K020243 · Eradlink, Inc. · LMA · Mar 5, 2002 · Radiology
Device Facts
Record ID
K020243
Device Name
LASERPRO 16
Applicant
Eradlink, Inc.
Product Code
LMA · Radiology
Decision Date
Mar 5, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2030
Device Class
Class 2
Indications for Use
The LaserPro 16 Laser Film Digitizer is intended to be used in a Radiological Laboratory or Doctors office to read and digitize gray scale images from conventional x-rays for transmission to PC computer. The receiving computer allows the technician to add patient identification, review the x-ray at various rotations and options and compress the image using industry accepted techniques to an average file size of less than one megabyte. Compressed images are encapsulated in DICOM (Digital Imaging and Communications in Medicine) format, which then may be transmitted via various communication protocols to other sites or archived by a PACS host.
Device Story
The eRadLink LaserPro 16 is a desktop laser film digitizer that converts conventional radiographic film into 16-bit digital grayscale images. The device uses a laser beam light source and a sealed optic path to scan film (2 to 14 inches wide; up to 52 inches long) driven by a clocked stepping motor. The system outputs digital data in DICOM format to a host PC. Operated by technicians in radiological labs or clinics, the device allows for patient ID entry, image rotation, and compression (JPEG/Wavelet) on the host computer. The digitized images facilitate remote transmission and archiving via PACS, enabling healthcare providers to review x-rays digitally, potentially improving workflow efficiency and image accessibility.
Clinical Evidence
Bench testing only. Performance was validated using Stoeffer T4110 gray-scale strips, linearity test patterns, and typical x-ray film samples to demonstrate conformance to specifications. No clinical data provided.
Technological Characteristics
Desktop laser digitizer; solid-state 35mW laser; fiber optic beam scan; 16-bit grayscale output; 116um spot size; 0.0-4.1 OD dynamic range. Connectivity via USB or Ethernet. Power: 85-264VAC. Compliant with UL260, CSA 22.2, TUV, IEC 601-1, and CISPR 11 Class B. Software supports DICOM, JPEG, and Wavelet compression.
Indications for Use
Indicated for use in radiological laboratories or doctors' offices to digitize conventional x-ray film for transmission to a PC, subsequent network transmission, or PACS archiving. No specific patient population or contraindications provided.
Regulatory Classification
Identification
A medical image digitizer is a device intended to convert an analog medical image into a digital format. Examples include systems employing video frame grabbers, and scanners which use lasers or charge-coupled devices.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std.). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
# KO20243
# Appendix A. 510(k) Summary of Safety and Effectiveness (See next page)
MAR 5 2002
{2}------------------------------------------------
Per Title 21 CFR 807 . 92, the following is the 510(k) Summary for the Laser Film Digitizer manufactured by Clinical Laser Systems and marketed by eRadLink Inc. under the trade-name eRadLink LaserPro 16:
| (1) | SUBMITTER: | eRadLink Inc.<br>22750 Hawthorne Blvd.<br>Torrance, California 90505-<br>Phone: 310-373-5673<br>Fax: 310-373-9763 |
|-----|------------|-------------------------------------------------------------------------------------------------------------------|
| | CONTACT: | Albert Kouba<br>Regulatory Affairs |
19529 Chaparral Circle Penn Valley, California 95946 Phone: 530-432-3791 530-432-4046 Fax:
نور دارنانو،
18 DEC 2001 SUBMISSION DATE:
#### DEVICE NAME: (2)
TRADE NAME: eRadLink LaserPro 16
COMMON NAME: Laser Film Digitizer
CLASSIFICATION NAME: Digitizer, Image, Radiological (per regulation 21 CFR 892.2030) (Class II device)
PRODUCT CODE:
Lumiscan 75 (510(k) number Not Known) PREDICATE DEVICE: (3) DI-2000 (Software application - 510(k) Number K980213) Lumisys Inc. 225 Humboldt Court Sunnyvale, California 94089
LMA
### DEVICE DESCRIPTION: (4)
The eRadLink LaserPro 16 is a digitizing scanner that converts radiographic film transparency images to digital format. This is accomplished by utilizing a laser beam light source and a proprietary sealed optic path. There are no internal lenses, mirrors, or electro-optic devices. The new technology provides superior image quality, requires no internal optics cleaning, no optical alignment and is inherently highly accurate and reliable.
Film, from a minimum of 2 inches to a maximum of 14 inches in width and, from a minimum of 2 inches to a maximum of over 52 inches in length is driven passed the scanning laser beam by a clocked, stepping motor. Scanned data is electronically converted from analog to 16 bit digital gray scale and transmitted to the host computer in DICOM format.
{3}------------------------------------------------
K020243
#### INTENDED USE OF DEVICE (ર)
The LaserPro 16 Laser Film Digitizer is intended to be used in a Radiological Laboratory or Doctors office to read and digitize gray scale images from conventional x-rays for transmission to PC computer. The receiving computer allows the technician to add patient identification, review the x-ray at various rotations and options and compress the image using industry accepted techniques to an average file size of less than one megabyte. Compressed images are encapsulated in DICOM (Digital Imaging and Communications in Medicine) format, which then may be transmitted via various communication protocols to other sites or archived by a PACS host.
### SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICE -(6) HARDWARE
Table - A illustrates the comparative equivalence of the LaserPro 16 to the Predicate device Hardware.
| Feature | eRadLink | Lumisys | |
|-----------------|-----------------------------------|--------------------------|--|
| Product Name | LaserPro 16 | Lumiscan 75 | |
| 510(k) Number | | Not Known | |
| Dimensions | 12"(h) x 18"(w) x 12"(d) | 16"(h) x 21"(w) x 29"(d) | |
| Weight | 30 lbs. (13.6 kg) | 90 lbs. (40.9 kg) | |
| Power | 85-264vac 2.8A 47-63 Hz | 110-120vac 1A 50/60 Hz | |
| Scan Size | 2" x 2" (min), 14" x 15' (max) | 7" x 7" (min), 14" x 36" | |
| Spot Size | 116 um | 100 um | |
| Dynamic Range | 0.0-4.1 OD | 0.5 - 3.8 OD | |
| Gray Scale | 8, 12, or 16 bits | 12 bits | |
| Digitizing Rate | 100 lines/sec (16 bit gray scale) | 115 lines/sec | |
| Laser | 35 mw Solid State | He-Ne Laser | |
| Beam Scan | Fiber Optics | Galvanometer | |
| Pixel/mm | 8.6 | 10.09 | |
| Interface | USB or Ethernet | PCISA or SCSI | |
# Table A - Hardware Comparison of LaserPro 16 to Predicate Devices
{4}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICE -(7) SOFTWARE
Table B illustrates the comparative equivalence of the LaserPro 16 to the Predicate device Software.
| Features | eRadLink | Lumisys |
|----------------------------------------------------|--------------------------------------------------------------|---------|
| Product Name | eRadLink Medical<br>Internet Image &<br>Management<br>System | DI-2000 |
| 510(k) Number | - | K980213 |
| Patient Information - Add/Modify/Delete | Y | Y |
| Fax Report | N1 | Y |
| Print Dicom Image | N1 | Y |
| Image Rotate and Flip | Y | Y |
| Black/White Inversion | Y | Y |
| Multiple Image Display | N2 | Y |
| Dicom Send/Receive/Echo | Y | Y |
| Dicom Query User/Provider | Y | Y |
| Dicom Retrieve User/Provider | Y | Y |
| Dicom Print | N3 | Y |
| JPEG Compression lossy/lossless | Y | Y |
| iJPEG Compression | Y | Y |
| Wavelet Compression | Y | Y |
| Industry Standard Digital Communication<br>Support | Y | Y |
| Color Images | N | Y |
| Cine Loop Viewing | N | Y |
| Dicom Removable Media Support | N | Y |
| Measurement Tools | N4 | Y |
| Communication protocols | Y5 | Y5 |
# Table B - Software Comparison of LaserPro 16 to Predicate Devices
In progress
2 Dicom Viewer
3 Display only
4 Dicom Viewer
5 ADSL, Cable and Analog Modems and Phone Lines, ATM, ISDN, FDDL, Ethernet, Token Ring
{5}------------------------------------------------
# (8) EFFECTIVENESS
Program testing and calibration using Stoeffer T4110 gray-scale strip, linearity test patterns and typical x-ray film samples has demonstrated the LaserPro's 16's conformance to its defined specifications.
1 6 5 1 1 2
## (9) SAFETY
The LaserPro 16 is powered by an external power supply, which is UL260, CSA 22.2, The LaserFro To is powered by an externa: power safety and Class B for EMT requirements.
TUV and IEC 601-1 approved for safety and CISPR 1 1 Class B for EMT requirements. TO V and IDC 001-1 approved for salesy anting range from 0 to 70° C, overvoltage and I catures mercure 4000 VIC ibelief, operating .5% maximum at full load.
The LaserPro 16 Scanner is currently under test for FCC Class A EMT, RFI and UL safety compliance.
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows a black and white logo. The logo consists of a circular border with text around the top half, and a symbol in the center. The symbol appears to be three stylized lines or curves, possibly representing a bird or a wave.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 5 2002
Mr. Albert J. Kouba Regulatory Affairs eRadLink, Inc. 19529 Chaparral Circle PENN VALLEY CA 95946-9443 Re: K020243
Trade/Device Name: LaserPro 16 Laser Digitizing Scanner Regulation Number: 21 CFR 892.2030 Regulation Name: Medical image digitizer Regulatory Class: II Product Code: 90 LMA Dated: January 22, 2002 Received: January 23, 2002
Dear Mr. Kouba:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: __eRadLink LaserPro 16
## Indications for Use
The LaserPro 16 is a desk top laser image digitizer intended to produce digital The LaserFlo 10 Is a desk top lastr mage orgy scale. The digital copies are a copies of radiological min in 10 roversonal computer (PC) where they may be transmitted to a rentiall based porsonal verspace (0) = ) networks to other medical facility sites.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
6
David G. Brown
Prescription Use (Per 21 CFR 901.109)
OR Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.