AXSYM N-ACETYLPROCAINAMIDE
Device Facts
| Record ID | K955422 |
|---|---|
| Device Name | AXSYM N-ACETYLPROCAINAMIDE |
| Applicant | Abbott Laboratories |
| Product Code | LAN · Clinical Toxicology |
| Decision Date | Apr 24, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3320 |
| Device Class | Class 2 |
Intended Use
The intended use of both assays is for the quantitative determination of N-acetylprocainamide in human serum or plasma (sodium heparin, tripotassium EDTA, potassium oxalate, and sodium citrate).
Device Story
Automated fluorescence polarization immunoassay (FPIA) for quantitative measurement of N-acetylprocainamide in human serum or plasma; used on AxSYM system; intended for clinical laboratory use; provides concentration data to clinicians for therapeutic drug monitoring; calibrators and controls used for system verification; performance validated via correlation with predicate TDx/TDxFLx assay.
Clinical Evidence
Correlation study comparing AxSYM N-Acetylprocainamide to predicate TDx/TDxFLx assay; n=205 samples; results: slope 1.05, y-intercept -0.1, correlation coefficient 0.993, standard error of Y estimate 0.081.
Technological Characteristics
Automated fluorescence polarization immunoassay (FPIA); utilizes Abbott AxSYM system; reagents, calibrators, and controls prepared gravimetrically; real-time stability testing for expiration dating.
Indications for Use
Indicated for the quantitative determination of N-acetylprocainamide in human serum or plasma to monitor therapeutic drug levels.
Regulatory Classification
Identification
A digoxin test system is a device intended to measure digoxin, a cardiovascular drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure appropriate therapy.
Predicate Devices
- TDx/TDxFLx N-Acetylprocainamide assay
Related Devices
- K955444 — ABBOTT AXSM PROCAINAMIDE ASSAY · Abbott Laboratories · Apr 2, 1996
- K971713 — N-ACETYLPROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM · Bayer Corp. · Jun 11, 1997
- K011620 — EMIT 2000 N-ACETYLPROCAINAMIDE ASSAY, MODEL OSR4N229 · Syva Co., Dade Behring, Inc. · Jun 18, 2001
- K032564 — DIMENSION N-ACETYLPROCAINAMIDE (NAPA) FLEX REAGENT CARTRIDGE METHOD, MODEL DF111 · Dade Behring, Inc. · Oct 31, 2003
- K971716 — PROCAINAMIDE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM · Bayer Corp. · Jun 27, 1997