Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug Urine Test Panel (Home Use)
K152122 · Alfa Scientific Designs, Inc. · DKZ · Jun 24, 2016 · Clinical Toxicology
Device Facts
Record ID
K152122
Device Name
Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug Urine Test Panel (Home Use)
Applicant
Alfa Scientific Designs, Inc.
Product Code
DKZ · Clinical Toxicology
Decision Date
Jun 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Intended Use
The Instant-View® Multi-Drug Urine Test Cup (Home Use) and Instant-View® Multi-Drug Urine Test Panel (Home Use) are rapid, qualitative immunoassays for the detection in human urine of one or more of the drugs at the cutoff concentrations listed below. These devices will detect up to 13 of the drugs below in any combination. The tests are not intended to distinguish prescription use or abuse of any drugs. These devices provide only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained. This test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. Tests for prescription drugs will yield preliminary positive results when these drugs are ingested at or above therapeutic doses. There are no uniformly recognized drug levels for prescription drugs in urine. This multi-drug of abuse urine test device shows the drug was or was not present at the cutoff level.
Device Story
Lateral flow chromatographic immunoassay; detects drugs in human urine. Device consists of 1-13 test strips; each strip contains conjugate pad with colloidal gold-labeled anti-drug antibodies and nitrocellulose membrane with test (T) and control (C) lines. User applies urine sample; capillary action moves sample across membrane. Competitive binding principle: drug in sample competes with conjugated drug antigen for antibody binding sites. Absence of T-line indicates presence of drug above cutoff; presence of T-line indicates drug absence or concentration below cutoff. C-line confirms proper liquid flow. Used in home setting by lay users. Provides preliminary results; requires confirmatory testing (e.g., GC/MS) for clinical decision-making. Benefits include rapid, accessible information for consumers regarding potential drug presence.
Clinical Evidence
Lay-user study (n=400) across three sites; participants (18 to >60 years) with no prior OTC test experience. Samples spiked with drugs at 0%, 50%, 75%, 125%, and 150% of cutoff concentrations. Results compared to GC/MS. Average interpretation agreement >96%. Survey data indicates high ease-of-use and understanding of labeling (Flesch-Kincaid grade level 7).
Technological Characteristics
Lateral flow immunochromatographic assay. Competitive binding principle. Components: conjugate pad with colloidal gold-labeled anti-drug antibodies; nitrocellulose membrane with immobilized drug-specific conjugate (T-line) and control line (C-line). Form factors: test cup and panel cassette. Storage: 2°C to 30°C. Shelf life: 24 months. No electronic components or software.
Indications for Use
Indicated for OTC qualitative detection of up to 13 drugs/metabolites in human urine (Amphetamine, Barbiturates, Buprenorphine, Benzodiazepine, Cocaine, Methamphetamine, Methadone, Phencyclidine, Tricyclic antidepressants, Cannabinoid, MDMA, Morphine, Oxycodone) at specified cutoff concentrations. Intended for use by consumers (e.g., parents) as a preliminary screening tool; not for distinguishing prescription use from abuse.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Instant-View Multi-Drug of Abuse Urine Test Devices/Cup (k063545)
Instant-View Multi-Drug of Abuse Urine Test Devices/Cassettes (k063545)
Related Devices
K063545 — MULTI-DRUG OF ABUSE URINE TEST,URINE CUP TEST, AMPHETAMINE URINE TEST(CASSETTE, DIP-STRIP), COCAINE, METHAMPHETAMINE · Alfa Scientific Designs, Inc. · May 7, 2007
K243365 — Healgen Accurate Muti-Drug Urine Drug Screen Cup; Healgen Accurate Home Muti-Drug Urine Test Cup · Healgen Scientific,, LLC · Dec 17, 2024
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE
A. 510(k) Number:
k152122
B. Purpose for Submission:
Over the counter clearance of a device previously cleared for prescription use only.
C. Measurand:
Amphetamine, Barbiturates, Buprenorphine, Benzodiazepine, Cocaine, Methamphetamine, Methadone, Phencyclidine, Tricyclic antidepressants, Cannabinoid, 3,4-Methylenedioxymethamphetamine (MDMA), Morphine, Oxycodone.
D. Type of Test:
Qualitative
E. Applicant:
ALFA Scientific Designs, Inc.
