ABBOTT AXSM PROCAINAMIDE ASSAY
Device Facts
| Record ID | K955444 |
|---|---|
| Device Name | ABBOTT AXSM PROCAINAMIDE ASSAY |
| Applicant | Abbott Laboratories |
| Product Code | LAR · Clinical Toxicology |
| Decision Date | Apr 2, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3320 |
| Device Class | Class 2 |
Indications for Use
The intended use of both assays is for the quantitative determination of procainamide in human serum or plasma (sodium heparin, tripotassium EDTA, potassium oxalate, and sodium citrate).
Device Story
Automated fluorescence polarization immunoassay (FPIA) for quantitative procainamide measurement in human serum or plasma; used on AxSYM system; operated by laboratory personnel; provides numerical concentration values to clinicians for therapeutic drug monitoring; supports clinical decision-making regarding procainamide dosage.
Clinical Evidence
Bench testing only. Correlation study comparing AxSYM Procainamide to TDx/TDxFLx predicate (n=205) yielded slope 0.98, y-intercept 0.1, and correlation coefficient 0.989.
Technological Characteristics
Automated fluorescence polarization immunoassay (FPIA). Reagents, calibrators, and controls prepared gravimetrically. Real-time stability testing performed for expiration dating.
Indications for Use
Indicated for the quantitative determination of procainamide in human serum or plasma for therapeutic drug monitoring.
Regulatory Classification
Identification
A digoxin test system is a device intended to measure digoxin, a cardiovascular drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure appropriate therapy.
Predicate Devices
- TDx/TDxFLx Procainamide assay
Related Devices
- K955422 — AXSYM N-ACETYLPROCAINAMIDE · Abbott Laboratories · Apr 24, 1996
- K971716 — PROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM · Bayer Corp. · Jun 27, 1997
- K032573 — DIMENSION PROCAINAMIDE (PROC) FLEX REAGENT CARTRIDGE METHOD, MODEL DF110 · Dade Behring, Inc. · Oct 31, 2003
- K963158 — ABBOTT AXSYM DIGITOXIN (MODIFY) · Abbott Laboratories · Sep 10, 1996