K021507 · Spine Vision, Inc. · KWP · Aug 1, 2002 · Orthopedic
Device Facts
Record ID
K021507
Device Name
PLUS PIVOT LINK UNIVERSAL SYSTEM
Applicant
Spine Vision, Inc.
Product Code
KWP · Orthopedic
Decision Date
Aug 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as a posterior, non-pedicle system of the noncervical spine, the PLUS™ Pivot Link Universal System is indicated for: degenerative disc disease (discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis fracture spinal stenosis deformities (i.e. scoliosis, kyphosis, lordosis) tumors failed previous fusion (pseudoarthrosis). The PLUS Pivot Link Universal System is also a pedicle screw system indicated for skeletally mature patients who: have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; receive fusions using autogenous bone graft only; have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and have the device removed after the development of a solid fusion. In addition, the PLUS Pivot Link Universal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: Degenerative spondylolisthesis with objective evidence of neurologic impairment fracture dislocation scoliosis kyphosis spinal tumor failed previous fusion (pseudoarthrosis).
Device Story
PLUS Pivot Link Universal System is a spinal instrumentation set comprising hooks, screws, rods, and connectors. Surgeons assemble these components into constructs tailored to specific spinal pathologies. Used in clinical/OR settings for posterior spinal fixation and stabilization. Provides mechanical immobilization to facilitate fusion in acute/chronic instabilities or deformities. Benefits patients by stabilizing spinal segments, correcting deformities, and supporting bone graft fusion. Components are implanted by surgeons; pedicle screws are intended for removal after solid fusion development.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Components made of stainless steel (ASTM F138, F1586), titanium alloy (ASTM F136), and commercially pure titanium (ASTM F67). System consists of rods, hooks, screws, and connectors for posterior spinal fixation. Mechanical assembly.
Indications for Use
Indicated for skeletally mature patients requiring spinal stabilization/immobilization for degenerative disc disease, spondylolisthesis (including severe Grades 3-4 at L5-S1), fractures, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumors, dislocations, and pseudoarthrosis. Used in thoracic, lumbar, and sacral spine.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
AUG 0 1 2002
Image /page/0/Picture/2 description: The image shows the logo for SpineVision. The logo consists of a stylized black icon above the word "SpineVision" in a sans-serif font. Below the company name is the tagline "Spine systems for specialists" in a smaller font. The logo is simple and modern, and the tagline suggests that the company specializes in spine-related medical devices or services.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF FIRM: | SpineVision, Inc.<br>3003 Summit Blvd., Suite 1500<br>Atlanta, GA 30319<br>Phone: 404-460-5077 |
|--------------------------------------|------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Lynnette Whitaker<br>Vice President, Regulatory Affairs<br>574-269-3701 |
| TRADE NAME: | PLUS™ Pivot Link Universal System |
| COMMON NAME: | Rod, Hook, and Screw Spinal Instrumentation |
| CLASSIFICATION: | 888.3050 Spinal Interlaminal Fixation Orthosis<br>888.3070 Pedicle Screw Spinal System |
| DEVICE PRODUCT CODE: | Product code: 87 KWP, MNH, MNI |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES: | PLUS Pivot Link Spinal System, SpineVision, Inc. |
## DEVICE DESCRIPTION AND INTENDED USE:
The PLUS Pivot Link Universal System spinal instrumentation consists of hooks, screws, rods, and connectors that can be assembled in a variety of constructs. The instrumentation is combined and placed to treat a variety of spinal disorders, with the constructs varying according to the nature of the spinal pathology. The PLUS System components are available in stainless steel according to ASTMs F138 and F1586, from titanium alloy complying with ASTM-F136, and from commercially pure Titanium complying with ASTM F67.
:00004
{1}------------------------------------------------
# INDICATIONS FOR USE
When used as a posterior, non-pedicle system of the noncervical spine, the PLUS™ Pivot Link Universal System is indicated for:
- degenerative disc disease (discogenic back pain with degeneration of the disc . confirmed by history and radiographic studies)
- spondylolisthesis ●
- fracture ●
- spinal stenosis
- deformities (i.e. scoliosis, kyphosis, lordosis)
- . tumors
- failed previous fusion (pseudoarthrosis) .
The PLUS Pivot Link Universal System is also a pedicle screw system indicated for skeletally mature patients who:
- have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra;
- receive fusions using autogenous bone graft only;
- have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and .
- have the device removed after the development of a solid fusion. .
In addition, the PLUS Pivot Link Universal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic. lumbar, and sacral spine:
- . Degenerative spondylolisthesis with objective evidence of neurologic impairment
- fracture ●
- dislocation ●
- � scoliosis
- kyphosis
- spinal tumor .
- failed previous fusion (pseudoarthrosis) .
## BASIS OF SUBSTANTIAL EQUIVALENCE:
The components of the PLUS Pivot Link Universal System are identical in design, material, and intended use to other spinal instrumentation system that have been cleared by FDA for posterior spinal use. Mechanical testing was performed to demonstrate the equivalence of the construct design to currently marketed spinal systems.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Inside the circle is a stylized image of three human profiles facing to the right, with lines extending above them to create a sense of upward movement or flight. The logo is presented in black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 01 2002
Ms. Lynnette Whitaker Vice President, Regulatory Affairs SpineVision, Incorporated 3003 Summit Boulevard, Suite 1500 Atlanta, Georgia 30319
Re: K021507
Trade/Device Name: PLUSTM Pivot Link Universal System Regulation Number: 888.3050 and 888.3070 Regulation Name: Spinal Interlaminal Fixation Orthosis, Pedicle Screw Spinal System and Spondylolisthesis Spinal Fixation Device System Regulatory Class: II Product Code: KWP, MNI, and MNH Dated: May 8, 2002 Received: May 9, 2002
Dear Ms. Whitaker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Ms. Lynnette Whitaker
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark N. Mellem
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number (if known) ≤ ≥ ≤ ≤ ≤ ≤ ○ 7
#### PLUS™ Pivot Link Universal System, Additional Components Device Name:
### Indications for Use:
When used as a posterior, non-pedicle system of the noncervical spine, the PLUS™ Pivot Link Universal System is indicated for:
- degenerative disc disease (discogenic back pain with degeneration of the disc confirmed by history . and radiographic studies)
- spondylolisthesis .
- fracture ●
- spinal stenosis .
- deformities (i.e. scoliosis, kyphosis, lordosis)
- tumors
- failed previous fusion (pseudoarthrosis)
The PLUS Pivot Link Universal System is also a pedicle screw system indicated for skeletally mature patients who:
- have severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra; .
- receive fusions using autogenous bone graft only; ●
- have the device fixed or attached to the lumbar and sacral spine (L3 to sacrum); and .
- have the device removed after the development of a solid fusion. .
In addition, the PLUS Pivot Link Universal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:
- Degenerative spondylolisthesis with objective evidence of neurologic impairment .
- fracture .
- dislocation .
- scoliosis
- kyphosis
- spinal tumor
- failed previous fusion (pseudoarthrosis)
| Concurrence of CDRH. Office of Device Evaluation |
|--------------------------------------------------|
|--------------------------------------------------|
Mark N. Milliken
Division Sign-Off Division of General. Restorative and Neurological Devices
510(k) Number K021507
Prescription Use
OR (Per 21 CFR 801.109) Over-The Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.