K031579 · Akers Laboratories, Inc. · JIH · Dec 19, 2003 · Clinical Toxicology
Device Facts
Record ID
K031579
Device Name
INSTAREAD LITHIUM SYSTEM
Applicant
Akers Laboratories, Inc.
Product Code
JIH · Clinical Toxicology
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3560
Device Class
Class 2
Indications for Use
The InstaRead™ Lithium System is intended to measure lithium blood levels. Measurements of lithium are used to aid in the management of individuals taking lithium for the treatment of mental disturbances, such as manic-depressive illness (bipolar disorder). The test is targeted for physician's office use, and may be used with whole blood, serum, or EDTA-plasma.
Device Story
InstaRead™ Lithium System is a tabletop quantitative diagnostic device for physician office use. It measures lithium levels in whole blood, serum, or EDTA-plasma. The system includes a blood separator cassette, reagent-filled cuvettes, calibrator, and a photometric reader (spectrophotometer). The blood separator uses a lectin-binding membrane to capture blood cells and deliver a precise sample volume. The reagent contains a porphyrin that binds to lithium, causing an increase in light absorbance at 505 nm. The reader measures this absorbance to quantify lithium concentration. Results are displayed to the clinician, who uses them to manage lithium therapy for patients with bipolar disorder. Results must be interpreted alongside clinical findings; levels above 1.5 mEq/L require corroboration. The device provides a point-of-care alternative to central laboratory testing.
Clinical Evidence
Clinical performance was evaluated across multiple studies. A 200-sample serum study compared the device to a laboratory system, showing a total correlation coefficient (r) of 0.962. A whole blood study (n=29) showed an r of 0.983. Fingerstick whole blood studies (n=40 and n=20) compared results against atomic absorption reference laboratory testing, demonstrating correlation coefficients of 0.928 and 0.994, respectively. Non-clinical bench testing confirmed sensitivity of 0.1 mEq/L and linearity between 0.1 and 2.5 mEq/L.
Technological Characteristics
The system utilizes colorimetric spectrophotometry with a porphyrin-based lithium-sensitive dye. It features a lectin-coated membrane for blood cell separation. The reader is a handheld, battery-powered photometric device measuring absorbance at 505 nm. The system is designed for point-of-care use. It is a standalone device with no described connectivity or software algorithm complexity beyond basic photometric signal processing.
Indications for Use
Indicated for the management and monitoring of lithium levels in psychiatric patients treated with lithium for manic-depressive illness (bipolar disorder). Suitable for use with whole blood, serum, or EDTA-plasma in physician office settings.
Regulatory Classification
Identification
A lithium test system is a device intended to measure lithium (from the drug lithium carbonate) in serum or plasma. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number: K031579
B. Analyte: lithium
C. Type of Test: quantitative fluorescence assay
D. Applicant: Akers Laboratories Inc.
E. Proprietary and Established Names: InstaRead™ Lithium System
F. Regulatory Information:
1. Regulation section: 21CFR862.3560, Lithium Test System
2. Classification: Class II
3. Product Code: JIH
4. Panel: Toxicology (91)
G. Intended Use:
1. Indication(s) for use: The InstaRead™ Lithium System is intended to measure lithium blood levels. Measurements of lithium are used to aid in the management of individuals taking lithium for the treatment of mental disturbances, such as manic-depressive illness (bipolar disorder). The test is targeted for physician's office use and may be used with whole blood, serum or EDTA-plasma.
2. Special condition for use statement(s): Results should be interpreted in conjunction with clinical findings and levels above 1.5 mEq/L should be corroborated.
3. Special instrument Requirements:
H. Device Description:
The test system is a tabletop system and includes a blood separator (cassettes), reagent-filled cuvettes, calibrator and test reader (spectrophotometer). It also includes fingerstick blood sampling supplies.
I. Substantial Equivalence Information:
1. Predicate device name(s): Johnson and Johnson Clinical Diagnostics Vitros Lithium System.
2. Predicate K number(s): K934106
{1}
Page 2 of 8
3. Comparison with predicate: Both devices measure lithium as an aid management of patients taking lithium. The predicate device uses serum or plasma samples; InstaRead™ can also accept whole blood as a sample type since it includes a blood cell separator as part of the system. InstaRead™ is indicated for use in physician's offices; the predicate is intended for laboratory use. Both devices use colorimetric spectrophotometry, although the reagents differ. The upper range of linearity for this device is to 2.5 mEq/L; the predicate device is linear to 4.0 mEq/L.
## J. Standard/Guidance Document Referenced (if applicable):
K. Test Principle: The device includes a blood cell separator component, reagents for lithium measurement and a photometric reader. The whole blood separator includes a membrane which is designed to capture blood cells by lectin binding as well as deliver a specific sample volume to be added to the lithium reagent. The reagent contains a porphyrin which, when bound to lithium, causes an increase in light absorbance at 505 nm. Absorbance is measured and displayed by the photometric reader.
