark object near the center of the lighter area, and a small, dark line at the bottom of the image.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 0 4 2003
Mr. Ehab M. Esmail Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002
Re: K024336
Trade/Device Name: MIIG™ II Regulatory Class: Unclassified Product Code: MQV Dated: February 11, 2003 Received: February 12, 2003
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Ehab M. Esmail
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" on the left, followed by the words "WRIGHT MEDICAL" stacked on top of "TECHNOLOGY" on the right. The "W" is made up of three thick, black, angled lines, and the text is in a simple, sans-serif font.
A Wright Medical Group Company
## MIIGTM II
### INDICATIONS STATEMENT
The MIIG™ paste is intended to be injected into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities, spine, and pelvis) and to cure in-situ. These open bone voids may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The MIIG™ paste provides a bone void filler that resorbs and is replaced with bone during the healing process.
The MIIGM paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g., K-Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.
The MIIGTM is provided sterile for single use only.
R. Mark McMillenn
Division Sign-Off
Sign-Off Division of Gene I, Restorative and Nearological Devices
510(k) Number K024336
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General Restorative Devices 510(k) Number
Prescription Use V (Per21 CFR 801.109) OR
Over-The Counter Use (Optional Format 1-2-96)
Image /page/4/Picture/16 description: The image shows the logo for Wright Medical Technology. The logo consists of a large, stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the logo is the text "A Wright Medical Group Company."
Image /page/4/Picture/17 description: The image shows a black and white drawing of a celestial body, possibly a planet or moon. The object has a textured surface with varying shades of gray and black, suggesting craters or other geological features. A bright, curved line is visible at the top, possibly indicating an atmosphere or ring system. The overall impression is that of a distant, somewhat desolate world.
INDICATIONS STATEMENT.doc
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.