DEN120011 · Proteus Biomedical, Inc. · OZW · Jul 10, 2012 · General Hospital
Device Facts
Record ID
DEN120011
Device Name
PROTEUS INGESTION CONFINMATION SYSTEMS
Applicant
Proteus Biomedical, Inc.
Product Code
OZW · General Hospital
Decision Date
Jul 10, 2012
Decision
DENG
Submission Type
Post-NSE
Regulation
21 CFR 880.6305
Device Class
Class 2
Indications for Use
The Proteus Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relative to gravity, and time-stamped, patientlogged events, including events signaled by swallowing the Ingestion Event Marker (IEM) accessory. The Proteus Personal Monitor enables unattended data collection for clinical and research applications. The Proteus Personal Monitor may be used in any instance where quantiffable analysis of event-associated heart rate, activity, and body position is desirable.
Device Story
System comprises ingestible event marker (IEM), wearable patch, and mobile software. IEM is grain-of-sand sized sensor attached to inert excipient tablet; swallowed by patient. Upon ingestion, IEM communicates via volume conduction (intrabody communication) to wearable patch. Patch collects physiological data (heart rate, activity, body angle) and logs ingestion events. Patch stores and wirelessly transmits data to mobile computing device. Software organizes and displays ingestion events for clinician/researcher review. Used in ambulatory settings; enables unattended data collection. Output allows providers to correlate medication ingestion with physiological/behavioral metrics, potentially improving adherence monitoring and clinical decision-making.
Clinical Evidence
Clinical data from 254 subjects over 3,811 subject-days and 11,655 ingestions. Primary endpoints: Positive Detection Accuracy (PDA) and Negative Detection Accuracy (NDA). Cumulative PDA 97.2% (95% CI); cumulative NDA 100% (95% CI). Safety profile: No unanticipated adverse device effects or severe adverse events related to the system. Minor skin irritation reported with patch use.
Technological Characteristics
System includes ingestible sensor (IEM) and wearable patch. IEM uses volume conduction (intrabody communication). Patch includes accelerometer for activity/body angle and ECG sensors. Biocompatibility per ISO 10993-5, -10, -11. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Patch is internally powered. Software organizes/displays data on mobile devices.
Indications for Use
Indicated for ambulatory patients requiring monitoring of heart rate, activity, body angle, and time-stamped ingestion events. Used for clinical and research applications where quantifiable analysis of event-associated physiological and behavioral metrics is required.
Regulatory Classification
Identification
An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intrabody communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.
Special Controls
In combination with the general controls of the FD&C Act, the Proteus Personal Monitor including Ingestion Event Marker is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible and non-toxic;
(2) Nonclinical, animal, and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device;
(3) Appropriate analysis and nonclinical testing must validate electromagnetic compatibility performance, wireless performance, and electrical safety; and
(4) Labeling must include a detailed summary of the nonclinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
{0}------------------------------------------------
#### EVALUATION OF AUTOMATIC CLASS III DESIGNATION (DE NOVO) FOR PROTEUS PERSONAL MONITOR INCLUDING INGESTION EVENT MARKER
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Ingestible Event Marker - An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intra-body communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.
#### NEW REGULATION NUMBER: 880.6305
CLASSIFICATION: II
PRODUCT CODE: OZW
#### BACKGROUND
## DEVICE NAME: PROTEUS PERSONAL MONITOR INCLUDING INGESTION EVENT MARKER
510(K): K113070
DATE OF 510(K) NSE DECISION: MAY 7, 2012
DATE OF DE NOVO PETITION: MAY 14, 2012
PETITIONER CONTACT: PROTEUS BIOMEDICAL, INC. 2600 BRIDGE PARKWAY, SUITE 101 REDWOOD CITY, CA 94065 Phone: 650-632-4031 Fax: 650-362-1860
#### PETITIONER'S RECOMMENDED CLASSIFICATION: II
#### INDICATIONS FOR USE
The Proteus Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relative to gravity, and time-stamped, patientlogged events, including events signaled by swallowing the Ingestion Event Marker (IEM) accessory. The Proteus Personal Monitor enables unattended data collection for clinical and research applications. The Proteus Personal Monitor may be used in any instance where quantiffable analysis of event-associated heart rate, activity, and body position is desirable.
