EBI XFIX DFS SYSTEM

K013378 · Ebi, L.P. · KTT · Nov 8, 2001 · Orthopedic

Device Facts

Record IDK013378
Device NameEBI XFIX DFS SYSTEM
ApplicantEbi, L.P.
Product CodeKTT · Orthopedic
Decision DateNov 8, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EBI® XFIX® DFS® System is intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

Device Story

Unilateral external fixation system; consists of bone screws and fixator frame. Bone screws inserted through skin/soft tissue into bone; fixator frame attached to screw shanks. Used for stabilization in orthopedic procedures like leg lengthening, osteotomies, arthrodesis, and fracture fixation. Operated by surgeons in clinical settings. Provides mechanical stabilization of bone segments to facilitate healing. Modified version includes additional Angulating Screw Clamp component.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Unilateral external fixation device. Materials: aluminum, stainless steel, titanium alloy. Components provided non-sterile. Mechanical fixation principle. No software or electronic components.

Indications for Use

Indicated for patients requiring external bone fixation for conditions including leg lengthening, osteotomies, arthrodesis, and fracture fixation.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # NOV 0 82001 #### 510(k) Summary K013378 This 510(k) Summary for the modified EBI® XFIX® DFS® System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act. - Contact Person: Patricia Flood 1. Sponsor: EBI, L.P. Telephone: (973) 299-9300,ext.3318 100 Interpace Parkway Parsippany, NJ 07054 October 11, 2001 Date prepared: EBI® XFIX® DFS® System 2. Proprietary Name: External Fixation Device Common Name: Classification Names: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories, 21 CFR 888.3030 - 3. Predicate or Legally Marketed Device: - EBI® XFIX® DFS® System EBI, L.P. (K953406) ● #### 4. Description of the device: The System consists of external fixation components and implantable bone screws. The EBI XFIX DFS System is utilized in the following manner: bone screws are inserted through the patient's skin and soft tissue and into the bone. The fixator frame of the EBI XFIX DFS System is attached to the shanks of the bone screws. The intended use and fundamental scientific technology have not changed from the previously cleared submission. This submission is only for the addition of an Angulating Screw Clamp to the System. ### 5. Intended Use: The EBI® XFIX® DFS® System is intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality. {1}------------------------------------------------ ### 6. Materials: The components of the System may be manufactured from materials such as aluminum, stainless steel, and titanium alloy. ## 7. Comparison of the technological characteristics of the device to predicate devices: The modified EBI® XFIX® DFS® System is substantially equivalent to the following predicate device: EBI XFIX DFS System (K953406) - The modified EBI XFIX DFS System is fabricated from . the same materials as the components of the currently marketed EBI XFIX DFS Fixation System. - The modified EBI XFIX DFS System and the currently marketed EBI . XFIX DFS System are both indicated for the treatment of bone conditions, including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality. - The bone screw clamps of the modified EBI XFIX DFS System, like the ● bone screw clamps currently marketed in the EBI XFIX DFS System, are designed for attachment to the bone screws. - The additional component of the EBI XFIX DFS System, like the ● components of the currently marketed EBI XFIX DFS System, is provided non-sterile. - There are no significant differences between the modified EBI XFIX DFS ● System and the currently marketed EBI XFIX DFS System. It is substantially equivalent* to the predicate device with regard to intended use, materials, and function. <sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate the product can be lawfully marketed without pre-market approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)] {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract bird or a series of flowing lines, representing the department's mission related to health and well-being. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 0 8 2001 Ms. Patricia Flood, RAC Senior Regulatory Affairs Specialist EBI, L.P. 100 Interpace Parkway Parsippany, New Jersey 07054 Re: K013378 Trade/Device Name: EBI® XFIX® DFS® System Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: October 11, 2001 Received: October 12, 2001 Dear Ms. Flood: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Ms. Patricia Flood, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susa Walker, us Image /page/3/Picture/5 description: The image shows a signature. The signature is in black ink and appears to be a stylized version of the letters 'fa'. The 'f' is formed with a large loop at the top and a descending stroke, while the 'a' is smaller and connected to the 'f'. Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # STATEMENT OF INDICATIONS FOR USE Page 1 of 510(k) Number (if known): k013378 Device Name: EBI® XFIX® DFS® System Indications For Use: The EBI® XFIX® DFS® System is a unilateral external fixation device intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Lisa Walk (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...