COBAS INTEGRA AMIKACIN MAB

K991597 · Roche Diagnostics Corp. · LGJ · Jul 22, 1999 · Clinical Toxicology

Device Facts

Record IDK991597
Device NameCOBAS INTEGRA AMIKACIN MAB
ApplicantRoche Diagnostics Corp.
Product CodeLGJ · Clinical Toxicology
Decision DateJul 22, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3035
Device ClassClass 2

Indications for Use

The cassette COBAS INTEGRA Amikacin MAB contains and in vitro diagnostic reagent system intended for use on the COBAS INTEGRA 700 analyzer for the quantitative determination of amikacin in human serum or heparinized plasma.

Device Story

In vitro diagnostic reagent system for quantitative amikacin measurement in human serum or heparinized plasma. Used on COBAS INTEGRA 700 analyzer. Operates via fluorescence polarization: light excitation of fluorophore-labeled drug; polarization of emitted light inversely proportional to rotation speed; macromolecule-bound tracer rotates slowly, maintaining polarization. Surface active agents dissociate drug from serum proteins. Analyzer calculates millipolarization units (mP) for calibrators and samples; performs nonlinear least squares regression to generate best-fit curve; interpolates sample concentration from curve. Used in clinical laboratories by trained personnel for therapeutic drug monitoring. Output assists clinicians in adjusting drug dosage to optimize patient outcomes.

Clinical Evidence

No clinical data provided; bench testing only. Improvements in analytical sensitivity and specificity were demonstrated through assay parameter optimization and antibody substitution.

Technological Characteristics

In vitro diagnostic reagent system; fluorescence polarization sensing principle. Uses mouse monoclonal antibody. Compatible with COBAS INTEGRA 700 analyzer. Software performs automated mP calculation and nonlinear least squares regression analysis for concentration interpolation.

Indications for Use

Indicated for the quantitative determination of amikacin in human serum or heparinized plasma for therapeutic drug monitoring.

Regulatory Classification

Identification

An amikacin test system is a device intended to measure amikacin, an aminoglycoside antibiotic drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of amikacin overdose and in monitoring levels of amikacin to ensure appropriate therapy.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 222 1999 . . K991597 | <b>510(k) Summary</b> | | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | 1) Submitter name, address, contact | Roche Diagnostics Corporation<br>9115 Hague Rd<br>Indianapolis, IN 46250<br>(317) 845-3362 | | | Contact person: Lisa M. Gerard | | | Date prepared May 7, 1999 | | 2) Device name | Proprietary name: COBAS INTEGRA Amikacin MAB | | | Common name: Enzymatic assay for the determination of Amikacin | | | Classification name: Amikacin test system | | 3) Predicate device | We claim substantial equivalence to the Roche COBAS INTEGRA Amikacin (K954992). | Continued on next page {1}------------------------------------------------ ## 510(k) Summary, Continued 4) Device description The Roche COBAS INTEGRA Amikacin MAB assay contains an in vitro diagnostic reagent system intended for use on the COBAS INTEGRA 700 analyzer for the quantitative determination of amikacin in human serum or heparinized plasma. The COBAS INTEGRA Amikacin MAB assay determinations are made on the COBAS INTEGRA 700 analyzer using the principle of fluorescence polarization. When a fluorescent molecule, or fluorophore, is irradiated with light of the proper wavelength (the excitation wavelength) some of the light is absorbed. Within a few nano-seconds the absorbed light is emitted, although at a longer wavelength (the emission wavelength). Whether or not the emitted light is polarized depends on the freedom of the fluorophore to rotate in solution. A small molecule, such as fluorescein, can rotate rapidly before light emission occurs, resulting in depolarization of the emitted light. In contrast, a fluorescent macromolecule, such as a fluorescein-labeled protein, will rotate much more slowly. Thus, in the time frame between excitation and emission, the macromolecule will have rotated only very slightly and the emitted light will be polarized. Fluorescence polarization is a reproducible function of the drug concentration, and is suitable for the quantitative determination of drug concentrations in serum for the purpose of therapeutic drug monitoring. Surface active agents are used to ensure dissociation of the drug from serum proteins and to prevent nonspecific binding of the tracer. After completion of the assay, the COBAS INTEGRA 700 will calculate automatically the millipolarization units (mP) of the tracer. After mP values have been calculated for the 6 calibrators, the system calculates a best-fit curve for the calibrators using a nonlinear lest squares regression analysis. The concentration of drug in each sample is then interpolated from this curve using its measured mP value. Continued on next page {2}------------------------------------------------ ## 510(k) Summary, Continued The cassette COBAS INTEGRA Amikacin MAB contains and in vitro 5) Intended use diagnostic reagent system intended for use on the COBAS INTEGRA 700 analyzer for the quantitative determination of amikacin in human serum or heparinized plasma. The Roche COBAS INTEGRA Amikacin MAB (AMIKM) is substantially 6) Comparison to the predicate equivalent to other products in commercial distribution intended for similar device use. Most notably, it is substantially equivalent to the currently marketed Roche COBAS INTEGRA Amikacin (AMIK)(K954992). The Roche COBAS INTEGRA Amikacin MAB assay contains an in vitro diagnostic reagent system intended for use on the COBAS INTEGRA 700 analyzer for the quantitative determination of amikacin in human serum or heparinized plasma. Modifications of the Roche COBAS INTEGRA Amikacin MAB assay include: changing rabbit polyclonal antibody with mouse monoclonal antibody . and . changing the assay parameters. The modification has resulted in improved: . analytical sensitivity and . specificity. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. JUL 22 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Lisa M. Gerard Regulatory Affairs Consultant Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, Indiana 46250-0457 Re: K991597 > Trade Name: Roche COBAS INTEGRA Amikacin Regulatory Class: II Product Code: LGJ Dated: May 7, 1999 Received: May 10, 1999 Dear Ms. Gerard: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement | 510(k) Number<br>(if known): | K991597 | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | COBAS INTEGRA Amikacin MAB | | Indications for<br>Use: | The cassette COBAS INTEGRA Amikacin MAB contains and in vitro<br>diagnostic reagent system intended for use on the COBAS INTEGRA<br>analyzer for the quantitative determination of amikacin in human serum or<br>heparinized plasma. | | | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON<br>ANOTHER PAGE IF NEEDED)<br>- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | | | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | Prescription Use V_<br>OR<br>Over-the-Counter Use _<br>(Per 21 CFR 801.109) | | | (Optional format 1-2-96) | | | | | | | | | (Division Sign-Off)<br>Division of Clinical Laboratory Devices | 510(k) Number_K 991 597 . . . . . :
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...