The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders. This product, MASTER, like its predicate device, the Navigator Pro, is a diagnostic device intended to be used as part of a set of audiometric test protocols. It is especially indicated for use in defining the configuration of the hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. It allows for the estimation of behavioral hearing threshold at various frequencies, through the use of ABR (Auditory Brainstem Response) or ASSR (Auditory Steady-State Response) test protocols. It is designed to be used as a diagnostic test procedure by individuals who are trained in the performance and interpretation of evoked potentials such as audiologists and physicians. The results of the test will be used by trained hearing health care professionals to make recommendations regarding appropriate intervention strategies. Bio-logic EP Systems can be used for patients of all ages, from children to adults, including infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
Device Story
Diagnostic system for recording/analyzing human physiological data for neurological/auditory diagnosis. Inputs: scalp electrode signals (evoked potentials). Operation: uses Navigator Pro hardware (1-channel) with MASTER software; generates auditory stimuli (continuous frequency/amplitude modulated tones for ASSR; tone bursts for ABR). Software processes signals to determine hearing thresholds; avoids 'spectral splatter' via ASSR. Used in clinics by audiologists/physicians. Output: visual display of response spectrum/thresholds. Clinical decision-making: assists in defining hearing loss configuration; informs intervention strategies. Benefits: provides objective hearing assessment for patients unable to provide reliable behavioral responses.
Clinical Evidence
No clinical data provided. Safety and effectiveness established via bench testing, hazard/risk analysis (Fault Tree Analysis per EN-1441), and adherence to ISO-9001, EN-46001, and FDA QSR design controls. Software modifications validated to ensure patient safety (e.g., double fault conditions for acoustic stimuli).
Technological Characteristics
Hardware: Navigator Pro (1-channel) with DSP. Stimuli: continuous frequency/amplitude modulated tones (ASSR) or tone bursts (ABR). Connectivity: Windows-based host computer. Software: MASTER application. Patient connection: scalp electrodes. Safety: medical-grade power supply, standard patient isolation circuits.
Indications for Use
Indicated for patients of all ages, including infants, children, and adults (including geriatric and cognitively impaired/uncooperative individuals) requiring auditory diagnostic testing to estimate behavioral hearing thresholds when traditional behavioral audiometry is unreliable.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Navigator and Traveler Evoked Potential Product (K930328)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/24 description: The image shows the logo for Bio-logic Systems Corp. The logo features the word "Bio-logic" in a bold, sans-serif font, with the word "SYSTEMS CORP" in a smaller font underneath. The logo is primarily black and white, with the text in white against a black background. There is a registered trademark symbol to the right of the word "Bio-logic".
1 2002 JUL
Koz/895
One Bio-logic Piaza Mundelein, IL 60060-3700 1-800-323-8326 Fax: 847-949-8615 www.bio-logic.com
# SECTION 2: SUMMARY AND CERTIFICATION
# 510(K) SUMMARY
Safety and effectiveness information concerning the MASTER device modification to the Biologic Evoked Potential product family is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
| PREPARED BY: | Bio-logic Systems Corp<br>One Bio-logic Plaza<br>Mundelein, IL 60060 |
|--------------|----------------------------------------------------------------------|
|--------------|----------------------------------------------------------------------|
TELEPHONE: (847)-949-5200
CONTACT PERSON: Norman E. Brunner
DATE ON WHICH THE SUMMARY WAS PREPARED: June 4, 2002
NAME OF DEVICE: Bio-logic MASTER Evoked Response System.
COMMON NAME: Evoked Response System.
CLASSIFICATION NAME: Evoked Response Auditory Stimulator (per CFR 882-1900).
PREDICATE DEVICE: Navigator Pro Evoked Potential device, reference 510(k) #K994149.
## DESCRIPTION OF THE DEVICE:
The Bio-logic Evoked Potential family of products is intended to be used for the recording and analysis of human physiological data for the purpose of neurological diagnosis and treatment of sensory disorders. The predicate device referenced above is the latest in a series of diagnostic systems of this type marketed by Bio-logic. Other related diagnostic devices in the Evoked Potential family include:
- 510(k) #K803226 Bio-logic Evoked Response Stimulators. 1.
- 510(k) #K842543 Bio-logic Evoked Potential System. 2.
- 510(k) #K844992 Bio-logic Portable Evoked Response System. 3.
- 510(k) #K862690 -- Bio-logic Traveler LT System. 4.
