Healgen Multi-Drug Urine Test Cup; Healgen Multi-Drug Urine Test Dip Card
Applicant
Healgen Scientific,, LLC
Product Code
JXM · Clinical Toxicology
Decision Date
Aug 2, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.3170
Device Class
Class 2
Intended Use
The Healgen Drug Test Cup and Dipcard are lateral flow chromatographic immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine at the following cut-off concentrations: Amphetamine (AMP) 1000 ng/mL, Secobarbital (BAR) 300 ng/mL, Buprenorphine (BUP) 10 ng/mL, Cocaine (COC) 300 ng/mL, Methamphetamine (MET) 1000 ng/mL, Methylenedioxymethamphetamine (MDMA) 500 ng/mL, Methadone (MTD) 300 ng/mL, Morphine (MOP) 300 ng/mL, Nortriptyline (TCA) 1000 ng/mL, Oxazepam (OPI) 300 ng/mL, Phencyclidine (PCP) 25 ng/mL, Propoxyphene (PPX) 300 ng/mL, Marijuana (THC) 50 ng/mL, and Oxycodone (OXY) 100 ng/mL. The device is intended for healthcare professional use, including professional use at point-of-care sites, to assist in the determination of drug compliance. It is not intended for over-the-counter sale. The device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or liquid chromatography/tandem mass spectrometry (LC/MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Device Story
Lateral flow immunochromatographic assay for qualitative detection of drugs of abuse in human urine. Device uses competitive binding principle; drug-free samples or samples below cutoff generate colored line in test region; positive samples (at or above cutoff) do not. Control line always appears. Available as multi-drug test cup or dip card (2-14 analytes). Used by lay users (OTC) or clinicians (prescription) for preliminary screening. Results require confirmation via GC/MS or LC/MS. Benefits include rapid, simultaneous screening for multiple substances to support clinical decision-making.
Clinical Evidence
Lay-user study conducted at three sites with 300 participants per format (ages 20 to >50). Samples spiked with drugs at negative, +/-75%, +/-50%, +/-25% of cutoff concentrations, confirmed by GC/MS. Results showed high percentage agreement across all analytes and concentrations. Flesch-Kincaid analysis of package inserts indicated a Grade Level of 7, confirming ease of use for lay persons.
Technological Characteristics
Lateral flow chromatographic immunoassay. Format: multi-drug test cup and dipcard. Analyte detection via competitive binding principle. Qualitative visual readout. No electronic components, software, or energy source required.
Indications for Use
Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, MDMA, Phencyclidine, Propoxyphene, Nortriptyline, Methadone) in human urine. Intended for both prescription and OTC use. Not for distinguishing between prescription use and abuse.
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Healgen Oxazepam and Morphine Tests (k142280)
Healgen Amphetamine and Oxycodone Tests (k143187)
Healgen Cocaine Test (k141647)
Healgen Marijuana and Methamphetamine Tests (k140546)
Healgen Secobarbital, Buprenorphine, and Methadone Tests (k150791)
Healgen MDMA and Phencyclidine Tests (k150096)
Healgen Nortriptyline, Propoxyphene, and EDDP Tests (k151348)
Healgen Amphetamine, Cocaine, and Methamphetamine Tests (k152269)
Related Devices
K250727 — AllTest Multi-Drug Rapid Urine Test Cup; AllTest Multi-Drug Urine Test Cup · Hangzhou AllTest Biotech Co., Ltd. · Apr 4, 2025
K201630 — AssureTech DOA Dipstick Screen Panel Tests, AssureTech DOA Integrated Cup Tests · Assure Tech. (Hangzhou) Co, Ltd. · Aug 12, 2021
K133968 — WONDFO MULTI-DRUG URINE TEST CUP, PANEL · Guangzhou Wondfo Biotech Co., Ltd. · Feb 25, 2014
K170049 — AssureTech Panel Dip Test, AssureTech Quick Cup Test · Assure Tech. (Hangzhou) Co, Ltd. · May 2, 2017