## L. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Within-day precision of spiked serum samples was evaluated at concentrations (determined by a reference laboratory method) spanning the range of 0-1.9 mEq/L. Seven determinations were made for each sample within one day. Results are tabulated below:
| Target concentration (mEq/L) | Range observed(mEq/L) | Standard deviation (mEq/L) |
| --- | --- | --- |
| 0.0 | 0.0 | 0.0 |
| 0.12 | 0.1-0.2 | 0.04 |
| 0.24 | 0.2-0.3 | 0.05 |
| 0.30 | 0.3 | 0.00 |
| 0.47 | 0.5-0.6 | 0.05 |
| 0.60 | 0.5-0.6 | 0.05 |
| 0.84 | 0.8-0.9 | 0.04 |
| 0.95 | 0.9 | 0.00 |
| 1.20 | 1.1-1.3 | 0.08 |
| 1.55 | 1.4-1.5 | 0.04 |
| 1.90 | 1.7-1.9 | 0.08 |
Precision over 5 days was evaluated using clinical serum samples at concentrations of 0.51 and 1.23 mEq/L, as determined by a reference laboratory. Each sample was analyzed by the ™ System10 times per day for 5 days (total n per sample=50). Results are tabulated below:
{2}
| Target level (mEq/L) | Mean concentrations observed (range over 5 days) | Standard deviations (range over 5 days) |
| --- | --- | --- |
| 0.51 | 0.5-0.6 | 0.06-0.07 |
| 1.23 | 1.3-1.4 | 0.05-0.08 |
Precision with EDTA plasma samples was evaluated over 3 days, using samples at concentrations of 0.3 and $0.7\mathrm{mEq / L}$ , as quantitated by a reference laboratory. Each sample was analyzed by the Instaread™ 10 times per day for 3 days (total n per sample = 30). Results are tabulated below:
| Target level (mEq/L) | Mean concentrations observed (range over 3 days) | Standard deviations (range over 3 days) |
| --- | --- | --- |
| 0.3 | 0.3-0.4 | 0.05-0.11 |
| 0.7 | 0.8-0.9 | 0.04-0.06 |
Within-run precision for whole blood samples was evaluated using two spiked EDTA whole blood samples. Two operators and two readers were incorporated in this evaluation. Results shown below were determined from 40 replicates for each level (20 replicates per reader). There was no observed effect of Operator or Instrument on precision. (These results are shown as 2 significant figures, however the marketed device reads to 1 significant figure.)
| Target Level (mEq/l)/Operator/Reader | Mean | SD | %CV |
| --- | --- | --- | --- |
| 0.5/Operator 1/Both readers | 0.41 | 0.05 | 11.8 |
| 1.2/Operator 1/Both readers | 1.02 | 0.08 | 7.4 |
| 0.5/Operator 2/Both readers | 0.38 | 0.05 | 13.7 |
| 1.2/Operator 2/Both readers | 1.03 | 0.09 | 8.9 |
Precision was also assessed over 20 days for 2 levels of lithium controls. The following results were obtained based on the average of all values over the 20 days:
| Mean value | SD | %CV |
| --- | --- | --- |
| 0.49 | 0.04 | 8.1 |
| 1.26 | 0.06 | 4.4 |
b. Linearity/assay reportable range:
{3}
The instrument reports values between 0-2.5 mEq/L lithium. Linearity was evaluated with samples prepared by diluting a concentrated LiCl solution into serum and EDTA-plasma samples. Twenty five prepared serum samples and 11 prepared plasma sample in the range between 0-2.5 mEq/L were evaluated. Results with serum are summarized below. Results with plasma show comparable linearity.

c. Traceability (controls, calibrators, or method):
Calibrators are prepared from commercially available material that is analyzed and assigned by atomic absorption spectroscopy. Calibrators are further evaluated using assay reagents on an automated analyzer.
Acceptance criteria are that recoveries of test standards and control standards (previously approved fill) must agree within $5\%$ .
Observed recoveries for accelerated stability testing (45 degrees C, 5.5 days, 60 degrees C for 28 hours) were between $99 - 102\%$ . Acceptance criteria are that recoveries should be within $10\%$ of "fresh" sample results. Real-time stability studies are ongoing.
External controls are available commercially.
d. Detection limit: Whole blood free of lithium was assayed 20 times within one day. Results ranged from 0 to $0.16\mathrm{mEq / L}$ (which will read as 0.2 on the marketed device.) The mean was .054 and $2\mathrm{SD} = 0.082$ Comparable results were obtained in testing with serum.
e. Analytical specificity: To evaluate potential interference from endogenous compounds, the following compounds were spiked into a commercially available serum control containing approximately $1\mathrm{mEq/L}$ lithium:
conjugated bilirubin up to 769 uMol/L,
triglycerides up to $22.6\mathrm{mMol / L}$
hemoglobin up to $1000\mathrm{mg / dL}$
{4}
Page 5 of 8
Less than 10% deviation was observed for bilirubin and triglycerides.
Less than 5% deviation was observed from hemoglobin up to 1000 mg/dL.