- 5. 510(k) #K930328 - Navigator and Traveler Evoked Potential Product.
{1}------------------------------------------------
The predicate device, the Bio-logic Evoked Potential system with Navigator Pro hardware, performs Evoked Potential recording and analysis functions, including up to 2 channels of data recording and numerous diagnostic protocols and modalities. This new MASTER Evoked Potential with Navigator Pro device performs many of these same functions in essentially the same ways with a one-channel version of the same hardware, but employs a new software application package with significant new capabilities over those of the predicate device. Through the use of the Auditory Steady State Response (ASSR) modality, which is a variant of the MLR / 40 Hz test modality used in the referenced systems already on the market, hearing threshold levels can quickly and reliably be determined based on physiological rather than behavioral means.
The Auditory Steady State Response test is an alternative to tone burst Auditory Brainstem Response (ABR) testing which is used to predict frequency-specific behavioral hearing thresholds particularly for patients who cannot provide a reliable behavioral response. The ASSR technique avoids one of the inherent pitfalls of the tone burst ABR technique which is the problem of "spectral splatter" (distortions in frequency-specific data due to the start/stop actions of the tone burst) of the short duration acoustic stimulus that is required for ABR measurement. The ASSR technique uses a continuous frequency and/or amplitude modulated tone as the stimulus and can combine several stimuli together simultaneously to assess responses to various frequencies all at the same time. The evoked response recorded from scalp electrodes is reflective of the frequency of the modulation envelope of the stimulus. Thus, assessment of the response spectrum can yield information about the presence of the response to stimuli of varving intensities in order to determine the response threshold. ASSR thresholds have a predictable relationship to behavioral thresholds that is dependent on stimulus frequency and the presence and degree of hearing loss.
The MASTER application software is a modification of software originally designed by and exclusively licensed from researchers at the Rotman Research Institute. Baverest Centre for Geriatric Care, at The University of Toronto. This software was designed originally for research purposes, and Bio-logic made significant modifications to create double fault conditions for any situation that could impact patient safety (e.g. presentation of prolonged, intense acoustic stimuli), to block the user from making parameter changes that are known to result in poor quality data, and to improve overall reliability, performance and ease of use. Additionally, the user interface has been simplified in terms of operation and information display so that only the relevant operations and data are available. More complex functions that do not have clinical relevance were eliminated from the software.
This system will be used to assist in defining the configuration of the hearing loss particularly in populations that are difficult to test using traditional behavioral audiometry. It is designed to be used as a diagnostic test procedure by individuals who are trained in the performance and interpretation of evoked potentials such as audiologists and physicians. The results of the test will be used by trained hearing health care professionals to make recommendations regarding appropriate intervention strategies.
{2}------------------------------------------------
## INTENDED USE:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders.
This product, MASTER, like its predicate device, the Navigator Pro, is a diagnostic device intended to be used as part of a set of audiometric test protocols. It is especially indicated for use in defining the configuration of the hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. It allows for the estimation of behavioral hearing threshold at various frequencies, through the use of ABR (Auditory Brainstern Response) or ASSR (Auditory Steady-State Response) test protocols. It is designed to be used as a diagnostic test procedure by individuals who are trained in the performance and interpretation of evoked potentials such as audiologists and physicians. The results of the test will be used by trained hearing health care professionals to make recommendations regarding appropriate intervention strategies.
Bio-logic EP Systems can be used for patients of all ages, from children to adults, including infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the prescription and supervision of a physician or other trained health care professional.
## SAFETY AND EFFECTIVENESS SUMMARY
To establish the safety and effectiveness of this modification to the Bio-logic Evoked Potential software, the modification was designed and incorporated into the product in accordance with the Bio-logic internal Product Development procedures which are intended to meet ISO-9001. EN-46001 and FDA QSR Design Control specifications. A detailed Hazard/Risk analysis for the EP family of products was performed using the Fault Tree analysis (FTA) approach, and a detailed Risk Assessment was written in accordance with EN-1441, the International Standard for Hazard/Risk analysis. An addendum to this Hazard/Risk file was written based on a review of this new MASTER software design.
