QUIKPAC II ONE STEP AMPHETAMINE TEST
Device Facts
| Record ID | K971109 |
|---|---|
| Device Name | QUIKPAC II ONE STEP AMPHETAMINE TEST |
| Applicant | Drial Consultants, Inc. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | May 14, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Indications for Use
The QuikPac II One Step Amphetamine Test is intended for professional medical/forensic screening of urine.
Device Story
Chromatographic absorbent device for qualitative urine drug screening; lateral flow immunoassay principle. Sample flows through membrane; drug/metabolites compete with immobilized drug conjugate for limited antibody sites. Labeled-antibody-dye conjugate binds free drug; complex competes with immobilized antigen in positive reaction zone. Absence of magenta band indicates drug presence >500 ng/ml; presence of magenta band in control zone confirms reagent/device function. Used in professional medical/forensic settings; operated by trained personnel. Visual interpretation of color bands by clinician; results provide preliminary screening data requiring GC/MS confirmation for positive results.
Clinical Evidence
Clinical trial of 308 urine samples compared QuikPac II to Syva EMIT® II. Results: 100% relative sensitivity, 100% relative specificity, 100% accuracy. All positive screening results confirmed by GC/MS. Discrepant analysis identified cross-reactivity with Ephedrine and Phenyl PropanolAmine (PPA).
Technological Characteristics
Chromatographic absorbent device; porous membrane support; antibody-dye conjugate; lateral flow immunoassay. Qualitative visual readout. No external energy source. Standalone test kit.
Indications for Use
Indicated for professional medical/forensic qualitative screening of human urine for the presence of Amphetamine and its metabolites at a cutoff concentration of 500 ng/ml.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Reference Devices
- Syva EMIT® II
Related Devices
- K973269 — QUIKSTRIP ONE STEP BENZODIAZEPINE TEST · Drial Consultants, Inc. · Feb 25, 1998
- K971218 — QUIKSTRIP ONE STEP AMPHETAMINE TEST · Drial Consultants, Inc. · May 30, 1997
- K980776 — ONE STEP URINE DRUG OF ABUSE: BENZODIAZEPINE TEST · Tcpi, Inc. · Apr 29, 1998
- K970724 — QUICKPAC II ONE STEP METHAMPHETAMINE TEST · Drial Consultants, Inc. · May 13, 1997
- K951853 — ONE STEP URINE DRUG OF ABUSE CANNABINOID TEST (THC) · Tcpi, Inc. · Aug 20, 1996