To evaluate potential interference from endogenous compounds, the following compounds were spiked into a commercially available serum control containing approximately 1 mEq/L lithium:
- N-acetylcysteine at 90 mg/dL,
- Carbamazepine at 12 ug/ml,
- Procainamide at 8 ug/ml,
- Quinidine at 6 ug/ml,
- Valproic Acid at 100 ug/ml.
Less than 10% assay bias was observed for samples containing these drugs.
To evaluate potential assay interference by other ions, the following were tested with assay reagents in solutions containing 1 mEq/L lithium:
- Sodium up to 200 mmol/L
- Potassium up to 8 mMol.L
- Calcium up to 4 mMol/L
- Magnesium up to 2 mMol/L
- Iron up to 200 uMol/L
- Zinc up to 250 uMol/L
- Copper up to 250 uMol/L
Less than 5% deviation from expected results was observed.
f. Assay cut-off: Not applicable. The device is quantitative.
2. Comparison studies:
a. Method comparison with predicate device:
Three types of samples were evaluated in terms of comparison to previously cleared devices. (1) One hundred serum samples were evaluated at each of two central laboratories and compared to results obtained with predicate devices. (2) Twenty nine spiked whole blood samples were evaluated at a physician's office and compared to the predicate device. (Samples were processed to plasma before they were evaluated with the predicate device). (3) Clinical fingerstick blood samples were evaluated at one physicians office in two separate studies. Subjects were consenting adults currently taking lithium under care of the physician at the site; there were no other specific inclusion/exclusion criteria. At that site, venous whole blood and fingerstick samples were collected from each patient. Fingerstick samples were measured by the Instaread™ Lithium Test and venous whole blood samples were processed to serum and evaluated at a reference laboratory by a predicate device. All samples were masked. Summaries of results of all the studies are summarized below:
{5}
| Sample type | Serum | Serum | Fingerstick | Fingerstick | Spiked whole blood |
| --- | --- | --- | --- | --- | --- |
| Site type | Central Lab | Central Lab | Physician's office | Physician's office | Physician's office |
| Sample # | 100 | 100 | 40 | 20 | 29 |
| Range (mEq/L) | 0-2.3 | 0-2.5 | 0-1.6 | 0-1.8 | 0.1-2.4 |
| Comparator | Vitros | Vitros | AA (matched serum) | AA (matched serum) | Vitros (matched plasma) |
| Slope (95%CI) | 0.97 | 1.025 | 0.85 (0.74-0.96) | 1.2 (1.13-1.26) | 0.83 (0.77-0.90) |
| Int (95%CI) | 0.03 (-0.04- 0.1) | -0.07 (-0.15-0.02) | 0.02 (-0.07-0.11) | 0 (-0.06-0.06) | 0.05 (-0.03-0.13) |
| R | 0.97 | 0.96 | 0.93 | 0.99 | 0.98 |
| Sy|x | 0.14 | 0.17 | 0.14 | 0.06 | 0.10 |
| Average bias relative to predicate | -0.01 | -0.03 | -0.09 | 0.16 | -.13 |
The graph below illustrates combined data for the two serum studies:

{6}
The graph below illustrates combined data for the two physician office fingerstick studies:

# 3. Clinical studies:
a. Clinical sensitivity: Not applicable. Not typically provided for this device type.
b. Clinical specificity: Not applicable. Not typically provided for this device type.
c. Other clinical supportive data (when $a$ and $b$ are not applicable):
4. Clinical cut-off: Not applicable. The device is for quantitative measurements.
5. Expected values/Reference range: The package insert recommends that clinicians should make their own assessments regarding patient status and lithium concentrations. Appropriate levels may vary among patients and may depend slightly on the specific assay used. Lithium values should be interpreted in conjunction with other clinical findings and values above 1.5 mEq/L should be corroborated. The package insert also cites the following. Mild to moderate adverse events may occur at concentrations between 1.5-2.5 mEq/L and moderate to severe reactions may occur at concentrations above 2.0 mEq/L, based on measurements 8-12 hrs post-dose.
# M. System Descriptions:
# 1. Modes of Operation:
The reader measures absorbance of a solution containing the separated sample and reagents.
{7}
Page 8 of 8
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Sample Identification:
None.
4. Specimen Sampling and Handling:
Fingerstick samples are applied to the whole blood separator. Specific volumes of sample (serum or plasma) are then collected in a pipette by the separator unit and added manually to the reagent.
5. Assay Types:
Chemical binding, Quantitative
6. Reaction Types:
Absorbance
7. Calibration: Calibration is performed by the instrument using an external calibrator provided to the user. Calibration information is stored in the instrument.
8. Quality Control: External quality control materials are commercially available.
N. Other Supportive Instrument Performance Characteristics Data Not Covered In The “L. Performance Characteristics” Section Of The SE Determination Decision Summary. The collection device was previously cleared under K030815.
O. Conclusion: I recommend that the Akers InstaRead™ Lithium System is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.