MULTIGENT GENTAMICIN

K060709 · Seradyn, Inc. · LCD · Jun 15, 2006 · Clinical Toxicology

Device Facts

Record IDK060709
Device NameMULTIGENT GENTAMICIN
ApplicantSeradyn, Inc.
Product CodeLCD · Clinical Toxicology
Decision DateJun 15, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3450
Device ClassClass 2

Indications for Use

The Multigent Gentamicin assay is intended for the quantitative determination of Gentamicin in human serum or plasma on the Architect C8000 System. The results obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

Device Story

Multigent Gentamicin assay is a homogeneous particle-enhanced turbidimetric immunoassay (particle agglutination) for use on the ARCHITECT c8000 System. Input: human serum or plasma samples. Principle: competitive binding; gentamicin in sample competes with gentamicin-coated microparticles for binding to anti-gentamicin monoclonal antibody. Output: absorbance inversely proportional to drug concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative drug levels to clinicians for monitoring therapeutic ranges and managing potential overdose, aiding in the prevention of renal and CNS toxicity.

Clinical Evidence

Bench testing only. Precision study (N=80 per control) showed total CVs of 2.15%–5.69%. Accuracy study (N=5 samples) demonstrated mean recovery of 100.31%. Linearity confirmed across range. Analytical sensitivity (LDD) is 0.09 µg/mL. Interference testing (bilirubin, hemoglobin, triglycerides, protein, HAMA, and common drugs) showed no significant impact. Method comparison against predicate (N=55) yielded R=0.996, slope=1.165, y-intercept=-0.719, demonstrating excellent correlation.

Technological Characteristics

Homogeneous particle-enhanced turbidimetric immunoassay. Two-reagent system (R1: anti-gentamicin monoclonal antibody; R2: gentamicin-coated microparticles). Uses Architect C8000 analyzer. Calibration via six-level calibrator set. Stabilizers include sodium azide. No specific ASTM material standards cited.

Indications for Use

Indicated for quantitative determination of gentamicin in human serum or plasma for monitoring therapy and diagnosing overdose.

