K012726 · Hou-Se Medical, Inc. · MEG · Oct 18, 2001 · General Hospital
Device Facts
Record ID
K012726
Device Name
PROSAFE SAFETY SYRINGE
Applicant
Hou-Se Medical, Inc.
Product Code
MEG · General Hospital
Decision Date
Oct 18, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SafePro* Safety Syringe is designed to gain access to a patient's vascular or intramuscular system for short term fluid sampling or infusion. The device is designed for single use and has a needlestick protection feature.
Device Story
SafePro Safety Syringe consists of syringe and needle assembly with integrated needle retraction mechanism. Used for vascular or intramuscular fluid sampling or infusion. Operated by healthcare professionals. After procedure completion, user activates retraction system to shield needle, reducing accidental needlestick injury risk. Device is single-use. Benefits include enhanced clinician safety via sharps protection.
Clinical Evidence
Bench testing and simulated use study conducted. 500 syringes evaluated by 50 participants. Results showed zero sharps injuries or safety mechanism failures. Biocompatibility testing met ISO 10993-1 and FDA G-95 requirements.
Technological Characteristics
Manual syringe with integrated needle retraction mechanism. Materials tested for biocompatibility per ISO 10993-1 and FDA G-95. Single-use, sterile device.
Indications for Use
Indicated for patients requiring vascular or intramuscular access for fluid sampling or infusion. Designed for single use with a needle retraction system to reduce accidental needlestick risk.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
B-D Luer Lok syringe
Retractable Technology, Inc. VanishPoint syringe
Submission Summary (Full Text)
{0}------------------------------------------------
Ko12726
OCT 1 8 2001
510(K) - SafePro* Safety Syringe
### SECTION VIII
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS for SafePro* Safety Syringe
#### 1. REGULATORY AUTHORITY
Safe Medical Device Act of 1990, CFR 807.92.
#### 2. CONTACT PERSON
Joseph J. Chang, Ph.D., PE HOU-SE MEDICAL, INC. U.S. Liaison Office P. O. Box 630374 Irving, TX 75063-0374
#### NAME OF MEDICAL DEVICE 3.
Classification Name: Syringe, Antistick Classification Code: 177 MEG Common/Usual Name: Syringe Proprietary Name: SafePro* Safety Syringe
#### DEVICE CLASSIFICATION 4.
The General Hospital Panel has classified Antistick Syringes (21CFR880.5600) into Class II, Special Controls under section 513 of the Act.
#### ડ. STATEMENT OF SUBSTANTIAL EQUIVALENCE
The SafePro* Safety Syringe is substantially equivalent to B-D Luer Lok syringe in terms of normal syringe function for fluid transport, and Retractable Technology, Inc. VanishPoint syringe for needlestick protection feature. See also the attached table.
{1}------------------------------------------------
510(K) - SafePro* Safety Syringe
#### 6. INTENDED USE
/
The SafePro* Safety Syringe is designed to gain access to a patient's vascular or intramuscular system for short term fluid sampling or infusion. The device is designed for single use and has a needlestick protection feature.
#### DESCRIPTION OF DEVICE 7.
The SafePro* Safety Syringe consists of a syringe assembly and a needle assembly. The device has a built-in safety feature to reduce the risk of accidental needlestick injuries.
#### 8. SUMMARY OF MATERIAL TESTING
The SafePro* Safety Syringe was tested for material safety and biocompatibility. Test results indicated that the SafePro* Safety Syringe meets ISO 10993-1 and US FDA G-95 requirements.
#### 9. SUMMARY OF SIMULATED USE STUDY
A total of 500 SafePro* Safety Syringes were evaluated by 50 participants. No sharps injuries or failure of the safety mechanism occurred. Successful completion of the study supports the claim that SafePro* Safety Syringe can reduce the risk of accidental needlestick injuries. The positive responses from the Evaluators regarding functional and performance aspects also indicated that the SafePro* Safety Syringe meets customer requirements.
#### CONCLUSION 10.
The results obtained from bench testing, material safety, and simulated use tests indicate that the SafePro* Safety Syringe is safe and effective for it's intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or abstract shapes. The logo is in black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joseph J. Chang, PHD, P.E. Chief Technology Officer Hou-se Medical, Incorporated
P.O. Box 630374 Irving, Texas 75063-0374
Re: K012726
Trade/Device Name: SafePro™ Safety Syringe Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: August 10, 2001 Received: August 15, 2001
Dear Dr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
OCT 1 8 2001
{3}------------------------------------------------
## Page 2 - Dr. Chang
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
## Applicant
HOU-SE MEDICAL, INC. Address in Taiwan Headquarter Office: 295, 7/F, Sec. 2, Ho-Ping E. Rd., Taipei, Taiwan, R. O. C. Zip Code: 106
Address in U.S. Liaison Office: P.O. Box 630374 Irving, TX 75063-0374
# 510(K) Number K Ol 2726
This is a new submission and FDA has not assigned a number at this time.
## Device Name
SafePro* Safety Syringe
## Indication for Use
A properly placed syringe provides access to a vein or tissue for fluid sampling or infusion. The safety syringe is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a needle retraction system which is activated after the infusion procedure is completed.
(Please don not write below this line-Continue on another page if needed) Concurrence of CDRH Office of Device Evaluation (ODE)
Prescription Use Per 21 CFR 801.109 (Optional Format 1-2-96)
OR Over-the-Counter **__**
*Sene Naveau for lXC*
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital D 510(k) Number
*Trademark