F. Proprietary and Established Names:
Instant-View Multi-Drug Urine Test Cup (Home Use)
Instant-View Multi-Drug Urine Test Panel (Home Use)
G. Regulatory Information:
| Product Code | Regulation Name | Regulation Section | Panel |
| --- | --- | --- | --- |
| DKZ | Amphetamine test system | 862.3100 | Toxicology (91) |
| JXM | Benzodiazepine test system | 862.3170 | |
| DIO | Cocaine and cocaine metabolite test system. | 862.3250 | |
| LDJ | Cannabinoid test system | 862.3870 | |
| DJC | Methamphetamine test system | 862.3610 | |
| LCM | Phencyclidine test system | Unclassified | |
| DIS | Barbiturate test system | 8862.3150 | |
| DJG | Opiate test system | 862.3650 | |
| DJR | Methadone test system | 862.3620 | |
| LFG | Tricyclic antidepressant drugs test system | 862.3910 | |
{1}
H. Intended Use:
1. Intended use(s):
Refer to Indications for Use
2. Indication(s) for use:
The Instant-View® Multi-Drug Urine Test Cup (Home Use) and Instant-View® Multi-Drug Urine Test Panel (Home Use) are rapid, qualitative immunoassays for the detection in human urine of one or more of the drugs at the cutoff concentrations listed below. These devices will detect up to 13 of the drugs below in any combination. The tests are not intended to distinguish prescription use or abuse of any drugs.
| Abbreviation | Test | Calibrator | Cut-off (ng/mL) |
| --- | --- | --- | --- |
| AMP | Amphetamine | d-Amphetamine | 1000 |
| BAR | Barbiturates | Secobarbital | 200 |
| BUP | Buprenorphine | Buprenorphine | 10 |
| BZD | Benzodiazepine | Oxazepam | 300 |
| COC | Cocaine | Benzoylecgonine | 300 |
| MDMA | MDMA or Ecstasy | Methylenedioxy-methamphetamine | 500 |
| MET | Methamphetamine | d-Methamphetamine | 1000 |
| MTD | Methadone | Methadone | 300 |
| MOR | Morphine/Opiate | Morphine | 2000 |
| OXY | Oxycodone | Oxycodone | 300 |
| PCP | Phencyclidine | Phencyclidine | 25 |
| TCA | Tricyclic Antidepressants | Nortriptyline | 1000 |
| THC | Marijuana | 11-nor-.69-THC-9-COOH | 50 |
These devices provide only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
This test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.
Tests for prescription drugs will yield preliminary positive results when these drugs are
{2}
ingested at or above therapeutic doses. There are no uniformly recognized drug levels for prescription drugs in urine. This multi-drug of abuse urine test device shows the drug was or was not present at the cutoff level.
3. Special conditions for use statement(s):
For Over the Counter Use.
4. Special instrument requirements:
Not applicable.
I. Device Description:
The Instant View Multi-Drug Urine Test Devices/Cup and the Instant View Multi-Drug Urine Test Devices/Panel Cassettes are one-step lateral flow chromatographic immunoassays. Each device consists of any combination of one to thirteen individual test strips(s) for the analytes(s) being tested. Each test strip in the device consists of 1) a conjugate pad containing colloidal gold coupled with the anti-drug antibodies and 2) nitrocellulose membrane containing a test line (T line) coated with the conjugated drug antigen and a control line (C line). The C line serves as an internal quality control of the system and appears as a burgundy -colored band during the test regardless of the presence of the drug.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Instant-View Multi-Drug of Abuse Urine Test Devices/Cup
Instant-View Multi-Drug of Abuse Urine Test Devices/Cassettes
2. Predicate 510(k) number(s):
k063545
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device k063545 |
| Intended Use | Qualitative detection of potential abuse of one or more drugs | Same |
| Test Principle | Chromatographic | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device k063545 |
| | immunoassay | |
| Test Strips | Each test strip in the device consists of 1) a conjugate pad containing colloidal gold coupled with the anti-drug antibodies and 2) nitrocellulose membrane containing a test line (T line) coated with the conjugated drug antigen and a control line (C line). | Same |
| Specimen | Urine | Same |
| Drug(s) detected per test strip | Each test strip detects 1 analyte | Same |
| Test formats | Cup and panel cassette | Same |
| Shelf life | 24 months | Same |
| Storage Conditions | 2°C to 30°C | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device k063545 |
| Number of analytes | 13 | 14 |
| Intended Use population | OTC use | Prescription use |
## K. Standard/Guidance Document Referenced (if applicable):
Not applicable
## L. Test Principle:
Each device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. Drugs, if present in concentrations below the cutoff level, will not saturate the binding sites of the antibody coated particles on the drug specific test strips. The goat-anti-rabbit IgG antibody-coated particles will then be captured by immobilized drug-specific conjugate. If the level of drug in the urine specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the urine specimen is above the cutoff, no T line develops.
