K021699 · Autolens, Inc. · LYL · Dec 16, 2002 · Ophthalmic
Device Facts
Record ID
K021699
Device Name
AUTOLENS CLEANING AND DISINFECTING UNIT
Applicant
Autolens, Inc.
Product Code
LYL · Ophthalmic
Decision Date
Dec 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The AUTOLENS ® Automatic Contact Lens Cleaning Accessory is indicated for the cleaning of soft (hydrophilic), rigid gas permeable (RGP) and hard (PMMA) contact lenses when used with AUTOLENS® Multipurpose Solution. This automatic contact lens cleaning accessory cleans contact lenses without digital rubbing. The AUTOLENS® Automatic Contact Lens Cleaning Accessory may be used as a receptacle for chemical disinfection with AUTOLENS® Multipurpose Solution. AUTOLENS® Multipurpose Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfecting and protein removal, storing soft (hydrophilic), rigid gas permeable (fluoro-silicone acrylate and silicone acrylate) and PMMA contact lenses as recommended by your eye care practitioner.
Device Story
System comprises battery-operated tumbler accessory and multipurpose solution. Input: contact lenses placed in molded plastic holder with 'R'/'L' identifiers. Operation: 1.5V motor rotates bottle assembly at ~10 RPM; tumbling action circulates solution around lenses to effect cleaning without manual rubbing. Output: cleaned, disinfected lenses. Used by patients at home. Clinical benefit: automated cleaning process replaces manual digital rubbing; provides receptacle for chemical disinfection. Healthcare provider role: recommends solution and lens compatibility.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility (cytotoxicity, ocular irritation, systemic toxicity) performed on plastic components; solution stability (pH and temperature) measured over 4-hour soak time; compatibility testing for wetting drops performed. All results indicated safety and stability.
Technological Characteristics
Materials: Polypropylene resin plastic. Energy: 1.5V DC (3 AA batteries). Mechanism: Motorized rotation (approx. 10 RPM) of bottle assembly. Form factor: Holder, bottle, cap, and base unit. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for cleaning, rinsing, chemical disinfection, and storage of soft (hydrophilic), rigid gas permeable (RGP), and hard (PMMA) contact lenses. Used by patients as an alternative to manual digital rubbing.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
SOFT MATE®, Automatic Contact Lens Cleaning Unit (K852386)
Acu-Clens Cleaning System for Contact Lenses (K964222)
Submission Summary (Full Text)
{1}------------------------------------------------
K02/699
# AUTOLENS® Inc. 1440 Atteberry Lane San Jose, California 95131-1410 Phone: (408) 432-0580
DEC 1 6 2002
## 510(k) Summary Statement
# SUBMITTER:
| Submitted on behalf of: | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name: | AUTOLENS®, Inc. |
| Address: | 1440 Atteberry Lane<br>San Jose, California 95131-1410 |
| Phone: | (408) 432-0580 |
| Fax: | (408) 432-8252 |
| CONTACT PERSON: | Richard E. Lippman, O.D., F.A.A.O.<br>Official Representative and Correspondent<br>R.P. Chiacchierini & Associates, LLC<br>15825 Shady Grove Road<br>Suite 30<br>Rockville, Maryland 20850 |
| Phone: | (240) 683-3738 |
| FAX: | (240) 683-9236 |
| DATE SUMMARY PREPARED: | November 15, 2002 |
| TRADE NAME: | AUTOLENS® System: Comprised of the<br>AUTOLENS® Automatic Contact Lens Cleaning<br>Accessory and AUTOLENS® Multipurpose<br>Solution |
COMMON NAME: CLASSIFICATION: contact lens cleaning accessory device Class II: LYL 886 LPN 886
ESTABLISHMENT REGISTRATION: PENDING
# SUBSTANTIALLY EQUIVALENT TO:
The AUTOLENS® System, comprised of the AUTOLENS® Automatic Contact Lens Cleaning Accessory and AUTOLENS® Multipurpose Solution is substantially equivalent to the following device accessories:
| K982115 | LensServer Automatic Contact Lens Cleaning Accessory cleared |
|----------|----------------------------------------------------------------------|
| | 1/13/99 |
| K974724: | Lens Comfort Contact Lens Accessory cleared 3/2/98 |
| K852386: | "SOFT MATE®, Automatic Contact Lens Cleaning Unit" cleared<br>9/4/85 |
| K964222: | Acu-Clens Cleaning System for Contact Lenses- Cleared 12/20/1996 |
: :
{2}------------------------------------------------
Each of these accessories was cleared in conjunction with contact lens solutions for cleaning and disinfection, comprising a contact lens cleaning system. The AUTOLENS® Automatic Contact Lens Cleaning System is indicated as an aid for cleaning as an accessory for all contact lenses when used with the AUTOLENS® Brand MULTI-PURPOSE Contact Lens Solution.