The Navigator Pro patient-connection hardware is unchanged with this modification. A onechannel version of this hardware is used with the MASTER program. There are no newlyintroduced hardware-related methods by which the patient can be harmed or injured through the use of this device. The same patient isolation methods are used in both products, and the Navigator Pro utilizes a medical-grade power supply. Direct hardware control of all Navigator Pro functions is provided from the Digital Signal Processor (DSP) and its program code located inside the Navigator Pro package, just as it is in the predicate device. By distributing the hardware-specific functions to the DSP, the Windows-based host computer program has fewer real-time demands, resulting is high reliability and performance.
{3}------------------------------------------------
The MASTER software does not make any final decisions that result in any direct forms of diagnosis or treatment. The system is designed to be used for auditory diagnostic testing by trained health care professionals. Audiologists and physicians perform evoked potential testing and interpret the results. The results of the test are then used by trained hearing health care professionals to make recommendations regarding appropriate intervention strategies.
The following table provides a summary comparison of the technological characteristics of the new MASTER device relative to the Bio-logic Evoked Potential with Navigator Pro predicate device. This is to demonstrate that this new MASTER device has no significant differences which would adversely affect product safety and effectiveness.
| Parameter for<br>Comparison | Similarity or Difference |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | No differences. |
| Patient Population | No differences. |
| Hardware<br>Configuration | No differences. |
| Computer Control<br>Software | The software for the MASTER is a new Windows-based program<br>designed specifically for the collection of Auditory Steady State<br>Response (ASSR) data. The predicate device was DOS-based and<br>performed a variety of ABR tests, including the 40 Hz test of which<br>ASSR is a variant. |
| Navigator Pro<br>Firmware | The Digital Signal Processor (DSP) code in the Navigator Pro hardware<br>box utilizes the same Loader program as in the predicate device. The<br>program downloaded from the host computer has changes in order to<br>perform the AM/FM modulated stimulation signals for ASSR testing. |
| Patient information<br>and tracking | No significant differences. |
| Safety Characteristics | No differences. The basic patient connection methods and isolation<br>circuits are the same for both products. |
| Product Labeling | Labeling on the Navigator Pro is unchanged. |
| Presentation of Data /<br>User Interface | Information displayed on the host computer screen has been completely<br>changed in the MASTER. The user interface is specifically designed to<br>allow the user to enter test protocol information and display results in a<br>manner more intuitive and familiar to Windows users, as opposed to the<br>DOS user interface of the predicate device. Also, the overall display of<br>data is less complex than that of the predicate device. |
| Patient connections<br>(transducers and<br>electrodes) | No differences |
| Anatomical sites | No differences. |
| Physical<br>Characteristics | No differences. |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: the department's name arranged in a circular fashion on the left and a stylized symbol on the right. The symbol features three abstract human profiles facing right, stacked on top of each other, creating a sense of depth and unity. The overall design is simple, clean, and professional, reflecting the department's role in public health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### . IUIL 1 2007
Bio-logic Systems Corporation Norman E. Brunner Vice President of Research and Development One Bio-Logic Plaza Mundelein, Illinois 60060 - 3700
Re: K021895
Trade Name: Bio-logic Master Evoked Response System Regulation Number: 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: II Product Code: GWJ Dated: June 7. 2002 Received: June 10, 2002
Dear Mr. Brunner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 – Mr. Norman E. Brunner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (Misbranding by reference to premarket notification( (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 6382041 or (301) 4436597 or at its Internet address HYPERLINK http://www.fda.gov/cdrh/dsma/dsmamain.html http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Stipt Rhodes
Celia M. Witten, Ph.D., M.D. Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
## 510(k) Number (if known): #K021895
Device Name: MASTER, Modification to Bio-logic Evoked Potential Product.
### Indications For Use:
The Bio-logic Evoked Potential (EP) product family is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of auditory and hearing-related disorders.
This product, MASTER, like its predicate device, the Navigator Pro, is a diagnostic device intended to be used as part of a set of audiometric test protocols. It is especially individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. It allows for the estimation of behavioral hearing threshold at various frequencies, through the use of ABR (Auditory Brainstem Response) or ASSR (Anditory Steady-State Response) test protocols. It is designed to be used as a diagnostic test procedure by individuals who are trained in the performance and interpretation of evoked potentials such as andiologists an physicians. The results of the test will be used by trained hearing health care professionals to make recommendations regarding appropriate intervention strategies.
Bio-logic EP Systems can be used for patients of all ages, from children to adults, including infants and geriaric patients. The use of the Bio-logic EP family of products is to be performed under the prescription of a physician or other trained health care professional.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hut Rhodes
Division of General, Restorative and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.