{1}------------------------------------------------
# LIMITATIONS
Prescription-use only
Caution: Do not wear (the Patch) during magnetic resonance imaging (MRI), cautery, and external defibrillation procedures.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The Proteus Personal Monitor, also called the "Patch", is a body-worn sensor that collects physiological and behavioral metrics including heart rate, activity, body angle and time-stamped user-logged events generated when a user marks an event by swallowing an Ingestion Event Marker (IEM) or by manually pressing an event marker button on the Patch. The Patch stores and wirelessly sends the IEM data to a general computing device.
The Proteus Personal Monitor Ingestion Event Marker system is comprised of three main subsystems; (1) the ingestion event marker (IEM), (2) the data recorder (Patch), and (3) the Proteus software.
- 1. Ingestion Event Marker (IEM)
The grain-of-sand sized IEM is designed to communicate the time-stamped confirmation of IEM device ingestion as a unique identifier to the Proteus Personal Monitor worn on the skin. The ingestion signal is communicated via volume conduction communication also known as intrabody communication. The IEM is attached to an inert pharmaceutical excipient tablet for ease of handling and swallowability.
2. Proteus Personal Monitor (Patch)
The Proteus Personal Monitor (Patch) receives, stores, and wirelessly sends ingestion confirmation data to a general computing device.
## 3. Software
The Proteus software is used to pair the Patch with a mobile computing device. The software organizes and displays ingestion events.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a diagram of a Proteus Personal Monitor (PPM) and an Ingestible Event Marker (IEM). The PPM is a small, wearable sensor that is placed on the skin. The IEM is a small, ingestible sensor that is swallowed. The IEM sends a signal to the PPM when it is ingested. The PPM then records the time and date of ingestion.
Figure 1: An overview of the Proteus Personal Monitor System - IEM (attached to an inert tablet carrier) and Patch, plus a display screen on a paired computing device (not pictured). Magnified view of IEM with attached excipient skirt is also displayed in graphic.
# SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS (IEM)
The petitioner conducted a series of tests to demonstrate that the patient-contacting components of the Proteus Personal Monitor demonstrated acceptable performance for its intended purpose, which included the tests indicated below.
ISO 10993-5 CYTOTOXICITY TESTING ISO 10993-10 IRRITATION TESTING ISO 10993-11 SYSTEMIC TOXICITY TESTING
Given the results of the biocompatibility testing, the petitioner conducted additional animal testing to assess materials toxicity, which included the following tests summarized in Table 1 below.
| | Testing Performed | Results |
|------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Chemical<br>Characterization | HPLC And Spectroscopic<br>Analysis Of Concentrated<br>Device Extracts | No Unintended Compounds<br>Detected Above Stringent ICH<br>Reporting Threshold For Drug<br>Impurities |
| Copper (Cu)<br>Toxicity | Risk Assessment By<br>Gradient Corp (Metal<br>Toxicology Experts) | No Risk Of Cu Toxicity With<br>Realistic Exposure |
| Cytotoxicity | Quantitative Studies With<br>Physiologic Device Extracts | Realistic Exposure Levels Are<br>Non-Cytotoxic |
TABLE 1. PRE-CLINICAL SAFETY TESTING
{3}------------------------------------------------
# IN VIVO STUDIES
The petitioner also performed forty-two (42) in-vivo studies, including rodent, canine and porcine models, to characterize device performance and safety. Porcine and canine animal models are frequently used in gastrointestinal (GI) device testing, and were chosen because of the similarities of their GI anatomy to that of a human. Efforts were made to include a wide range of body size in the non-clinical experiments, with body weight ranging from 25 to 95 kg. The purpose of this inclusion criterion is to provide an opportunity to investigate the potential effects of body size on the performance of the system. Canine testing was also performed to validate that device egestion occurred as well as additional rodent testing and literature review to assess toxicology of materials. A summary of the studies conducted are provided in Table 2.
