TRI TEST REAGENT CD4 FIT C/CD8 PE/CD3 PERCP WITH TRUCOUNT ABSOLUTE COUNT TUBES
K971205 · Becton Dickinson Immunocytometry Systems · GKZ · Dec 1, 1997 · Hematology
Device Facts
Record ID
K971205
Device Name
TRI TEST REAGENT CD4 FIT C/CD8 PE/CD3 PERCP WITH TRUCOUNT ABSOLUTE COUNT TUBES
Applicant
Becton Dickinson Immunocytometry Systems
Product Code
GKZ · Hematology
Decision Date
Dec 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
For in vitro diagnostic use to identify and enumerate absolute counts of T lymphocyte (CD3+), helper/inducer T lymphocyte (CD3+CD4+) and suppressor/cytotoxic T lymphocyte (CD3+CD8+) subsets in blood.
Device Story
System consists of TriTEST monoclonal antibody reagent (CD4 FITC/CD8 PE/CD3 PerCP) and TRUCOUNT Absolute Count Tubes. Input: erythrocyte-lysed whole blood sample. Process: sample added to counting bead pellet; stained with fluorochrome-labeled antibodies; analyzed via flow cytometry. Principle: fluorescence gating identifies T lymphocyte subsets (CD3+, CD3+CD4+, CD3+CD8+); absolute counts calculated by comparing reagent-positive events to bead events. Used in clinical laboratories by trained personnel. Output: absolute counts of T lymphocyte subsets. Clinical utility: monitoring immunodeficiency and autoimmune disease status.
Clinical Evidence
Multi-site clinical studies (Cleveland Clinic, Johns Hopkins, etc.) compared TriTEST/TRUCOUNT to FACSCount. Accuracy demonstrated equivalence. Stability studies established sample handling requirements (staining within 24-48 hours). Within-specimen reproducibility showed acceptable performance. Linearity confirmed across 16,700 to 200 lymphocytes/μL. Cross-platform study showed good correlation with Coulter systems, though <20% bias noted, necessitating user validation.
Technological Characteristics
Three-color direct immunofluorescence reagent (CD4 FITC, CD8 PE, CD3 PerCP) and counting bead tubes. Requires flow cytometer with 488 nm laser and specific fluorescence detection channels (515-545 nm, 562-607 nm, >650 nm). Uses fluorescence gating for cell identification. Software-based analysis of list mode data for absolute count calculation.
Indications for Use
Indicated for use with flow cytometers (488 nm laser; detection ranges 515-545 nm, 562-607 nm, >650 nm) and erythrocyte-lysed whole blood to identify/enumerate CD3+, CD3+CD4+, and CD3+CD8+ lymphocyte absolute counts. Used to characterize/monitor immunodeficiency (e.g., HIV) and autoimmune diseases.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{1}
H471205
# Attachment F
## Summary of Safety and Effectiveness
DEC - 1 1997
### Submitter Information (21 CFR 807.92(a)(1))
Submitter: Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, CA 95131-1807
Contact: Anna Longwell, Esq.
Director, Regulatory Affairs - Corporate
(408) 954-2254
Summary date: March 31, 1997
### Name of Device and Classification (21 CFR 807.92(a)(2))
Name: Becton Dickinson TriTEST™ reagent CD4 FITC/CD8 PE/CD3 PerCP; TRUCOUNT™ Absolute Count Tubes
Classification: Class II
### Predicate Device (21 CFR 807.92(a)(3))
The BDIS TriTEST™ CD4 FITC/CD8 PE/CD3 PerCP reagent with TRUCOUNT Absolute Count Tubes is substantially equivalent to FACSCount (cleared to market under K933486).
### Description of the Device (21 CFR 807.92(a)(4))
The BDIS TriTEST CD4 fluorescein isothiocyanate (FITC)/CD8 phycoerythrin (PE)/CD3 peridinin chlorophyll protein (PerCP) reagent is a three-color, direct immunofluorescence reagent. When used with TRUCOUNT Absolute Count Tubes it is used for identifying and enumerating absolute counts of T lymphocytes (CD3+), helper/inducer T lymphocytes (CD3+CD4+) and suppressor/cytotoxic T lymphocytes (CD3+CD8+) in erythrocyte-lysed whole blood (LWB). The Becton Dickinson TriTEST/TRUCOUNT system for immunophenotyping consists of a flow cytometer (either from BDIS or from another manufacturer), conjugated monoclonal reagent (TriTEST CD4 FITC/CD8 PE/CD3 PerCP) and TRUCOUNT Absolute Count Tubes.
The process to obtain T lymphocyte subset absolute counts includes: 1) obtaining a whole blood sample, 2) adding a precise volume of whole blood directly to the Absolute Count Tube, 3) cell-surface antigen staining with three-color monoclonal antibody reagents, 4) erythrocyte lysis, and 5) flow cytometric acquisition and analysis of list mode data. Analysis for absolute counts requires that the bead region be identified and the events in this region counted. The proportion of reagent positive events to bead events (P) is computed. The absolute count is P x (beads/pellet)/(volume of blood sample).
301
{2}
302
# Summary of Safety and Effectiveness
When monoclonal antibody reagents are added to human whole blood, the fluorochrome-labeled antibodies bind specifically to antigens on the surface of leucocytes, thus identifying lymphocyte populations. The patient blood sample is added to the counting bead pellet and is treated with fluorochrome-labeled antibodies and the erythrocytes are lysed with FACS® Lysing Solution. The flow cytometer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cytometer and the stained cells and beads fluoresce when excited by a laser beam.
