AllTest Multi-Drug Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel
Applicant
Hangzhou AllTest Biotech Co., Ltd.
Product Code
NFT · Clinical Toxicology
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Intended Use
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of: [Table of cutoffs]. AllTest Multi-Drug Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use. The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
Device Story
Lateral flow immunochromatographic assay for qualitative detection of drugs of abuse in human urine; competitive binding principle. Urine sample absorbed via capillary action; mixes with drug-specific monoclonal antibody conjugate; flows across pre-coated membrane. Negative result: colored test line appears (drug level below cutoff). Positive result: no colored test line (drug level at/above cutoff). Control line confirms proper procedure. Used in OTC or clinical settings; operated by lay users or healthcare providers. Provides preliminary screening results; requires confirmatory testing via GC/MS or LC/MS. Assists in identifying potential drug presence for clinical consideration.
Clinical Evidence
Bench testing only. Precision/reproducibility studies performed over 25 days with three lots. Analytical specificity/interference evaluated against various metabolites and common compounds. Method comparison study conducted with 80 clinical samples per drug compared to LC-MS/MS. Lay-person study (n=140) conducted at three sites; participants demonstrated high agreement across various drug concentrations relative to cutoffs.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding. Materials include chromatographic absorbent device, drug monoclonal antibody (mouse) conjugate, and drug-protein conjugate. Ready-to-use panel format; sealed with desiccant in aluminum pouch. No electronic components or software. Stability: 2-30°C for 24 months.
Indications for Use
Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, MDMA, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana, Fentanyl) in human urine. Intended for OTC and prescription use. Not for distinguishing between prescription use and abuse. Preliminary results only; requires GC/MS or LC/MS confirmation.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K180878 — BIO-VENTURE Rapid Amphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Oxazepam Test Cassette for OTC Use, BIO-VENTURE Rapid Cocaine Test Cassette for OTC Use, BIO-VENTURE Rapid Methamphetamine Test Cassette for OTC Use, BIO-VENTURE Rapid Morphine Test Cassette for OTC Use, BIO-VENTURE Rapid Marijuana Test Cassette for OTC Use, BIO-VENTURE Rapid Amphetamine Test Cassette for Rx Use, BIO-VENTURE Rapid Oxazepam Test Cassette for Rx Use, BIO-VENTURE Rapid Cocaine Test Cassette for Rx · Shanghai Venture Bio-Tech Co., Ltd. · Dec 17, 2018
K182701 — Wondfo T-Cup Multi-Drug Urine Test Cup · Guangzhou Wondfo Biotech Co., Ltd. · Dec 3, 2018
K243996 — AssureTech Panel Dip Tests; AssureTech Quick Cup Tests; AssureTech Multi-drug Urine Test Panel; AssureTech Multi-drug Urine Test Cup · Assure Tech., LLC · Feb 7, 2025
K182530 — BIOEASY Multi-Drug Test Cup · Shenzhen Bioeasy Biotechnology Co., Ltd. · Nov 9, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hangzhou Alltest Biotech Co.,Ltd % Jenny Xia, Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877
Re: K242540
Trade/Device Name: AllTest Multi-Drug Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, NGM, OAW, NFW Dated: August 23, 2024 Received: August 26, 2024
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Joseph A. Digitally signed by Kotarek -S Date: 2024.09.27
> Kotarek -S Date: 2024.09.20 Joseph Kotarek Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242540
Device Name
AllTest Multi-Druq Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel
Indications for Use (Describe)
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off leve |
|--------------------------------------------------------------|--------------|
| Amphetamine (AMP) | 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP/OPI) | 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Fentanyl(FYL) | 1 ng/mL |
| AllTost Multi-Drug Urine Test Panel can be a single drug tes | |
rug Urine Test Panel can be a single druq test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Rapid Urine Test Panel tests are competitive binding, lateral flow
immunochromatographic assays for qualitative and simultaneous detection of Amphetamine,
Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine,
Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off (ng/mL) |
|---------------------|---------------|-----------------|
| Amphetamine (AMP) | d-Amphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Oxazepam (BZO) | Oxazepam | 300 |
{3}------------------------------------------------
| Cocaine (COC) | Benzoylecgonine | 150 |
|--------------------------------------|-----------------------------------|------|
| Methamphetamine (MET) | d-Methamphetamine | 500 |
| Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
| Morphine (MOP/OPI) | Morphine | 300 |
| Methadone (MTD) | Methadone | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Nortriptyline (TCA) | Nortriptyline | 1000 |
| Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
| Fentanyl(FYL) | Fentanyl | 1 |
AllTest Multi-Drug Rapid Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinquish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY K242540
- August 23, 2024 Date: 1. Hangzhou AllTest Biotech Co., Ltd. 2. Submitter: Plant Bldg. 3, 4, 5, No. 550 Yinhai Street, Baiyang Street, Hangzhou ETDZ, Jianggan District 3. Contact person: Jenny Xia LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 301-525-6856 Email: jxia@lsi-consulting.org
- 4. Device Name: AllTest Multi-Drug Urine Test Panel AllTest Multi-Drug Rapid Urine Test Panel Classification: Class II ਨ.
