DATASCOPE'S 8FR. ALT B IAB; DATASCOPE'S PROFILE 8FR. ALT B IAB; DATASCOPE'S TRUE SHEATHLESS DL 9.5 FR. IAB; DATASCOPE'S

K020257 · Datascope Corp. · DSP · Apr 24, 2002 · Cardiovascular

Device Facts

Record IDK020257
Device NameDATASCOPE'S 8FR. ALT B IAB; DATASCOPE'S PROFILE 8FR. ALT B IAB; DATASCOPE'S TRUE SHEATHLESS DL 9.5 FR. IAB; DATASCOPE'S
ApplicantDatascope Corp.
Product CodeDSP · Cardiovascular
Decision DateApr 24, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3535
Device ClassClass 2
AttributesTherapeutic

Intended Use

The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: Refractory unstable angina . Impending infarction Acute MI . . Refractory ventricular failure Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) . . Cardiogenic shock . Support for diagnostic, percutaneous revascularization, and interventional procedures. Ischemia related intractable ventricular arrhythmias ● . Septic shock . Intraoperative pulsatile flow generation ● . Weaning from bypass Cardiac support for non-cardiac surgery . Prophylactic support in preparation for cardiac surgery . Post surgical myocardial dysfunction/low cardiac output syndrome ● . Myocardial contusion Mechanical bridge to other assist devices . . Cardiac support following correction of anatomical defects

Device Story

Intra-aortic balloon (IAB) catheter placed in descending aorta; improves cardiovascular function via counterpulsation. Used in clinical settings (OR, ICU) by physicians. Device provides mechanical circulatory support; assists in managing cardiac failure, shock, and surgical recovery. Output is hemodynamic support; helps stabilize patient hemodynamics; reduces myocardial workload. No software or algorithm involved.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Intra-aortic balloon catheter; mechanical circulatory support device. No software, electronics, or energy source integral to the catheter itself (catheter is a passive component of the IAB pump system).

Indications for Use

Indicated for patients requiring cardiovascular support due to refractory unstable angina, impending or acute MI, refractory ventricular failure, complications of acute MI, cardiogenic shock, septic shock, myocardial contusion, or post-surgical myocardial dysfunction. Also indicated for support during diagnostic/interventional procedures, intraoperative pulsatile flow generation, weaning from bypass, non-cardiac surgery support, prophylactic cardiac surgery support, mechanical bridging to other assist devices, and support following correction of anatomical defects.