| | Testing Performed | Results |
|-------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical safety | Excretion and GI injury studies in canines | Ingested IEMs reliably excreted<br>Supra-normal doses of IEMs do not inflict<br>any clinically significant injuries |
| Electrical safety | Tissue stimulation in canines | No abnormal ECG morphology or<br>arrhythmia |
| In vivo toxicity | 14-day rat oral gavage study with physiologic device extracts | No evidence of toxicity in any dosing<br>groups, including max dose group<br>(equivalent to 30,000 IEMs/day), based<br>upon clinical observations, hematology,<br>serum chemistries and histopathology. |
| | Canine oral toxicology study | No evidence of IEM toxicity, based upon<br>clinical observations and GI tract<br>histopathology. No changes in blood<br>levels of IEM inorganic materials<br>following exposure. |
| | Rodent oral toxicology study | No evidence of IEM toxicity—even in<br>highest dosing group, which received the<br>weight-adjusted equivalent of 30,000<br>IEMs/day—based upon clinical<br>observations, hematology, coagulation<br>tests, blood chemistries, necropsy, and<br>comprehensive histopathology. |
| | IEM copper (Cu) human<br>health assessment, general<br>use | Practical-use scenario (15 IEMs ingested<br>simultaneously, daily or twice-daily)<br>poses no risk of copper toxicity. Extreme-<br>use scenario (30 IEMs ingested<br>simultaneously, daily) poses no risk of<br>systemic toxicity, but transient, non-<br>systemic gastric upset could result at this<br>dose. This concentration dependent effect |
# TABLE 2. PERFORMANCE AND SAFETY TESTING
{4}------------------------------------------------
| IEM copper human health<br>assessment, chronic use in<br>a compromised population<br>(renal transplant patients) | would be mitigated by intake with a meal.<br>Post-operative renal transplant patients are<br>not at greater risk than the normal<br>population from Cu toxicity associated<br>with chronic ingestion of four IEMs/day.<br>There is no scientific basis to believe that<br>the physiological response to Cu in IEM-<br>enabled medicines will differ from the<br>physiological response to Cu in food. |
|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Quantitative cytotoxicity | Corroborates conclusion of IEM Cu<br>human health assessment. |
| Additional chemical<br>characterizations | No unintended compounds detected above<br>reporting threshold for new drug<br>substances, a stringent standard that was<br>adapted for analysis of the IEM device. |
# ELECTROMAGNETIC COMPATIBILITY (EMC) AND ELECTRICAL SAFETY
Electromagnetic compatibility and electrical safety testing were performed to FDA recognized standards. All applicable tests passed.
| Testing<br>Category | Test Descriptions | Reference | Results |
|------------------------------|---------------------------------------------------------------------------|---------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Electrical<br>safety testing | Power input | IEC 60601-1, Sub-<br>clause 7.1 | Not applicable,<br>because the unit is<br>internally powered |
| | Limitation of voltage<br>and/or energy | IEC 60601-1, Sub-<br>clause 15 b | Not applicable,<br>because the unit is<br>internally powered |
| | Protective earthing,<br>functional earthing and<br>potential equalization | IEC 60601-1, Sub-<br>clause 18 f | Not applicable,<br>because the unit is<br>internally powered<br>and has a non-<br>conductive<br>enclosure |
| | Earth leakage current | IEC 60601-1, Sub-<br>clause 19.4 f | Not applicable,<br>because the unit is<br>internally powered |
| | Enclosure leakage<br>current | IEC 60601-1, Sub-<br>clause 19.4 g | Not applicable,<br>because the unit is<br>internally powered<br>and has a non-<br>conductive<br>enclosure |
| | Patient leakage current | IEC 60601-1, Sub-<br>clause 19.4 h, 6 | Passed, 0 μA<br>measured |
TABLE 3. EMC AND ELECTRICAL SAFETY TESTING
{5}------------------------------------------------
| | Patient leakage with<br>mains voltage on F-<br>type isolated applied<br>parts | IEC 60601-1, Sub-clause 19.4 h.6 | Passed, 0 $\u03bc$ A r.m.s measured |
|--------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| | Patient auxiliary<br>current | IEC 60601-1, Sub-clause 19.4 j | Passed, maximum<br>of < $1$ $\u03bc$ A r.m.s measured |
| | Dielectric voltage<br>withstand | IEC 60601-1, Sub-clause 20.4 | Passed |
| | Reversed battery<br>connection | IEC 60601-1, Sub-clause 56.7 | Not applicable;<br>battery is not user-accessible |
| | Overflow, spillage,<br>leakage, cleaning | IEC 60601-1, Sub-clause 44 | Passed |
| | Creepage distances and<br>air clearances | IEC 60601-1, Sub-clause 57.10 | Passed |
| EMC testing | Group 1 Class B –<br>Radiated emissions | per EN 60601-1-2,<br>EN 55011 (CISPR<br>11) | Passed |
| IEM EMC<br>testing | Electromagnetic<br>compatibility (EMC) | IEC 60601-1-2:2007<br>6.2.3 Radiated RF<br>electromagnetic<br>fields Part 4-3:<br>Testing and<br>measurement<br>techniques –<br>Radiated, radio-<br>frequency,<br>electromagnetic field<br>immunity test | Passed |
# ADDITIONAL DEVICE CHARACTERIZATION AND PERFORMANCE TESTING
The sponsor performed testing to characterize the performance of the ingestible disc antenna used in the IEM (Table 4) and simulation testing signal reception performance of the Patch data recorder (Table 5).