The three-color reagent permits identification of T lymphocyte subsets using fluorescence gating instead of forward scatter gating. This three-color reagent allows direct gating on the CD3+ population using a combination of fluorescence and side scatter parameters. By gating on the CD3+ population, a maximum number of T lymphocytes may be captured in the gate.
## Intended Use (21 CFR 807.92(a)(5))
For in vitro diagnostic use to identify and enumerate absolute counts of T lymphocyte (CD3+), helper/inducer T lymphocyte (CD3+CD4+) and suppressor/cytotoxic T lymphocyte (CD3+CD8+) subsets in blood.
## Indications for Use
- For use with any flow cytometer equipped with a 488 nm laser and capable of detection in the ranges: 515-545 nm, 562-607 nm, and >650 nm
- For use with erythrocyte lysed whole blood
- For in vitro diagnostic use
- To identify and enumerate absolute counts of CD3+ and CD3+CD4+ and CD3+CD8+ lymphocytes
- To characterize and monitor some forms of immunodeficiency, such as in HIV infected individuals
- To characterize and monitor some forms of autoimmune diseases
## Clinical Utility
The determination of T lymphocyte (CD3+), helper/inducer T lymphocyte (CD3+CD4+) and suppressor/cytotoxic T lymphocyte (CD3+CD8+) has been found useful in monitoring some forms of immunodeficiency and autoimmune disease.
## Comparison to Predicate Device (21 CFR 807.92(a)(6))
The TriTEST/TRUCOUNT system (for absolute counts) is substantially equivalent to the FACSCount system for CD3+, CD3+CD4+ and for CD3+CD8+, (cleared to market under K933486). Both the TriTEST reagent with TRUCOUNT tubes and the predicate device yield equivalent results for the same analytes, and both are intended for use as an in vitro diagnostic test using a flow cytometer-based instrument and recommended computer hardware and software. Results demonstrate that the products yield essentially equivalent performance characteristics.
CD4/CD8/CD3/TRUCOUNT 510(k) Notification
Attachment F: Summary of Safety and Effectiveness
{3}
# Summary of Safety and Effectiveness
Performance Data (21 CFR 807.92(b)(2))
Performance of the product was established by testing at Cleveland Clinic, Johns Hopkins Hospital, Institute of Tropical Medicine, University of North Carolina Hospital, and at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California.
Several studies were performed:
- Accuracy data demonstrated the TriTEST/TRUCOUNT product's equivalence to FACSCount.
- A stability study was conducted to assess the time effect relating to age of blood (time-from-draw) and the time effect relating to the age of the stain (time-from-sample preparation), as well as the combined effect of both. Results indicate that for the calculation of absolute counts, blood specimen should be stained within 24 hours of draw, and analysis of the stained samples should occur within 24 hours or alternatively, the blood specimen can be stained within 48 hours of draw and analyzed within 6 hours.
- Within-specimen reproducibility was performed at BDIS; 10 replicates from 1 high, 1 medium, and 1 low were assessed. Within-specimen reproducibility was also performed at 3 clinical sites; 3 aliquots from each donor were assessed. Results demonstrated acceptable within-sample reproducibility.
- Linearity was determined using blood samples from 3 normal donors diluted to 5 concentrations, ranging from 16,700 to 200 lymphocytes/μL and from 31,000 to 2,500 WBC/μL. Results indicate a linear response over this range.
- Cross reactivity of these clones is reported in the literature. Conjugation and product formulation have not changed their specificity.
- Results from a cross platform reproducibility study for determining absolute counts using TriTEST CD4 FITC/CD8 PE/CD3 PerCP reagent with TRUCOUNT Absolute Count Tubes indicated a small (<20%) non-zero bias, but good correlation between results on a Becton Dickinson flow cytometer versus a Coulter. Therefore, users will be advised that they must validate performance characteristics for absolute counts, as required under CLIA regulations (42 CFR 493.1202 and 493.1213).
Performance Data - Conclusions (21 CFR 807.92(b)(3))
The results of the clinical studies demonstrate that the device is as safe and effective as the predicate device.
CD4/CD8/CD3/TRUCOUNT 510(k) Notification
Attachment F: Summary of Safety and Effectiveness
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Anna Longwell
Director, Regulatory Affairs
BECTON DICKINSON IMMUNOCYTOMETRY SYSTEMS
2350 Qume Drive
San Jose, CA 95131-1807
DEC - 1 1997
Re: K971205
Trade Name: Becton Dickinson TriTEST Reagent CD4 FITC/CD8 PE/CD 3 PerCP; TruCOUNT Absolute Count Tubes
Regulatory Class: II
Product Code: GKZ, 81
Dated: March 31, 1997
Received: April 1, 1997
Dear Ms. Longwell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions.
Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K971205
Device Name:
## Indications For Use:
- For use with any flow cytometer equipped with a 488 nm laser and capable of detection in the ranges: 515-545 nm, 562-607 nm, and >650 nm
- For use with erythrocyte lysed whole blood
- For *in vitro* diagnostic use
- To identify and enumerate absolute counts of CD3+ and CD3+CD4 and CD3+CD8+ lymphocytes
- To characterize and monitor some forms of immunodeficiency, such as in HIV infected individuals
- To characterize and monitor some forms of autoimmune diseases
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
[Signature: Peter E. Malin]
(Division Sign-Off)
Division of Clinical Laboratory Devices K971205
510(k) Number
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.