IMMUNE CELL FUNCTION ASSAY

K013169 · Cylex, Inc. · NID · Apr 2, 2002 · Hematology

Device Facts

Record IDK013169
Device NameIMMUNE CELL FUNCTION ASSAY
ApplicantCylex, Inc.
Product CodeNID · Hematology
Decision DateApr 2, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K013169 · Apr 2, 2002IMMUNE CELL FUNCTION ASSAYCylex, Inc.Clinical cohort of transplant recipients; Clinical cohort of apparently healthy adultsThe study evaluated the Cylex Immune Cell Function Assay in a multi-center clinical setting to compare ATP concentration results between immunosuppressed transplant patients and healthy controls, establishing clinical performance characteristics.Transplant recipients; Immunosuppressive therapy; Multi-center study; Clinical performance

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Multi-center study of Cylex Immune Cell Function Assay; Multi-center clinical evaluation78 transplant recipients (immunosuppressed) and 44 apparently healthy adults; Sample Size: 122 total (63 transplant subjects and 40 healthy adults included in final analysis)Becton Dickinson TriTest/MultiTest CD4 count by flow cytometryATP concentration (ng/mL) as an indicator of immune cell function

Indications for Use

The Cylex Immune Cell Function Assay measures the concentration of ATP from circulating CD4 cells following in vitro stimulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anticoagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell mediated immune response in populations undergoing immunosuppressive therapy for organ transplant.

Device Story

Device measures ATP concentration in CD4 cells to assess cell-mediated immune function. Input: heparin-anticoagulated whole blood. Process: in vitro stimulation with phytohemagglutinin (PHA) triggers ATP synthesis in responsive cells; antibody-coated magnetic particles immunoselect CD4 cells; lysis reagent releases intracellular ATP; luciferase-luciferin reaction produces light proportional to ATP concentration. Output: light measurement (emission max 562 nm) via luminometer, converted to ATP concentration (ng/mL) using calibration curve. Used in clinical laboratory settings by trained personnel. Results assist clinicians in evaluating immune status of transplant patients on immunosuppressive therapy, potentially aiding in management of immunosuppression.

Clinical Evidence

Multi-center study of 44 healthy adults and 78 transplant recipients. Evaluated Cylex ICF Assay vs. flow cytometry CD4 counts. Results showed statistically significant differences between healthy and transplant populations for both assays (p < 0.0001). Mean ATP levels were 464 ng/mL (healthy) vs. 304 ng/mL (transplant).

Technological Characteristics

Uses antibody-coated magnetic particles for CD4 cell immunoselection. Detection via bioluminescence (luciferin/luciferase reaction) measured by luminometer (562 nm emission). Requires heparin-anticoagulated whole blood. Standalone assay system.