| Product Code<br>Target Drug | Regulation Section | Panel |
|---------------------------------------------------------------------------------------|---------------------------------------------------------|------------|
| NFT<br>Amphetamine (AMP) | 862.3100, Amphetamine Test System | Toxicology |
| PTH<br>Secobarbital (BAR) | 862.3150, Barbiturate Test System | Toxicology |
| NGL<br>Buprenorphine (BUP)<br>Fentanyl (FYL)<br>Morphine (MOP/OPI)<br>Oxycodone (OXY) | 862.3650, Opiate Test System | Toxicology |
| NFV<br>Oxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFY<br>Cocaine (COC) | 862.3250, Cocaine and cocaine<br>metabolite test system | Toxicology |
| PTG<br>Methadone (MTD) | 862.3620, Methadone Test System | Toxicology |
| NGG<br>Methylenedioxymethamphetamine<br>(MDMA)<br>Methamphetamine (MET) | 862.3610,<br>Methamphetamine Test System | Toxicology |
| NGM<br>Phencyclidine (PCP) | Unclassified | Toxicology |
| QAW<br>Nortriptyline (TCA) | 862.3910 Tricyclic antidepressant drugs<br>test system | Toxicology |
{5}------------------------------------------------
| NFW<br>Cannabinoids (THC) | 862.3870, Cannabinoids Test System | Toxicology |
|---------------------------|------------------------------------|------------|
|---------------------------|------------------------------------|------------|
#### 6. Predicate Devices:
AllTest Multi-Drug Urine Test Cup (K241428)
#### 7. Intended Use
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|--------------------------------------|---------------|
| Amphetamine (AMP) | 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP/OPI) | 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Fentanyl(FYL) | 1 ng/mL |
AllTest Multi-Drug Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.
{6}------------------------------------------------
AllTest Multi-Drug Rapid Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine,
Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off<br>(ng/mL) |
|------------------------------------------|---------------------------------------|--------------------|
| Amphetamine (AMP) | d-Amphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Oxazepam (BZO) | Oxazepam | 300 |
| Cocaine (COC) | Benzoylecgonine | 150 |
| Methamphetamine<br>(MET) | d-Methamphetamine | 500 |
| Methylenedioxymetham<br>phetamine (MDMA) | d,l-<br>Methylenedioxymethamphetamine | 500 |
| Morphine (MOP/OPI) | Morphine | 300 |
| Methadone (MTD) | Methadone | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Nortriptyline (TCA) | Nortriptyline | 1000 |
| Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
| Fentanyl (FYL) | Fentanyl | 1 |
AllTest Multi-Drug Rapid Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.
#### Device Description 8.
AllTest Multi-Drug Urine Test Panel and AllTest Multi-Drug Rapid Urine Test Panel are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a panel format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
{7}------------------------------------------------
| Similarities | | |
|--------------------------------------|--------------------------------------------------------------------------------------------------|------------------------|
| Item | Candidate Device<br>(K242540) | Predicate<br>(K241428) |
| Intended use | Qualitative detection of drugs of abuse in urine. For prescription use or over-the-counter use | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assay based on antigen-antibody reaction | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Target Drug<br>and Cut Off<br>Values | Target Drugs | Cutoff (ng/mL) Same |
| | Amphetamine(AMP) | 500 |
| | Secobarbital (BAR) | 300 |
| | Buprenorphine (BUP) | 10 |
| | Oxazepam (BZO) | 300 |
| | Cocaine (COC) | 150 |
| | Methylenedioxymethamphetamine (MDMA) | 500 |
| | Methamphetamine (MET) | 500 |
| | Morphine (MOP/OPI) | 300 |
| | Methadone (MTD) | 300 |
| | Oxycodone (OXY) | 100 |
| | Phencyclidine (PCP) | 25 |
| | Nortriptyline (TCA) | 1000 |
| | Cannabinoids (THC) | 50 |
| | Fentanyl (FYL) | 1 |
| Differences | | |
| Configuration | Test Panel | Test Cup |
#### Substantial Equivalence Information 0
## 10. Standard/Guidance Document Reference (if applicable)
None referenced.
