The New Comfort Contact Lens Care Accessory is intended for use in conjunction with contact lens solutions as an accessory in the cleaning of contact lenses. The New Comfort Contact Lens Care Accessory is indicated as aid for cleaning as an accessory for soft hydrophilic lenses and gas permeable lenses when used with the appropriate Allergan® Soft Mate® solutions such as the Consept® Cleaning and Disinfection System which is comprised of the Allergan® Soft Mate® Consept®-1 Cleaning and Disinfecting solution and Allergan® Soft Mate® Consept®-2 Neutralizing and Rinsing Solution or Spray. The New Comfort Contact Lens Care Accessory is indicated for use for gas permeable solutions such as Allergan® Gas Permeable Daily Cleaner and Allergan® ComfortCare GP Wetting and Soaking Solution. The New Comfort Contact Lens Care Accessory may be used for a receptacle for chemical disinfection.
Device Story
Device is an ultrasonic contact lens cleaning accessory. Input: contact lenses submerged in a reservoir with cleaning solution. Operation: piezo-electric crystal generates ~100 kHz ultrasonic waves, creating cavitation bubbles that implode on lens surfaces to remove deposits. Device includes housing, lens baskets, reservoir, electronics, and transducer. Powered by standard wall outlet via UL-approved transformer. User places lenses in baskets, adds solution, and initiates a 3-minute cleaning cycle via start button. LED indicator glows during operation and shuts off automatically at cycle end. Used in home setting by contact lens wearers. Benefits: provides automated mechanical cleaning aid to supplement chemical disinfection, potentially improving lens hygiene and comfort.
Clinical Evidence
Bench testing only. Protocol assessed effectiveness and impact on lenses (Group 1, 3, 4, and gas permeable). 30 cycles performed per lens type. Endpoints: physical/optical parameters (color, clarity, power, diameter, base curve) and presence of deposits/residue. Results: no changes in parameters; no deposits, residue, or damage observed.
Technological Characteristics
Materials: ABS housing/baskets, passivated/electropolished stainless steel reservoir. Principle: Piezo-electric crystal transducer (100 kHz). Energy: 110/120V AC via wall transformer. Form factor: Reservoir with lens baskets. Connectivity: None. Sterilization: Not applicable (device is a cleaning accessory).
Indications for Use
Indicated for cleaning soft hydrophilic and gas permeable contact lenses when used with specified Allergan® solutions; also serves as a receptacle for chemical disinfection. No specific age or gender restrictions.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Sola/Barnes-Hind® Hydra-Mat® (P 810017 and P 840066)
Sola/Barnes-Hind® Soft Mate® Automatic Cleaning Unit (K 852386)
Clensatron® 700 CL (K 884414)
Visonic Dome™ (K 902306)
Lensonic (K 921615)
Lens Comfort Contact Lens Care Accessory (K 962112)
Submission Summary (Full Text)
{1}------------------------------------------------
K 974724
## 510(k) SUMMARY (807.92(c))
| [807.92(a)(1)]<br>Prepared: | December 8, 1997 | | |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|--------------|
| Reference: | 510(k) Summary | | MAR - 2 1998 |
| Submission by: | James Demestichas | | |
| Entity: | Lens Comfort, Inc.<br>912 Broad Street<br>Elizabeth, New Jersey 07280<br>973-633-6726<br>973-633-7231 Fax | | |
| [807.92(a)(2)]<br>Classification<br>Name and Code: | Not Classified<br>Accessory to Contact Lens Solutions<br>Product Code (86LYL) | | |
| Common or Usual<br>Name: | Ultrasonic Contact Lens Care Accessory | | |
Trade or Proprietary
New Comfort Contact Lens Care Accessory Name:
[807.92(a){3}]
The Sola/Barnes-Hind® Hydra-Mat® (PMA Nos. P 810017 and Equivalence: P 840066); the Sola/Barnes-Hind® Soft Mate® Automatic Cleaning Unit (K 852386); the Clensatron® 700 CL (K 884414); the Visonic Dome™ (K 902306); the Lensonic (K 921615); and Lens Comfort Contact Lens Care Accessory (K 962112) all have been given market clearance to be used as an accessory for contact lens care.
[807.92(a)(4)]
Device _
The New Comfort Ultrasonic Contact Lens Care Accessory Description: uses a piezo-electric crystal to generate a wave form, in a reservoir containing a specific volume of fluid. Cavitation in fluid results in the generation of microscopic bubbles that implode upon the lens surface. The matrix of frequency, duration of ultrasound, and the control of temperature rise in the fluid medium leads to cleaning when a specific contact lens solution is present. The New Comfort device uses standard power (110 v/120 v). There is a four (4) foot cord and a UL approved wall transformer.
