CEMVAC Ultra pre-packed with DePuy 1 Bone Cement is indicated for the fixation of prostheses to living bone in orthopaedic musculoskeletal surgical procedures for rheumatoid arthritis, traumatic arthritis, osteoporosis, avascular necrosis, collagen disease, severe joint destruction secondary to trauma or other conditions, and revision of previous arthroplasty.
Device Story
In-syringe vacuum mixing system pre-packed with bone cement powder; includes syringe barrel, mixing rod, locking plate, filter/vacuum hose, vacuum pump adapter, monomer cartridge, and disposal bag. Used in orthopaedic surgery to prepare bone cement while controlling monomer fumes. Physician operates device to mix cement components; vacuum system reduces exposure to vapors. Output is prepared bone cement for prosthesis fixation. Benefits include improved safety via fume control and standardized mixing.
Clinical Evidence
Bench testing only. Testing compared compressive, flexural, and impact strengths of DePuy 1 Bone Cement mixed via CEMVAC Ultra versus traditional bowl and spatula methods. Results demonstrated comparable mechanical properties. No clinical data provided.
Technological Characteristics
Vacuum mixing system; pre-packed syringe barrel with bone cement powder and monomer cartridge. Components include mixing rod, locking plate, filter/vacuum hose, and vacuum pump adapter. Single-use. Materials identical to previously cleared DePuy 1 Bone Cement. Mechanical mixing principle.
Indications for Use
Indicated for fixation of prostheses to bone in orthopaedic musculoskeletal surgeries including rheumatoid arthritis, traumatic arthritis, osteoporosis, avascular necrosis, collagen disease, severe joint destruction, and revision arthroplasty.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
## JUL 2 4 2002
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| 510(K) SPONSOR: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>P.O. Box 988<br>Warsaw, Indiana 46581-0988 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER: | DePuy International Ltd<br>Trading as DePuy CMW<br>Cornford Road<br>Blackpool, Lancashire<br>FY 4 4QQ, England |
| 510(K) CONTACT: | Janet Johnson, RAC<br>DePuy Orthopaedics, Inc.<br>Group Leader, Regulatory Submissions<br>(574) 371-4907; FAX (574) 371-4987<br>E-mail: jjohnso7@dpyus.jnj.com |
| TRADE NAME: | CEMVAC Ultra Pre-packed with DePuy 1 Bone Cement |
| COMMON NAME: | Pre-packed Bone Cement Dispenser |
| CLASSIFICATION: | Cement Dispenser, Class I Exempt per 21 CFR 888.4200<br>PMMA Bone Cement: Class II per 21 CFR 888.3027 |
| DEVICE CODE: | KIH pre-packed with LOD |
| EQUIVALENT DEVICES: | DePuy 1 Bone Cement - P960001/Supplement 3<br>Cemex System Gun Application Version - K000943<br>Howmedica Surgical Simplex P Radiopaque Bone Cement Pre-<br>Packed in ACM and Mix Evac II - K002652 |
#### DEVICE DESCRIPTION AND INTENDED USE:
CEMVAC Ultra pre-packed with DePuy 1 Bone Cement is a in-syringe vacuum mixing system prepacked with bone cement powder. The subject device is for single use and helps control monomer fumes during the preparation of the cement. The subject device pre-packed with DePuy 1 Bone Cement is available in a single 60g or 100g unit and also in unit packs of 5. Each single unit (device) pack consists of 1) syringe barrel filled with bone cement powder; 2) central mixing rod; 3) locking plate; 4) filter/vacuum hose; 5) vacuum pump adapter; 6) monomer cartridge with glass ampoules containing bone cement liquid; and 7) plastic disposal bag.
CEMVAC Ultra pre-packed with DePuy 1 Bone Cement is indicated for the fixation of prostheses to living bone in orthopaedic musculoskeletal surgical procedures for rheumatoid arthritis, traumatic arthritis, osteoporosis, avascular necrosis, collagen disease, severe joint destruction secondary to trauma or other conditions, and revision of previous arthroplasty.
{1}------------------------------------------------
02 1499
page 2 of 2
#### BASIS OF SUBSTANTIAL EQUIVALENCE:
With the exception of the primary packaging of the powder component of the DePuy 1 Bone Cement in the CEMVAC Ultra device, the DePuy 1 Bone Cement is identical in materials (formulation), and indications for use as the currently marketed DePuy 1 Bone Cement. On testing, DePuy 1 Bone Cement powder has been found to be compatible and stable in the CEMVAC Ultra device. Comparable DePuy 1 Bone Cement compressive, flexural and impact strengths were obtained for bowl and spatula and CEMVAC Ultra mixed cements.
CEMVAC Ultra pre-packed with DePuy 1 Bone Cement has similar indications for use as other bone cements currently marketed in the United States. These predicate devices include:
- 1) Cemex System Gun Application Version marketed by Tecres,
- 2) Howmedica Surgical Simplex P Radiopaque Cement Pre-Packed in ACM or Mix Evac II marketed by Stryker Corporation.
All three pre-filled mixing systems are pre-packed with bone cement and are intended to be used for the fixation of artificial joints prostheses to host bone.
Based on similarities of design, materials, intended use, and testing, DePuy believes that the CEMVAC Ultra pre-packed with DePuy 1 Bone Cement is substantially equivalent to the Cemex System Gun Application Version by Tecres, Howmedica Surgical Simplex P Radiopaque Cement Pre-Packed in ACM or Mix Evac II marketed by Stryker Corporation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 2 4 2002
Ms. Janet G. Johnson, RAC DePuy Orthopaedics, Inc. 700 Orthopaedic Drive P.O. Box 988 Warsaw, Indiana 4681-0988
Re: K021499
Trade Name: CEMVAC Ultra Pre-packed with DePuy 1 Bone Cement Regulation Number: 21 CRF 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: LOD Dated: May 8, 2002 Received: May 9, 2002
Dear Ms Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Janet G. Johnson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Mark N Millner
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use:
| 510(k) Number (if known): | " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
Device Name: CEMVAC Ultra Pre-packed with DePuy 1 Bone Cement
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Indications for Use:
CEMVAC Ultra pre-packed with DePuy 1 Bone Cement is indicated for the fixation of prostheses to Uning bone in orthopaedic musculoskeletal surgical procedures for rheumatoid arthritis, traumatic arthritis, osteoporosis, avascular necrosis, collagen disease, severe joint destruction secondary to trauma or other conditions, and revision of previous arthroplasty.
Concurrence of CDRH, Office of Device Evaluation
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Mark N. Millenn
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number.
Prescription Use X_OR
Over-The-Counter U
Over-The-Counter Use ___________ (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.