OZW · Ingestible Event Marker

General Hospital · 21 CFR 880.6305 · Class 2

Overview

Product CodeOZW
Device NameIngestible Event Marker
Regulation21 CFR 880.6305
Device ClassClass 2
Review PanelGeneral Hospital

Identification

An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intrabody communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the Proteus Personal Monitor including Ingestion Event Marker is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible and non-toxic; (2) Nonclinical, animal, and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device; (3) Appropriate analysis and nonclinical testing must validate electromagnetic compatibility performance, wireless performance, and electrical safety; and (4) Labeling must include a detailed summary of the nonclinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.

Recent Cleared Devices (8 of 8)

RecordDevice NameApplicantDecision DateDecision
K251088Otsuka Digital Feedback DeviceOtsuka America Pharmaceutical, Inc.Jun 30, 2025SESE
K223463Otsuka Digital Feedback Device-RWOtsuka America Pharmaceutical, Inc.Aug 11, 2023SESE
K183052ID-Cap SystemEtectrx, Inc.Dec 6, 2019SESE
K150494Proteus Digital Health Feedback DeviceProteus Digital Health, Inc.Jun 27, 2015SESE
K133263PROTEUS PATCH, PROTEUS INGESTIBLE SENSOR (ACCESSORY)Proteus Digital Health, Inc.Feb 7, 2014SESE
K131524PROTEUS PATCH INCLUDING INGESTIBLE SENSORProteus Digital Health, Inc.Jun 23, 2013SESE
K131009PROTEUS PATCH, INGESTIBLE SENSOR (ACCESSORY)Proteus Digital Health, Inc.May 7, 2013SESE
DEN120011PROTEUS INGESTION CONFINMATION SYSTEMSProteus Biomedical, Inc.Jul 10, 2012DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...