## 11. Test Principle
AllTest Multi-Drug Urine Test Panel or AllTest Multi-Drug Rapid Urine Test Panel is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.
When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When a drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test
{8}------------------------------------------------
Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.
When sample drug levels are at or above the detection level of the tree drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.
## 12. Performance Characteristics
#### A. Analytical performance
#### a. Precision/Reproducibility:
Precision studies were carried out for samples with concentrations of +100% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Samples were prepared by spiking target drugs in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test panels. The results obtained are summarized in the following tables:
| Drug | Lot<br>Number | +100%<br>cutoff | +75%<br>cutoff | +50%<br>cutoff | +25%<br>cutoff | Cutoff | -25%<br>cutoff | -50%<br>cutoff | -75%<br>cutoff | -100%<br>cut-off |
|-------------|---------------|-----------------|----------------|----------------|----------------|---------|----------------|----------------|----------------|------------------|
| AMP<br>500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| BAR<br>300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| BUP 10 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| BZO<br>300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| MDMA<br>500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| MET<br>500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
{9}------------------------------------------------
| MOP/ | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
|-------------|-------|--------|--------|--------|--------|---------|--------|--------|--------|--------|
| OPI<br>300 | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| MTD<br>300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| OXY<br>100 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| PCP 25 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| TCA<br>1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| THC<br>50 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| COC<br>150 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 28+/22- | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 29+/21- | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 49+/1- | 26+/24- | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| FYL 1 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 20-/30+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
## b. Linearity/assay reportable range:
Not applicable. This device is intended for qualitative use only.
## c. Stability:
The device is stable at 2-30°C for 24 months based on real time stability study.
## d. Analytical specificity/Interference:
To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine samples were tested using three lots of the device.
Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.
| Drug/Cutoff | Compound | Minimum concentration required to obtain a positive result | % Cross-Reactivity |
|--------------------------------|----------------------------------------------------------|------------------------------------------------------------|--------------------|
| | | (ng/mL) | |
| | Hydroxyamphetamine | 5000 | 10% |