{2}------------------------------------------------
## Lens Comfort, Inc.
The New Comfort device consists of the housing, lens baskets, a reservoir, electronics and a transducer. All components that come in contact with contact lenses or solutions are of medical grade composition and have been used in legally marketed predicate devices in the United States. The housing is injection molded of ABS. The baskets and supporting structure are injection molded from ABS. The reservoir is deep drawn stainless steel, is passivated and electropolished. The electronic circuitry has been used in other cleaning applications in other industries for years. The single crystal is affixed directly to the bottom of the reservoir with an epoxy generally used in cleaning applications in the ultrasound industry.
## [807.92(a)(5)]
The New Comfort Contact Lens Care Accessory is intended for Intended Use: use in conjunction with contact lens solutions as an accessory in the cleaning of contact lenses.
The New Comfort Contact Lens Care Accessory is indicated as aid for cleaning as an accessory for soft hydrophilic lenses and gas permeable lenses when used with the appropriate Allergan® Soft Mate® solutions such as the Consept® Cleaning and Disinfection System which is comprised of the Allergan® Soft Mate® Consept® 1 Cleaning and Disinfecting solution and Allergan® Soft Mate® Consept®-2 Neutralizing and Rinsing Solution or Spray. The New Comfort Contact Lens Care Accessory is indicated for use for gas permeable solutions such as Allergan® Gas Permeable Daily Cleaner and Allergan® ComfortCare GP Wetting and Soaking Solution. The New Comfort Contact Lens Care Accessory may be used for a receptacle for chemical disinfection.
[807.92(a){6)] Comparison of Technological The predicate devices are: Characteristics:
| 1. | Sola/Barnes-Hind® Hydra-Mat® | P 810017 |
|----|---------------------------------------------------------|----------|
| | | P 840066 |
| 2. | Sola/Barnes-Hind® Soft Mate®<br>Automatic Cleaning Unit | K 852386 |
| 3. | Clensatron® 700 CL | K 884414 |
| 4. | Visonic Dome™ | K 902306 |
| 5. | Lensonic | K 921615 |
| 6. | Lens Comfort | K 962112 |
#### Similarities:
All predicate devices have referenced Sola/Barnes-Hind® Solutions (now Allergan) in their submissions and specifically the Soft Mate® Concept -1 and 2 Cleaning and Disinfection System for use with soft lenses. Lens Comfort and New Comfort
{3}------------------------------------------------
references the same solutions and data.
The predicate devices use medical grade materials that are in contact with the contact lenses or the solutions and have been previously used in ophthalmic medical devices legally marketed in the United States. Materials in New Comfort are the same or similar as predicate devices.
All predicate devices have a fluid reservoir in which lenses are submerged. The New Comfort device has a similar reservoir and submerges lenses for cleaning.
All predicate devices have plastic lens baskets into which lenses are encased for submersion while cleaning is in process. New Comfort has plastic lens baskets.
All predicate devices operate on standard power from a wall receptacle. New Comfort has a cord with a UL approved transformer that plugs into a wall socket.
Three predicate devices Lensonic. Visonic Dome™ and Lens Comfort use piezoelectric crystals in the generation of cavitation as does New Comfort.
## Differences:
New Comfort operates at a frequency of approximately 100 Khz and yields approximately 100,000 vibrations per second. The Hydra-Mat® is hand operated. The SoftMate® Automatic Cleaning Unit vibrates a small fraction of that of any of the predicate devices. The Clensatron® 700CL is belt driven and yields 300 cycles a minute. The Visonic Dome™ is ultrasonic and generates approximately 40,000 vibrations a second. The Lensonic device is ultrasonic and generates approximately 60,000 vibrations per second. The Lens Comfort device operates at a frequency of approximately 65 Khz.
New Comfort operates in a specific and controlled range of end temperatures. The predicate devices have end temperatures running from ambient to above 55° C. The New Comfort device end temperature is in the middle temperature range of the predicate devices.
New Comfort uses .5 to 1 watts of power per milliliter of fluid and the predicate devices range from 1 watt to 8 watts
New Comfort, Lens Comfort, Clensatron® and Lensonic® all have wall transformers while the Visonic Dome™ has no cord and plugs directly into an electrical outlet.
