The Benzodiazepines assay is used for the qualitative analysis of benzodiazepines in human urine with a cutoff of 200 ng/mL. For use in clinical laboratories. Measurements obtained by this device are used in the diagnosis of benzodiazepines use or overdose. The Benzodiazepines assay is calibrated with lormetazepam and will detect a variety of benzodiazepines and their metabolites. The Benzodiazepines assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GCMS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Device Story
In vitro diagnostic assay for qualitative detection of benzodiazepines in human urine; 200 ng/mL cutoff. Principle: homogeneous enzyme immunoassay; competition between specimen drug and G6PDH-labeled drug for antibody binding sites. Enzyme activity inversely proportional to drug concentration; conversion of substrate to NADH measured via spectrophotometry. Used in clinical laboratories by trained personnel. Output: preliminary analytical result; requires confirmation by GC/MS. Assists clinicians in identifying benzodiazepine use or overdose.
Clinical Evidence
Bench testing only. Method comparison studies using AEROSET System showed 99% agreement with predicate (Emit II) and 87% agreement with GC/MS. Clinical specimens ranged from 57.5 to 13,174.0 ng/mL. Precision studies (within-run and total) performed with five levels of material; total %CV ranged from 1.10% to 1.92%. Limit of detection (sensitivity) is 15 ng/mL.
Technological Characteristics
Homogeneous enzyme immunoassay; G6PDH-labeled drug competition principle. Spectrophotometric detection of NADH absorbance change. Calibrated with lormetazepam. 200 ng/mL cutoff. Designed for use on automated clinical chemistry analyzers (e.g., AEROSET System).
Indications for Use
Indicated for the qualitative analysis of benzodiazepines in human urine in clinical laboratory settings to assist in the diagnosis of benzodiazepine use or overdose. Not for over-the-counter use.
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
## MAR 1 3 2002
# <013143
Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive MS 1-8 Irving, Texas 75038
Regulatory Specialist Affairs Regulatory Affairs
November 21, 2001 Date of Preparation of this Summary Device Trade or Proprietary Name: Benzodiazepines Device Common/Usual Name or Class me: Benzodiazepines Classification Number/Class: JXM/Class II
This summary of 510(k) safety and effectivene formation is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.
The assigned 510(k) number is: K01314
#### Test Description:
The Benzodiazepines assay is an in vitro diagnostic assay for the qualitative analysis of benzodiazepines in human urinc. The assay is a homogeneous enzyme immunoassay with a 200 ng/mL cutoff. The assay is based on competition between arug in the specimen and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity concentration in the specimen can be measured in decreases upon binding to the antibody, so the drug terms of enzyme activity. Active enzyme conve D to NADH, resulting in an absorbance change that can be measured spectrophotometric
{1}------------------------------------------------
### Substantial Equivalence:
The Benzodiazepines assay is substanti ent to the Emit II Benzodiazepine assay (K910376) on the SYVA-30R Analyzer.
Both assays yield similar Performance (
Similarities:
- . Both assays are in vitro immunoass
- י Both assays can be used for the qualitative at alysis of Benzodiazepines.
- + Both assays yield similar results.
- en drug in the specimen and drug labeled with the Both assays are based on the competitio . enzyme glucose-6-phosphate dehydrogenase G6PDH) for antibody binding sites.
#### Differences:
- There is a difference between the as .
- is a qualitative and semiquantitative assay. . Benzodiazepines is a qualitative assa
#### Intended Use:
ve analysis of benzodiazepines in human urine with The Benzodiazepines assay is used for the qualita a cutoff of 200 ng/mL. For use in clinical labora
The Benzodiazepines assay is calibrated w zepam and will detect a variety of benzodiazepines and their metabolites.
#### Performance Characteristics:
Comparative performance studies were conducted using the AEROSET® System. The Benzodiazepines assay method comparison yielded acceptable correlation with the Emit II Benzodiazepine assay on the SYVA-30R Analyzer. The concordance table shows: 99% agreement. The Benzodiazepines ussay method comparison yielded acceptable correlation with GC/MS. The concordance table shows 87% agreement with GCMS. The clinical specimens tested ranged from 57.5 to 13,174.0 ng/mL. Precision studies were conducted using the Benzodiazepin say. A within-run and total precision study was
{2}------------------------------------------------
%CV for Verifier I is 1.36%, The total %CV for the performed using five levels of material. The tota Cutoff Calibrator is 1.48%. The total %CV for Verifier II is 1.10%. The total %CV for the - 25% Control of Cutoff Calibrator and the +25% Control of Cutoff Calibrator samples are 1.92% and 1.60% respectively. The Benzodiazepines cutoff is 200 ng mL. The limit of detection (sensitivity) of the Benzodiazepines assay is 15 ng/mL. These data demonstrate that the performance of the Benzodiazepines assay is substantially equivalent to the performance of the Emit II Benzodiazepines assay on the SYVA-30R Analyzer.
#### Conclusion:
The Benzodiazepines assay is substantially equivalent to the Emit II Benzodiazepine assay on the SYVA-30R Analyzer as demonstrated by results obtained in the studies.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 1 3 2002
Ms. Linda Morris Senior Regulatory Affairs Specialist Abbott Laboratories 1921 Hurd Dr. Irving. Texas 75038
Re: k013143
> Trade/Device Name: Benzodiazephines Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiaphines test system Regulatory Class: Class II Product Code: JXM Dated: November 26, 2001 Received: November 28, 2001
Dear Ms. Morris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket I nits lotter will and the program al equivalence of your device to a legally marketed prealicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (1) see contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K013143
Benzodiazepine Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
gualitative analysis of benzodiazepines in human The Benzodiazepines assay is t in clinical laboratories. Measurements obtained by this urine with a cutoff of 200 ng/m1 for use ent of benzodiazepines use or overdose. device are used in the diagnosis at
with lormetazepam and will detect a variety of The Benzodiazepines assay is calibrated benzodiazepines and their metabolites
The Benzodiazepines assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GCMS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive s are used.
a) Laboratory Devices
ONTINUE ON ANOTHER PAGE IF NEEDED) (PLEASE DO NOT WRITE BELOW TI
Prescription Use (Per 21 CFR 801.109)
of Device Evaluation (ODE) Concurrence of C
Over-The-Counter Use_
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.