Regulatory Classification

Identification

A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k060709 B. Purpose for Submission: Clearance of a new device C. Measurand: Gentamicin D. Type of Test: Turbidimetric Immunassay E. Applicant: Seradyn, Inc. F. Proprietary and Established Names: Multigent Gentamicin G. Regulatory Information: 1. Regulation section: 21 CFR § 862.3450, Gentamicin test system. 2. Classification: Class II 3. Product code: LCD, enzyme immunoassay, gentamicin 4. Panel: Toxicology (91) H. Intended Use: 1. Intended use(s): The Multigent Gentamicin assay is intended for the quantitative determination of Gentamicin in human serum or plasma on the Architect C8000 System. The results obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy. 2. Indication(s) for use: See Intended Use. {1} 3. Special conditions for use statement(s): For prescription use. 4. Special instrument requirements: The ARCHITECT c8000 System. I. Device Description: The Multigent Gentamicin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti-gentamicin monoclonal antibody and R2: gentamicin-coated microparticles. A six-level set of Multigent Gentamicin Calibrators (A through F) is used to calibrate the assay. J. Substantial Equivalence Information: 1. Predicate device name(s): Abbott TDx/TDxFLx Gentamicin 2. Predicate 510(k) number(s): k904226 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Intended for the quantitative determination of gentamicin in human serum or plasma. | A reagent system for the quantitative measurement of gentamicin, an antibiotic drug in human serum or plasma. | | Calibration | Six levels | Six levels | | Sample Types | Plasma and Serum | Plasma and Serum | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | Homogeneous particle-enhanced turbidimetric immunoassay (particle agglutination). | Florescence Polarization Immunoassay (FPIA) technology. | | Instrument Required | Abbott ARCHITECT c8000 System. | Abbott TDx/TDxFLx | K. Standard/Guidance Document Referenced (if applicable): {2} CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach CLSI EP9-A2, Evaluation of Precision Performance of Quantitative Measurement Methods ## L. Test Principle: In particle agglutination assays, the degree of agglutination is inversely proportional to the quantity of free drug in the reaction cell. If no drug is present in the sample, the antibodies in the Multigent Gentamicin Antibody Reagent (R1) will bind only to the bound drug on the particle which will cause it to agglutinate and will result in higher absorbance. If increased amount of competing drug is present in the sample, this will result in decreased binding of bound drug by the antibody, resulting in a relative decrease in particle agglutination, which will result in lower absorbance. Therefore, the absorbance is inversely proportional to the concentration of drug. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Three control levels (Low Control, Mid Control, and High Control) were evaluated on the ARCHITECT c8000 and, after 2 hours within the same day, the samples were re-run in duplicate, over 20 days resulting in a total of 80 replicates for each control. Calibration was performed initially and recalibration was performed when the controls did not recover within labeled ranges. Results for precision are summarized below. Precision on the ARCHITECT c8000 | | Low Control | Mid Control | High Control | | --- | --- | --- | --- | | N | 80 | 80 | 80 | | Mean, μIU/mL | 2.68 | 6.47 | 9.41 | | | SD (%CV) | SD (%CV) | SD (%CV) | | Within Run | 0.08(2.93) | 0.07(1.09) | 0.14(1.53) | | Between Run | 0.06(2.17) | 0.07(1.07) | 0.07(0.70) | | Between Day | 0.12(4.37) | 0.12(1.91) | 0.13(1.34) | | Total | 0.15(5.69) | 0.16(2.44) | 0.20(2.15) | #### b. Linearity/assay reportable range: The reportable range of the assay is $0.34 - 10.0\ \mu\mathrm{g/mL}$ (highest calibrator). The sponsor evaluated the recovery of samples across most of the measuring range. Two groups of samples were used ranging from 1.72 to $6.88\ \mu\mathrm{g/mL}$ and 0.25 to $8.00\ \mu\mathrm{g/mL}$. Samples were prepared by spiking and dilution. The {3} samples were run in triplicate. The percent recovery for each sample was determined by dividing the mean observed result by the theoretical value. Results are summarized below. | | Theoretical Conc. (μg/mL) | Rep 1 | Rep 2 | Rep 3 | Mean Recovered Conc. | SD | %CV | % Recovery | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Study 1 | 6.88 | 7.14 | 7.31 | 7.14 | 7.20 | 0.098 | 1.36 | 104.60% | | | 5.16 | 5.20 | 5.23 | 5.39 | 5.27 | 0.102 | 1.94 | 102.20% | | | 3.44 | 3.50 | 3.65 | 3.70 | 3.62 | 0.104 | 2.88 | 105.14% | | | 1.72 | 1.59 | 1.67 | 1.66 | 1.64 | 0.044 | 2.66 | 95.35% | | | Mean Percent Recovery | | | | | | | 101.82% | | Study 2 | 0.25 | 0.23 | 0.28 | 0.26 | 0.26 | 0.025 | 9.82 | 102.67% | | | 1.00 | 1.02 | 0.99 | 0.98 | 1.00 | 0.021 | 2.09 | 99.67% | | | 2.25 | 2.24 | 2.22 | 2.21 | 2.22 | 0.015 | 0.69 | 98.81% | | | 4.50 | 4.44 | 4.55 | 4.42 | 4.47 | 0.070 | 1.57 | 99.33% | | | 8.00 | 8.12 | 7.99 | 8.15 | 8.09 | 0.085 | 1.05 | 101.08% | | | Mean Percent Recovery | | | | | | | 100.31% | Linear regression analysis