New Comfort, Lens Comfort, Lensonic® and Clensatron® all have a start button. New Comfort, Lens Comfort and Lensonic® have an (LED) that glows and then automatically shuts off after the end of the 10 minute cycle. The Hydra-Mat® has no time indicators. The Soft Mate® Automatic Cleaning unit has a timer but it
{4}------------------------------------------------
#### Lens Comfort, Inc.
correlates to the vibration cycle and there is no light indicator. The Clensatron® has a timer to stop the revolutions of the basket in the reservoir, but it has no light/time indicator to define for the user the end of a cycle. The Visonic Dome™ has a light that shuts off when the ultrasound stops, which coincides with the end of a 10 minute cycle. Lensonic has a light that shuts off at the end of 10 minutes.
To create cavitation, the Visonic Dome™ uses a piezo-electric crystal, affixed to the bottom of the stainless steel reservoir. Lensonic uses two piezo-electric crystals combined with a multi-part transducer affixed to a polycarbonate tank. New Comfort and Lens Comfort uses a piezo-electric crystal affixed to the bottom of a stainless steel reservoir.
The Visonic Dome™ runs ultrasound for 10 minutes. Lensonic® runs ultrasound for 3 plus minutes. New Comfort and Lens Comfort runs ultrasound for approximately 3 minutes.
## [807.92(b)(1)]
Evaluation of the New Comfort, based on a protocol used by the industry to assess the effectiveness of systems and solutions designed to clean contact lenses, was conducted by an optometrist who is director of optometric services in a large ophthalmology practice.
A protocol was obtained from the Office of Device Evaluation (ODE) that has been used by the industry to assess the effectiveness of devices and cleaning solutions and their impact, if any, on the contact lenses. Lenses were selected from Group 1, Group 3, Group 4 and the Gas Permeable class. They were soaked in an artificial tear solution. Physical and optical parameters were measured before the lenses were treated with the artificial tear solution and then after each cycle. Thirty (30) cycles were performed on all lenses.
No changes were noted in color, clarity, power, diameter or base curve. The lenses all appeared clear and no deposits or residue or damage were noted on anv of the lenses.
The Company did not have access to comparative data from the predicate devices other than Lens Comfort, but believe that the test results recorded by the investigator were sufficient to support the premise that the Company's device performs to the standards intended by the protocol.
#### [807.92(b)(2)]
Clinical data is not needed and has not been submitted with this 510(k).
{5}------------------------------------------------
# [807.92(b)(3)]
The data gathered from the cycling tests, strongly supports substantial equivalence between the New Comfort and Lens Comfort Device as well as predicate devices classified as contact lens care accessories (86 LYL). Cycling tests produced no physical or optical parameter changes after 30 cycles. There was no reported deposits or residual film observed on test lenses after the tests were concluded. There was no damage to lenses observed. The New Comfort device referencees the same solutions as all the predicate devices.
{6}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 2 1998
Mr. James Demestichas Regulatory Affairs Officer Lens Comfort, Inc. 912 North Broad St. Elizabeth, NJ 07208
K974724 Re:
> Trade Name: The New Comfort Contact Lens Care Accessory (Ultrasonic Cleaning Accessory) Regulatory Class: Unclassified Product Code: LYL Dated: December 8, 1997 Received: December 18, 1997
Dear Mr. Demestichas:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
#### Page 2 - Mr. James Demestichas
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Alberl Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________
510(K) Number (if Known) _ K 974724
Device Name: Lens Comfort Contact Lens Accessory
Indications For Use:
The New Comfort Contact Lens Care Accessory is intended for use in conjunction with contact lens solutions as an accessory in the cleaning of contact lenses.
The New Comfort Contact lens Care Accessory is indicatod as aid for cleaning as an accessory for soft hydrophilic lenses and gas pormoablo lonsos whon usod with the appropriate Allergan® Soft Mate® solutions such as the Consept® Cleaning and Disinfoction Systom which is comprisod of the Allergan® Soft Mato® Consept®-1 Cleaning and Disinfecting solution and Allergen® Soft Mate® Consept - 2 Neutralizing and Rinsing Solution or Spray. The New Comfort Contact Lens Care Accessory is indicated for use for gas permeable solutions such as Allergan® Gas Permeable Daily Cleaner and Allergan® ComfortCare GP Wetting and Soaking Solution. The New Comfort Contact Lens Care Accessory may be used for a receptacle for chemical disinfection.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E.S. Co. Ph.D. 8
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number, ... K-874724 ... . . . . . . . . . . . . . .
Prescription Use (per 21 CFR 801.109) OR
Over-The-Countor Uso _V
(Optional Format 1-2-96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.