K193480 — BIOEASY Multi-Drug Test Cup · Shenzhen Bioeasy Biotechnology Co., Ltd. · Jan 31, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized caduceus, which is a symbol of medicine. The caduceus is made up of three intertwined snakes and a staff.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
HEALGEN SCIENTIFIC LLC C/O JOE SHIA LSI INTERNATIONAL INC. 504 EAST DIAMOND AVE. SUITE I GAITHERSBURG MD 20877
August 2, 2017
Re: k163704
Trade/Device Name: Healgen Multi-Drug Urine Test Cup Healgen Multi-Drug Urine Test Dip Card Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiazepine test system Regulatory Class: II Product Code: JXM, DJG, DIO, LDJ, LCM, LAF, DKZ, DIS, DJR, JXN, LFG Dated: June 26, 2017 Received: July 03, 2017
Dear Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Donald J. St Pierre -S
Donald J. St. Pierre Acting Director Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### Device Name
Healgen Multi-Drug Urine Test Cup Healgen Multi-Drug Urine Test Dip Card
#### Indications for Use (Describe)
Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine. Phencyclidine. EDDP. Nortriptyline and Methadone in human urine at the cutoff concentrations of:
| Drug(Identifier) | Cut-off level |
|--------------------------------|-------------------------|
| Amphetamine | 500 ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine | 150 ng/mL |
| Cannabinoids | 50 ng/mL |
| Methamphetamine | 500 ng/mL |
| Morphine | 300 ng/mL or 2000 ng/mL |
| Oxycodone | 100 ng/mL |
| Secobarbital | 300 ng/mL |
| Buprenorphine | 10 ng/mL |
| Methylenedioxy-methamphetamine | 500 ng/mL |
| Phencyclidine | 25 ng/mL |
| Methadone | 300 ng/mL |
| Propoxyphene | 300 ng/mL |
| Nortriptyline | 1000 ng/mL |
Configuration of the Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
- 1. Date: December 18, 2016
- 2. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX77401
- 3. Contact person: Jianqiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
- 4. Device Name: Healgen Multi-Drug Urine Test Cup Healgen Multi-Drug Urine Test Dip Card
Classification: Class II
| Product Code | Regulation Section | Panel |
|-------------------------------------------|-------------------------------------------------------------------|------------|
| DKZ<br>Amphetamine | 21 CFR § 862.3100, Amphetamine Test System | Toxicology |
| LDJ<br>Cannabinoids | 21 CFR § 862.3870, Cannabinoids Test System | Toxicology |
| DIO<br>Cocaine | 21 CFR § 862.3250, Cocaine and Cocaine Metabolites<br>Test System | Toxicology |
| LAF<br>Methamphetamine | 21 CFR § 862.3610, Methamphetamine Test<br>System | Toxicology |
| DJG<br>Morphine | 21 CFR § 862.3650, Opiate Test System | Toxicology |
| JXM<br>Oxazepam | 21 CFR § 862.3170, Benzodiazepine Test System | Toxicology |
| DJG<br>Oxycodone | 21 CFR § 862.3650, Opiate Test System | Toxicology |
| DIS<br>Secobarbital | 21 CFR § 862.3150, Barbiturate Test System | Toxicology |
| DJG<br>Buprenorphine | 21 CFR § 862.3650, Opiate Test System | Toxicology |
| LAF<br>Methylenedioxy-<br>methamphetamine | 21 CFR § 862.3610, Methamphetamine Test System | Toxicology |
{5}------------------------------------------------
| LCM<br>Phencyclidine | Enzyme Immunoassay Phencyclidine | Toxicology |
|----------------------|----------------------------------------------------------------------------|------------|
| DJR<br>Methadone | 21 CFR § 862.3620, Methadone Test System | Toxicology |
| JXN<br>Propoxyphene | 21 CFR § 862.3700, Propoxyphene Test System | Toxicology |
| LFG<br>Nortriptyline | 21 CFR, 862.3910 Tricyclic antidepressant drugs<br>test system Test System | Toxicology |
## 5. Description of the device:
Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test DipCard are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Propoxyphene, Nortriptyline and Methadone in human urine samples. Healgen Multi-Drug devices detect each of analytes on different strips.
