SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip Card

K201120 · Safecare Biotech (Hangzhou) Co., Ltd. · NFT · May 27, 2020 · Clinical Toxicology

Device Facts

Record IDK201120
Device NameSAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip Card
ApplicantSafecare Biotech (Hangzhou) Co., Ltd.
Product CodeNFT · Clinical Toxicology
Decision DateMay 27, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3100
Device ClassClass 2

Indications for Use

SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cutoffs]. Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use. SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cutoffs]. Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

Device Story

Single-use in vitro diagnostic devices (Dip Card or Cup formats) for qualitative detection of drugs of abuse in human urine. Operates via competitive binding, lateral flow immunochromatographic assay; urine sample migrates via capillary action. If drug concentration is below cutoff, monoclonal antibody-coated particles bind to immobilized drug-conjugate, forming a visible colored line. If drug concentration exceeds cutoff, binding sites are saturated, preventing line formation. Used by lay users in home or point-of-care settings. Results are visual; positive results are preliminary and require confirmation via GC/MS or LC/MS. Provides rapid screening to assist in identifying potential drug presence.

Clinical Evidence

No clinical studies performed. Evidence consists of analytical performance (precision, interference, specificity, pH/specific gravity effects) and a lay-user study. Lay-user study (n=310 per format) evaluated performance across various drug concentrations relative to cutoffs, confirming high accuracy (typically 85-100% correct results) compared to LC/MS. Flesch-Kincaid analysis confirmed 7th-grade reading level for instructions.

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle. Single-use dip card or cup format. Materials include monoclonal mouse antibodies, drug-conjugates, and desiccant. No electronic components or software. Stable at 4-30°C for 24 months.

Indications for Use

Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, MDMA, Phencyclidine, Methadone, Nortriptyline, Propoxyphene, EDDP) in human urine. Intended for over-the-counter use. Not intended to distinguish between prescription use and abuse.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) Number: k201120 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...