K981705 · American Bio Medica Corp. · DKZ · Aug 27, 1998 · Clinical Toxicology
Device Facts
Record ID
K981705
Device Name
'RAPID DRUG SCREEN' 8 PANEL-LOW VOLUME
Applicant
American Bio Medica Corp.
Product Code
DKZ · Clinical Toxicology
Decision Date
Aug 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Indications for Use
Rapid Drug Screen 8-Panel low volume test is a one-step, lateral flow immunoassay for the simultaneous detection of eight abused substances in urine (each assay occupies a separate channel). Rapid Drug Screen 8-Panel low volume test is intended for use in the qualitative detection of the following 8 drugs of abuse in human urine at the following levels: d-Amphetamine 750 ng/ml; Barbiturates 300 ng/ml; Benzodiazepines 300 ng/ml; Benzoyl ecognine 225 ng/ml; Cannabinoids (11-nor-9-carboxy-delta-9-THC) 50 ng/ml; Methamphetamine 1000 ng/ml; Opiates (codeine) 225 ng/ml; (morphine-3-glucuronide) 225 ng/ml; Phencyclidine (PCP) 19 ng/ml. Rapid Drug Screen 8-Panel low volume test is intended for use by professional laboratories and physicians offices in a clinical setting. The assays are easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary analytical test result for use in determining the need for additional or confirmatory testing (i.e., gas chromatography/mass spectrometry (GC/MS)). Rapid Drug Screen 8-Panel low volume test provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Device Story
Rapid Drug Screen 8-Panel is a one-step lateral flow immunoassay; detects 8 drugs of abuse in human urine simultaneously; each assay occupies a separate channel. Used in clinical settings by professional laboratory staff or physicians. Device provides preliminary qualitative results; requires confirmatory testing (e.g., GC/MS) for definitive identification. Assists clinicians in identifying potential substance abuse; supports decision-making regarding further diagnostic or clinical intervention.
Clinical Evidence
No clinical data provided; device is a qualitative screening immunoassay requiring confirmatory testing.
Technological Characteristics
Lateral flow immunoassay; multi-channel test panel; qualitative detection of 8 drug analytes in urine; professional use; non-automated.
Indications for Use
Indicated for the qualitative screening of 8 drugs of abuse (Amphetamine, Barbiturates, Benzodiazepines, Benzoyl ecognine, Cannabinoids, Methamphetamine, Opiates, PCP) in human urine. Intended for professional use in clinical laboratories and physician offices. Provides preliminary results requiring confirmatory testing via GC/MS.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K980872 — RAPID DRUG SCREEN' 8 PANEL · American Bio Medica Corp. · Mar 20, 1998
K021006 — STARTOX DRUG OF ABUSE SCREENING TEST · Starplex Scientific, Inc. · Jun 10, 2002
K032575 — FIRST SIGN DRUG OF ABUSE SCREENING TEST · W.H.P.M., Inc. · May 19, 2004
K033566 — MICROMEDIC DRUGS OF ABUSE PANEL TEST (9), CATALOG NUMBER 07RD-7062 · Rapid Diagnostics, Inc. · May 17, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three human profiles, which are connected and appear to be facing to the right.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG 27 1998
John B. Dubeck, Esq. American Bio Medica Corporation C/O Keller and Heckman LLP 1001 G Street, NW, Suite 500W Washington, D.C. 20001
Re : K981705 Rapid Drug Screen 8-Panel Low Volume II Regulatory Class: Product Code: DKZ, DIO, DJG, DIS, LDJ, LCL, JXM Dated: August 19, 1998 Received: August 19, 1998
Dear Mr. Dubeck:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Page
1 of 1
510(k) NUMBER (IF KNOWN) :
"Rapid Drug Screen" 8 Panel low volume test DEVICE NAME:
## INDICATIONS FOR USE:
"Rapid Drug Screen" 8-Panel low volume test is a one-step, lateral flow immunoassay for the simultaneous detection of eight abused substances in urine (each assay occupies a separate channel). "Rapid Drug Screen" 8-Panel low volume test is intended for use in the qualitative detection of the following 8 drugs of abuse in human urine at the following levels:
| d-Amphetamine | 750 ng/ml |
|--------------------------------|------------|
| Barbiturates | 300 ng/ml |
| Benzodiazepines | 300 ng/ml |
| Benzoyl ecognine | 225 ng/ml |
| Cannabinoids | |
| (11-nor-9-carboxy-delta-9-THC) | 50 ng/ml |
| Methamphetamine | 1000 ng/ml |
| Opiates (codeine) | 225 ng/ml |
| (morphine-3-glucuronide) | 225 ng/ml |
| Phencyclidine (PCP) | 19 ng/ml |
"Rapid Drug Screen" 8-Panel low volume test is intended for use by professional laboratories and physicians offices in a clinical setting. The assays are easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary analytical test result for use in determining the need for additional or confirmatory testing (i.e., · gas chromatography/mass spectrometry (GC/MS)).
"Rapid Drug Screen" 8-Panel low volume test provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div align="center">AH An AWM</div> |
|-----------------------------------------|-------------------------------------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K981705 |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.