A positive urine sample will not generate a colored-line for the specific drug tested in the designated region. A negative urine specimen or a urine sample containing Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Propoxyphene, Nortriptyline and Methadone at the concentration below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a test control, a color line will always appear at the control region.
## 6. Predicate Devices:
K142280; Healgen Oxazepam and Morphine Tests (Dip Card, Cup) K143187; Healgen Amphetamine and Oxycodone Tests (Dip Card, Cup) K141647; Healgen Cocaine and Morphine Tests (Dip Card, Cup) K140546; Healgen Marijuana and Methamphetamine Tests (Dip Card, Cup) K150791; Healgen Secobarbital, Buprenorphine and Methadone Tests (Cup, Dip Card) K150096; Healgen Methylenedioxy-methamphetamine and Phencyclidine Tests (Dip Card, Cup) K151348; Healgen Nortriptyline and Propoxyphene Tests (DipCard, Cup) K152269; Healgen Amphetamine, Cocaine and Methamphetamine Tests (Dip Card, Cup)
- 7. Intended Use / Indications for Use
Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Propoxyphene, Nortriptyline and Methadone in human urine at the cutoff concentrations of:
| Drug(Identifier) | Cut-off level |
|------------------|---------------|
| Amphetamine(AMP) | 500 ng/mL |
{6}------------------------------------------------
| Oxazepam (OXA) | 300 ng/mL |
|---------------------------------------|------------------------|
| Cocaine (COC) | 150 ng/mL |
| Cannabinoids (THC) | 50 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Morphine (MOR) | 300ng/mL or 2000 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Methylenedioxy-methamphetamine (MDMA) | 500 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Propoxyphene | 300 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
Configuration of the Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.
- 8. Substantial Equivalence Information
Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card are a modified product format derived from the previously FDA-cleared Healgen single DOA Tests. A summary comparison of features of the Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card and the predicate devices is provided in the following Table
| Item | New Devices | Predicate devices (K142280,<br>K143187, K141647, K140546,<br>K150791, K150096, K151348,<br>K152269) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Indication(s) for use | For the qualitative determination of<br>Amphetamine, Oxazepam, Cocaine,<br>Cannabinoids, Methamphetamine, Morphine,<br>Oxycodone, Secobarbital, Buprenorphine,<br>Methylenedioxy-methamphetamine,<br>Phencyclidine, Propoxyphene, Nortriptyline<br>and Methadone in human urine.<br><br>The configurations of the New Devices are<br>available in any combination of the above<br>tests. | For the qualitative determination of<br>Amphetamine, Oxazepam, Cocaine,<br>Cannabinoids, Methamphetamine,<br>Morphine, Oxycodone,<br>Secobarbital, Buprenorphine,<br>Methylenedioxy-methamphetamine,<br>Phencyclidine, Propoxyphene,<br>Nortriptyline and Methadone in<br>human urine.<br><br>The configurations of the Predicate<br>devices are only available in single<br>drug test. |
| Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based on the<br>principle of antigen antibody<br>immunochemistry. | Same |
| Type Of Test | Immunoassay principles that rely on<br>antigen- antibody interactions to indicate<br>positive or negative result | Same |
| Results | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Cut Off Values | Amphetamine (AMP): 500 ng/ml<br>Oxazepam (OXA):300 ng/ml<br>Cocaine(COC): 150 ng/ml<br>Cannabinoids (THC):50 ng/ml<br>Methamphetamine (MET): 500 ng/ml<br>Morphine (MOR): 300ng/mL or 2000 ng/ml<br>Oxycodone(OXY) : 100 ng/ml<br>Secobarbital (BAR): 300 ng/ml<br>Buprenorphine (BUP): 10 ng/ml<br>Methylenedioxy-methamphetamine(MDMA):<br>500 ng/ml<br>Phencyclidine (PCP): 25 ng/ml<br>Methadone (MTD): 300 ng/ml<br>Propoxyphene: 300 ng/ml<br>Nortriptyline (TCA): 1000 ng/ml | Same |
| Configurations | Cup, Dip Card | Same |
| Intended Use | OTC Use & Prescription Use | Same |
{7}------------------------------------------------
{8}------------------------------------------------
Healgen Multi-Drug Urine Test Cup is a multi-drug test that offers any combination from 2 to 14 drugs of abuse tests while the predicate devices are single-drug test. And the Healgen Multi-Drug Urine Test Dip Card is the same as the test dip card format of the predicate devices except that the Healgen Multi-Drug Urine Test Dip Card is a multi-drug test that offers any combination from 2 to 14 drugs of abuse tests while the predicate devices are single-drug test.
