CARBAMAZEPINE FLEX REAGENT CARTRIDGE
Device Facts
| Record ID | K962820 |
|---|---|
| Device Name | CARBAMAZEPINE FLEX REAGENT CARTRIDGE |
| Applicant | Dade Intl., Inc. |
| Product Code | KLT · Clinical Toxicology |
| Decision Date | Sep 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3645 |
| Device Class | Class 2 |
Indications for Use
The CRBM FLEX™ reagent cartridge used on the Dimension® clinical chemistry system is an in vitro diagnostic test intended to measure carbamazepine, an anti-convulsant drug, in plasma or serum. Measurements obtained by this device are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to ensure appropriate therapy.
Device Story
In vitro diagnostic reagent cartridge for Dimension® clinical chemistry system; measures carbamazepine in human serum/plasma. Uses Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA) technique; latex particle-carbamazepine conjugate reacts with carbamazepine-specific monoclonal antibody. Operated by laboratory personnel on automated analyzer. Output provides quantitative carbamazepine concentration; assists clinicians in therapeutic drug monitoring and overdose management.
Clinical Evidence
Method comparison study using 82 samples tested on Dimension® system vs. aca® discrete clinical analyzer. Results: slope 0.99, intercept 0.06, correlation coefficient 0.979, range 0-18.6 µg/mL.
Technological Characteristics
Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA); utilizes latex particle-carbamazepine conjugate and monoclonal antibody; prepackaged reagent cartridge format for automated clinical chemistry analyzer.
Indications for Use
Indicated for patients requiring measurement of carbamazepine levels in plasma or serum for diagnosis/treatment of overdose or therapeutic drug monitoring.
Regulatory Classification
Identification
A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.
Predicate Devices
- aca® carbamazepine analytical test pack
Related Devices
- K011648 — RANDOX CARBAMAZEPINE · Randox Laboratories, Ltd. · Jan 15, 2002
- K123518 — ABBOTT CARBAMAZEPINE ASSAY · Microgenics Corporation · Jun 5, 2013
- K964068 — CARBAMAZEPINE ASSAY FOR THE TECHNICON IMMUNO 1 SYSTEM · Bayer Corp. · Dec 13, 1996
- K042808 — ADVIA IMS CARBAMAZEPINE METHOD · Bayer Healthcare, LLC · Feb 8, 2005