The control line (C line) serves as an internal quality control. The control line should always appear as a burgundy-colored band regardless of the presence of the drug, if enough sample volume has been added to the test and if the sample has correctly migrated up the test strip. Testing is based on the principle of a competitive immunochemical reaction between a chemically labeled drug (drug-protein conjugate) and the drug or drug metabolites which may be present in the urine sample competing for the limited antibody binding sites.
{4}
M. Performance Characteristics (if/when applicable):
This multi-drug test was cleared for prescription use only in k063545. Twelve analytes of the candidate devices were previously cleared for OTC use in k131645. The candidate devices contain test strips identical to k063545 and k131645. Specimen matrix, test formats, and cut-off concentrations and analytes tested are also identical.
Lay user testing for buprenorphine, the only analyte not previously cleared for OTC use, is described in section 3.c. below.
1. Analytical performance:
a. Precision/Reproducibility:
See k063545
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
See k063545
d. Detection limit:
See k063545
e. Analytical specificity:
For exogenous interference, see k063545.
For cross-reactivity information for Amphetamine, Cocaine, Methamphetamine and Oxycodone, see k063545. For cross reactivity information for Buprenorphine, see k060527..
For interference studies to evaluate the potential of interference between drug analytes with the exception of Buprenorphine, see k131645.
f. Assay cut-off:
See k063545
{5}
6
2. Comparison studies:
a. Method comparison with predicate device:
See k063545
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
A lay user study was performed at three intended user sites with 400 lay persons testing cup and panel devices. A total of 210 females and 190 males tested the devices. They had diverse educational and professional backgrounds and ranged in age from 18 to >60 years. The participants had no previous experience with an over the counter (OTC) drug test. Urine samples were prepared at the following concentrations: negative, ± 50%, ± 25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package inserts for the test cup and the test panel and performed the test independently without additional help.
The lay user study results are summarized in the tables below:
Summary of Results for the Cup Format
| Drug | Cutoff Concentration% (ng/ml) | Number of studies | Correctly interpreted | Incorrectly interpreted | % agreement |
| --- | --- | --- | --- | --- | --- |
| AMP | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (750) | 10 | 8 | 2 | 80% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
| BAR | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (150) | 10 | 9 | 1 | 90% |
| | 125% (250) | 10 | 9 | 1 | 90% |
| | 150% (300) | 30 | 30 | 0 | 100% |
{6}
| Drug | Cutoff Concentration% (ng/ml) | Number of studies | Correctly interpreted | Incorrectly interpreted | % agreement |
| --- | --- | --- | --- | --- | --- |
| BUP | 0% (0) | 20 | 20 | 0 | 100% |
| | 50% (5) | 60 | 60 | 0 | 100% |
| | 75% (7.5) | 60 | 56 | 4 | 93.3% |
| | 125% (12.5) | 120 | 109 | 11 | 90.8% |
| | 150% (15) | 140 | 140 | 0 | 100% |
| BZD | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 9 | 1 | 90% |
| | 125% (375) | 10 | 9 | 1 | 90% |
| | 150% (450) | 30 | 30 | 0 | 100% |
| COC | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 9 | 1 | 90% |
| | 125% (375) | 10 | 8 | 2 | 80% |
| | 150% (450) | 30 | 30 | 0 | 100% |
| MET | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (750) | 10 | 9 | 1 | 90% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
| MTD | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 8 | 2 | 80% |
| | 125% (375) | 10 | 10 | 0 | 100% |
| | 150% (450) | 30 | 30 | 0 | 100% |
| PCP | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (12.5) | 10 | 10 | 0 | 100% |