gave the following: Study 1: Observed $= 1.0657$ (Expected) - 0.15; $\mathrm{r} = 0.9989$ Study 2: Observed $= 1.01$ (Expected) - 0.024; $\mathrm{r} = 0.9999$ c. Traceability, Stability, Expected values (controls, calibrators, or methods): The Multigent Gentamicin Calibrators the assay utilizes are currently marketed and were cleared under k981706. Calibration is stable for 28 days for this assay. d. Detection limit: The Limit of Quantitation (LoQ) (defined by the sponsor as the analytical concentration at which the precision of the assay [%CV] is less than $20\%$ and the accuracy is within $10\%$ ) was determined by diluting a patient sample with known concentration gentamicin with human serum (negative for gentamicin) to achieve an approximate concentration of $0.5 \mu \mathrm{g} / \mathrm{mL}$ . This sample was further diluted with negative serum to achieve approximate gentamicin concentrations of 0.4, 0.3, 0.2, and $0.1 \mu \mathrm{g} / \mathrm{mL}$ . All samples were then run on an ARCHITECT c8000 System. The data submitted supports a LoQ of 0.33 $\mu \mathrm{g} / \mathrm{mL}$ . The sponsor will claim a higher LoQ of $0.34 \mathrm{mg} / \mathrm{dL}$ in the labeling. Results of this study are summarized below. {4} | N | 20 | 20 | 20 | 20 | 20 | 20 | | --- | --- | --- | --- | --- | --- | --- | | Mean, μIU/mL | 0.00 | 0.10 | 0.20 | 0.30 | 0.40 | 0.50 | | | SD (%CV) | SD (%CV) | SD (%CV) | SD (%CV) | SD (%CV) | SD (%CV) | | Total | 0.025(56.04) | 0.039(30.26) | 0.027(10.94) | 0.050(14.89) | 0.034(7.89) | 0.038(7.37) | The Least Detectable Does (LDD) or analytical sensitivity (the lowest amount of analyte in a sample that can be distinguished from zero with 95% confidence) was determined. The sponsor claims a LDD of 0.1 µg/mL. e. Analytical specificity: Serum samples with gentamicin stock concentrations of 2.46 and 3.44 µg/mL were spiked with bilirubin (20 mg/dL), hemoglobin (2000 mg/dL), triglyceride (1691 mg/dL), total protein (12 g/dL), and Rheumatoid Factor (582 IU) and run in triplicate (hemoglobin in duplicate). The endogenous substances above were found not to interfere within the sponsor’s percent recovery acceptance criteria of 100±10%. HAMA Type-1 and Type-2 samples were also spiked with 3.33 µg/mL gentamicin. HAMA Type-1 and Type-2 were found not to interfere (percent recovery within 100±10%). For the Common Co-Administered Drugs study, two amounts of gentamicin were present in the samples used: 1.5 µg/mL and 3.5 µg/mL. The following nine test samples had 1.5 µg/mL of gentamicin present during testing with a range of 1.55 µg/mL to 1.59 µg/mL: Theophylline, Cyclosporin, Rifampicin, Acetaminophen, Ibuprofen, Penicillin V, Phenylbutazone, Acetylsalicylic acid and Metronidazole. All other samples had 3.5 µg/mL of gentamicin present during testing with a range of 3.51 µg/mL to 3.88 µg/mL. The interference results are listed in the package insert. Cross-reactivity of 50.35% was observed for sisomicin. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Sixty-eight purchased clinical samples were tested (ranging from 0.22 µg/mL to 9.97 µg/mL). A Passing-Bablok regression analysis was performed comparing device results to predicate results. Results of the analysis gave the following linear regression statistics: Device = 1.122(Predicate) - 0.445; r = 0.996 {5} b. Matrix comparison: A study was conducted to determine the performance of the assay for both serum and plasma samples containing gentamicin. Blood was drawn from ten healthy donors (with no gentamicin therapy) for each tube type listed below: - Plastic K2 EDTA tube - Glass K3 EDTA tube - Glass plasma separator lithium heparin tube - Glass sodium heparin tube - Glass lithium heparin tube - Glass serum separator tube - Plastic tube with clot activator - Glass tube with no additives - Plastic tube with no additives. The serum or plasma was removed from the collection tubes and aliquoted into new tubes for testing (which is the same procedure the customer would use). Serum or plasma from each tube was then spiked with gentamicin. The samples were analyzed on the ARCHITECT c8000 analyzer in duplicate with baseline results obtained on day zero for each type of tube. Results support that the collection tubes evaluated show no adverse effects on gentamicin (recovery within 100 +/- 10%). 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The sponsor cites a therapeutic range for gentamicin for moderate infections as 2 - 8 µg/mL. Peak serum levels of 5 - 10 µg/mL have been shown to cause renal {6} and CNS toxicity. Recommended trough levels are < 2 µg/mL to avoid associated neprotoxicity. The reference ranges are referenced from: Tietz N. Clinical Guide to Laboratory Tests. 2nd ed. Philadelphia, PA: WB Saunders; 1990;742. Nordstrom L, Banck G, Belfrage S, et al. Prospective study of the ototoxicity of gentamicin. Acta Pathol Microbiol Immunol Scand (B) (Supp 241) 1973:58-61. Dayal VS, Smith EL, McCain WG. Cochlear and vestibular gentamicin toxicity. A clinical study of systemic and topical usage. Arch Otolaryngol 1974;100:338-40 ## N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. ## O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...