All of the predicate cleaning units act by similar tumbling action or ultrasonic agitation of lenses by directing the flow of contact lens solution around the contact lenses while immersed in the solution. The accessory devices all work in conjunction with approved contact lens cleaning and disinfection solutions. The AUTOLENS® Automatic Contact Lens Cleaning System functions in an equivalent manner. The Directions for Use describe the sequential actions of the tumbling cleaning device, a rinse step, and a storage soaking step for disinfection in AUTOLENS® Multipurpose Solution.
#### DESCRIPTION of the DEVICE:
The device is a holder, cleaning and disinfection system for contact lenses. It consists of a contact lens holder, bottle, cap and base. A separate battery operated tumbler device is integral to the system. It is designed to slowly tumble the bottle assembly including contact lens solution to effect cleaning of contact lenses.
The lens holder is a single piece molded plastic device including hinged snap closure covers to contain each contact lens separately. An easily identifiable, raised letter "R" and "L" is included adjacent each closure receptacle to allow clear identification for right and left contact lenses placed within the lens holder. The lens holder contains openings to sufficiently allow the free flow of solution to circulate on all surfaces of the contact lens during the cleaning process.
The bottle is made of molded plastic and is of oval shape to facilitate the flow of solution over the lens holder and lenses during operation. The bottle is of sufficient size to contain the lens holder and 8 ml. of contact lens solution to cover the contact lenses at all times during operation and storage. The bottle is designed to hold the contact lens holder securely to ensure its rotation as the bottle spins during operation. The bottle is designed with a fluid-tight, screw closure cap. The lens holder, bottle, cap and base are made of polypropylene resin plastic.
The tumbler device is a battery powered unit operating at 1.5 volts utilizing three "AA" batteries. A plastic housing is molded to contain a small electrical motor, tumbler drive, on/off switch, batteries and a battery door to easily replace batteries. A cradle is molded within the shape of the housing to contain the assembled lens holder bottle. The drive shaft is connected to the motor inside the base housing.
When the device is turned on at the switch, the battery operated electrical motor turns the extended drive shaft. The bottle assembly resting upon the drive shaft turns in unison. The mechanical action of the bottle assembly on the drive shaft provides for a rotation of approximately 10 RPM.
4
{3}------------------------------------------------
#### INDICATIONS FOR USE:
The AUTOLENS ® Automatic Contact Lens Cleaning Accessory is indicated for the cleaning of soft (hydrophilic), rigid gas permeable (RGP) and hard (PMMA) contact lenses when used with AUTOLENS® Multipurpose Solution. This automatic contact lens cleaning accessory cleans contact lenses without digital rubbing. The AUTOLENS® Automatic Contact Lens Cleaning Accessory may be used as a receptacle for chemical disinfection with AUTOLENS® Multipurpose Solution.
AUTOLENS® Multipurpose Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfecting and protein removal, storing soft (hydrophilic), rigid gas permeable fluoro-silicone acrylate and silicone acrylate) and PMMA contact lenses as recommended by vour eve care practitioner.
#### ACTIONS:
The AUTOLENS ® System provides for cleaning of soft (hydrophilic), rigid gas permeable (RGP), or hard (PMMA) contact lenses by means of the automatic slow rotation tumbling action of the contact lenses immersed in AUTOLENS® MULTI-PURPOSE SOLUTION within a bottle container, followed by a rinsing step and soaking and disinfection step in AUTOLENS® Multipurpose Solution.