| | 125% (31.25) | 10 | 9 | 1 | 90% |
| | 150% (37.5) | 30 | 30 | 0 | 100% |
| TCA | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (500) | 10 | 10 | 0 | 100% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
| THC | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (25) | 10 | 10 | 0 | 100% |
| | 125% (62.5) | 10 | 8 | 2 | 80% |
| | 150% (75) | 30 | 30 | 0 | 100% |
| XTC | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (250) | 10 | 10 | 0 | 100% |
| | 125% (625) | 10 | 9 | 1 | 90% |
| | 150% (750) | 30 | 30 | 0 | 100% |
| MOR | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (100) | 10 | 10 | 0 | 100% |
| | 125% (2500) | 10 | 8 | 2 | 80% |
| | 150% (3000) | 30 | 30 | 0 | 100% |
| OXY | 0% (0) | 350 | 350 | 0 | 100% |
{7}
8
| Drug | Cutoff Concentration % (ng/ml) | Number of studies | Correctly interpreted | Incorrectly interpreted | % agreement |
| --- | --- | --- | --- | --- | --- |
| AMP | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (750) | 10 | 9 | 1 | 90% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| BAR | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (150) | 10 | 9 | 1 | 90% |
| | 125% (250) | 10 | 8 | 2 | 80% |
| | 150% (300) | 30 | 30 | 0 | 100% |
| BUP | 0% (0) | 20 | 20 | 0 | 100% |
| | 50% (5) | 60 | 60 | 0 | 100% |
| | 75% (7.5) | 60 | 55 | 5 | 91.7% |
| | 125% (12.5) | 120 | 112 | 8 | 93.3% |
| BZD | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 9 | 1 | 90% |
| | 125% (375) | 10 | 9 | 1 | 90% |
| COC | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 9 | 1 | 90% |
| | 125% (375) | 10 | 9 | 1 | 90% |
| | 150% (450) | 30 | 30 | 0 | 100% |
| MET | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (750) | 10 | 8 | 2 | 80% |
| | 125% (1250) | 10 | 10 | 0 | 100% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
| MTD | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (225) | 10 | 9 | 1 | 90% |
| | 125% (375) | 10 | 9 | 1 | 90% |
| | 150% (450) | 30 | 30 | 0 | 100% |
| PCP | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (12.5) | 10 | 10 | 0 | 100% |
| | 125% (31.25) | 10 | 9 | 1 | 90% |
Summary of Results for the Panel Format
| Drug | Cutoff Concentration % (ng/ml) | Number of studies | Correctly interpreted | Incorrectly interpreted* | % agreement |
| --- | --- | --- | --- | --- | --- |
| AMP | 0% (0) | 350 | 350 | 0 | 100% |
| | 75% (750) | 10 | 9 | 1 | 90% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
{8}
| Drug | Cutoff Concentration% (ng/ml) | Number of studies | Correctly interpreted | Incorrectly interpreted* | % agreement |
| --- | --- | --- | --- | --- | --- |
| | 150% (37.5) | 30 | 30 | 0 | 100% |
| TCA | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (500) | 10 | 10 | 0 | 100% |
| | 125% (1250) | 10 | 9 | 1 | 90% |
| | 150% (1500) | 30 | 30 | 0 | 100% |
| THC | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (25) | 10 | 10 | 0 | 100% |
| | 125% (62.5) | 10 | 8 | 2 | 80% |
| | 150% (75) | 30 | 30 | 0 | 100% |
| XTC | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (250) | 10 | 10 | 0 | 100% |
| | 125% (625) | 10 | 9 | 1 | 90% |
| | 150% (750) | 30 | 30 | 0 | 100% |
| MOR | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (100) | 10 | 10 | 0 | 100% |
| | 125% (2500) | 10 | 9 | 1 | 90% |
| | 150% (3000) | 30 | 30 | 0 | 100% |
| OXY | 0% (0) | 350 | 350 | 0 | 100% |
| | 50% (150) | 10 | 10 | 0 | 100% |
| | 125% (375) | 10 | 8 | 2 | 80% |
| | 150% (450) | 30 | 30 | 0 | 100% |
No invalid test results were obtained during the lay user study.
All participants completed questionnaires after they completed the testing. The majority of participants indicated they could understand the labeling without difficulty.
A Flesh-Kincaid reading analysis revealed that package inserts for both formats had a reading grade level of 7.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{9}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.