| | (+/-)-<br>Methylenedioxyamphetamine(MD<br>A) | 25 | 2000% |
| | D,L-Amphetamine | 500 | 100% |
| | D-Amphetamine | 500 | 100% |
| | Diethylstilbestrol | >100000 | <0.5% |
| | L-Amphetamine | 500 | 100% |
| | Phentermine | 500 | 100% |
| | β-Phenylethylamine | >100000 | <0.5% |
| | Tyramine | >100000 | <0.5% |
| | p-Hydroxynorephedrine | >100000 | <0.5% |
| | D,L-Norephedrine | >100000 | <0.5% |
| AMP 500 | p-Hydroxyamphetamine | 5000 | 10% |
| | D-Methamphetamine | >100000 | <0.5% |
| | L-Methamphetamine | >100000 | <0.5% |
| | Ephedrine hydrochloride | >100000 | <0.5% |
| | (+/-)3,4-<br>Methylenedioxymethamphetamine<br>(MDMA) | 100000 | 0.5% |
| | Phenylpropanolamine | >100000 | <0.5% |
| | Benzphetamine | >100000 | <0.5% |
| | L-Ephedrine | >100000 | <0.5% |
| | L-Epinephrine | >100000 | <0.5% |
| | D,L-Epinephrine | >100000 | <0.5% |
| | (+/-)3,4-<br>Methylenedioxyethylamphetamine<br>(MDEA) | 100000 | 0.5% |
| | Alphenal | 300 | 100% |
| | Amobarbital | 300 | 100% |
| BAR 300 | Aprobarbital | 300 | 100% |
| | Butabarbital | 500 | 60% |
| | Butethal | 200 | 150% |
| | Cyclopentobarbital | 500 | 60% |
| | Pentobarbital | 200 | 150% |
| | Phenobarbital | 200 | 150% |
| | Secobarbital | 300 | 100% |
| | Butalbital | 2000 | 15% |
| | Barbital | 300 | 100% |
| | Buprenorphine | 10 | 100% |
| | Buprenorphine-3-D-Glucuronide | 20 | 50% |
| | Norbuprenorphine | 20 | 50% |
| | Norbuprenorphine-3-D- | | 50% |
| | Glucuronide | 20 | |
| BUP 10 | Morphine | >100000 | <0.01% |
| | Oxymorphone | >100000 | <0.01% |
| | Hydromorphone | >100000 | <0.01% |
| | Codeine | >100000 | <0.01% |
| | Nalorphine | >100000 | <0.01% |
| | a-Hydroxyalprazolam | 10000 | 3% |
| | Alprazolam | 200 | 150% |
| | Bromazepam | 1000 | 30% |
| | Chlordiazepoxide | 300 | 100% |
| | Clobazam | 200 | 150% |
| | Clonazepam | 200 | 150% |
| | Clorazepate Dipotassium | 300 | 100% |
| | Desalkylflurazepam | >100000 | <0.3% |
| | Diazepam | 300 | 100% |
| BZO 300 | Estazolam | 100 | 300% |
| | Flunitrazepam | 300 | 100% |
| | D,L-Lorazepam | 1000 | 30% |
| | Midazolam | 1000 | 30% |
| | Nitrazepam | 200 | 150% |
| | Norchlordiazepoxide | 200 | 150% |
| | Nordiazepam | 200 | 150% |
| | Oxazepam | 300 | 100% |
| | R.S-Lorazepam glucuronide | 1000 | 30% |
| | Temazepam | 100 | 300% |
| | Triazolam | 300 | 100% |
| | Demoxepam | 2000 | 15% |
| | Flurazepam | 300 | 100% |
| | Delorazepam | 2000 | 15% |
| | Lormetazepam | 100 | 300% |
| COC 150 | Benzoylecogonine | 150 | 100% |
| | Cocaethylene | 250 | 60% |
| | Cocaine hydrochloride | 125 | 120% |
| | Ecgonine | >100000 | <0.2% |
| | Norcocaine | >100000 | <0.2% |
| | Ecgonine methyl ester | >100000 | <0.2% |
| | (+/-)3,4-Methylenedioxy-n-<br>ethylamphetamine<br>(MDEA) | 250 | 200% |
| | (+/-)-<br>Methylenedioxyamphetamine(MD<br>A) | 1000 | 50% |
| | (±)-MDMA | 500 | 100% |
| | L-Methamphetamine | >100000 | <0.5% |
| MDMA 500 | d-methamphetamine | >100000 | <0.5% |
| | d-amphetamine | >100000 | <0.5% |
| | l-amphetamine | >100000 | <0.5% |
| | d,l-Amphetamine | >100000 | <0.5% |
| | d,l-Methamphetamine | >100000 | <0.5% |
| | Phentermine | >100000 | <0.5% |
| | Phenylephrine | >100000 | <0.5% |
| | Ephedrine | >100000 | <0.5% |
| | Pseudoephedrine | >100000 | <0.5% |
| MET 500 | (+/-)3,4-Methylenedioxy-n-<br>ethylamphetamine<br>(MDEA) | 6000 | 8.3% |
| | (±)-MDMA | 1000 | 50% |
| | D-Methamphetamine | 500 | 100% |
| | L-Methamphetamine | 10000 | 5% |