The accessory contact lens cleaning system is recommended for use with AUTOLENS ® Brand MULTI-PURPOSE SOLUTION to clean, rinse, disinfect and store soft (hydrophilic), hard (PMMA), or rigid gas permeable (RGP) contact lenses. The AUTOLENS® Brand MULTI-PURPOSE SOLUTION has been licensed from a commercially cleared product as private label manufacturer under the AUTOLENS ® Brand label. Labeling instructions and labels have been provided for the private label licensed solution.
A minimum of four hours operational use is recommended to effect the cleaning of contact lenses. The AUTOLENS ® System effectively cleans contact lenses automatically without the need for digital rubbing or manual hand cleaning.
### TOXICITY TESTING
A battery of toxicology tests were conducted on the plastic components of the system to determine the potential for problems with biocompatibility. The tests included Cytotoxicity, Ocular Irritation, and Systemic Toxicity. The results of each of the tests conducted on the device components did not indicate any response that would raise concern regarding the safety of the components for biocompatibility when tested in laboratory animals.
### EFFECTIVENESS TESTING for the AUTOLENS®
A bench testing protocol was employed to test the AUTOLENS® device with the appropriately recommended Multi-Purpose contact lens solution for a measure of solution pH and temperature gain in order to ascertain solution stability over the 4 hour recommended soak-time for cleaning. The protocol called for measuring three (3) test devices with solution immersed in the lens holder bottle. In all cases, the final pH of the test
{4}------------------------------------------------
solution in the chamber was the same as the pH at the start of the procedure concluding that no change of solution was witnessed during the operation of the device under the directions of use.
A measure of the temperature at the start and completion of the procedure was made to determine whether any heat transfer occurred to raise the temperature beyond that which would be safe for stability of the supporting cleaning and disinfection solution. The results indicate that the temperature rise was minimal.
Additional compatibility testing for recommended wetting drops and the AUTOLENS® Brand MULTI-PURPOSE Solution was conducted. The results indicate that the solutions were deemed compatible and without precipitate formation.
### CLINICAL INFORMATION
Clinical studies were deemed unnecessary as the solution used in the AUTOLENS® System is already cleared for marketing as a cleaning, rinsing, disinfection, and storage solution for soft, RGP, and hard contact lenses.
#### LABELING
The AUTOLENS® System is provided to the user with instructions for use. Additionally, the Package Insert for the AUTOLENS® Brand MULTI-PURPOSE Solution is provided. These materials are also available from the company. The address is as follows:
> AUTOLENS® Inc. 1440 Atteberry Lane San Jose, California 95131-1410 Phone: (408) 432-0580
> > 사용 - 서비스 - [사
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing the human services aspect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is black and white.
ood and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
Autolens. Inc. c/o Richard E. Lippman, O.D. R.P. Chiacchierini and Associates 15825 Shady Grove Rd., Suite 30 Rockville, MD 20850
Re: K021699
Trade/Device Name: Autolens System comprised of the Autolens Automatic Contact Lens Cleaning Accessory and Autolens Brand Multipurpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LYL: LPN Dated: November 15, 2002 Received: November 18, 2002
Dear Dr. Lippman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Richard E. Lippman, O.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
### Indications Statement
510(k) Number (if known) Device Name:
K021699 AUTOLENS® System comprised of the AUTOLENS® Automatic Contact Lens Cleaning Accessory and AUTOLENS® Multipurpose Solution
#### Indications for Use:
The AUTOLENS ® Automatic Contact Lens Cleaning Accessory is indicated for the cleaning of soft (hydrophilic), rigid gas permeable (RGP) and hard (PMMA) contact lenses when used with AUTOLENS® Multipurpose Solution. This automatic contact lens cleaning accessory cleans contact lenses without digital rubbing. The AUTOLENS® Automatic Contact Lens Cleaning Accessory may be used as a receptacle for chemical disinfection with AUTOLENS® Multipurpose Solution.
AUTOLENS® Multipurpose Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfecting and protein removal, storing soft (hydrophilic), rigid gas permeable (fluoro-silicone acrylate and silicone acrylate) and PMMA contact lenses as recommended by your eye care practitioner.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over -The-Counter Use
(Optional Format 1-2-96)
Mina Smith
Division Sign Division of Cionthalmic Ear. Nose and Throat Devises
510(k) Number K021649
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.