K991039 · Saf-T-Med, Inc. · FMF · May 21, 1999 · General Hospital
Device Facts
Record ID
K991039
Device Name
EASYSAFE RETRACTING NEEDLE SYRINGE
Applicant
Saf-T-Med, Inc.
Product Code
FMF · General Hospital
Decision Date
May 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The primary intended use of the EasySafe® Retracting Needle Syringe is to inject fluid into, or withdraw fluid from the body. The secondary intended uses are: 1. to help prevent sharps injuries when using the product for its primary intended use, and; 2. to help reduce wasted medication when using the product for its primary intended use.
Device Story
EasySafe Retracting Needle Syringe; manual syringe device for intramuscular fluid injection or withdrawal. Features integrated needle retraction mechanism to mitigate sharps injuries and reduce medication waste. Operated by clinicians in clinical settings. Device functions as standard syringe; retraction mechanism activated post-injection to shield needle. Benefits include enhanced clinician safety and improved medication delivery efficiency.
Clinical Evidence
Two simulated use studies involving >1,500 injections/withdrawals performed by clinicians. All procedures successfully completed with no reported injuries. Focus group of 20 clinicians reported no impediments to use. Bench testing confirmed compliance with ISO performance standards, graduation accuracy, and needle guard puncture resistance. Biocompatibility and sterilization (SAL 10^-6) validated by contract laboratories.
Technological Characteristics
Manual retracting needle syringe. Materials biocompatibility validated. Performance compliant with ISO standards for graduation accuracy and needle guard puncture resistance. Sterilization via validated process to SAL 10^-6. No electronic components or software.
Indications for Use
Indicated for patients requiring intramuscular fluid injection or fluid withdrawal. No specific age or gender restrictions noted. Contraindications not specified.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
K973792 — EASYSAFE RETRACTING NEEDLE SYRINGE · Saf-T-Med, Inc. · Jan 29, 1998
K180417 — Self-destruction Safety Syringes for Single Use; Sterile Hypodermic Syringes for Single Use; Sterile Hypodermic Needles for Single Use; Sterile Safety Hypodermic Needles for Single Use · Berpu Medical Technology Co., Ltd. · Aug 9, 2018
K980069 — VANISHPOINT SYRINGE · Retractable Technologies, Inc. · Feb 12, 1998
K221308 — Disposable Syringe with permanently attached needle, Safety Syringe with permanently attached needle · Jiangxi Maidikang Medical Devices Co., Ltd. · Oct 17, 2022
K042102 — YIXIN SAFETY SYRINGE · Jiangxi Sanxin (Eversincere) Medical Devices Group · Mar 8, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K991039
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION [21 CFR 807.92]
| 510(k) Number: | Pending |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | EasySafe™ Retracting Needle Syringe |
| Common Name: | Retracting Needle Syringe |
| Manufacturer: | Saf-T-Med, Inc.<br>1250 South Grove, Suite 200<br>Barrington, IL 60010 |
| Predicate Device Information: | A claim of substantial equivalence is made to:<br>USMI SafeSnap™ Syringe, K925039<br>BD syringe (pre-Ammendment device) |
| Device Claims: | For the injection of fluid intramuscularly (IM) into<br>the body, while helping to reduce the risk of sharps<br>injuries and reducing the amount of wasted<br>medication. |
In compliance with the requirements of the Safe Medical Device Act (SMDA) of 1990, the following information is submitted as a summary of safety and effectiveness information for this 510(k) premarket notification.
## Effectiveness
In two recent simulated use studies, over 1,500 injections/fluid withdrawals were administered by a wide variety of clinicians to test the effectiveness of the product in injecting fluids into or withdrawing fluids from a patient. All injections/withdrawals were successfully completed.
Bench testing of the product confirms its compliance with the applicable performance standards established by the International Standards Organization (ISO).
In focus groups twenty clinicians indicated that they saw no impediments to being able to effectively administer injections to patients using the product.
## Safety
In two recent simulated use studies, over 1,500 injections/fluid withdrawals were administered by a wide variety of clinicians to test the safety of the product when used to inject fluids into or withdraw fluids from a patient. No injuries of any kind were reported.
Bench testing of the product confirms its comparability with legally-marketed predicate devices for a variety of safety factors, including but not limited to graduation accuracy and needle guard puncture resistance.
{1}------------------------------------------------
Biocompatability testing was performed by outside contract laboratories, with the results indicating that, from a biocompatability standpoint, the product is safe for its intended purpose.
A variety of design factors are intended to improve upon patient and clinician safety during use of the product. While sample size requirements have precluded a broad statistical study of the impact of such factors on injury rates, clinicians have reacted in a unanimous fashion to those factors.
A detailed risk analysis was performed on the product and, based on information currently available, the risk of harm from the product appears to be relatively low.
Sterilization validation was conducted by an outside contract laboratory. Stenlization to a SAL of 10° was confirmed during the tests. Pyrogenicity was also evaluated and acceptable results obtained.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center of the logo is a stylized symbol consisting of three curved lines that resemble a person or figure. The symbol is meant to represent the department's mission of protecting the health of all Americans and providing essential human services.
MAY 2 1 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. James Erbs President Saf-T-Med, Incorporated 1250 South Grove, Suite 200 Barrington, Illinois 60010
Re : K991039 Trade Name: EasySafe® Retracting Needle Syringe Regulatory Class: II Product Code: FMF March 25, 1999 Dated: March 29, 1999 Received:
Dear Mr. Erbs:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Erbs
this response to your premarket notification Please note: submission does not affect any obliqation you miqht have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labelinq requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
> Sincerely yours, unotis
t Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): _________
Device Name: EasySafe® Retracting Needle Syringe
Indications For Use:
The primary intended use of the EasySafe® Retracting Needle Syringe is to inject fluid into, or withdraw fluid from the body. The secondary intended uses are:
- 1. to help prevent sharps injuries when using the product for its primary intended use, and;
- 2. to help reduce wasted medication when using the product for its primary intended use.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OH
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Paltone Cucente
(Division Sign-Off Division of Dental, Info and General Hospital D 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.