| | Fenfluramine | 50000 | 1% |
| | p-Hydroxymethamphetamine | 250 | 200% |
| | D,L-Methamphetamine | 250 | 200% |
| | β-Phenylethylamine | >100000 | <0.5% |
| | Mephetermine | 25000 | 2% |
| | Methoxyphenamine hydrochloride | 75000 | 0.7% |
| | L-Amphetamine | 100000 | 0.5% |
| | D-Amphetamine | 100000 | 0.5% |
| | D,L-Amphetamine | 250 | 200% |
| | Chloroquine | 25000 | 2.0% |
| | Ephedrine hydrochloride | >100000 | <0.5% |
| | (+/-)3,4-<br>Methylenedioxyamphetamine<br>(MDA) | >100000 | <0.5% |
| | Trimethobenzamide | >100000 | <0.5% |
| | l-phenylephrine | >100000 | <0.5% |
| | (1R,2S)-(-)-Ephedrine | >100000 | <0.5% |
| | Procaine hydrochloride | >100000 | <0.5% |
| | Phentermine | >100000 | <0.5% |
| | Pseudoephedrine | >100000 | <0.5% |
| | 6-acetylmorphine | 300 | 100% |
| | Codeine | 250 | 120% |
| | Dihydrocodeine | 5000 | 6% |
| | EthylMorphine | 200 | 150% |
| | Heroin | 500 | 60% |
| | Hydrocodone | 1000 | 30% |
| MOP/OPI<br>300 | Hydromorphone | 10000 | 3% |
| | Levorphanol tartrate | 1500 | 20% |
| | Morphine | 300 | 100% |
| | Nalorphine hydrochloride | 300 | 100% |
| | Thebaine | 20000 | 1.5% |
| | s-Monoacetylmorphine | 300 | 100% |
| | Morphine-3-β-d-glucuronide | 1000 | 30% |
| | 6-Monoacetylmorphine (6-MAM) | 300 | 100% |
| | Codeine-6-β-D-glucuronide | 25000 | 1.2% |
| | Morphine-6-β-D-glucuronide | 1000 | 30% |
| | 6-Acetylcodeine | 25000 | 1.2% |
| | Normorphine | 50000 | 0.6% |
| | Oxycodone | >100000 | <0.3% |
| | Oxymorphone | >100000 | <0.3% |
| | Norcodeine | 25000 | 1.2% |
| | Procaine hydrochloride | >100000 | <0.3% |
| | Norpropoxyphene | >100000 | <0.3% |
| MTD 300 | (±)-Methadone | 300 | 100% |
| | EDDP | >100000 | <0.3% |
| | EMDP | >100000 | <0.3% |
| | LAAM | >100000 | <0.3% |
| | Alpha Methadol | >100000 | <0.3% |
| | Doxylamine | >100000 | <0.3% |
| | Disopyramide | 6000 | 5.0% |
| | Esomeprazole | 10000 | 3.0% |
| | Pheniramine | 100000 | 0.3% |
| OXY 100 | Ethyl Oxycodone | 100 | 100% |
| | Hydrocodone | 50000 | 0.2% |
| | Hydromorphone | 25000 | 0.4% |
| | Levorphanol tartrate | 50000 | 0.2% |
| | Naloxone hydrochloride | 50000 | 0.2% |
| | Naltrexone hydrochloride | >100000 | <0.1% |
| | Oxycodone | 100 | 100% |
| | Oxymorphone | 100 | 100% |
| | Oxymorphone-3β-D-glucuronide | >100000 | <0.1% |
| | Noroxycodone | 1000 | 10% |
| | Noroxymorphone | 25000 | 0.4% |
| | Dihydrocodeine | 12500 | 0.8% |
| | Codeine | 100000 | 0.1% |
| | Morphine | >100000 | <0.1% |
| | Acetylmorphine | >100000 | <0.1% |
| | Buprenorphine | >100000 | <0.1% |
| | Ethylmorphine | >100000 | <0.1% |
| | Thebaine | >100000 | <0.1% |
| | 6-acetylmorphine | >100000 | <0.1% |
| PCP 25 | PCP (Phencyclidine) | 25 | 100% |
| | 4-Hydroxyphencyclidine | 15 | 167% |
| | Pheniramine | 50000 | 0.05% |
| TCA 1000 | Amitriptyline | 1500 | 66.7% |
| | Chlorpheniramine | >100000 | <1% |
| | Clomipramine | 1000 | 100% |
| | Cyclobenzaprine Hydrochloride | >100000 | <1% |
| | Desipramine | 300 | 333.3% |
| | Doxepine | 10000 | 10% |
| | Duloxetine | >100000 | <1% |
| | Imipramine | 500 | 200% |
| | Norclomipramine | 250 | 400% |
| | Nordoxepine | >100000 | <1% |
| | Nortriptyline hydrochloride | 1000 | 100% |
| | Promazine | 500 | 200% |
| | Trimipramine | 1500 | 66.7% |
| | Maprotiline | >100000 | <1% |
| | Promethazine hydrochloride | 50000 | 2% |
| THC 50 | 11-nor-Δ8-THC-9-COOH | 20000 | 0.25% |
| | (-)-11-nor-9-carboxy-Δ9-THC | 50 | 100% |
