MEDIGROUP KEY TUBE

K013017 · Janin Group, Inc. · FJS · Dec 6, 2001 · Gastroenterology, Urology

Device Facts

Record IDK013017
Device NameMEDIGROUP KEY TUBE
ApplicantJanin Group, Inc.
Product CodeFJS · Gastroenterology, Urology
Decision DateDec 6, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5630
Device ClassClass 2

Indications for Use

The Key Tube is designed to be used to implant the Ash Advantage™ catheter. No usage change from original Ash Advantage catheter 510(k) 991042.

Device Story

Key Tube™ is a stainless steel, quasi-tubular (U-shaped) surgical accessory used to facilitate the implantation of the Ash Advantage™ peritoneal dialysis catheter. The device is operated by a physician during peritoneoscopic or laparoscopic procedures. It functions as a guide to ensure proper placement of the catheter's T-junction against the parietal peritoneum and the distal cuff within the rectus sheath, thereby avoiding incision of the rectus sheath. The device is a mechanical tool; it does not process data or provide automated outputs. It benefits patients by ensuring accurate, minimally invasive catheter placement.

Clinical Evidence

Clinical data derived from the Ash Advantage™ catheter clinical trials (510(k) 991042). A total of 14 catheters were implanted in 13 patients using the prototype Key Tube via peritoneoscopic or laparoscopic methods. Results demonstrated successful implantation in all cases, with the Key Tube functioning as designed to position the T-junction against the parietal peritoneum and the distal cuff within the rectus sheath, avoiding rectus sheath incision.

Technological Characteristics

Stainless steel, quasi-tubular (U-shape) device; approximately 200 mm long and 8.0 mm diameter. Mechanical surgical accessory; no energy source, connectivity, or software components.

Indications for Use

Indicated for use in the implantation of the Ash Advantage™ catheter in patients requiring peritoneal dialysis.

Regulatory Classification

Identification

(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.

Special Controls

*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ KO13017 Page 10f1 ## DEC 0 6 2001 #### 7.1. 510(k) Summary of Safety and Effectiveness 7.1.1 Basic Data | Date Prepared: | September 6, 2001 | |----------------|--------------------------| | Company: | MEDIGROUP, Inc | | Contact: | John A. Navis, President | | Phone: | (630) 428-4143 | | Fax: | (630) 428-4148 | | E-mail: | jnavis@medigroupinc.com | 7.1.2 Device Information: > Classification Name: Common Name: Peritoneal Catheter Accessory Catheter Implantation Device #### 7.1.3 Predicate Device This device (Key Tube™) is an adaptation (not replacement) for an existing tool, Cuff Implantor®, currently produced by Medigroup and included in two of its Y-TEC® kits, VP-112 and VP-210, 510(k)883303A and adapted in April 7, 1993 letter to FDA. An early prototype Key Tube version was designed and used during the Ash Advantage™ clinical trials [510(k)991042]. - Ash Advantage™ catheter clinical trials were conducted in three dialysis centers. 7.1.4 A total of fourteen catheters were implanted in thirteen patients via the peritoneoscopic or laprascopic method (doctor's choice). All implanting doctors used the prototype version of the Key Tube™ to implant the Ash Advantage™ catheter. Clinical results were that all catheters were successfully implanted; the Key Tube™ functioned as designed. That is, the T junction of the Ash Advantage™ was against the parietal peritoneum; the distal cuff was within the rectus sheath; incision of the rectus sheath was avoided. - 7.1.5 Device Description and Intended Use Stainless steel, quasi tubular (U shape) device, approximately 200 mm long and 8.0 mm diameter. It is designed to be used to implant the Ash Advantage catheter. {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that converge and then separate into wavy lines at the bottom. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## DEC 0 6 2001 Mr. John A. Navis President Medigroup, Inc. 505 Weston Ridge Drive NAPERVILLE IL 60563-3932 Re: K013017 Trade/Device Name: Key Tube™ (KT-2580) Regulation Number: 21 CFR §876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: 78 FJS Dated: September 6, 2001 Received: September 7, 2001 Dear Mr. Navis: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(t) premexice is substantially equivalent (for the indications ferenced above and nave determined ly marketed predicate devices marketed in interstate 101 use stated in the cholosure) to regars actment date of the Medical Device Amendments, or to conmente province to May 20, 1978, the enaounce with the provisions of the Federal Food, Drug, devices mat have been reculted in avee approval of a premarket approval application (PMA). alla Cosment Act (Act) that do not require appro the general controls provisions of the Act. The 1 ou may, diciolor, manov of the Act include requirements for annual registration, listing of general controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (300 above) and existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Dri 3 issualite of our device complies with other requirements of the Act that I DA has made a determinations administered by other Federal agencies. You must of any I coord statutes and roguirements, including, but not limited to: registration and listing Compry with an the Act 3 requirements, models. good manufacturing practice requirements as set manie (21 OF IC Part 801), faceming (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality bystems (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notification. The PDA midmis of successification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 811), please II you desire specific advice for your de rest of ollowing numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on and prosses of the regulation entitled, "Misbranding Office of Comphanee at (301) 39 ( 1057) Part 807.97). Other general information on by reletence to premarket notification (2) obtained from the Division of Small Manufacturers, your responsibilities the Act Her May of other of other (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### 510(k) Number: K013017 Device Name: Medigroup Key Tube™ Indications for Use: The Key Tube is designed to be used to implant the Ash Advantage™ catheter. No usage change from original Ash Advantage catheter 510(k) 991042. # (PLEASE DO NOT WRITE BELOW THIS LINE ~ CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) David le. Segerson Division of Ophthalmic, Anesthesia, and Neurological Devices 510(k) Number K013017
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...