#### 9. Performance
Performance data of precision, cut-off, interference, specificity and method comparison of the devices were reported in K142280, K143187, K141647, K140546, K15079, K150096, K151348 and K152269.
#### 10.Lay User Verification
A lay user study was performed at three intended user sites with 300 lay persons for each format of the device. The lay users had diverse educational and professional backgrounds and ranged in age from 20 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and Each participant was provided with the package insert, blind labeled samples and test blind-labeled. devices.
| Drugs | % of Cutoff | Number of samples | Lay person results | | The percentage agreement (%) |
|-----------------|--------------|-------------------|--------------------|-----------------|------------------------------|
| | | | No. of Positive | No. of Negative | |
| Methamphetamine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Cocaine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Cannabinoids | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | | | | | |
| | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Morphine | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Oxazepam | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 17 | 3 | 85% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Amphetamine | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Oxycodone | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Methadone | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75%Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Secobarbital | -100%Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Buprenorphine | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Phencyclidine | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| MDMA | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Propoxyphene | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| Tricyclic | -50% Cutoff | 160 | 0 | 160 | 100% |
| Antidepressants | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | | | | | |
#### Comparison between GC/MS and Lav Person Results for Dip Card with MOP 300
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| | +75% Cutoff | 20 | 20 | 0 | 100% |
|----------------------------------------------------------------------------|--------------|-------------------|------------------------------------|------------------------------------|------------------------------|
| Drugs | % of Cutoff | Number of samples | Lay person results No. of Positive | Lay person results No. of Negative | The percentage agreement (%) |
| Methamphetamine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Cocaine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Cannabinoids | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Morphine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Oxazepam | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Comparison between GC/MS and Lay Person Results for Dip Card with MOP 2000 | | | | | |
| | | | | | |
| Amphetamine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Oxycodone | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Methadone | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Secobarbital | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 0 | 20 | 100% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Buprenorphine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Phencyclidine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| MDMA | -25% Cutoff | 20 | 2 | 18 | 90% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Propoxyphene | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| Tricyclic<br>Antidepressants | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
{12}------------------------------------------------
{13}------------------------------------------------
## Comparison between GC/MS and Lay Person Results for Cup with MOP 300
| Drugs | % of Cutoff | Number<br>of samples | Lay person results | | The<br>percentage<br>agreement<br>(%) |
|------------------------------|--------------|----------------------|--------------------|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| Methamphetamine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| Cocaine | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| | | | | | |
| Cannabinoids | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 3 | 17 | 85% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| Morphine | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| Oxazepam | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| Amphetamine | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 18 | 2 | 90% |
| Oxycodone | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | -25% Cutoff | 20 | 1 | 19 | 95% |
| | +25% Cutoff | 20 | 19 | 1 | 95% |
| Methadone | +50% Cutoff | 40 | 40 | 0 | 100% |
| | +75% Cutoff | 20 | 20 | 0 | 100% |
| | -100% Cutoff | 20 | 0 | 20 | 100% |
| | -75% Cutoff | 20 | 0 | 20 | 100% |
| | -50% Cutoff | 160 | 0 | 160 | 100% |
| | | | | | |
| | -25% Cutoff | 20 | 1 | 19 | તેરજ |
| | +25% Cutoff | 20 | 1 d | l | તેરજ |
| | +50% Cutoff | 40 | 40 | 0…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.