| | (±)-11-nor-9-Carboxy-Δ9-THC | 50 | 100% |
| | 11-nor-Δ9-THC -carboxy<br>glucuronide | 50 | 100% |
| | 11-hydroxy-Δ9-<br>Tetrahydrocannabinol | 5000 | 1.0% |
| | Δ8- Tetrahydrocannabinol | 100000 | 0.05% |
| | Δ9- Tetrahydrocannabinol | 100000 | 0.05% |
| | Cannabinol | >100000 | <0.05% |
| | Cannabidiol | >100000 | <0.05% |
| FYL 1 | Fentanyl | 1 | 100% |
| | Acetyl fentanyl | 1 | 100% |
| Acrylfentanyl | 1 | 100% | |
| ω-1-Hydroxyfentanyl | 20000 | 0.005% | |
| Isobutyryl fentanyl | 1 | 100% | |
| Ocfentanil | 2.5 | 40% | |
| Butyryl fentanyl | 2.5 | 40% | |
| Furanyl fentanyl | 5 | 20% | |
| Valeryl fentanyl | 10 | 10% | |
| (±) β-hydroxythiofentanyl | 2 | 50% | |
| 4-Fluoro-isobutyrylfentanyl | 50 | 2% | |
| Para-fluorobutyryl fentanyl | 4 | 25% | |
| Para-fluoro fentanyl | 3 | 33.3% | |
| (±)-3-cis-methyl fentanyl | 50 | 2% | |
| Carfentanil | 2 | 50% | |
| Sufentanil | 10 | 10% | |
| Alfentanil | 5000 | 0.02% | |
| Despropionyl fentanyl (4-ANPP) | 2500 | 0.04% | |
| Cyclopropylfentanyl | 3 | 33.3% | |
| Remifentanil | >100000 | <0.001% | |
| Norfentanyl | >100000 | <0.001% | |
| Acetyl norfentanyl | >100000 | <0.001% | |
| Norcarfentanil | >100000 | <0.001% | |
| 6-Acetyl morphine | >100000 | <0.001% | |
| Amphetamine | >100000 | <0.001% | |
| Buprenorphine | >100000 | <0.001% | |
| Buprenorphineglucuronide | >100000 | <0.001% | |
| Codeine | >100000 | <0.001% | |
| Dextromethorphan | >100000 | <0.001% | |
| Dihydrocodeine | >100000 | <0.001% | |
| EDDP | >100000 | <0.001% | |
| EMDP | >100000 | <0.001% | |
| Fluoxetine | >100000 | <0.001% | |
| Heroin | >100000 | <0.001% | |
| Hydrocodone | >100000 | <0.001% | |
| Hydromorphone | >100000 | <0.001% | |
| | | | |
| | Ketamine | >100000 | <0.001% |
| | Levorphanol | >100000 | <0.001% |
| | Meperidine | >100000 | <0.001% |
| | Methadone | >100000 | <0.001% |
| | Morphine | >100000 | <0.001% |
| | Morphine-3-glucuronide | >100000 | <0.001% |
| | Naloxone | >100000 | <0.001% |
| | Naltrexone | >100000 | <0.001% |
| | Norbuprenorphine | >100000 | <0.001% |
| | Norcodeine | >100000 | <0.001% |
| | Norketamine | >100000 | <0.001% |
| | Normeperidine | >100000 | <0.001% |
| | Normorphine | >100000 | <0.001% |
| | Noroxycodone | >100000 | <0.001% |
| | Oxycodone | >100000 | <0.001% |
| | Oxymorphone | >100000 | <0.001% |
| | Pentazocine (Talwin) | >100000 | <0.001% |
| | Pipamperone | >100000 | <0.001% |
| | Risperidone | >100000 | <0.001% |
| | Tapentadol | >100000 | <0.001% |
| | Thioridazine | >100000 | <0.001% |
| | Tilidine | >100000 | <0.001% |
| | Tramadol | >100000 | <0.001% |
| | Tramadol-O-Desmethyl | >100000 | <0.001% |
| | Tramadol-N-Desmethyl | >100000 | <0.001% |
| | Isotonitaze | >100000 | <0.001% |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.
Compounds that show no interference at a concentration of 100μ.g/mL are summarized in the following table.
| 3-Hydroxytyramine | Digoxin | Nitroglycerin |
|-------------------|---------------------|---------------|
| Acetaminophen | Diphenhydramine HCl | Nordoxepin |
{18}------------------------------------------------
| Acetone(1000mg/dL) | Disopyramide(except MTD test) | Norethindrone |
|---------------------------------|-----------------------------------------------|----------------------------------------|
| Acetophenetidin | DL-Tryptophan | Norfentanyl |
| Acetylsalicylic acid…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.