Regulatory Classification

Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device; (ii) Software verification, validation, and hazard analysis must be performed; (iii) The device must be demonstrated to be biocompatible; (iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; (v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and (vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device. (2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation. (c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020257 # APR 2 4 2002 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ## FOR ## DATASCOPE Intra-Aortic Balloon Catheter Systems (Prepared in accordance with 21 CFR Part 807.92) Pursuant to Section 513(I)(3)(A) of the Food, Drug, and Cosmetic act, Datascope Corp. is required to submit this Premarket Notification either an "…adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Datascope Corp. chooses to submit a summary of information respecting safety and effectiveness. #### GENERAL INFORMATION A. | Submitter: | Datascope Corp. | |-----------------|------------------------------| | | Cardiac Assist Division | | Address: | 15 Law Drive | | | Fairfield, NJ 07004 | | Contact Person: | JoAnn Wolf | | | Regulatory Affairs Associate | #### B. DEVICE INFORMATION | Generic Name: | Intra-Aortic Balloon (IAB) Catheter | |----------------------|------------------------------------------------------------------------------| | Trade Name: | Datascope Intra-Aortic Balloon (IAB) Catheter | | Classification Name: | Intra-Aortic Balloon (IAB) Catheters are classified under 21 CFR<br>870.3535 | Product Code: 74DSP #### ﺰ PREDICATE DEVICE INFORMATION Datascope's Intra-Aortic Balloon Catheters Indications for Use are substantially equivalent to the following marketed devices: | K013326 - | Datascope 8Fr. 25, 34 & 40cc Alt B IAB | |-----------|------------------------------------------------------------------------------------------------------------------------------------------| | K003598 - | Datascope Profile 8Fr. 34 & 40cc Alt B IAB | | K980780 - | Datascope Percor STAT-DL 9.5Fr. 25, 34 & 40cc Alt B IAB | | K964987 - | Percor STAT-DL 9.5Fr. 25 & 40cc IABs for Optional Sheathless Insertion<br>and Percor STAT-DL 10.5Fr. 40 & 50cc IABs for Sheath Insertion | | K965236 - | Datascope System 96 Intra-Aortic Balloon Pump | | K002256 - | Arrow ACAT 2 Intra-Aortic Balloon Pump | All of the above referenced premarket notifications were found substantially equivalent by FDA. {1}------------------------------------------------ #### D. DEVICE DESCRIPTION/INTENDED USE The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: - Refractory unstable angina . - Impending infarction - Acute MI . - . Refractory ventricular failure - Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) . - . Cardiogenic shock - . Support for diagnostic, percutaneous revascularization, and interventional procedures. - Ischemia related intractable ventricular arrhythmias ● - . Septic shock - . Intraoperative pulsatile flow generation - Weaning from bypass - Cardiac support for non-cardiac surgery - . Prophylactic support in preparation for cardiac surgery - Post surgical myocardial dysfunction/low cardiac output syndrome ● - . Myocardial contusion - Mechanical bridge to other assist devices . - . Cardiac support following correction of anatomical defects #### E. TECHNOLOGICAL CHARACTERISTICS Datascope's enhanced Indications for Use for IAB Systems are substantially equivalent to the predicate devices. Supporting published data establish that the expanded Indications for Use is substantially equivalent to the predicate devices and does not affect the safety and effectiveness of the device. #### F. NON-CLINICAL TESTS N/A #### G. CLINICAL TESTS N/A #### H. CONCLUSIONS Based on the information presented in this 510(k) premarket notification, Datascope's expanded Indications for Use for the IAB Catheter System are considered substantially equivalent to the predicate devices' Indications for Use for their currently marketed IAB Catheter System. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 4 2002 Ms. JoAnn Wolf Regulatory Affairs Associate Datascope Corporation Cardiac Assist Division 15 Law Drive Fairfield, NJ 07004 Re: K020257 Trade Name: Datascope's 8Fr. 25, 34 & 40cc IAB; True Sheathless DL 9.5Fr. 25, 34 & 40cc IAB; Percor STAT-DL 9.5Fr. 34 & 40cc IAB for Optional Sheathless Insertion; Percor STAT-DL 10.5Fr. 50cc IAB Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon Catheter Regulatory Class: Class III (three) Product Code: DSP Dated: January 23, 2002 Received: January 24, 2002 Dear Ms. Wolf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. {3}------------------------------------------------ ## Page 2 - Ms. JoAnn Wolf If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase of actived and I Dr unation that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I other battler and states and station of limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaince nodicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Rödl & Partner Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # 510(k) Number (if known): Known): Known) Datascope's 8Fr. 25, 34 & 40cc IAB; True Sheathless DL 9.5Fr., 25, 34 & 40cc IAB; Device Name: Percor STAT-DL 9.5Fr. 34 & 40cc IAB for Optional Sheathless Insertion; Percor STAT-DL 10.5Fr. 50cc IAB Indications for Use: ## Intra-Aortic Balloon Catheter System to be used for the following: - Refractory unstable angina . - . Impending infarction - Acute MI . - . Refractory ventricular failure - Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) ● - Cardiogenic shock ● - Support for diagnostic, percutaneous revascularization, and interventional procedures. . - Ischemia related intractable ventricular arrhythmias . - . Septic shock - Intraoperative pulsatile flow generation ● - . Weaning from bypass - . Cardiac support for non-cardiac surgery - Prophylactic support in preparation for cardiac surgery . - Post surgical myocardial dysfunction/low cardiac output syndrome . - . Myocardial contusion - Mechanical bridge to other assist devices . - Cardiac support following correction of anatomical defects . This information can be found in Section 3 - Attachment I of our Premarket Notification under the section titled "II. Indications". # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use OR (Optional Format 1-2-96) Division of Cardiovascular & Respiratory Devices 510(k) Number K000257
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%