TABLE 4. INGESTIBLE DISC ANTENNA TESTING
| Test Description | Result |
|------------------------------------------------------------------|--------|
| Mechanical strength - immersion in SGF<br>for 10 minutes at 37°C | (b) (4 |
| Residual solvent of the disc material | |
| Friability - immersion in SGF (10 mins)<br>then SIF at 37°C | |
| Electrical properties of disc | |
{6}------------------------------------------------
| Test | Test Description | Result |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| High Frequency (HF)<br>Signal Chain<br>Performance Test | A “body simulation” network was<br>interposed between the signal<br>source and the Data Recorder | The passband is<br>substantially flat<br>between 10Hz and<br>80Hz |
| Low Frequency (LF)<br>Signal Chain<br>Performance Test | A patient simulator was attached<br>to the inputs and the amplitude of<br>the output was measured by an<br>oscilloscope | The passband is<br>substantially flat<br>between 2Hz and<br>100Hz |
| Accelerometer<br>Performance Tests | Rotating the accelerometer with<br>respect to gravity<br>--<br>Acceleration measurement from a<br>subject during a steady walk | Measured values<br>agreed well with the<br>applied values R2<br>=.99<br>--<br>The acceleration<br>traces appear to be of<br>a subject walking. |
| ECG Performance<br>Testing | The algorithm was tested against<br>all 48 test files from the MIT-BIH<br>arrhythmia database.<br>--<br>The PROMITTER substudy was<br>conducted on the Proteus campus<br>and enrolled 5 healthy volunteer<br>subjects | The Median was<br>99.7% detection with<br>a 5.9% standard<br>deviation<br>--<br>ECG results and<br>accuracy was 99.4%<br>for chest location and<br>99.2% for xyphoid<br>location |
| Respiratory Rate<br>Performance Testing | R-wave amplitude is modulated by<br>the respiratory cycle, a stationary<br>subject was instructed to breath<br>regularly at a rate of 6 breaths/min | Device measurement<br>result of 6<br>breaths/minute |
TABLE 5. DATA RECORDER TESTING
## MAGNETIC RESONANCE (MR) COMPATIBILITY
No testing has been conducted to demonstrate whether the device is MR compatible. The labeling has included a Caution that the user should not wear the Patch during magnetic resonance imaging (MRI).
## SOFTWARE
The petitioner provided a description of software development processes, software hazard analysis and device system performance testing. The Patch software was reviewed in K093976 in conformance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005.
{7}------------------------------------------------
# SUMMARY OF CLINICAL INFORMATION
The Proteus system has been used by > 250 patients who participated in >3,800 cumulative days of system use involving >11,500 cumulative IEM ingestions as summarized in Tables 6 and 7 below. The studies characterized the safety and performance of the Proteus system. The safety of the system was characterized by recording all patient adverse events (AEs) noted during the study whether device related or not. The key measures of system performance (positive detection accuracy (PDA) and negative detection accuracy (NDA)) characterize the ability of the system to properly detect and register IEM ingestions.