Indications for Use

Indicated for the detection of cell-mediated immunity in immunosuppressed populations, specifically those undergoing immunosuppressive therapy for organ transplant.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K 013169 # APR 0 2 2002 March 21, 2002 ## 510(K) SUMMARY ## Cylex Inc. Immune Cell Function Assay Submitted by: Cylex Inc. 8980-1 Old Annapolis Road Columbia, MD 21045 Contact: Dr. Judy Britz Name of Device: | Trade Name: | Immune Cell Function Assay | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | CD4 Cell Stimulation Assay | | Classification Name: | Automated Differential Cell Counter | | Predicate Device: | Becton Dickinson TriTest™ CD4 FITC/CD8 PE/CD3<br>PerCP Reagent;<br>Becton Dickinson MultiTest™ CD3 FITC/CD8 PE/CD4<br>PerCP/CD4 APC Reagent | #### Device Description: Device Describitor. Intended Use: The Cylex Immune Cell Function Assay measures the concentration of ATP intended Ose. The Gylox filmulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anti-coagulated whole Indicator of inindile cell function: "This models on the assay is used for the detection of cell blood dailing a luminoneter and lablem. Include going immunosuppressive therapy for organ transplant. Test Description: The Cylex Immune Cell Function Assay detects cell-mediated immunity in Test Description: The Oylex Mininens Collinitians Colling incubation, increased ATP writhesis occurs within the cells that respond to the stimulant phytohemagglutinin (PHA). synthesis occurs whall the oblished in the absence of stimulant for the purpose of assessing Concarrently, whole blood to mouslanal antibody coated magnetic particles are added to immunoselect CD4 cells from both the stimulated and non-stimulated wells. After washing the selected CD4 cells on a magnet tray, Lysis Reagent is added to release intracellular ATP. selected CD4 cells of a magnet tray, byels nought the released ATP produces light according to the following equation: Mo Oxyluciferin +AMP +Pyrophosphate +CO2 + Light Luciferin +ATP +O2 Luciferase The amount of light measured by a luminometer (emission maximum 562 nm) is propritional to The amount of light meadrou by a lantration of ATP (ng/mL) is calculated from a calibration the concentration of ATT : The concerners to characterize the cellular immune function of the sample. {2}------------------------------------------------ #### 510(K) SUMMARY #### Cylex Inc. Immune Cell Function Assay (cont.) ### Substantial Equivalence: The Cylex Inc. Immune Cell Function Assay has been found to be substantially equivalent to the Becton Dickinson TriTest™ CD4 FITC/CD8 PE/CD3 PerCP Reagent (K971205) and MultiTest™ CD3 FITC/CD8 PE/CD45 PerCP/CD4 ACP Reagent (K974360). All assays differentiate CD4 cells: the Cylex assay determines the responsiveness of those cells and the Becton Dickinson assays count the number of those cells. A multi-center study was conducted on freshly drawn blood collected from 44 apparently healthy adults and 78 transplant recipients (17 at discharge from the hospital and 61 post-discharge follow-up). The samples were evaluated with the Cylex Immune Cell Function Assay. The apparently healthy adult population consisted of 11% (5) females, 86% (38) males and 3% (1) unknown, with an age range of 20 - 60 years. The ethnicity of the population was 80% (35) African American, 16% Caucasian (7), and 4% (2) other or unknown. The transplant population consisted of 33% (26) females and 67% (52) males, with an age range of 20 - 64 years. The ethnicity of the population was 15% (12) African American, 74% Caucasian (58), 10% (8) other or unknown. The organs transplanted were 55% (43) liver, 36% (28) kidney, 4% (3) pancreas, and 5% (4) multiple organs. For purposes of these calculations, samples with %CV > 20% (after application of the outlier rule) were not included. The means of the two populations were found to be statically different; the results are summarized in the following table. | Population | Statistic | Cylex ICF Assay<br>(ATP ng/mL) | Comparator Assay<br>CD4 Count by Flow<br>Cytometry<br>(cells/µL) | |----------------------------------------------|-----------------|--------------------------------|------------------------------------------------------------------| | Apparently Healthy<br>(Non-immunosuppressed) | n | 40 | 40 | | | Mean | 464 | 746 | | | SD | 145 | 431 | | | Median | 443 | 654 | | | Range of Values | 243-967 | 130-2659 | | Transplant<br>(immunosuppressed) | n | 63 | 63 | | | Mean | 304 | 542 | | | SD | 163 | 423 | | | Median | 293 | 503 | | | Range of Values | 58-759 | <68*-1904 | Summary of Clinical Trial Results *Lower linear limit of the flow cytometry assay NOTE: The means of the Cylex ICF Assay results for the two populations are statistically significantly different (p <0.0001). The means of the comparator assay CD4 count results for the two populations are also statistically significantly different (p < 0.0001). {3}------------------------------------------------ ### 510(K) SUMMARY ## Cylex Inc. Immune Cell Function Assay (cont.) | | Patient Status | | | |---------------------------|------------------------|---------------------------------|-------| | Cylex ICF Assay<br>Result | Transplant<br>Subjects | Apparently<br>Healthy<br>Adults | Total | | ≤225 ATP ng/mL | 24 (38%) | 0 (0%) | 24 | | 226 - 524 ATP<br>ng/mL | 33 (52%) | 27 (67%) | 60 | | ≥525 ATP ng/mL | 6 (10%) | 13 (33%) | 19 | | Total | 63 | 40 | 103 | | | Patient Status | | | |--------------------|------------------------|---------------------------------|-------| | Total CD4<br>Count | Transplant<br>Subjects | Apparently<br>Healthy<br>Adults | Total | | <410 cells/µL | 29 (46%) | 8 (20%) | 37 | | ≥410 cells/µL | 34 (54%) | 32 (80%) | 66 | | Total | 63 | 40 | 103 | The percentage of transplant patients with an assay result <525 ATP ng/mL was 90% (57/63, The percentage of transplant patients). The percentage of transplant subjects having an assay result ≤225 ATP ng/ml was 38% (24/63, 95% confidence interval 26.1 – 51.2%). The result =220 XTP Tighth was 00% (2525 ATP ng/mL was 10% (6/63, 95% confidence interval 3.6 - 19.6%). The percentage of apparently healthy adults having an assay result <525 ATP ng/mL was 67% The percentage of apparently houlthy additioning and securities of apparently healthy adults (21140, 90% ounned finer of one of one of a sure of apparently healthy adults having an assay result ≥525 ATP ng/mL was 33% (13/40, 95% confidence interval 18.6 – 49.1%). *U.S. Patent No. 5,773,232 {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract design featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped with wings. Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993 Cylex Inc. c/o Ms. Judi Smith, Principal, Judi Smith, LLC 8980-I Old Annapolis Road Columbia, MD 21045 SEP 2 1 2010 Re: k013169 Trade/Device Name: Cylex Immune Cell Function Assay Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: NID Dated: January 14, 2002 Received: January 15, 2002 Dear Ms. Smith: This letter corrects our substantially equivalent letter of April 2, 2002. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements {5}------------------------------------------------ . of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Maria M. Chan, Ph. D. Maria M. Chan. Ph. D Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## k013169 Cylex Corp. | Div/Branch | Last Name | Date | Div/Branch | Last Name | Date | |------------|------------|------------|------------|-----------|------| | DIHD/IMDB | Reeves | 09/21/2010 | | | | | DIHD/IMPB | PHILIP | 9-21-2010 | | | | | DIHD/IMDB | CARRINGTON | 9/21/2010 | | | | | Date of Update | By | Description of Update | |----------------|---------------|-------------------------------| | 7/27/09 | Brandi Stuart | Added Updates to Boiler Table | | 8/7/09 | Brandi Stuart | Updated HFZ Table | {7}------------------------------------------------ Page 1 of 1 510(k) Number (if known): 013169 Device Name: Immune Cell Function Assay Indications For Use: The Cylex Immune Cell Function Assay measures the concentration of ATP from circulating CD4 cells following in vitro stimulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anticoagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell mediated immune response in populations undergoing immunosuppressive therapy for organ transplant. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Reena Philip **Division Sign-Off** Office of In Vitro Diagnostic Device Evaluation and Safety 510K k013/69 ) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) {8}------------------------------------------------ 510(k) Number (if known): K0i31i69 Device Name: Immune Cell Function Assay Indications For Use: The Cylex Immune Cell Function Assay detects cell-mediated immunity (CMI) by measuring the concentration of ATP from CD4 cells following stimulation. This measurement is made on heparin anti-coagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell-mediated immunity in an immunosuppressed population. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dmu H rision Sign Division of Clinical Laboratory Devices KG13169 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...