The cumulative average of PDA across all conducted studies is 97.2% (95% CI). The cumulative average of NDA across all conducted studies is 100% (95% CI).
| TABLE 6. RESULTS OF HUMAN CLINICAL TESTING | | | |
|----------------------------------------------------------------------------------------------------------------------------------------|--------|-----------------------------|-------------------------------|
| Cumulative Clinical Experience | N | | |
| Number of subjects wearing the<br>Proteus Personal Monitor (Patch) | 254 | | |
| Number of subjects ingesting IEM | 219 | | |
| Number of subject/days | 3,811 | | |
| Number of IEM ingestions | 11,655 | | |
| No unanticipated adverse device effects, no severe adverse events<br>related to or possibly related to Proteus Personal Monitor System | | | |
| Non-serious AEs--92% mild, 8%<br>moderate | | | |
| Adverse Event (AE) – Ingestible<br>Sensor | | Rate as<br>% of<br>subjects | Rate as %<br>of<br>ingestions |
| At least one AE | 0 | 0 | 0 |
| At least one severe AE | 0 | 0 | 0 |
| Discontinued due to AE | 0 | 0 | 0 |
| Adverse Events | | | |
| Nausea/vomiting | 4 | 1.8% | 0.0% |
| Related | 1 | 0.5% | 0.0% |
| Constipation | 2 | 0.9% | 0.0% |
| Anxiety | 1 | 0.5% | 0.0% |
| Asthma attack | 1 | 0.5% | 0.0% |
| Abdominal cramping | 1 | 0.5% | 0.0% |
| Non-cardiac chest pain | 1 | 0.5% | 0.0% |
| Bitter taste in mouth | 1 | 0.5% | 0.0% |
| Adverse Event – PPM (510(k)<br>cleared component) | | | |
| Localized skin irritation and<br>inflammation | 45 | 17.7% | NA |
| Discontinued due to skin irritation | 7 | 2.8% | NA |
TABLE 6. RESULTS OF HUMAN CLINICAL TESTING
{8}------------------------------------------------
# TABLE 7. SUMMARY OF HUMAN CLINICAL TESTING
| Overall System Performance | |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| 219 subjects of 254 subjects,<br>inclusive of PPM-only users (IEM<br>ingestion) | 99.3% Detection accuracy<br>100% Correct identification<br>No SAEs / UADEs related to system |
| 11,655 ingestions | |
| 3810 subject-days of system<br>utilization | |
| Maximum daily ingestion: 34 IEMs | |
| Maximum system utilization: 42<br>days | |
# LABELING
Labeling includes all information required for the safe and effective use of the device as outlined in 801.109, including a detailed summary of the non-clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
## RISKS TO HEALTH
Table 8 below identifies the risks to health that may be associated with use of Ingestible Event Markers and the measures recommended to mitigate these risks.
| Identified Risks | Recommended Mitigation Measures |
|---------------------------------|-----------------------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility Testing<br>Labeling (dose limits) |
| Systemic toxicity | Toxicology Testing<br>Labeling (dose limits) |
| Electromagnetic incompatibility | Electromagnetic Compatibility Testing<br>Wireless testing<br>Labeling |
| Electrical safety issues | Electrical Safety Testing<br>Labeling |
| Electrical/Mechanical failure | Non-clinical Performance Testing |
| Failure to mark event | Non-clinical Performance Testing<br>Clinical Evaluation |
| Failure to excrete | Animal Testing |
| Usability | Human Factors Testing<br>Labeling |
Table 8 Risk/Mitigation Measures
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the Proteus Personal Monitor including Ingestion Event Marker is subject to the following special controls:
{9}------------------------------------------------
- 1. The device must be demonstrated to be biocompatible and non-toxic;
- 2. Non-clinical, animal and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device;
- 3. Appropriate analysis and non-clinical testing must validate electromagnetic compatibility (EMC) performance, wireless performance, and electrical safety; and
- 4. Labeling must include a detailed summary of the non-clinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
#### BENEFIT/RISK DETERMINATION
The Benefit/Risk Determination for the Proteus Personal Monitor finds that although the benefits realized by the use of the device system are small, the risks posed by the device system are also small and pose little to no risk to the patient when Special Controls are met and are outweighed by the benefits of the device system.
#### CONCLUSION
The de novo petition for the Proteus Personal Monitor including Ingestion Event Marker is granted and the device is classified under the following:
Product Code: OZW Device Type: Proteus Personal Monitor including Ingestion Event Marker Class: II Regulation: 21 CFR 880.6305
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.